K091807 · Porex Surgical, Inc. · JEY · Jun 28, 2010 · Dental
Device Facts
Record ID
K091807
Device Name
MEDPOR FIXATION SYSTEM
Applicant
Porex Surgical, Inc.
Product Code
JEY · Dental
Decision Date
Jun 28, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4760
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Porex Fixation System for maxillofacial surgery is intended for rigid fixation in the repair of maxillofacial fractures and in maxillofacial reconstruction. The Porex Fixation System for maxillofacial surgery is intended for single use.
Device Story
Porex Fixation System comprises unalloyed titanium plates and titanium alloy bone screws used for rigid fixation in maxillofacial surgery. System includes surgical stainless steel instruments for screw installation. Plates are grade 4 titanium (ASTM F67); screws are 6-4 ELI titanium (ASTM F136). Screws provided in 1.5mm diameter/4mm length; rescue screws are green color-anodized, 1.8mm diameter/4mm length. Device is single-use and sterilized via ethylene oxide. Used by surgeons in clinical settings for fracture repair and reconstruction.
Clinical Evidence
No clinical data.
Technological Characteristics
Materials: Grade 4 titanium plates (ASTM F67), 6-4 ELI titanium screws (ASTM F136), surgical stainless steel instruments. Sterilization: Ethylene oxide. Form factor: Plates and screws for maxillofacial fixation.
Indications for Use
Indicated for patients requiring rigid fixation for maxillofacial fracture repair or maxillofacial reconstruction.
Regulatory Classification
Identification
A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.
K023665 — MODIFICATION TO BIOPLATE RIGID BONE PLATING SYSTEM FOR CRANIOMAXILLOFACIAL SURGERY · Bioplate, Inc. · Nov 22, 2002
K101835 — MEDPOR FIXATION SYSTEM-CRANIAL, MODELS 80080, 80079 · Porex Surgical, Inc. · Dec 3, 2010
K160363 — OPTIMUS CMF SYSTEM · Osteonic Co., Ltd. · Oct 20, 2016
K182758 — MCI - CMF System · Mci Medical Concept Innovation, Inc. · Dec 5, 2019
K112457 — LEFORTE SYSTEM BONE PLATE & SCREW · Arkin Consulting Group · May 18, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
POREX
SURGICAL PRODUCTS GROUP
K091807
JUN 2 8 2010
2
### 510(k) SUMMARY
Date of Summary: June 21, 2010
### Manufacturer and Submitter:
Porex Surgical, Inc. 15 Dart Road Newnan, GA 30265 Tel: (678) 479-1610 Fax: (678) 479-4495
Contact: Stephanie Fullard E-mail: stephanie.fullard@porex.com
Trade Name: Porex Fixation System for Maxillofacial Surgery
Common Name: Plates and Screws
Class: II, 21 CFR 872.4760 - Plate, Fixation, Bone II, 21 CFR 872.4880 - Screw, Fixation, Intraosseous
Product Code: JEY Subsequent Product Code: DZL
### Substantially equivalent to:
- 1) Synthes (USA) Synthes Craniofacial Plates K021642
- 2) Synthes (USA) SMF Ti Alloy Bone Screws K963546
- 3) Stryker Leibenger Universal CMF System K022185
ર - I
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for POREX Surgical Products Group. The logo is in black and white, with the word "POREX" in large, bold letters. Below the word "POREX" is the phrase "SURGICAL PRODUCTS GROUP" in smaller letters. The logo is simple and professional, and it is likely used on the company's products and marketing materials.
### Device Description:
The Porex Fixation System is a plate and screw system to be used in maxillofacial surgery and is provided in various configurations. The components that comprise the systems are composed of unalloyed titanium plates, titanium alloy bone screws and surgical stainless steel instruments for the installation of the screws. The plates are manufactured of grade 4 titanium and adhere to the American Society of Testing Materials (A.S.T.M.) F67 standard. The screws are manufactured of 6-4 ELI titanium that meets the A.S.T.M F136 standard. The screws are provided in 1.5mm diameter and 4mm length. The rescue screws are green color annodized and provided in 1.8mm diameter and 4mm length. The Porex Fixation Systems are sterilized by ethylene oxide sterilization.
### Indications for Use:
The Porex Fixation System for maxillofacial surgery is intended for rigid fixation in the repair of maxillofacial fractures and in maxillofacial reconstruction. The Porex Fixation System for maxillofacial surgery is intended for single use.
### Technological Characteristics:
The Porex Fixation System has the same intended use and is manufactured from substantially the same materials as the predicate devices. The technological characteristics of the design are substantially equivalent to the predicate devices and any slight differences raise no new issues of safety and effectiveness.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of two main elements: a circular seal on the left and a stylized symbol on the right. The circular seal contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. The symbol on the right is a stylized representation of a human figure, with three overlapping lines forming the body and head.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Stephanie Fullard, RAC Regulatory Affairs Manager Porex Surgical, Incorporated 15 Dart Road Newman, Georgia 30265
JUN 2 8 2010
Re: K091807
Trade/Device Name: Porex Fixation System For Maxillofacial Surgery Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: II Product Code: JEY Dated: June 21, 2010 Received: June 23, 2010
Dear Ms. Fullard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
### Page 2- Ms. Fullard
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Nfor
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
### Enclosure
{4}------------------------------------------------
K091807
# Indications for Use
510(k) Number (if known): K091807
Device Name: Porex Fixation System For Maxillofacial Surgery
Indications for Use: The Porex Fixation System for maxillofacial surgery is intended for rigid fixation in the repair of maxillofacial fractures and in maxillofacial reconstruction. The Porex Fixation System For Maxillofacial Surgery is intended for single use.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
(Division Sign-Off) Division of Anesthesiology, General HospitalPage 1 of 1 Infection Control, Dental Devices
510(k) Number: Ko 9 1807
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.