← Product Code [ELC](/submissions/DE/subpart-e%E2%80%94surgical-devices/ELC) · K203302

# GentleWave X (K203302)

_Sonendo, Inc. · ELC · Feb 5, 2021 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/ELC/K203302

## Device Facts

- **Applicant:** Sonendo, Inc.
- **Product Code:** [ELC](/submissions/DE/subpart-e%E2%80%94surgical-devices/ELC.md)
- **Decision Date:** Feb 5, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.4850
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

The Sonendo GentleWave X System is intended to prepare, clean, and irrigate teeth indicated for root canal therapy. When used with the Sonendo GentleWave Molar Procedure Instrument or Molar CrossFire™ Procedure Instrument, the System is indicated for 1st and 2nd molar teeth. When used with the Sonendo Gentle Wave Anterior/Premolar Procedure Instrument, the System is indicated for anterior and premolar teeth. When used with the Sonendo Gentle Wave Premolar CrossFire™ Procedure Instrument, the system is indicated for premolar teeth.

## Device Story

System prepares, cleans, and irrigates root canals using hydroacoustic waves and fluid motion generated by sonic energy. Comprised of console, foot pedal, and single-use sterile handpieces (molar, anterior/premolar, CrossFire variants). Delivers 3% Sodium Hypochlorite (NaOCl) and 8% EDTA solutions into pulp chamber. Used by dental professionals in clinical settings. Console manages fluid delivery; larger reservoirs in X model increase treatment cycle capacity. Output is cleaned root canal system. Benefits include efficient cleaning of complex root canal anatomy. Device operation is identical to predicate; performance verified via bench testing.

## Clinical Evidence

No clinical data. Substantial equivalence demonstrated via bench testing, including thermal, cavitation, hydroacoustic, cleaning, apical pressure, and apical extrusion tests, alongside human factors evaluation, biocompatibility, and electrical/EMC safety testing.

## Technological Characteristics

Ultrasonic scaler (Class II, ELC). Materials: medical grade. Energy: sonic energy for hydroacoustic wave generation. Connectivity: standalone console. Sterilization: Gamma (handpieces). Software: moderate level of concern. Standards: ISO 10993-1/5/10/11, IEC 60601-1, IEC 60601-1-2, IEC 80601-2-60, IEC 62366-1.

## Regulatory Identification

An ultrasonic scaler is a device intended for use during dental cleaning and periodontal (gum) therapy to remove calculus deposits from teeth by application of an ultrasonic vibrating scaler tip to the teeth.

## Predicate Devices

- Sonendo GentleWave System ([K160905](/device/K160905.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 5, 2021

Sonendo, Inc Lyle VerPlanck Director of Regulatory Affairs & Quality Assurance 26061 Merit Circle, Suite 102 Laguna Hills, California 92653

Re: K203302

Trade/Device Name: GentleWave X Regulation Number: 21 CFR 872.4850 Regulation Name: Ultrasonic Scaler Regulatory Class: Class II Product Code: ELC Dated: November 6, 2020 Received: November 9, 2020

### Dear Lyle VerPlanck:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Confidential

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known)

Device Name GentleWave X

Indications for Use (Describe)

The Sonendo Gentle Wave X System is intended to prepare, clean, and irrigate teeth indicated for root canal therapy. When used with the Sonendo GentleWave Molar Procedure Instrument or Molar CrossFire™ Procedure Instrument, the System is indicated for 1st and 2nd molar teeth. When used with the Sonendo Gentle Wave Anterior/Premolar Procedure Instrument, the System is indicated for anterior and premolar teeth. When used with the Sonendo Gentle Wave Premolar CrossFire™ Procedure Instrument, the system is indicated for premolar teeth.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

## 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

| General Information                   |                                                                                                                                                        |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number:                        | K203302                                                                                                                                                |
| Date Summary Prepared:                | February 3, 2021                                                                                                                                       |
| Submitted by:                         | Sonendo, Inc.<br>22061 Merit Circle, Suite 102<br>Laguna Hills, CA 92653                                                                               |
| Establishment Registration<br>Number: | 3010817521                                                                                                                                             |
| Contact Person:                       | Lyle VerPlanck<br>Director of Regulatory Affairs & Quality Assurance<br>Phone: (949) 667-7260<br>FAX: (949) 305-5201<br>E-mail: lverplanck@sonendo.com |
| Reason for Premarket<br>Notification: | New model of GentleWave System                                                                                                                         |
| Trade Name:                           | GentleWave X                                                                                                                                           |
| Regulation Name:                      | Ultrasonic Scaler                                                                                                                                      |
| Review Panel  :                       | Dental                                                                                                                                                 |
| Classification Regulation:            | 872.4850                                                                                                                                               |
| Product Code:                         | ELC                                                                                                                                                    |
| Regulatory Class:                     | Class II                                                                                                                                               |
| Predicate Device:                     | Sonendo GentleWave System via K160905                                                                                                                  |

{4}------------------------------------------------

# Device Description

The Sonendo GentleWave X System is a medical device intended to prepare, clean and irrigate root canals. The Sonendo GentleWave X System is comprised of a Console, and disposable single-use Handpieces and accessories. The Handpiece is offered for molar, anterior and premolar treatments. A Molar Handpiece (original or CrossFire™ model) is intended to be used on 1st and 2nd molar teeth. An Anterior/Premolar (APM) Handpiece is intended to be used on anterior and pre-molar teeth and CrossFire™ Premolar Handpiece is intended for pre-molar teeth only.

The Sonendo GentleWave X System delivers a stream of treatment fluids into the pulp chamber of the tooth. The treatment fluids delivered to the tooth are 3% Sodium Hypochlorite (NaOCl) solution and an 8% Ethylenediaminetetraacetic acid (EDTA) utilized in traditional endodontic treatments.

The design of this device incorporated the consensus standards listed in Table 1.

| Recognition<br>Number | Standard<br>Designation<br>Number | Standard Title                                                                                                   |
|-----------------------|-----------------------------------|------------------------------------------------------------------------------------------------------------------|
| 2-258                 | ISO 10993-1                       | Biological evaluation of medical devices - Part 1:<br>Evaluation and testing within a risk management<br>process |
| 2-245                 | ISO 10993-5                       | Biological evaluation of medical devices - Part 5: Tests<br>for in vitro cytotoxicity                            |
| 2-174                 | ISO 10993-10                      | Biological evaluation of medical devices - Part 10: Tests<br>for irritation and skin sensitization               |
| 2-255                 | ISO 10993-11                      | Biological evaluation of medical devices - Part 11: Tests<br>for systemic toxicity                               |
| 19-4                  | IEC 60601-1                       | Medical electrical equipment - Part 1: General<br>requirements for basic safety and essential performance        |
| 19-8                  | IEC 60601-1-2                     | Medical electrical equipment - Part 1-2: General<br>requirements for basic safety and essential perform          |
| 4-262                 | IEC 80601-2-60                    | Particular requirements for the basic safety and essential<br>performance of dental equipment                    |
| 5-114                 | IEC 62366-1                       | Medical devices - Part 1: Application of usability<br>engineering to medical devices                             |

Table 1 - Recognized Consensus Standards

{5}------------------------------------------------

# Indications for Use

The Sonendo GentleWave X System is intended to prepare, clean, and irrigate teeth indicated for root canal therapy. When used with the Sonendo GentleWave Molar Procedure Instrument orMolar CrossFire™ Procedure Instrument, the System is indicated for 1st and 2nd molar teeth. When used with the Sonendo GentleWave Anterior/Premolar Procedure Instrument, the System is indicated for anterior and premolar teeth. When used with the Sonendo GentleWave Premolar CrossFire™ Procedure Instrument, the system is indicated for premolar teeth.

## Substantial Equivalence Summary and Discussion

The table below includes a comparison between the new device and the predicate device:

| Characteristic                       | Predicate Device                                                                                                                                                                                                                                                                                                                                                                        | Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Clearance #                   | K160905                                                                                                                                                                                                                                                                                                                                                                                 | TBD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Console Model                        | GentleWave®                                                                                                                                                                                                                                                                                                                                                                             | GentleWave™ X                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Indications for Use                  | The Sonendo GentleWave System<br>is intended to prepare, clean, and<br>irrigate teeth indicated for root<br>canal therapy. When used with<br>the Sonendo GentleWave Molar<br>Handpiece, the System is indicated<br>for 1st and 2nd molar teeth.<br>When used with the Sonendo<br>GentleWave Anterior/Premolar<br>Handpiece, the System is indicated<br>for anterior and premolar teeth. | The Sonendo GentleWave X System<br>is intended to prepare, clean, and<br>irrigate teeth indicated for root canal<br>therapy. When used with the<br>Sonendo GentleWave Molar<br>Procedure Instrument or Molar<br>CrossFire™ Procedure Instrument,<br>the System is indicated for 1st and<br>2nd molar teeth. When used with<br>the Sonendo GentleWave<br>Anterior/Premolar Procedure<br>Instrument, the System is indicated<br>for anterior and premolar teeth.<br>When used with the Sonendo<br>GentleWave Premolar CrossFire™<br>Procedure Instrument, the system is<br>indicated for premolar teeth. |
| New Contraindications                | ---                                                                                                                                                                                                                                                                                                                                                                                     | Primary dentition<br>Patients with sensitivity to NaOCl<br>and / or EDTA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Function                             | Preparation, cleaning and irrigation<br>of root canal                                                                                                                                                                                                                                                                                                                                   | Identical as predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Characteristic                       | Predicate Device                                                                                                                                                                                                                                                                                                                                                                        | Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Principal of Operation               | Generation of hydroacoustic waves<br>and fluid motion via application of<br>sonic energy                                                                                                                                                                                                                                                                                                | Identical as predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Treatment Site                       | Root Canal                                                                                                                                                                                                                                                                                                                                                                              | Identical as predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Components                           | Control Unit<br>Irrigation reservoirs<br>Foot pedal<br>Handpiece<br>Hand piece Accessories                                                                                                                                                                                                                                                                                              | Identical as predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Duration of treatment<br>application | 7 minutes or less                                                                                                                                                                                                                                                                                                                                                                       | Identical as predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Treatment fluid<br>Concentration     | Fixed                                                                                                                                                                                                                                                                                                                                                                                   | Identical as predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Fluid Containers                     | 1 × 500 mL NaOCl irrigation<br>solution<br>1 × 250 mL EDTA irrigation<br>solution<br>1 × 2000 mL distilled water for<br>irrigation                                                                                                                                                                                                                                                      | 1 × 1360 mL NaOCl irrigation<br>solution<br>1 × 540 mL EDTA irrigation<br>solution<br>1 × 2650 mL distilled water for<br>irrigation                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Material                             | Medical grade material                                                                                                                                                                                                                                                                                                                                                                  | Identical as predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Biocompatibility                     | Biocompatible per 10993                                                                                                                                                                                                                                                                                                                                                                 | Identical as predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Handpiece models                     | Molar Handpiece (original model),<br>Anterior/Premolar Handpiece                                                                                                                                                                                                                                                                                                                        | Molar Handpiece (original or<br>CrossFire™ model ),<br>Anterior/Premolar Handpiece,<br>Premolar CrossFire™ model<br>Handpiece                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Handpiece Sterilization              | Sterile (Gamma)<br>Single use<br>Sterility Assurance Level (SAL) 10-6                                                                                                                                                                                                                                                                                                                   | Identical as predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Consumable Shelf Life                | 1 year                                                                                                                                                                                                                                                                                                                                                                                  | Identical as predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |

Table 1 - Comparison of Predicate and Proposed Devices

{6}------------------------------------------------

The proposed device has the same intended use, function, principal of operation, control mechanism, sterilization and shelf life as the predicate device. The indications for use is slightly changed to add new models of the handpiece within the scope of the predicate device indications 

{7}------------------------------------------------

for use. To highlight that the use of GentleWave X system is limited to treat permanent teeth only, contraindication for primary dentition teeth is added. Furthermore, to minimize any potential adverse reaction, the device is contraindicated for use on patients with a known sensitivity to NaOCl and/ or EDTA. The new handpiece models are added to increase manufacturability and there is no change in the handpiece operating principal. The new Console model has increased capacity, with larger bottle sizes to facilitate more treatments cycles. The new model is designed to improve customer interface, product reliability and manufacturability. Performance testing is completed to show the equivalency between the proposed and predicate devices.

## Non-Clinical Performance Testing

All necessary performance testing has been conducted on the Sonendo GentleWave System X to assure substantial equivalence to the predicate devices and to demonstrate the device performs as intended. The following bench testing in accordance with standards listed in Table 1 was performed and passed to support the substantial equivalence of the subject device:

- Performance Verification
	- Thermal test O
	- Cavitation test O
	- Hydroacoustic test O
	- Cleaning test O
	- Apical pressure test o
	- Apical extrusion test o
- Human Factor Evaluation per the FDA Guidance Document "Applying Human Factors and Usability ● Engineering to Medical Devices"
- Software Validation for moderate level of concern per the FDA "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"
- . Biocompatibility
	- Cytotoxicity, sensitization, irritation, material mediated pyrogenicity, and acute systemic 0 toxicity
- Electromagnetic Compliance and Electrical Safety

## Clinical Data

No clinical data is submitted in support of this Premarket Notification.

## Conclusion

The Sonendo GentleWave X System is a new model of the existing Sonendo GentleWave® System. The Sonendo GentleWave X System is substantially equivalent to the Sonendo GentleWave System cleared under premarket notification K160905.

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/ELC/K203302](https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/ELC/K203302)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
