← Product Code [EGS](/submissions/DE/subpart-e%E2%80%94surgical-devices/EGS) · K222518

# FX Contra (K222518)

_Nakanishi, Inc. · EGS · Nov 17, 2022 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/EGS/K222518

## Device Facts

- **Applicant:** Nakanishi, Inc.
- **Product Code:** [EGS](/submissions/DE/subpart-e%E2%80%94surgical-devices/EGS.md)
- **Decision Date:** Nov 17, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.4200
- **Device Class:** Class 1
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Intended Use

FX15 / FX15m / FX25 / FX25m FX Contra is intended for the following application(s): Caries removal, cavity and crown preparation, removal of dental restorations (fillings and prostheses), finishing and polishing of teeth and dental restorations. FX57 / FX57m FX Contra is intended for the following application(s): Prophylaxis treatment of the surface of teeth and dental restorations.

## Device Story

FX Contra is a series of reusable contra-angle dental handpieces (FX15, FX15m, FX25, FX25m, FX57, FX57m) used in healthcare/hospital settings by dental professionals. Devices connect to ISO 3964-compliant electronic micromotors or air motors to receive rotational energy. Energy is transmitted via internal shafts and gears to a head housing, which accepts ISO 1797-1 burs (for preparation/restoration models) or ISO 13295 cups/brushes (for prophylaxis models). Operation is based on rotatory movement; clinicians use the handpiece to perform caries removal, cavity preparation, and polishing. The device is constructed of stainless steel and aluminum and is autoclave-sterilizable. Output is mechanical rotation at the tool tip, enabling dental procedures. Benefits include efficient removal of decayed matter and surface finishing.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including performance testing (ISO 14457), sterilization validation (ISO 17665-1), cleaning validation, and biocompatibility testing (cytotoxicity, sensitization, irritation, acute systemic toxicity, and pyrogenicity per ISO 10993).

## Technological Characteristics

Reusable contra-angle handpieces; stainless steel and aluminum construction. ISO 3964-compliant coupling; ISO 1797-1 or ISO 13295 tool interface. Rotatory movement driven by external electronic micromotor or air motor. Max drive speed 20,000-40,000 min-1; max rotation speed 5,000-40,000 min-1 depending on model. Sterilization via steam autoclave.

## Regulatory Identification

A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.

## Predicate Devices

- SANAO Dental Handpieces ([K132356](/device/K132356.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 17, 2022

Nakanishi Inc. % Yulia Nikova Regulatory Affairs Manager Ken Block Consulting LLC 800 E Campbell Road, Suite 202 Richardson. Texas 75081

Re: K222518

Trade/Device Name: FX Contra Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: Class I, reserved Product Code: EGS Dated: August 19, 2022 Received: August 19, 2022

Dear Yulia Nikova:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Michael E. Adjodha -S

Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K222518

Device Name FX Contra

Indications for Use (Describe)

FX15 / FX15m / FX25 / FX25m FX Contra is intended for the following application(s): Caries removal, cavity and crown preparation, removal of dental restorations (fillings and prostheses), finishing and polishing of teeth and dental restorations.

FX57 / FX57m

FX Contra is intended for the following application(s): Prophylaxis treatment of the surface of teeth and dental restorations.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### 510(k) Summary

K222518

| Submitter:          | NAKANISHI INC.<br>700 Shimohinata<br>Kanuma Tochigi 322-8666 Japan                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:     | NAKANISHI INC.<br>Mr. Masaaki Kikuchi<br>General Manager, Quality/Regulatory Affairs Dept.<br>700 Shimohinata<br>Kanuma Tochigi 322-8666 Japan<br>TEL: +81-289-64-7277<br>FAX: +81-289-62-9738<br>m-kikuchi@nsk-nakanishi.co.jp                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Date Prepared:      | November 16, 2022                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Submission Type:    | Traditional 510(k)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Subject Device:     | Manufacturer: NAKANISHI INC.<br>Trade/Device Name: FX Contra<br>Common Name: Handpiece, contra- and right-angle attachment, dental<br>Regulation Number: 21 CFR 872.4200<br>Regulation Name: Dental Handpiece and Accessories<br>Regulatory Class: Class I<br>Product Code: EGS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Predicate Device:   | Clearance: K132356 dated January 31, 2014<br>Manufacturer: SciCan GmbH<br>Trade/Device Name: SANAO Dental Handpieces<br>Common Name: Handpiece, contra- and right-angle attachment, dental<br>Regulation Number: 21 CFR 872.4200<br>Regulation Name: Dental Handpiece and Accessories<br>Regulatory Class: Class I<br>Product Code: EGS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Device Description: | The FX Contra includes six models: FX15, FX15m, FX25, FX25m, FX57, and FX57m. The FX Contra handpieces are reusable contra-angle handpieces, driven by an ISO 3964 compliant electronic micromotor or air motor, that are intended for use in a healthcare facility/hospital setting by a trained health care professional in the field of general dentistry. The FX Contra handpieces are designed for general applications to prepare dental cavities for restorations (FX15/FX15m/FX25/FX25m) and for prophylactic teeth cleaning (FX57/FX57m).<br><br>The FX Contra handpieces feature ISO 3964 compliant handpiece coupling connecting them to a dental unit through either a micromotor or an air motor. By these means the handpieces receive the energy for the gear for dental treatment. The maximum drive speed is 40,000 min-1 for FX15, FX15m, FX25, and FX25m; and 20,000 min-1 for FX57 and FX57m. The maximum rotation speed is 10,000 min-1 for FX15 and FX15m; 40,000 min-1 for FX25 and FX25m; and 5,000 min-1 for FX57 and FX57m.<br><br>The FX Contra handpieces are composed mainly of stainless steel and aluminum and can be sterilized in a steam sterilizer (autoclave). |

# NAKANISHI INC.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image features the logo of NSK, a company specializing in motion and control technologies. The logo consists of the letters 'NSK' in a bold, sans-serif font, with a stylized line extending from the top of the 'N' over the 'S' and 'K'. The logo is set against a dark blue background, providing a strong contrast that makes the logo easily visible.

| 510(k) Summary                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | K222518<br>The FX Contra handpieces are intended to be used with ISO 1797-1 compliant burs<br>(FX15, FX15m, FX25 and FX25m) or ISO 13295compliant cups or brush (FX57 and<br>FX 57m).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                 |                         |  |  |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|-------------------------|--|--|
| Principle of<br>Operation /<br>Mechanism of<br>Action:                                                                                                                                                                                                                                                                                                                                                                                                                           | The principle of operation of the FX Contra handpieces is based on rotatory<br>movement. The contra-angle FX Contra handpieces are driven by energy of<br>electronic micromotor or an air motor, which is converted into mechanical energy<br>through the motor. This mechanical energy is transferred to the drivetrains in the<br>contra-angle handpiece in the form of a rotary movement via an ISO 3964 compliant<br>coupling. The force is transmitted through shafts and gear components to another<br>coupling interface located in the head housing, which is designed for safe insertion<br>and removal of ISO 1797 or ISO 13295 compliant tools. The mechanical energy is<br>passed on to the cutting or polishing part of these tools and thus to the procedure site. |                 |                         |  |  |
| Statement of<br>Intended Use:                                                                                                                                                                                                                                                                                                                                                                                                                                                    | FX15 / FX15m / FX25 / FX25m<br>FX Contra is intended for the following application(s):<br>Caries removal, cavity and crown preparation, removal of dental restorations (fillings<br>and prostheses), finishing and polishing of teeth and dental restorations.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                 |                         |  |  |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | FX57 / FX57m<br>FX Contra is intended for the following application(s):<br>Prophylaxis treatment of the surface of teeth and dental restorations.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                 |                         |  |  |
| Summary of<br>Technological<br>Characteristics:                                                                                                                                                                                                                                                                                                                                                                                                                                  | The intended use, technological characteristics, and functionality are substantially<br>equivalent between the subject FX Contra handpieces and the predicate device. The<br>performance characteristics are also equivalent to the predicate device. Both devices<br>are intended to be used by dental health care professionals (e.g., dentists,<br>periodontists). Both devices are intended for multiple patients use and sterilized by<br>autoclave sterilization. Additionally, the FX Contra handpieces and the predicate<br>device have the same technological characteristics:                                                                                                                                                                                          |                 |                         |  |  |
| Power source<br>●                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                 |                         |  |  |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Principle of operation<br>●                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                 |                         |  |  |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Push-button auto-chuck<br>●                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                 |                         |  |  |
| As presented in the brief comparison table below, several differences with respect to<br>technological characteristics were identified between the subject device and the<br>predicate, including but not limited to the transmission gear ratios, compatible burs,<br>and material composition. Performance testing was conducted to support substantial<br>equivalence and demonstrate that these differences do not raise different questions of<br>safety and effectiveness. |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                 |                         |  |  |
| A brief summary of the technological characteristics in comparison to those of the<br>predicate device is as follows:                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                 |                         |  |  |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Subject Device: | Predicate Device:       |  |  |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Trade/Device Name                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | FX Contra       | SANAO Dental Handpieces |  |  |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 510(k) Submitter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | NAKANISHI INC.  | SciCan GmbH             |  |  |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | [Number]                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | K222518         | K132356                 |  |  |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Device Class                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Class I         | Class I                 |  |  |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Product Code                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | EGS             | EGS                     |  |  |

{5}------------------------------------------------

510(k) Summary

K222518

| Regulation                   | 21 CFR 872.4200                                                                                                                                                                                                                                                                     | K222518<br>21 CFR 872.4200                                                                                                                                                                                                                                                                                                                              |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use          | FX15 / FX15m / FX25 /<br>FX25m<br>FX Contra is intended for<br>the following application(s):<br>Caries removal, cavity and<br>crown preparation, removal<br>of dental restorations<br>(fillings and prostheses),<br>finishing and polishing of<br>teeth and dental<br>restorations. | This medical device is only<br>intended for dental<br>treatment in the area of<br>dentistry. It is intended to be<br>used for the following<br>applications:<br>- SANAO 200 L / 40 ST:<br>The removal of decayed<br>matter, cavity and crown<br>preparations, the removal of<br>fillings and surface<br>finishing of tooth and<br>restoration surfaces. |
|                              | FX57 / FX57m<br>FX Contra is intended for<br>the following application(s):<br>Prophylaxis treatment of the<br>surface of teeth and dental<br>restorations.                                                                                                                          | - SANAO 40 / 40 L / 10 /<br>10 L Cavity preparations,<br>caries excavation,<br>endodontics, surface<br>finishing of tooth and<br>restoration surfaces.<br>- SANAO PSI / PSO:<br>prophylaxis treatment.                                                                                                                                                  |
| Application                  | General dentistry and<br>prophylaxis                                                                                                                                                                                                                                                | General dentistry and<br>prophylaxis                                                                                                                                                                                                                                                                                                                    |
| Model Numbers                | General Application<br>FX15 / FX15m / FX25 /<br>FX25m                                                                                                                                                                                                                               | General Application<br>SANAO 10 / SANAO 10 L<br>/ SANAO 40 / SANAO 40<br>L / SANAO 200L / SANAO<br>40 ST                                                                                                                                                                                                                                                |
|                              | Prophylaxis<br>FX57 / FX57m                                                                                                                                                                                                                                                         | Prophylaxis<br>SANAO PSI / SANAO PSO                                                                                                                                                                                                                                                                                                                    |
| Power Source                 | Electronic Micromotor or<br>Air Motor                                                                                                                                                                                                                                               | Electronic Micromotor or<br>Air Motor                                                                                                                                                                                                                                                                                                                   |
| Handpiece Chuck<br>Mechanism | Push Button<br>FX15 / FX15m / FX25 /<br>FX25m                                                                                                                                                                                                                                       | Push Button<br>SANAO 10 / SANAO 10 L<br>/ SANAO 40 / SANAO 40<br>L / SANAO 200 L                                                                                                                                                                                                                                                                        |
|                              | Not Applicable<br>FX57 / FX57m                                                                                                                                                                                                                                                      | Not Applicable<br>SANAO PSI / SANAO PSO<br>Mechanical<br>SANAO 40 ST                                                                                                                                                                                                                                                                                    |
| Burs / Cup or Brush          | ISO 1797 Type 1 (2.334 -<br>2.35mm)                                                                                                                                                                                                                                                 | ISO 1797 Type 1 (2.35mm)                                                                                                                                                                                                                                                                                                                                |
|                              |                                                                                                                                                                                                                                                                                     | K222518                                                                                                                                                                                                                                                                                                                                                 |
|                              | FX15 / FX15m / FX25 /<br>FX25m                                                                                                                                                                                                                                                      | SANAO 10 / SANAO 10 L<br>/ SANAO 40 / SANAO 40<br>L                                                                                                                                                                                                                                                                                                     |
|                              |                                                                                                                                                                                                                                                                                     | ISO 1797 Type 2 (2.35mm)                                                                                                                                                                                                                                                                                                                                |
|                              |                                                                                                                                                                                                                                                                                     | SANAO 40 ST                                                                                                                                                                                                                                                                                                                                             |
|                              |                                                                                                                                                                                                                                                                                     | ISO 1797 Type 3 (1.60mm)                                                                                                                                                                                                                                                                                                                                |
|                              |                                                                                                                                                                                                                                                                                     | SANAO 200 L                                                                                                                                                                                                                                                                                                                                             |
|                              |                                                                                                                                                                                                                                                                                     | ISO 13295 Type 3<br>SANAO PSI                                                                                                                                                                                                                                                                                                                           |
|                              | ISO 13295 Type 5<br>FX57 / FX57m                                                                                                                                                                                                                                                    | ISO 13295 Type 5<br>SANAO PSO                                                                                                                                                                                                                                                                                                                           |
|                              |                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                         |
| Transmission Ratio           | 1:1<br>FX25 / FX25m                                                                                                                                                                                                                                                                 | 1:1<br>SANAO 40 / SANAO 40 L<br>/ SANAO 40 ST                                                                                                                                                                                                                                                                                                           |
|                              | 4:1<br>FX15 / FX15m / FX57 /<br>FX57m                                                                                                                                                                                                                                               | 5:1<br>SANAO 10 / SANAO 10 L<br>/ SANAO PSI / SANAO<br>PSO                                                                                                                                                                                                                                                                                              |
|                              |                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                         |
|                              |                                                                                                                                                                                                                                                                                     | 1:5<br>SANAO 200 L                                                                                                                                                                                                                                                                                                                                      |
|                              |                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                         |
| Materials                    | Stainless Steel<br>Aluminum                                                                                                                                                                                                                                                         | Stainless Steel<br>PEEK                                                                                                                                                                                                                                                                                                                                 |

# NAKANISHI INC.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image features the logo for NSK, a company specializing in motion and control technologies. The logo consists of the letters "NSK" in a bold, sans-serif typeface, rendered in white. The background is a solid, dark blue color, providing a strong contrast that makes the logo stand out.

510(k) Summary

### Summary of Non-Clinical Performance Testing:

The FX Contra handpieces are developed and produced under considerations of all applicable technical standards, internal specifications, and FDA guidance documents. The product's conformance with applicable international and internal standards was verified in the course of bench testing.

Tests were performed on the subject device including verification/validation testing to internal functional specifications which demonstrated that the device is substantially equivalent. Specifically,

- -Sterilization has been validated in conformance to the FDA recognized consensus standard AAMI/ANSI/ISO 17665-1:2006 "Sterilization of Health Care Products - Moist Heat - Part 1: Requirements for the Development, Validation, and Routine Control of a Sterilization Process for Medical Devices."
- -Performance has been demonstrated in conformance to ISO 14457:2017 "Dentistry - Handpieces and Motors."
- Cleaning validation was provided demonstrating that the FX Contra handpieces complies with the FDA requirements stated in Guidance for the Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling -Guidance for Industry and Food and Drug Administration Staff.

Biocompatibility Biocompatibility evaluations were selected in accordance with AAMI/ANSI/ISO Testing: 10993-1: 2018 "Biological evaluation of medical devices - Part 1: Evaluation and

### NAKANISHI

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for NSK, a Japanese manufacturer of bearings. The logo is white and is set against a blue background. The letters "NSK" are in a bold, sans-serif font, with a stylized line above the letters.

#### 510(k) Summary K222518 testing" and FDA Guidance Use on International Standard ISO 10993, "Biological evaluation of medical devices – Part 1: Evaluation and Testing " and included: ● Cytotoxicity per ISO10993-5 Sensitization per ISO10993-10 ● Irritation per ISO10993-10 ● Acute Systemic Toxicity per ISO 10993-11 ● Pyrogenicity per ISO 10993-11 ● Clinical Testing: Clinical testing was not required as non-clinical testing is believed to be sufficient for a determination of substantial equivalence of the FX Contra handpieces. Conclusion: NAKANISHI INC. considers the FX Contra handpieces to be substantially equivalent to the predicate device listed above. This conclusion is based on the similarities in primary intended use, principles of operation, functional design, and established medical use.

## NAKANISHI ING

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/EGS/K222518](https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/EGS/K222518)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
