← Product Code [JEY](/submissions/DE/subpart-e%E2%80%94surgical-devices/JEY) · K092743

# THE LEIBINGER ADVANCE INTERNAL MIDFACE DISTRACTION SYSTEM (K092743)

_Stryker Craniomaxillofacial · JEY · Jan 15, 2010 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/JEY/K092743

## Device Facts

- **Applicant:** Stryker Craniomaxillofacial
- **Product Code:** [JEY](/submissions/DE/subpart-e%E2%80%94surgical-devices/JEY.md)
- **Decision Date:** Jan 15, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.4760
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Intended Use

The Leibinger Advance Internal Midface Distraction System is intended to be used in conditions of the cranium and mid-face for which osteotomy and segmental advancement are indicated.

## Device Story

Internal bone distraction system for cranial and midfacial reconstruction. Components include distractor frame, connection screws, plates, removable activation rod with protective sheath, and activation key. Device surgically implanted to stabilize osteotomy site; gradual lengthening achieved via manual activation of rod by clinician or patient using key. Used in clinical settings for patients requiring bone advancement. Benefits include controlled, gradual bone lengthening to correct skeletal deformities. No electronic or software components.

## Clinical Evidence

No clinical data provided; substantial equivalence based on design, intended use, and mechanical characteristics compared to predicate devices.

## Technological Characteristics

Metallic bone distraction system. Components: distractor frame, plates, activation rod, and key. Mechanical principle: gradual distraction via manual activation. No energy source, software, or electronic connectivity. Sterilization method not specified.

## Regulatory Identification

A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.

## Predicate Devices

- Cohen Distractor ([K972154](/device/K972154.md))
- Synthes Midface Distractor ([K022005](/device/K022005.md))
- Vazquez-Diner Intraoral Distractor ([K964649](/device/K964649.md))
- Molina Orbital Malar Distractor ([K003883](/device/K003883.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K092743

# 510(k) Summary of Safety and Effectiveness

JAN 1 5 2010

| Proprietary Name:                  | The Leibinger Advance Internal Midface<br>Distraction System                                                                                                                        |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name:                       | Bone Distraction System                                                                                                                                                             |
| Classification Name and Reference: | Smooth or Threaded metallic bone fixation fastener<br>21 CFR §888.4760                                                                                                              |
| Proposed Regulatory Class:         | Class II                                                                                                                                                                            |
| Product Codes:                     | JEY - Bone Plate                                                                                                                                                                    |
| For Information contact:           | Stephanie Fitts<br>Director, Regulatory Affairs and Compliance<br>Howmedica Osteonics.Corp.<br>325 Corporate Drive<br>Mahwah, NJ 07430<br>Phone: (201) 831-5405 Fax: (201) 831-4405 |
| Date Prepared:                     | September 4, 2009                                                                                                                                                                   |

Description:

The Leibinger Advance Internal Midface Distraction System is a distraction system consisting of the following major components: distractor frame, which incorporates connection screws for the plates, a removable activation rod, plates, and an activation key. The plates and frame initially stabilize and then gradually distract the osteotomy. The removable activation rod, covered with a protective sheath, is connected to the frame and provides the point of attachment for the activation key used to initiate the distraction of the osteotomy.

{1}------------------------------------------------

### Intended Use:

The Leibinger Advance Internal Midface Distraction System is intended to be used in conditions of the cranium and mid-face for which osteotomy and segmental advancement are indicated.

#### Indications:

Treatment of cranial or midfacial conditions for which reconstructive osteotomy and segment advancement are indicated. The indications include conditions such as syndromic craniosynostosis (e. g. Apert, Crouzon, Pfeiffer, Antley Bixler) and midfacial retrusion. The device is intended to provide temporary stabilization and gradual lengthening of facial bones of the cranium and the midface. It is not intended to be used in the mandible.

#### Substantial Equivalence:

The Leibinger Advance Internal Midface Distraction System is substantially equivalent to other commercially available cranial and midface distraction systems. The following devices are examples of predicate systems: Cohen Distractor (K972154), Synthes Midface Distractor (K022005), Vazquez-Diner Intraoral Distractor (K964649) and Molina Orbital Malar Distractor (K003883).

{2}------------------------------------------------

Food and Drug Administration 19905 New Hampshire Avenue Dacumem Control Room -WO68-0666 Silver Spring, MD 20997-9002

Stephanie M. Fitts, Ph.D. Director, Regulatory Affairs and Regulatory Compliance Howmedica Osteonics Corporation Stryker Craniomaxillofacial 325 Corporate Drive Mahwah, New Jersey 07430

JAN 1 5 2010

Re: K092743

Trade/Device Name: The Leibinger Advance Internal Midface Distraction System Regulation Number: 21 CFR 872:4760 Regulation Name: Bone Plate Regulatory Class: II Product Code: JEY Dated: December 29, 2009 Received: December 4, 2009

Dear Dr. Fitts:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Dr. Fitts

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Anthom U. Nwin

Watson, BS, MS, MBA Anthony D. Director Division of Anesthesiology, General Hospital, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K092743 Deficiency Response Page 3 December 29, 2009

## Indications for Use Statement

510(k) Number (if known): K092743

Device Name: The Leibinger Advance Internal Midface Distraction System

Indications for Use:

Treatment of cranial or midfacial conditions for which reconstructive osteotomy and segment advancement are indicated. The indications include conditions such as syndromic craniosynostosis (e. g. Apert, Crouzon, Pfeiffer, Antley Bixler) and midfacial retrusion. The device is intended to provide temporary stabilization and gradual lengthening of facial bones of the cranium and the midface. It is not intended to be used in the mandible.

Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

RSBetz DDS for Dr. K.P. Muliy (Acting)
(Division Sign-Off)

Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K092743

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/JEY/K092743](https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/JEY/K092743)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
