← Product Code [JEY](/submissions/DE/subpart-e%E2%80%94surgical-devices/JEY) · K013039

# INION CPS 2.0 ORTHOGNATHIC PLATE (K013039)

_Inion , Ltd. · JEY · Oct 5, 2001 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/JEY/K013039

## Device Facts

- **Applicant:** Inion , Ltd.
- **Product Code:** [JEY](/submissions/DE/subpart-e%E2%80%94surgical-devices/JEY.md)
- **Decision Date:** Oct 5, 2001
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 872.4760
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The Inion CPS" Orthognathic Plate is intended for use in trauma and reconstructive procedures in the craniofacial skeleton, midface, maxilla and mandible (in conjunction with appropriate maxillomandibular fixation) as a part of the Inion CPS™ 1.5/2.0/2.5 Bioabsorbable Fixation System. a) Fractures of the cranium, midface, maxilla and mandible. b) Infant craniofacial surgery (i.e. craniosynostosis, congenital malformations. c) LeFort (I, II, III) osteotomies. d) Pediatric reconstructive procedures. e) Orthognathic or reconstructive procedures of the cranium, midface, maxilla or mandible. f) Craniotomy flap fixation. The Inion CPS™ 2.0 Orthognathic Plate is not intended for use in and is contraindicated for: Mandibular tumor resection; Active or potential infection; Patient conditions including limited blood supply, insufficient quantity or quality of bone; and where patient cooperation cannot be guaranteed (e.g., alcoholism, drug abuse). The system is not intended for use in the mandible without appropriate maxillomandibular fixation.

## Device Story

Resorbable bone plating system for craniofacial, midface, maxilla, and mandible fixation; used in trauma and reconstructive surgery. Device consists of plates made of polylactic acid/trimethylenecarbonate copolymer. Operates as mechanical fixation hardware; implanted by surgeons during procedures. Provides structural stabilization of bone segments; facilitates healing in fractures, osteotomies, and congenital malformation corrections. Line extension of existing bioabsorbable fixation system.

## Clinical Evidence

No clinical data provided; substantial equivalence based on technological characteristics and design similarity to predicate.

## Technological Characteristics

Resorbable bone plate; material: polylactic acid/trimethylenecarbonate copolymer. Form factor: various shapes for craniofacial/maxillofacial application. Sterilization: same as predicate.

## Regulatory Identification

A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.

## Predicate Devices

- Inion CPS™ 1.5/2.0/2.5 Bioabsorbable Fixation System ([K010352](/device/K010352.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K013039

E

# September/7/2001

# Special 510(k) Summary for the Inion CPS™ 2.0 Orthognathic Plate

# ADMINISTRATIVE INFORMATION

Manufacturer´s Name:

Inion Ltd. Lääkärinkatu 2 FIN-33520 Tampere

Contact Person: Hanna Marttila Regulatory Affairs Coordinator Lääkärinkatu 2 FIN-33520 Tampere Phone: +358 3 230 6600 +358 3 230 6601 Fax: -

### DEVICE NAME

Classification Name: Common/Usual Name: Trade Name:

Bone Plate Bone Plating System Inion CPSTM 2.0 Orthognathic Plate

## ESTABLISHMENT REGISTRATION NUMBER

Inion Ltd. has not yet obtained an Establishment Registration Number.

#### DEVICE CLASSIFICATION AND PRODUCT CODE

As shown in 21 CFR 872.4760 bone plates are classified as Class II. Bone Plates have been assigned Product Code 76 JEY.

#### PREDICATE DEVICES

(1) Inion CPS™ 1.5/2.0/2.5 Bioabsorbable Fixation System (K010352)

{1}------------------------------------------------

### INTENDED USE

The Inion CPS" Orthognathic Plate is intended for use in trauma and reconstructive procedures in the craniofacial skeleton, midface, maxilla and mandible (in conjunction with appropriate maxillomandibular fixation) as a part of the Inion CPS™ 1.5/2.0/2.5 Bioabsorbable Fixation System.

a) Fractures of the cranium, midface, maxilla and mandible.

b) Infant craniofacial surgery (i.e. craniosynostosis, congenital malformations.

c) LeFort (I, II, III) osteotomies.

d) Pediatric reconstructive procedures.

e) Orthognathic or reconstructive procedures of the cranium, midface, maxilla or mandible.

f) Craniotomy flap fixation.

The Inion CPS™ 2.0 Orthognathic Plate is not intended for use in and is contraindicated for: Mandibular tumor resection; Active or potential infection; Patient conditions including limited blood supply, insufficient quantity or quality of bone; and where patient cooperation cannot be guaranteed (e.g., alcoholism, drug abuse). The system is not intended for use in the mandible without appropriate maxillomandibular fixation.

#### DEVICE DESCRIPTION

The Inion CPS" 2.0 Orthognathic Plates are provided in different shapes similar to predicate device. The Inion CPS™ 2.0 Orthognathic Plate is made of resorbable polylactic acid/trimethylenecarbonate copolymer.

# EQUIVALENCE TO MARKETED PRODUCTS

The Inion CPS™ Orthognathic Plate is a line extension to currently marketed Inion CPS™ 1.5/2.0/2.5 Bioabsorbable Fixation System. Inion CPS™ 2.0 Orthognathic Plate has the same technological characteristics as the Inion CPS™ System identified above. Both will be offered in the same materials and with the same packaging and sterility options. Both the plates have the same intended use and principles of operation and there is no change in safety or efficacy.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three curved lines representing its body and wings. The eagle's head is facing left. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Public Health Service

OCT - 5 2001

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Hanna Marttila Regulatory Affairs Inion Limited Laakarinkatu 2 Tampere, FINLAND

Re: K013039

Trade/Device Name: Inion CPS 2.0 Orthognathic Plate Regulation Number: 872.4760 Regulation Name: Bone Plating System Regulatory Class: II Product Code: JEY Dated: September 7, 2001 Received: September 10, 2001

Dear Ms. Marttila:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements 

{3}------------------------------------------------

Page 2 - Ms. Marttila

of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613 . Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Susan Kunner

Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### STATEMENT OF INDICATIONS FOR USE D

# Applicant: Inion Ltd. 510(k) Number: Device Name: Inion CPS ™ 2.0 Orthognathic Plate

Indications For Use:

### Indications:

- A. General indications: The Inion CPS™ 2.0 Orthognathic Plate is intended for use in trauma and reconstructive procedures in the craniofacial skeleton, midface, maxilla and mandible (in conjunction with appropriate maxillomandibular fixation) as a part of the Inion CPS™ 1.5/2.0/2.5 Bioabsorbable Fixation System.
- B. Specific indications:
	- Fractures of the cranium, midface, maxilla and mandible. ●
	- Infant craniofacial surgery (i.e. craniosynostosis, congenital ● malformations).
	- . LeFort (I, II, III) osteotomies.
	- Pediatric reconstructive procedures. .
	- Orthognathic or reconstructive procedures of the cranium, midface, . maxilla or mandible.
	- Craniotomy flap fixation. ●

#### Contraindications:

The Inion CPS™ 2.0 Orthognathic Plate is not intended for use in and is contraindicated for:

- 1. Mandibular tumor resection
- 2. Active or potential infection
- 3. Patient conditions including limited blood supply, insufficient quantity or quality of bone; and where patient cooperation cannot be guaranteed (e.g., alcoholism, drug abuse)
- 4. DO NOT USE in the mandible without appropriate maxillomandibular fixation.

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Per 21 CFR 801.109)

(Optional Format 1-2-96)

Susan Runne

(Division Sian-Off Division of Dental, Infection Co and General Hospita 510(k) Number

Special 510(k)

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/JEY/K013039](https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/JEY/K013039)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
