K093084 · Rolence Ent., Inc. · EFB · Dec 16, 2009 · Dental
Device Facts
Record ID
K093084
Device Name
ROLENCE DENTAL HIGH SPEED HANDPIECE, RHP
Applicant
Rolence Ent., Inc.
Product Code
EFB · Dental
Decision Date
Dec 16, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4200
Device Class
Class 1
Attributes
Therapeutic
Intended Use
The ROLENCE Dental High Speed Handpiece, model RHP is intended for removing carious material, reducing hard tooth structure, cavity preparation, finishing tooth preparations and restorations and polishing teeth. CAUTION: U.S. Federal law restricts the use of this device to licensed professionals.
Device Story
Air-powered dental handpiece; utilizes high-speed air turbine for rotary instrumentation. Used by dentists/licensed professionals in clinical settings for tooth structure removal, cavity preparation, and polishing. Device connects to dental unit air supply; operator controls speed/engagement via foot pedal. Mechanical action facilitates precise dental procedures; benefits patient through efficient removal of decay and restoration finishing. No electronic or software components.
Clinical Evidence
Bench testing only. Performance verified against ISO 7785-1:1997 standards for high-speed air turbine handpieces.
Technological Characteristics
Air-powered turbine handpiece. Design compliant with ISO 7785-1:1997. Quality management system compliant with ISO 13485:2003. Mechanical operation; no software or electronic components.
Indications for Use
Indicated for dental patients requiring removal of carious material, reduction of hard tooth structure, cavity preparation, finishing of preparations/restorations, and tooth polishing. Restricted to use by licensed professionals.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
Predicate Devices
JINDELL High Speed Air Turbine Handpiece, models: SW, SP, SU, ETU, MU (K062740)
Related Devices
K062740 — JINDELL HIGH SPPED AIR TURBINE HANDPIECE, MODELS SW, SP, SU, ETU, MU · Jindell Medical Instruments Co., Ltd. · Nov 2, 2006
K033213 — CODENT DENTAL AIR-POWERED HANDPIECE, MODEL HPS · Codent Technical Industry Co., Ltd. · Feb 6, 2004
K062947 — HIGH SPEED AIR TURBINE HANDPIECE, MODELS SUPER-AIR AND SUPER-QD SERIES · Peng Lim Enterprise Co., Ltd. · Dec 14, 2006
K071561 — HI-SPEED TURBINE HANDPIECES, MODEL GSO1-B,GSO2-B,GSO3-B,GSO1-D,GSO2-D · Beijing North Pole Dental Handpiece Co., Ltd. · Oct 12, 2007
K061581 — EXCEL HIGH SPEED DENTAL HANDPIECE · Orthosource, Inc. · Aug 16, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
-K093084
Rolence Enterprise Inc. 18-3 Lane 231 PuChung Rd. ChungLi 320 Taiwan R.O.C. Tel: +886-3-463 1999 ext 10 Fax: +886-3-463 1997
# 5. 510(K) SUMMARY
DEC 1 6 2009
ROLENCE Dental High Speed Handpiece, Model: RHP
| 510K: | |
|----------------------------|--------------------------------------------------|
| Submitted by: | ROLENCE Enterprise Inc. |
| | No.18-3, Lane 231, Pu Chung Rd., Chungli, 32083, |
| | Taiwan, ROC |
| Contact person: | Dr. Jen, Ke-Min |
| | No.58, Fu-Chiun Street, Hsin-Chu City, 30067, |
| | Taiwan, ROC |
| | Tel: 886-3-5208829 fax: 886-3-5209783 |
| | E-mail: ceirs.jen@msa.hinet.net |
| Date Summary Prepared: | September 18, 2009 |
| Name of the Device: | Dental Air-Powered Handpiece |
| Classification: | Dental Air-Powered Handpiece ( class I medical |
| | device; 21 CFR 872.4200 ) |
| | Product code: EFB |
| | Panel: 72 |
| Proprietary name: | ROLENCE Dental High Speed Handpiece, model RHP |
| Predicate Device: | JINDELL High Speed Air Turbine Handpiece, |
| | models: SW, SP, SU, ETU, MU; 510K No – K062740 |
| Statement of Intended Use: | The ROLENCE Dental High Speed Handpiece, |
| | model RHP is intended for removing carious |
| | material, reducing hard tooth structure, cavity |
| | preparation, finishing tooth preparations and |
| | restorations and polishing teeth. |
| | CAUTION: U.S. Federal law restricts the use of |
| | this device to licensed professionals. |
P6
{1}------------------------------------------------
### Rolence Enterprise Inc. 18-3 Lane 231 PuChung Rd. ChungLi 320 Taiwan R.O.C. Tel: +886-3-463 1999 ext 10 Fax: +886-3-463 1997
- Description of Device Design: According to ISO7785-1:1997, Dental Handpiece . -- Part 1: High-speed air turbine handpieces; and ISO13485:2003, Medical Device Quality Management System to complete the device design steps.
- Risks Analysis Method: According to ISO14971:2007, we completed the Risks ● Management Report for the ROLENCE Dental High Speed Handpiece.
- Performance Data: The claim of substantial equivalence is based on . comparisons of formulations and intended uses of the ROLENCE Dental High Speed Handpiece, model RHP and its claimed predicate.
And we also refer to the ISO7785-1 to verify the function test for the ROLENCE Dental High Speed Handpiece.
- Conclusion: Based on the information in the notification, ROLENCE . Enterprise Inc. believes that the Dental High Speed Handpiece is substantially equivalent to the claimed predicate JINDELL High Speed Air Turbine Handpiece, models: SW, SP, SU, ETU, MU, 510K No. K062740.
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle or other bird with outstretched wings.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002
DEC 1 6 2009 ROLENCE Enterprise Incorporated C/O Mr. Shu-Chen Cheng ROC Chinese-European Industry Research Society 2064 Tamarin Drive Columbus, Ohio 43235
Re: K093084
Trade/Device Name: Rolence Dental High Speed Handpiece Model RHP Regulation Number: 21CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: I Product Code: EFB Dated: September 20, 2009 Received: October 1, 2009
Dear Mr. Cheng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2- Mr. Cheng
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/SafetyReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
. for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
Rolence Enterprise Inc. 18-3 Lane 231 PuChung Rd. ChungLi 320 Taiwan R.O.C. Tel: +886-3-463 1999 ext 10 Fax: +886-3-463 1997
## Indications for Use
510 (K) Number ( If Known ): _ K (J 9 3084
ROLENCE Dental High Speed Handpiece, Model: RHP ● Device Name:
Indications for Use :
- · ROLENCE Dental High Speed Handpiece, model: RHP is intended for removing carious material, reducing hard tooth structure, cavity preparation, finishing tooth preparations and restorations and polishing teeth.
• ROLENCE Dental High Speed Handpiece carries the following label:
CAUTION: U.S. Federal law restricts the use of this device to licensed professionals.
Prescription Use
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Kevin Muly for MSR
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K093084
Page 1 of l
ાર
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.