PREMIUM SLOW SPEED; 20K AIR MOTOR, HEADS, CONTRA-ANGLES, STRAIGHT NOSECONES

K083101 · Sable Industries, Inc. · EFB · Feb 23, 2009 · Dental

Device Facts

Record IDK083101
Device NamePREMIUM SLOW SPEED; 20K AIR MOTOR, HEADS, CONTRA-ANGLES, STRAIGHT NOSECONES
ApplicantSable Industries, Inc.
Product CodeEFB · Dental
Decision DateFeb 23, 2009
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 872.4200
Device ClassClass 1
AttributesTherapeutic

Intended Use

Sable Industries Premium Slow Speed Dental Handpieces & Accessories are indicated for use in general dentistry for cleaning and polishing of teeth.

Device Story

Air-powered dental handpieces and gear-driven attachments used in general dentistry. System includes 20K air motors, contra-angle heads, and straight nosecones. Inputs: compressed air and water supply via Midwest-type 4-hole tubing. Operation: air-driven motor rotates burs at speeds up to 20,000-40,000 RPM; 1:1 gear ratio. Fiber optic version provides illumination. Used by dentists/dental professionals in clinical settings for tooth cleaning and polishing. Output: mechanical rotation of dental burs for debris removal and surface treatment. Benefits: enables precise dental procedures through controlled rotation and irrigation.

Clinical Evidence

Bench testing only. Conformity with ISO 13294, ISO 7785-2, ISO 9168, ISO 3964, ISO 17665-1, ISO 11134, and ISO 10993-1 standards demonstrates safety and performance equivalence.

Technological Characteristics

Air-powered dental handpieces; 1:1 gear ratio; 20,000-40,000 RPM operating speed; 42-43.5 PSI (3 Bar) pressure; Midwest-type 4-hole tubing interface; fiber optic illumination (25,000 LUX); integrated locking mechanisms; compatible with ISO 1797-1 burs; moist heat sterilization per ISO 17665-1/ISO 11134; biocompatibility per ISO 10993-1.

Indications for Use

Indicated for use in general dentistry for cleaning and polishing of teeth in patients requiring dental prophylaxis.

Regulatory Classification

Identification

A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ KOS3bl ### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS Sable Industries FEB 2 3 2009 Premium Slow Speed Dental Handpieces & Accessories 510(k) Summary Page 1 of 4 #### I. Submitter Information 510(k) Owner: Sable Industries, Inc. 100 Campbell Avenue, Suite 5 Kitchener, Ontario N2H 4X8 Canada Establishment Registration: 3002105090 Jacob Jagt - QA Manager (519) 579-9323 (office) (519) 579-9324 (fax) Manufacturer Contact Person: United States Agent / U.S. Consultant: GlobalReg Compliance Associates, Inc. 16576 W. 53rd Way Golden, CO 80403 (303) 956-8595 (direct) (303) 828-0835 (fax) Email: info@globalregcompliance.com Date Summary Prepared: October 6, 2008 #### II. Name of Device - Trade Name: Premium Slow Speed Dental Handpieces & Accessories . - Common Name: Air-Powered Dental Handpieces & Gear-Driven Attachments - Classification Name: Dental Handpiece & Accessories #### III. Predicate Devices | Subject Device | Predicate Device(s) | |-------------------------------------------------------------------------------------|-------------------------------------------------------------| | Premium Slow Speed 20K Air Motor<br>w/ Internal Air & Water (p/n 1610101) | Bien Air Low Speed Motors<br>(K983183) | | Slow Speed & Premium Slow Speed Heads<br>Ball Bearing Switch Latch<br>(p/n 1610801) | Bien Air Straight and Contra-<br>Angle Handpieces (K983183) | {1}------------------------------------------------ ## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS Sable Industries Premium Slow Speed Dental Handpieces & Accessories 510(k) Summary Page 2 of 4 | Subject Device | Predicate Device(s) | |-----------------------------------------------------------------------------------------------------------|-------------------------------------------------------------| | Slow Speed & Premium Slow Speed Heads<br>Ball Bearing Push Button<br>(p/n's 1610802, 1600411, 1600508) | Bien Air Straight and Contra-<br>Angle Handpieces (K983183) | | Premium Slow Speed Contra-Angles<br>1:1 Fiber Optic Ball Bearing w/ Internal Air<br>& Water (p/n 1610299) | Bien Air Straight and Contra-<br>Angle Handpieces (K983183) | | Premium Slow Speed Contra-Angles<br>1:1 Ball Bearing w/ Internal Air & Water<br>(p/n 1610201) | Bien Air Straight and Contra-<br>Angle Handpieces (K983183) | | Premium Slow Speed Straight Nosecones<br>1:1 Ball Bearing w/ Internal Air & Water<br>(p/n 1610301) | Bien Air Straight and Contra-<br>Angle Handpieces (K983183) | #### IV. Device Description & Technological Characteristics A. Premium Slow Speed 20K Air Motor w/ Internal Air & Water: The Sable Premium Slow Speed Air Motor is used in conjunction with accessories such as contra-angle and straight attachments that are manufactured in accordance with ISO 7785-2. This premium slow-speed air-driven motor is capable of running up to a speed of 20,000 RPM at 43.5 PSI (3 Bar). It does not require a separate coupler and interfaces with Midwest-type 4-hole tubing. The motor has an integrated locking mechanism for easy connection and removal of any attachment. The motor can deliver both chip air and water to assist with the removal of debris from the cutting area in the oral cavity when used in conjunction with an ISO designed attachment with internal air and water lines. Slow Speed & Premium Slow Speed Heads (push button and switch latch) B. The Sable Premium Slow Speed Heads are used in conjunction with attachments (Contra-angles and straight) and slow speed motors manufactured in accordance with -ISO-7785-2 and ISO 13294 respectively. These heads have an integrated locking mechanism for easy connection and removal of any attachment. These heads can be used with a variety of different latch burs conforming to the specifications of ISO 1797-1" +These heads are capable of running up to a speed of 40,000 RPM at 42 PSI (3 Bar) and have a 1:1 gear/drive ratio. These heads may be used with attachments capable of delivering both chip air and water when used in conjunction with an ISOdesigned air or electric motor with internal air and water. {2}------------------------------------------------ ### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS Sable Industries Premium Slow Speed Dental Handpieces & Accessories 510(k) Summary Page 3 of 4 - Premium Slow Speed Contra-Angles (1:1 Fiber Optic and Non-Fiber Optic C. Ball Bearing w/ Internal Air & Water): The Sable Premium Slow Speed Contra-Angles are used in conjunction with components such as slow speed heads and slow speed motors manufactured in accordance with ISO 7785-2-and ISO 13294, respectively. The fiber optic version produces 25,000 LUX for a shadow-less view of the oral cavity. These contraangles have an integrated locking mechanism for easy "snap-on" connection and removal from the motor. These contra-angles can be used with Sable premium slow speed heads which are attached / removed by a "quick-twist" locking mechanism on the contra Angle. These contra-angles are capable of running up to a speed of 40,000 RPM at 43.5 PSI (3 Bar) and have a 1:1 gear/drive ratio. These contra-angles deliver both chip air and water when used in conjunction with an ISO-designed air or electric motor with internal air and water. - D. Premium Slow Speed Straight Nosecones (1:1 Ball Bearing w/ Internal Air & Water): The Sable Premium Slow Speed Straight Nosecones are used in conjunction with slow speed motors manufactured in accordance with ISO 13294 These nosecones have an integrated locking mechanism for easy connection and removal from the motor and can be used with a variety of different burs conforming to the specifications of ISO 1797-1. These nosecones are capable of running up to a speed of 40.000 RPM at 42 PSI (3 Bar) and have a 1:1 gear/drive ratio. These nosecones deliver both chip air and water when used in conjunction with an ISO-designed air or electric motor with internal air and water. #### V. Intended Use Sable Industries Premium Slow Speed Dental Handpieces & Accessories are indicated for use in general dentistry for cleaning and polishing of teeth. #### Non-Clinical Performance/Safety Data VI. The performance and safety of the Sable Industries Premium Slow Speed Dental Handpieces & Accessories are based upon conformity with applicable aspects of the following standards: - ISO 13294 Dental Handpieces Dental Air Motors ◆ - ISO 7785-2 Dental Handpieces -- Part 2: Straight & Geared Angle Handpieces {3}------------------------------------------------ ## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS Sable Industries Premium Slow Speed Dental Handpieces & Accessories 510(k) Summary Page 4 of 4 - ISO 9168 Dental Handpieces -- Hose Connections . - ISO 3964 Dental Handpieces Coupling Dimensions - ISO 17665-1:2006 Sterilization of health care products -- Moist . heat -- Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices - ISO 11134 Sterilization of health care products Requirements . for validation and routine control -- Industrial moist heat sterilization - ISO 10993-1 Biological Evaluation of Medical Devices Part-1: . Evaluation and Testing Conformity with the above standards demonstrates that the Sable Industries Premium Slow Speed Dental Handpieces & Accessories are as safe, as effective, and perform as well as or better than the predicate devices. ## END OF 510(K) SUMMARY XX {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle. ### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service FEB 2 3 2009 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Sable Industries, Incorporated C/o Mr. Kevin Randall President GlobalReg Compliance Associates, Incorporated 16576 West 53rd Way Golden, Colorado 80403 Rc: K083101 Trade/Device Name: Premium Slow Speed Dental Handpieces and Accessories Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: I Product Code: EFB, EGS Dated: January 27, 2009 Received: February 20, 2009 Dear Mr. Randall: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ ### Page 2 -- Mr. Randall Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Susan Kunes Ginette Y. Michaud, M.D. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health ### Enclosure {6}------------------------------------------------ # 510(k) Number (if known): KO831D| Device Name: ## Premium Slow Speed Dental Handpieces & Accessories Indications For Use: Sable Industries Premium Slow Speed Dental Handpieces & Accessories are indicated for use in general dentistry for cleaning and polishing of teeth. ## PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED- Concurrence of CDRH, Office of Device Evaluation (ODE) Over-The-Counter Use Prescription Use OR (Optional Format 1-2-96) 0000 (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices xvi Ko83 O 510(k) Number: _
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...