← Product Code [ELC](/submissions/DE/subpart-e%E2%80%94surgical-devices/ELC) · K130137

# BIOSONIC SUVI PIEZEOELECTRIC SCALER AND POLISHER (K130137)

_Coltene/Whaledent · ELC · Aug 23, 2013 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/ELC/K130137

## Device Facts

- **Applicant:** Coltene/Whaledent
- **Product Code:** [ELC](/submissions/DE/subpart-e%E2%80%94surgical-devices/ELC.md)
- **Decision Date:** Aug 23, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.4850
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Intended Use

The BioSonic Suvi ultrasonic scaler is intended for use in dental applications such as supra and subgingival scaling, periodontal therapy, endodontic procedures and cavity preparation. The BioSonic Suvi Air Polisher is intended for use in the cleaning, surface preparation and polishing of teeth by the projection of water, air and dental powders onto the tooth surface. The device removes dental plaque, soft deposits, and surface stains from pits, grooves, interproximal spaces or smooth surfaces of teeth. The BioSonic Suvi Air Polisher is also indicated for the nonsurgical removal of subgingival plaque in pockets up to 5 mm after initial periodontal treatment.

## Device Story

BioSonic Suvi is a portable, stand-alone dental device combining an ultrasonic piezoelectric scaler and an air polisher. Scaler uses piezoelectric crystals to convert electrical signals into mechanical vibrations, transmitted via a resonator to a stainless steel tip; vibrations break up tartar/plaque, while water irrigation flushes debris. Polisher uses pressurized air, water, and dental powder to remove stains/plaque and polish tooth surfaces. Operated by dental professionals in clinical settings. Models include scaler-only or combination units, with options for tap water or medicament dispenser systems. LED lights provide visibility. Output allows clinicians to perform periodontal therapy and surface cleaning, improving oral hygiene and patient dental health.

## Clinical Evidence

Bench testing only. Device meets IEC 60601-1 (general safety) and EN 60601-1-2 (electromagnetic compatibility) standards.

## Technological Characteristics

Portable stand-alone unit. Materials: Polysulfone (LED lens), stainless steel (tips/nozzle), silicone (grip), ABS (housing), PEEK (handpiece), polyurethane (tubing). Piezoelectric sensing/actuation. Energy: AC power (100, 115, 230 Vac, 50-60 Hz). Features: LED illumination, optional medicament dispenser. Dimensions: 270 x 140 x 165 mm.

## Regulatory Identification

An ultrasonic scaler is a device intended for use during dental cleaning and periodontal (gum) therapy to remove calculus deposits from teeth by application of an ultrasonic vibrating scaler tip to the teeth.

## Predicate Devices

- EMS Electro Medical Systems EMS Air-Flow Master Piezon ([K110173](/device/K110173.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K130137

Coltene/Whaledent BioSonic Suvi Safety Summary (Modified July 15, 2013)

# "" COLTENE®

235 Ascot Parkway Cuyahoga Falls, OH 44223 Unites States of America Tel: +1-330-916-8800

#### "510(K) SUMMARY" AS REQUIRED BY SECTION 807.92(c) Modified July 15, 2013

#### 510(k) Owner's Name, Address, Telephone Number, Fax Number, Contact Person and Date Prepared.

Coltene/Whaledent, Inc. 235 Ascot Parkway Cuyahoga Falls, Ohio 44223 Phone: (330) 916-8800 (330) 917-7077 Fax:

Contact Person:

# AUG 2 3 2013

Edward A. Kroll President, Spectre Solutions, Inc. and Representative Consultant for Spinergy, Inc. 5905 Fawn Lane Cleveland, Ohio 44141 Phone: (440) 546-9810 Fax: (440) 546-9124

Date Prepared: July 15, 2013

# Name of Device

- Trade Name: BioSonic Suvi Piezoelectric Scaler and Polisher
	- . Common Name: Scaler
	- Classification Name: ●
		- O Scaler, Ultrasonic

# Predicate Device

EMS Electro Medical Systems EMS Air-Flow Master Piezon (K110173)

# Device Description:

The BioSonic Suvi Piezoelectric Scaler and Air Polisher (BioSonic Suvi) is a combination scaler and polisher device for use in dental care. It combines the functions of an ultrasonic scaler and dental air polisher into a single device.

| Volume | Page Number |
|--------|-------------|
| 4      | 4-1         |

{1}------------------------------------------------

The scaler is used for removal of tartar or calculus on teeth and other dental work where ultrasonic vibration is beneficial. The polisher unit is used for removal of plaque, cleaning discolored teeth and other dental work where air polishing is beneficial.

The BioSonic Suyi will be offered in (4) four different models. These are the BioSonic Suvi Premier, BioSonic Suvi Premier Plus, BioSonic Suvi Elite and the BioSonic Suvi Elite Plus. The Premier and Premier Plus are scalers only. The Elite and Elite Plus are combination scaler and polisher units

The Premier and Plus versions differ in whether they include a standard tap water supply attachment or an optional medicament dispenser system. The standard tap water versions connect directly to a tap water supply using a hose connection. The medicament dispenser system provides either medicine or clean water using a medicament bottle which connects to the side of the unit.

The medicament versions utilize an internal air compressor to pressurize the medicament. During use, compressed air forces fluids from the bottle, through the hose, to the hand piece and the tip/nozzle. Thus medication can be provided if desired.

#### Device Function

#### Scaler:

The scaler includes a power unit and ultrasonic hand piece with scaler tip. A mechanical resonator within the scaler tip vibrates when power is supplied to the hand piece. When held against the tooth, sound waves bounce off the tooth and cause the tartar and plaque to break up. Water coming from the scaler then washes away the broken debris.

#### Polisher:

The polisher includes a power unit, a water line, an air/powder line and a hand piece nozzle. It functions by spraying pressurized water mixed with a polishing agent onto the surface of the teeth. This mixture washes away residue and plaque while removing stains and neutralizing acidic conditions in the mouth.

#### Scientific Concepts

#### Scaler:

The scaler utilizes the scientific concept of piezoelectricity whereby materials, such as piezoelectric crystals, have the ability to generate a voltage when a mechanical force is applied to them. The scaler hand piece contains a mechanical resonator that is coupled to the scaler tip. The resonator is driven by a power electronic unit.

The conversion of the electrical signal to mechanical movement is achieved by a piezoelectric ceramic plate within the hand piece. The resulting movement of the ceramic plate is transferred to a metal resonator which is brought into resonance by selecting the correct frequency. This movement is transferred to the scaler tip causing the outer part of the tip to vibrate.

# Polisher:

| Volume |  | Page Number |
|--------|--|-------------|
| 4      |  | 4-2         |

{2}------------------------------------------------

The polisher incorporates the basic principle of using pressurized air, mixed with water and a polishing agent as a method of cleaning and polishing. This mixture removes extrinsic stains, dental plaque and soft debris while simultaneously polishing tooth surfaces.

# Significant Physical and Performance Characteristics

Design:

- . Portable stand-alone device
- . Combination scaler and polisher in one unit
- Equipped with LED lights that provide for optimal visibility ◆
- Equipped with either a tap water connector or quick-connect water/medicament . dispenser.

Materials:

| Component                           | Material        |
|-------------------------------------|-----------------|
| LED lens                            | Polysulfone     |
| Scaler tips and air polisher nozzle | Stainless steel |
| Handpiece Grip                      | Silicone        |
| Housing and top of housing          | ABS             |
| Handpiece (external)                | PEEK            |
| Water line tubing                   | Polyurethane    |
| Air line tubing                     | Polyurethane    |

Physical Properties:

Elite Plus Premier Premier Plus ltem Elite 270 x 140 x 165 270 x 140 x 270 x 140 x 165 270 x 140 x 165 *Dimensions 165mm mm mm mm 3400 g 3400 g 2900 g 2900 g Weight 100, 115, 230 Vac, Same Voltage Same Same 50-60 Hz

* Without bottle and powder container attachments

# Intended Use/Indications for Use

The BioSonic Suvi ultrasonic scaler is intended for use in dental applications such as supra and subgingival scaling, periodontal therapy, endodontic procedures and cavity preparation.

The BioSonic Suvi Air Polisher is intended for use in the cleaning, surface preparation and polishing of teeth by the projection of water, air and dental powders onto the tooth surface. The device removes dental plaque, soft deposits, and surface stains from pits, grooves, interproximal

| Volume | Page Number |
|--------|-------------|
| 4      | 4-3         |

{3}------------------------------------------------

Coltene/Whaledent BioSonic Suvi Safety Summary (Modified July 15, 2013)

> spaces or smooth surfaces of teeth. The BioSonic Suvi Air Polisher is also indicated for the nonsurgical removal of subgingival plaque in pockets up to 5 mm after initial periodontal treatment.

#### Predicate Device Comparison

The BioSonic Suvi Piezoelectric Scaler and Polisher (BioSonic Suvi) is substantially equivalent to the EMS Electro Medical Systems SA EMS Air Flow Master Piezon (EMS Air Flow) Ultrasonic Scaler and Polisher Unit. This device was granted marketing clearance by FDA on April 7, 2011 under 510(k) Accession Number K110173.

Both of these products are electrically powered, electromechanical devices designed for use in dental applications. They are constructed from the same basic materials, incorporate the same operational principles and both use AC power as their power source. Additionally, both devices include a footswitch, an Ultrasonic Scaler and a Polisher.

Both devices are intended for use in the cleaning and polishing of teeth by the projection of water, air, and dental powders onto the tooth surface. Additionally, both devices remove dental plaque, soft deposits, and surface stains from pits, grooves, interproximal spaces, or smooth surfaces of teeth. Finally, both can be used for a variety of dental applications.

#### Performance Data: (Non-clinical Testing)

The BioSonic Suvi Piezoelectric Scaler and Polisher has been tested to and meets the requirements of;

- . IEC 60601-1/EN 60601-1 Medical electrical equipment, Part 1: General requirements for basic safety and essential performance
- . EN 60601-1-2 Medical electrical equipment, Parts 1-2 : General requirements for safety-Collateral standard: Electromagnetic compatibility - Requirements and tests

#### Conclusion:

The BioSonic Suvi Piezoelectric Scaler and Polisher is substantially equivalent to its predicate device. It has the same indications for use, is constructed from the same basic materials and incorporates the same operational principles.

| Volume | Page Number |
|--------|-------------|
| 4      | 4-4         |

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - (USA)" arranged around the perimeter. Inside the circle is a stylized image of a bird-like figure, possibly representing a human figure with outstretched arms, or a stylized eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

August 23, 2013

Coltene/Whaledent C/O Mr. Edward Kroll President Spectre Solutions. Incorporated 5905 Fawn Lane CLEVELAND OH 44141

Re: K130137

> Trade/Device Name: BioSonic Suvi Piezoelectric Scaler and Polisher Regulation Number: 21 CFR 872.4850 Regulation Name: Dental Scaler, Dental Polisher Regulatory Class: II Product Code: ELC, EFB Dated: July 22, 2013 Received: July 31, 2013

Dear Mr. Kroll:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2- Mr. Kroll

{5}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note

the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/Medica]Devices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Mary S. Runner -S

Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

#### Indications for Use

510(k) Number (if known): TBD

Device Name: BioSonic Suvi Piezoelectric Scaler and Polisher

Indications For Use:

The BioSonic Suvi ultrasonic scaler is intended for use in dental applications such as supra and subgingival scaling, periodontal therapy, endodontic procedures, cavity preparation, and restorative dentistry.

The BioSonic Suvi Air Polisher is intended for use in the cleaning, surface preparation and polishing of teeth by the projection of water, air and dental powders onto the tooth surface. The device removes dental plaque, soft deposits, and surface stains from pits, grooves, interproximal spaces or smooth surfaces of teeth. The BioSonic Suvi Air Polisher is also indicated for the nonsurgical removal of subgingival plaque in pockets up to 5 mm after initial periodontal treatment.

Prescription Use X (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 |
|--------------------------------------------------------------------|
| Andrew I. Steen -S<br>2013.08.22 08:53:55 -04'00:<br>FDA           |
| (Division Sign-Off)                                                |
| Division of Anesthesiology, General Hospital                       |
| Respiratory, Infection Control and                                 |
| Dental Devices                                                     |
| 510(k) Number: K130137                                             |

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/ELC/K130137](https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/ELC/K130137)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
