OSTEOMEDICS RESORBABLE SMALL FIXATION SYSTEM, OSTEOSORB
K022035 · Osteomedics, Inc. · JEY · Jan 17, 2003 · Dental
Device Facts
Record ID
K022035
Device Name
OSTEOMEDICS RESORBABLE SMALL FIXATION SYSTEM, OSTEOSORB
Applicant
Osteomedics, Inc.
Product Code
JEY · Dental
Decision Date
Jan 17, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4760
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Intended Use
The Osteomedics® Resorbable Small Fixation System is intended for use in selective trauma or for reconstructive procedures in the midface and craniofacial skeleton; craniofacial surgery; reconstructive procedures, and selective orthognathic surgery of the maxilla and chin. The Osteomedics® Resorbable Small Fixation System aids in healing in conjunction with appropriate postoperative immobilization. The use is in both adult and pediatric patients' cases involving trauma or reconstruction.
Device Story
Osteomedics® Resorbable Small Fixation System (OsteoSorb®) comprises resorbable plates, screws, bone fasteners, and meshes; manufactured from poly (L-lactide-co-D, L-lactide) 70:30. Device used by surgeons in clinical settings for stabilization/fixation of craniofacial/midface bone fractures or osteotomies. System provides mechanical support during bone healing; material resorbs over time. Surgeon selects appropriate plate configuration/length; attaches to bone using screws/fasteners. Benefits include elimination of need for secondary removal surgery associated with permanent metal implants.
Clinical Evidence
No clinical data provided; substantial equivalence based on design, material, and operational similarities to legally marketed predicate devices.
Technological Characteristics
Material: poly (L-lactide-co-D, L-lactide) 70:30. Form factor: plates, screws, bone fasteners, and meshes of varying configurations and lengths. Principle: mechanical fixation of bone segments. Resorbable polymer composition.
Indications for Use
Indicated for adult and pediatric patients requiring stabilization and fixation of small bones of the craniofacial skeleton due to trauma, reconstructive procedures, or orthognathic surgery of the maxilla and chin.
Regulatory Classification
Identification
A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.
Predicate Devices
Bionx® BiosorbFx® 1.5/2.0 Bioabsorbable Fixation System
BiosorbFx® 2.0/2.4 Orthognathic / Mandibular Fix
Stryker Leibinger Wurzburg Titanium Mini Bone Plates and Bone Screws
K974554 — SYNTHES RESORBABLE FIXATION SYSTEM · Synthes (Usa) · Feb 24, 1998
K070737 — XSORB BIOABSORBABLE CRANIOFACIAL BONE FIXATION SYSTEM & ACCESSORIES · Prosurg, Inc. · Jul 25, 2007
K251441 — Inion CPS 1.5 Baby Bioabsorbable Fixation System; Inion CPS 1.5/2.0/2.5 Bioabsorbable Fixation System · Inion OY · Nov 7, 2025
K992905 — CODMAN CRANIOSORB ABSORBABLE FIXATION SYSTEM · Codman & Shurtleff, Inc. · Aug 1, 2000
Submission Summary (Full Text)
{0}------------------------------------------------
## JAN 1 7 2003
K022035
## 9. 510 (k) Summary
Device Name: Osteomedics® Resorbable Small Fixation System, OsteoSorb® Classification Name: Bone Plate, Bone Screw Device Regulation Number: 76JEY (21 CFR 872.4760, 21 CFR 872.4880), Bone Plate:
The Osteomedics® Resorbable Small Fixation System, OsteoSorb® consists of a series of plates, screws, bone fasteners and meshes, which are made from poly (L-lactide-co-D, Llactide) 70: 30. The plates are designed in varying configurations and lengths, which are attached to the bone using screw or bone fastener fixation.
The Osteomedics® Resorbable Small Fixation System, OsteoSorb® is intended for use in selective trauma or for reconstructive procedures in the midface and craniofacial skeleton; craniofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin. The use is in both adult and pediatric patients cases involving trauma or reconstruction.
The substantial equivalence of device is based on similarities in intended use, design and operational principals to the Bionx® BiosorbFx® 1.5/2.0 Bioabsorbable Fixation System and the BiosorbFx® 2.0/2.4 Orthognathic / Mandibular Fix and Stryker Leibinger Wurzburg Titanium Mini Bone Plates and Bone Screws The material used in the manufacture of the Osteomedics® Resorbable Small Fixation System is poly (L-lactide-co-D, L-lactide) and is substantially equivalent to that used in the Bionx® Bioabsorbable Fixation System. The material used for Stryker Leibinger Wurzburg Titanium Mini Bone Plates and Bone Screws is Commercial Pure Titanium.
The basic operational principals, site preparation, relative indications and contraindications are similar for Osteomedics® Resorbable Small Fixation System and the, Bionx® Bioabsorbable Fixation System and Stryker Leibinger Wurzburg Titanium Mini Bone Plates and Bone Screws. This will provide stabilization and fixation of small bones of the craniofacial skeleton due to fracture or osteotomy.
## Official Contact Person:
Albert Enavati President Osteomedics Inc. 809 Carter Lane Paramus. NJ 07652 Tel: (201) 444-7306 Fax: (201) 444-7395 E-mail: ostcomedic a aol.com
6/18/2002
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around it. Inside the circle is an abstract symbol resembling a bird or a stylized human figure, composed of curved lines.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 1 7 2003
Mr. Albert Enayati President Osteomedics, Incorporated 809 Carter Lane Paramus, New Jersey 07652
Re: K022035
Trade/Device Name: Osteomedics® Resorbable Small Fixation System, OsteoSorb® Regulation Number: 872.4760 Regulation Name: Bone Plate Regulatory Class: II Product Code: JEY Dated: October 28, 2002 Received: November 1, 2002
Dear Mr. Enayati:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
## Page 2 -- Ms. Enayati
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
L. Ulatowski
Timothy A. Ulatowski Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{3}------------------------------------------------
Indications for use
510 (k) Number ( if known):___________________________________________________________________________________________________________________________________________________
Device Name: Osteomedics® Resorbable Small Fixation System, OsteoSorb®
Indications for use:
The Osteomedics® Resorbable Small Fixation System is intended for use in selective trauma or for The ostructive procedures in the midface and craniofacial skeleton; craniofacial surgery; reconstructive recombines, and selective or hognathic surgery of the maxilla and chin. The Osteomedics® Resorbable procedures, and selectric or ling healing in conjunction with appropriate postoperative immobilization. The use is in both adult and pediatric patients' cases involving trauma or reconstruction.
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
OR
CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)
Prescription use V
OVER - THE - COUNTER USE_
(Per 21 CFR 801.109)
(Optional Format 1-2-96)
Susan Russe
on of Anesthesiology, General Hospital, Infection Control, Dental De
510(k) Number: K000035
6/18/2002
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.