← Product Code [JEY](/submissions/DE/subpart-e%E2%80%94surgical-devices/JEY) · K100779

# SUSTAIN SPACE MAINTENANCE SYSTEM (K100779)

_Medtronic Sofamor Danek USA, Inc. · JEY · Dec 13, 2010 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/JEY/K100779

## Device Facts

- **Applicant:** Medtronic Sofamor Danek USA, Inc.
- **Product Code:** [JEY](/submissions/DE/subpart-e%E2%80%94surgical-devices/JEY.md)
- **Decision Date:** Dec 13, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.4760
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Intended Use

The ARCHITEX™ Space Maintenance System is indicated for use as temporary implants to stabilize and support autograft, autograft extenders, allograft, and bone void fillers and/or fractured bone segments with or without bone plates or titanium mesh in bony defects of oral maxillofacial anatomy.

## Device Story

ARCHITEX™ Space Maintenance System comprises titanium mesh implants and various screws (Socket Preservation, Tenting, Mesh Fixation) used in oral-maxillofacial surgical reconstruction and dental regeneration. System maintains space during bone grafting and supports soft tissue until bone formation. Operated by clinicians in a surgical setting. Components provide mechanical stabilization for bone graft materials or fractured bone segments. Device benefits patient by facilitating bone regeneration in oral-maxillofacial defects. System includes convenience placement instruments.

## Clinical Evidence

No clinical data. Bench testing only. Performance testing included dimensional/material analysis, static axial pullout (ASTM F543-02), static subsidence, static cantilever bending, and static removal torque/torque to failure (ASTM F543-02). Results confirmed subject devices met all acceptance criteria and demonstrated equivalent strength characteristics to predicate devices.

## Technological Characteristics

Materials: Medical grade titanium/titanium alloy (ASTM F136-08e1, ISO 5832-3, ASTM F67-06) and medical grade stainless steel (ASTM F899-09). Components: Titanium mesh, socket preservation screws, tenting screws, mesh fixation screws. Form factor: Various shapes/sizes for oral-maxillofacial reconstruction. Non-sterile. No software or electronic components.

## Regulatory Identification

A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.

## Predicate Devices

- TiMesh® System ([K062348](/device/K062348.md))
- ACE Tru-FIX™ Implant System ([K080074](/device/K080074.md))
- Straumann Dental Implant System ([K071585](/device/K071585.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# ARCHITEX™ Space Maintenance System 510(k) Summarv

### December 2010

DEC 1 3 2010

1. Company: Medtronic Sofamor Danek USA 1800 Pyramid Place Memphis, Tennessee 38132 Telephone: (901) 396-3133 (901) 346-9738 Fax:

Contact: Ryan Massey, Sr. Regulatory Affairs Specialist

### II. Proposed Proprietary Trade Name: ARCHITEX™ Space Maintenance System

- 011. Classification Name(s): Implants, Endosseous, Root-Form per 21 CFR 872.4760; Product Code(s): DZE, JEY
#### IV. Description:

ARCHITEX™ Space Maintenance System is intended for use in oral-maxillofacial surgical reconstruction and dental regeneration procedures; the implants contained in the System are for maintaining space during bone grating procedures and/or to support soft tissue until bone formation.

The ARCHITEX™ Space Maintenance System consists of a variety of shapes and sizes of screws (Socket Preservation, Tenting, and Mesh Fixation Screws) and titanium mesh implants components as well as placement instruments components. These instruments supplied are general class I instruments added as a convenience for doctors to use to place the provided implants.

The ARCHITEX™ Space Maintenance System includes a selection of Socket Preservation screws which are designed to aid in extraction socket grafting while simultaneously supporting the original gingival margins and papilla.

The ARCHITEX™ Space Maintenance System components are fabricated from medical grade stainless steel and medical grade titanium or titanium alloy. Medical grade titanium and titanium alloy may be used together. The subject implants components will be manufactured from medical grade titanium allov described by such standards as ASTM F 136-08el or ISO 5832-3, and/or from medical grade unalloyed titanium described by such standard as ASTM F67-06. The placement instruments components will be manufactured from medical grade stainless steel as described in ASTM F8PP-09. The ARCHITEX™ Space Maintenance System is sold non-sterile.

The purpose of this 510(k) application is to seek marketing clearance for the ARCHITEX™ Space Maintenance System to be used as a root-form endosseous dental implants system.

{1}------------------------------------------------

#### IV. Indications for Use:

The ARCHITEX™ Space Maintenance System is indicated for use as temporary implants to stabilize and support autograft, autograft extenders, allograft, bone void fillers and/or fractured bone segments with or without bone plates or titanium mesh in bony defects of oral maxillofacial anatomy.

The Indications for Use Statement referenced above does not differ from the listed predicate devices.

### V. Identification of the Legally Marketed Predicate Devices Use to Claim Substantial Equivalence:

The design features, materials, and indications for use of the ARCHITEX™ Space Maintenance System incorporates a variety of functional aspects from the following predicate devices: TiMesh® System (K062348, SE 09/08/06), ACE Tru-FIX™ Implant System (K080074, SE 02/14/08), and Straumann Dental Implant System (K071585, SE 07/11/07),

| ARCHITEX™ Implant             | Predicate                                    | Substantial equivalence                                                                                                                                                         |
|-------------------------------|----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Titanium Mesh                 | TiMesh® (K062348)                            | Geometrically, materially,<br>and mechanically<br>equivalent                                                                                                                    |
| Titanium Mesh Screws          | TiMesh® (K062348)                            | Geometrically, materially,<br>and mechanically<br>equivalent.                                                                                                                   |
| Tenting Screws                | ACE Tru-FIX™ (K080074)                       | Geometrically, materially,<br>and mechanically<br>equivalent. Bone pull-out<br>strength and screw<br>subsidence equivalent to<br>predicate.                                     |
| Socket Preservation<br>Screws | ACE Tru-FIX™ (K080074)                       | Bone interfacing portion of<br>screw geometrically,<br>materially, and<br>mechanically equivalent.<br>Bone pull-out strength and<br>screw subsidence<br>equivalent to predicate |
|                               | Straumann Dental Implant<br>System (K071585) | Head portion of screw<br>geometrically, materially,<br>and mechanically<br>equivalent to predicate<br>abutments.                                                                |

### VI. Brief Discussion of the Non-Clinical Tests Submitted

For a determination of substantial equivalence, the following analysis and bench performance tests were performed on Subject Devices and Predicate Devices:

- > Analysis of Dimensional and Material Features
- ·> Statis Axial Pullout Testing in accordance with ASTM F543-02

{2}------------------------------------------------

- > Statis Subsidence Testing (To date there is no industry standards)
- > Statis Cantilever Bending Testing (To date there is no industry standards)
- > Static "Removal Torque" and "Torque to Failure" testing in accordance with ASTM F543-02

#### VII. Conclusions Drawn from the Non-Clinical Tests

Results of material and dimensional analysis have demonstrated that ARCHITEX™ Space Maintenance System implants are equivalent to the predicate devices in terms of bone/screw interface features.

When compared with predicate devices, results of bench performance testing indicated all acceptance criteria were met, and demonstrated the subject Tenting Screws and Socket Preservation Screws have equivalent static axial pullout and subsidence strength and that the subject Titanium Mesh has equivalent static cantilever bending strength.

The conclusions drawn from the analysis and performance testing, along with the intended use of the subject devices demonstrate that the ARCHITEX™ Space Maintenance System is substantially equivalent to the listed predicate devices.

{3}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

DFC 1 3 2010

Ms. Michelle Obenauer Regulatory Affairs Manager Medtronic Sofamor Danek USA, Incorporated 1800 Pyramid Place Memphis. Tennessee 38132

Re: K100779

Trade/Device Name: ARCHITEX™ Space Maintenance System Regulation Number: 21CFR.872.4760 Regulation Name: Bone Plate Regulatorý Class: II Product Code: JEY, DZE Dated: November 30, 2010 Received: December 1, 2010

Dear Ms. Obenauer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

### Page 2- Mr. Obenauer

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), a olease go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Wh for

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and

Radiological Health

Enclosure

{5}------------------------------------------------

## 510(k) Number (if known): K100779

## Device Name: ARCHITEX™ Space Maintenance System

### Indications for Use:

The ARCHITEX™ Space Maintenance System is indicated for use as temporary implants to stabilize and support autograft, autograft extenders, allograft, and bone void fillers and/or fractured bone segments with or without bone plates or ritanium mesh in bony defects of oral maxillofacial anatomy.

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

_Yes_X__OR Over-The-Counter Use No Prescription Use Per 21 CFR 801.109

Susan Reamer

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K100776

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/JEY/K100779](https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/JEY/K100779)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
