High speed air turbine handpiece

K251389 · Foshan Topmed Dental Co., Ltd. · EFB · Jan 12, 2026 · Dental

Device Facts

Record IDK251389
Device NameHigh speed air turbine handpiece
ApplicantFoshan Topmed Dental Co., Ltd.
Product CodeEFB · Dental
Decision DateJan 12, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.4200
Device ClassClass 1
AttributesTherapeutic

Intended Use

High-speed air turbine handpiece is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth.

Device Story

High-speed air turbine handpiece; air-powered dental instrument for drilling, grinding, and repairing. Composed of head, handle, and connector. Used in dental clinics by dental professionals. Operates at 300,000–400,000 rpm; utilizes push-button chuck for bur retention. Requires routine maintenance with specified lubricant (PANA SPRAY Plus). Sterilized via steam autoclave. Output is mechanical rotation for dental procedures; assists clinicians in tooth preparation and restoration. Benefits patient by enabling precise removal of carious material and finishing of dental restorations.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including performance testing per ISO 14457:2017 and biocompatibility testing per ISO 10993-1, 10993-5, 10993-10, 10993-11, and 10993-23.

Technological Characteristics

Air-powered turbine handpiece; materials: stainless steel, brass. Dimensions/form factor: head, handle, connector. Operational speed: 300,000–400,000 rpm. Operating pressure: 0.20-0.25 MPa. Chuck type: push-button. Sterilization: steam autoclave. Compliance: ISO 14457:2017, ISO 10993 series.

Indications for Use

Indicated for dental professionals for removal of carious material, excess filling material, cavity and crown preparation, finishing of restorations, root canal preparation, and tooth polishing.

Regulatory Classification

Identification

A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION January 12, 2026 Foshan Topmed Dental Co., Ltd % Salon Chen System engineer IMD Medical & Drug technology service institutions Room 308, Building 11 No. 23 Jinqu Road, Wanjiang District Dongguan, Guangdong 523932 China Re: K251389 Trade/Device Name: High speed air turbine handpiece Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece And Accessories Regulatory Class: Class I, reserved Product Code: EFB Dated: May 5, 2025 Received: May 5, 2025 Dear Salon Chen: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251389 - Salon Chen Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K251389 - Salon Chen Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, ![img-0.jpeg](img-0.jpeg) Bobak Shirmohammadi -S For Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} High-speed air turbine handpiece Page 8 of 28 | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251389 | ? | | Please provide the device trade name(s). | | ? | | High-speed air turbine handpiece (TM-T1B2、TM-T1M4、TM-T2B2、TM-T2M4、TM-T3B2、TM-T3M4、TM-T4B2、TM-T4M4、TM-T6B2、TM-T6M4) | | | | Please provide your Indications for Use below. | | ? | | High-speed air turbine handpiece is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} K251389 - 510(k) Summary 1. Submitter's Identification: - Company Name: Foshan Topmed Dental Co., Ltd - Address: Room 401, Floor 4, Block C, No.7, Gui dan road side section, Luo cun Street, Shi shan Nanhai District, Foshan City - Phone: +86-757-81803362 - Fax: +86-757-81803362 - Contact Person (Title): Liang Guo chan (General Manager) - E-mail: TOPMED03 @TOPMEDDDENTAL.COM - Date of Preparation: January 7, 2026 2. Common Name and Classification: - Device Classification Name: Dental Handpiece and accessories - Classification Product Code: EFB - Regulation Number: 21CFR 872.4200 - Class: 1 - Review Panel: Dental - Device Name: High-speed air turbine handpiece - Model: TM-T1B2、TM-T1M4、TM-T2B2、TM-T2M4、TM-T3B2、TM-T3M4、TM-T4B2、TM-T4M4、TM-T6B2、TM-T6M4 3. Predicate Device Information: - 510(k) Number: K170229 - Device Classification Name: Dental - Sponsor: Guangdong JINME Medical Technology Co., Ltd. - Classification Product Code: EFB - Regulation Number: 21CFR 872.4200 - Class: 1 - Review Panel: Dental - Trade/Proprietary Name: Dental High-speed Turbine Handpiece Page 1 {5} Common Name: Dental Handpiece and Accessories # 4. Application Correspondent > Company Name: IMD Medical & Drug technology service institutions Phone: +86-18613190779 Fax: +86-755-62809168 > Contact Person (Title): Salon Chen (System engineer) E-mail: 33999439@qq.com Address: Room 308, Building 11, No. 23 Jinqu Road, Wanjiang District, Dongguan City, Guangdong Province, China # 5. Device Description High-speed Turbine Air Handpiece is an instrument for drilling, grinding, repairing utilized in dental clinics. It composed of a head, a handle and a connector. The subject device is intended to be used for dental professional use only. Lubricant should be used during routine maintenance (e.g. after each patient use and prior to sterilization). The user must buy and use specified lubricant type "PANA SPRAY Plus" manufactured by NAKANISHI INC (cleared in K163483). # 6. Indications for Use High-speed air turbine handpiece is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth. # 7. Comparison to the predicate device Table 1 General Comparison | Elements of Comparison | Proposed Device | Predicate Device | Judgment | | --- | --- | --- | --- | | Company Name | Foshan Topmed Dental Co., Ltd | Guangdong JINME Medical Technology Co., Ltd. | / | | Device Name | High-speed air turbine handpiece | Dental High-speed Turbine Handpiece | / | {6} Page 3 | Classification Product Code | EFB | EFB | Same | | --- | --- | --- | --- | | Regulation | 21CFR 872.4200 | 21CFR 872.4200 | Same | | Classification Name | Dental | Dental | Same | | Class | 1 | 1 | Same | | Prescription or OTC | Prescription Use | Prescription Use | Same | | Intended Use | High-speed air turbine handpiece is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth. | Dental High-speed Turbine Handpiece is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth. | Same | Table 2 Safety factor & Performance Comparison | Safety factor & Performance | Proposed Device | Predicate Device | Judgment | | --- | --- | --- | --- | | Operational Modes | Air-powered | Air-powered | Same | | Water Spray | Single/Triple | Single/Triple | Same | | Type of Chunk | Push button | Push button Latch-type chuck | Same | | Composition of Main Materials | Stainless Steel, Brass | Stainless Steel, Brass, Titanium | Analysis 1 | | Bur Extraction Force | The pulling force of the test bar from the mechanical collet shall not be less than 22N | 28N | Analysis 2 | | Operating Pressure | 0.20-0.25Mpa | 177KPa ~ 301KPa | Analysis 2 | | Motor Speed | 300,000rpm ~ 400,000rpm | 300,000rpm ~ 400,000rpm | Same | {7} | Sterilization | Steam autoclave method | Steam autoclave method | Same | | --- | --- | --- | --- | | Compliance Standards | Complied with ISO 10993-5, ISO 10993-10, ISO 10993-23, ISO14457 | Complied with ISO 10993-5, ISO 10993-10, ISO14457 | Same | | Lubricant | The specified lubricant, type “PANA SPRAY Plus” manufactured by NAKANISHI INC (Cleared in K163483), must be used during routine maintenance. | The specified lubricant, type “PANA SPRAY Plus” manufactured by NAKANISHI INC (Cleared in K163483), must be used during routine maintenance. | Same | Review of Differences: Analysis 1 The main materials of the subject device are different from those in the predicate devices. However, the subject and predicate devices follow ISO 10993-1. The subject device was tested in accordance with the methods outlined in ISO 10993-5:2009, ISO 10993-10:2021, and ISO 10993-23:2021. These tests were selected on characterization of the components in accordance with ISO 10993-1:2018 (Mucosal membrane device, surface contacting, limited contact duration (<24 hours)). Analysis 2 Although the subject devices are different from the predicate device in Operating Pressure and Bur Extraction Force; the subject devices conform with ISO 14457:2017 Dentistry - Handpieces and Motors. The difference in operating pressure and bur extraction force does not affect substantial equivalence between the subject and predicate devices. Page 4 {8} Page 5 8. Discussion of Non-clinical Tests Performed for Determination of Substantial Equivalence: A series of non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards. 1) Basic Safety and Essential Performance Performance test according to 14457:2017 standard 2) Biocompatibility testing ISO 10993-1:2018 Biological evaluation of medical devices - Parts 1: Evaluation and testing. ISO 10993-5:2019 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity ISO 10993-10:2021 Biological evaluation of medical devices - Parts 10: Tests for skin sensitization ISO 10993-11:2017 Biological evaluation of medical devices, Part 11: Tests for systemic toxicity ISO 10993-23:2021 Biological evaluation of medical devices - Parts 10: Tests for irritation. All the test results demonstrate High-speed air turbine handpiece meets the requirements of its pre-defined acceptance criteria and intended uses, and is substantially equivalent to the predicate device. 9. Clinical Tests Performed No clinical test data was used to support the decision of substantial equivalence. 10. Conclusion The subject devices have all features of the predicate devices. The few differences were evaluated and do not affect the safety and effectiveness of the subject devices compared to the predicate. Thus, the subject devices are substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...