← Product Code [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX) · K234085

# LUNE PureHygiene (K234085)

_Enamel Pure · GEX · Mar 25, 2024 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/GEX/K234085

## Device Facts

- **Applicant:** Enamel Pure
- **Product Code:** [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX.md)
- **Decision Date:** Mar 25, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

Enamel Pure LUNE PureHygiene is indicated for the following in dental hard tissue: • Reduction of bacterial level (decontamination)

## Device Story

Mobile, cart-based dental laser system; comprises base console, pneumatic footswitch, and handpieces. Uses pulsed CO2 laser energy to reduce/remove plaque and embedded carbonate from dental enamel. Operated by dental professionals in clinical settings. Laser energy delivery via articulating arm and handpiece. Healthcare providers use the device to perform decontamination of dental hard tissue; benefits include bacterial reduction. Handpieces constructed from medical-grade Radel.

## Clinical Evidence

No clinical data. Bench testing only, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), laser safety (IEC 60601-2-22 & IEC 60825-1), software verification/validation, design verification (tooth temperature, bacterial reduction, user validation), reprocessing validation, and biocompatibility (ISO 10993-1).

## Technological Characteristics

Powered laser surgical instrument (Class II, GEX). Laser source: CO2 (treatment), Diode (aiming). Class IV treatment laser. Delivery: Articulating arm and handpiece (medical-grade Radel). Power: 100V-240V, 50/60 Hz. Features: Main power switch, remote interlock, emergency stop, footswitch activation. Sterilization: Steam autoclave (handpiece).

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Epic Pro/Hygiene Biolase ([K193486](/device/K193486.md))

## Reference Devices

- Solea Convergent Dental ([K221761](/device/K221761.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 25, 2024

Enamel Pure % Dhaval Saraiya Regulatory Affairs and Quality Assurance Consultant Omnee Strategic Solutions, Inc. 7 Desrosiers Landing South Grafton, Massachusetts 01560

Re: K234085

Trade/Device Name: LUNETM PureHygiene Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: December 22, 2023 Received: December 26, 2023

Dear Dhaval Saraiya:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device. please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

# Michael E. Adjodha -S

Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K234085

Device Name

LUNE™ PureHygiene

Indications for Use (Describe)

| Enamel Pure LUNE PureHygiene is indicated for the following in dental hard tissue: |
|------------------------------------------------------------------------------------|
| • Reduction of bacterial level (decontamination)                                   |

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Enamel Pure. The logo consists of a diamond shape that is split into two triangles, with the top triangle being green and the bottom triangle being blue. A curved line wraps around the diamond. To the right of the diamond is the word "ENAMELPure" in a serif font, with "ENAMEL" in green and "Pure" in blue. Underneath the word "ENAMELPure" is a curved blue line.

## K234085 - 510(k) Summary

In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the LUNE PureHygiene 510(k) premarket notification.

| Sponsor:             | Enamel Pure Inc.<br>Nathan Monty<br>17 Briden Street<br>Worcester, MA 01605                                                                                                                                                                        |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:      | Dhaval S.<br>Omnee Strategic Solutions, Inc.<br>Regulatory/Quality Consultant<br>Email: omneestrategicsol@gmail.com                                                                                                                                |
| Date:                | Mar 21, 2024                                                                                                                                                                                                                                       |
| Subject Device:      | Trade Name: LUNE™ PureHygiene<br>Common Name: Dental Laser System<br>Classification Name: Laser surgical instrument for<br>use in general and plastic surgery and in<br>dermatology - (21 CFR 878.4810)<br>GEX – Powered Laser Surgical Instrument |
| Predicate Device(s): | Primary Predicate: K193486 Epic Pro/Hygiene Biolase<br>Reference Device: K221761 Solea Convergent Dental                                                                                                                                           |

#### Purpose and Device Description:

The purpose of this submission is to request clearance for the new LUNE PureHygiene. The subject LUNE PureHygiene is a mobile, cart-based dental treatment system comprised of a base console, a pneumatic footswitch, and handpieces that uses a pulsed CO2 laser energy to reduce/remove plaque and embedded carbonate from dental enamel.

### Intended Use and Indications for Use:

Enamel Pure LUNE PureHygiene is indicated for the following in dental hard tissue:

- . Reduction of bacterial level (decontamination)
#### Summary of Technological Characteristics:

The rationale for substantial equivalence is based on consideration of the following characteristics:

- Intended Use: The intended use is similar to the intended use cleared in K193486. ●

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Enamel Pure. The logo features a green and blue diamond shape on the left, with the text "EnamelPure" in green next to it. A blue wave-like line is underneath the text.

- Indications for Use: The indications for use are similar to the indications for use cleared in K193486. ●
- Design Features: The design features are similar to those in currently marketed devices cleared in K193486 and K221761.

#### Summary of Performance Data (Nonclinical and/or Clinical):

#### Non-Clinical Tests: ●

- Electrical Safety Testing per IEC 60601-1 o
- O EMC Testing per IEC 60601-1-2
- Laser Safety Testing per IEC 60601-2-22 & IEC 60825-1 o
- O Software Verification and Validation
- Design Verification tooth temperature, bacterial reduction, and user validation O
- Reprocessing Validation (cleaning and sterilization) per the FDA Reprocessing Guidance o
- Biocompatibility Assessment per ISO 10993-1 and the FDA Biocompatibility Guidance o

#### Clinical Tests: ●

- o N/A
#### Substantial Equivalence

The LUNE PureHygiene has been shown to be substantially equivalent to the predicate device. Results of the non-clinical tests indicate that the device will perform within the intended uses.

| Property or<br>Characteristic | Proposed Device<br>LUNE PureHygiene                                                                                                             | Predicate Device<br>Epic Pro/Hygiene<br>(K193486)                                                                                                                                                        | Reference Device<br>Solea<br>(K221761)                                                                                                                                                                                                                                                                                                                                                         |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FDA Product<br>Code(s)        | GEX                                                                                                                                             | GEX                                                                                                                                                                                                      | GEX                                                                                                                                                                                                                                                                                                                                                                                            |
| Regulation<br>Number          | 21 CFR 878.4810                                                                                                                                 | 21 CFR 878.4810                                                                                                                                                                                          | 21 CFR 878.4810                                                                                                                                                                                                                                                                                                                                                                                |
| Classification<br>Name        | Powered Laser Surgical Instrument                                                                                                               | Powered Laser Surgical<br>Instrument                                                                                                                                                                     | Powered Laser Surgical Instrument                                                                                                                                                                                                                                                                                                                                                              |
| Indications for<br>Use        | Enamel Pure LUNE PureHygiene<br>is indicated for the following in<br>dental hard tissue:<br>● Reduction of bacterial level<br>(decontamination) | Biolase Epic Pro/Hygiene is<br>indicated for the following:<br>● Reduction of bacterial level<br>(decontamination) and<br>inflammation<br>Please refer to K193486 for a<br>complete list of indications. | The Solea system is indicated for<br>the following:<br>● Aiding in the reduction of<br>mineral loss in dental enamel<br>● Ablation of hard tissue for<br>caries removal and cavity<br>prevention<br>● Incision, excision, vaporization,<br>coagulation and hemostasis of<br>soft tissue in the oral cavity<br>● Cutting, shaving, contouring<br>and resection of oral osseous<br>tissue (both) |
| Operating<br>Voltage          | 100V-240V                                                                                                                                       | 100V-240V                                                                                                                                                                                                | 100V-240V                                                                                                                                                                                                                                                                                                                                                                                      |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Enamel Pure. The logo features a green and blue diamond shape with a curved line around it on the left side. The text "ENAMELPURE" is written in green and blue on the right side of the diamond.

| Property or<br>Characteristic | Proposed Device<br>LUNE PureHygiene      | Predicate Device<br>Epic Pro/Hygiene<br>(K193486)           | Reference Device<br>Solea<br>(K221761)      |
|-------------------------------|------------------------------------------|-------------------------------------------------------------|---------------------------------------------|
| Frequency                     | 50/60 Hz                                 | 50/60 Hz                                                    | 50/60 Hz                                    |
| Main Control                  | Main Power Switch                        | Main Power Switch                                           | Main Power Switch                           |
| Remote<br>Interruption        | Remote Interlock - Rear                  | Remote Interlock - Rear                                     | Remote Interlock - Rear                     |
| Disable Control               | Emergency Stop Button                    | Emergency Stop Button                                       | Emergency Stop Button                       |
| Laser Source                  | Treatment - CO2<br>Aiming - Diode laser  | Treatment - Semi-conductor<br>diode<br>Aiming - Diode laser | Treatment - CO2<br>Aiming - Diode laser     |
| Laser<br>Activation           | Footswitch                               | Footswitch                                                  | Footswitch                                  |
| Laser<br>Classification       | Treatment – Class IV<br>Aiming - Class 2 | Treatment – Class IV<br>Aiming - Class 2                    | Treatment – Class IV<br>Aiming - Class 2/3R |
| Delivery System               | Articulating Arm and handpiece           | Fiber optic cable, handpiece and<br>disposable tips         | Articulating Arm and handpiece              |
| Handpiece                     | Medical grade Radel                      | Stainless Steel                                             | Stainless Steel                             |
| Sterilization<br>Method       | Steam Autoclave (Handpiece only)         | Steam Autoclave (Handpiece<br>only)                         | Steam Autoclave (Handpiece only)            |

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/GEX/K234085](https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/GEX/K234085)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
