← Product Code [EBW](/submissions/DE/subpart-e%E2%80%94surgical-devices/EBW) · K141482

# KI-20 (K141482)

_Saeyang Microtech Co., Ltd. · EBW · Apr 24, 2015 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/EBW/K141482

## Device Facts

- **Applicant:** Saeyang Microtech Co., Ltd.
- **Product Code:** [EBW](/submissions/DE/subpart-e%E2%80%94surgical-devices/EBW.md)
- **Decision Date:** Apr 24, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.4200
- **Device Class:** Class 1
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Intended Use

For use in a wide range of dental procedures including; endodontic surgeries, such as drilling into the tooth canal, and general dentistry, such as removing carious material from the dentin.

## Device Story

Ki-20 is a dental handpiece system for surgery and general dentistry. System components: DC 28V E-type motor, handpieces, control box, foot switch, irrigation pump. Operation: clinician uses control box/foot switch to regulate motor speed, direction, and torque; irrigation system provides cooling water. Device facilitates drilling and material removal. Output: mechanical rotation of dental tools. Used in dental clinics by dentists. Benefits: precise control of speed/torque for endodontic and restorative procedures.

## Clinical Evidence

Bench testing only. Testing included ISO 3964 (coupling), ISO 7494-1 (dental units), ISO 14457 (handpieces/motors), electrical safety (IEC 60601-1 series), EMC, sterilization validation (ISO 11135/17665), software validation, usability testing (IEC 62366), and temperature rise testing.

## Technological Characteristics

System includes DC 28V E-type motor, control box, foot switch, irrigation pump. Materials: Stainless steel and BSBM. Connectivity: Electronic foot control. Sterilization: Validated per ISO 11135/17665. Software: Validated per FDA guidance. Standards: ISO 3964, ISO 7494-1, ISO 14457, IEC 60601-1, IEC 60601-1-2.

## Regulatory Identification

A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.

## Predicate Devices

- Endo E Class ([K123608](/device/K123608.md))

## Reference Devices

- TRAUS SIP10 ([K123695](/device/K123695.md))
- X-CUBE ([K092758](/device/K092758.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized image of three faces in profile, stacked on top of each other. The faces are rendered in a dark color, contrasting with the white background. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 24, 2015

SAEYANG MICROTECH CO., LTD. c/o Ms. Priscilla Chung LK Consulting Group USA, Inc. 2651 E Chapman Ave., Suite 110 Fullerton, CA 92831

Re: K141482 Trade/Device Name: Ki-20 Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: I Product Code: EBW Dated: March 20, 2015 Received: March 23, 2015

Dear Ms. Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tina
Kiang -S

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) k141482

Device Name

Ki-20

Indications for Use (Describe)

For use in a wide range of dental procedures including; endodontic surgeries, such as drilling into the tooth canal, and general dentistry, such as removing carious material from the dentin.

| Type of Use (Select one or both, as applicable)                                                   |                                                                                                  |
|---------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| <div> <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary (k141482)

This summary of 510(k) is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date: Apr 24, 2015

### 1. 510K Applicant / Submitter:

Saeyang Microtech Co., Ltd. 348, Seongseo-ro, Dalseo-gu, Daegu 704-900 REPUBLIC OF KOREA Tel: +82-53-582-9000~2 Fax: +82-53-581-9003

### 2. Submission Contact Person

LK Consulting Group USA, Inc. 2651 E Chapman Ave Ste 110, Fullerton, CA 92831 Priscilla Juhee Chung Phone: 714.202.5789 Fax: 714-409-3357 Email: juhee.c@lkconsultinggroup.com

### 3. Device

- . Proprietary Name: Ki-20
- Common Name: Dental Handpiece and Accessories ●
- Classification Name: Controller, Foot, Handpiece and Cord .
- Regulatory Class: Class I ●
- Product Code: EBW ●
- Classification Regulation: 21 CFR 872.4200 .

### 4. Predicate Device

#### Primary Predicate Device: . Endo E Class by Saeyang Microtech Co., Ltd. (K123608)

### Reference Predicate Devices: . TRAUS SIP10 by Saeshin Precision Co., Ltd. (K123695) X-CUBE by Saeshin Precision Co., Ltd. (K092758)

{4}------------------------------------------------

## 5. Description:

The Ki-20 is designed to use in dental surgery. It consists of an E-type motor (DC 28V Operation Motor), handpieces, a control box, a foot switch and other accessories. It is designed that the speed and direction of the handpiece can be controlled by the control box and the foot switch. The irrigation tube and the pump are used to supply the cooling water for the successful surgery.

### 8. Indications for Use

For use in a wide range of dental procedures including: endodontic surgeries, such as drilling into the tooth canal, and general dentistry, such as removing carious material from the dentin.

### 9. Substantial Equivalence Discussion:

Ki-20 is substantially equivalent to Endo E Class (K123608), TRAUS SIP10 (K123695) and X-CUBE (K092758). The following comparison table is presented to demonstrate substantial equivalence.

Ki-20 does not have a new intended use and it shows equivalent specifications with the predicate devices in most of parameters. The handpiece rotation speed of the subject device encompasses the subject device as well.

Torque setting range applied to the motor of the subject device is different from the predicate device; however, it is within what was cleared in the reference devices.

|                        | Candidate<br>Device                                                                                  | Primary<br>Predicate<br>Device                                                                       | Reference<br>Predicate<br>Device                                                                       | Reference<br>Predicate<br>Device                                                                             | Micromotor<br>drive                      | as drilling into<br>the tooth canal,<br>and general<br>dentistry, such<br>as removing<br>carious material<br>from the dentin.<br>Electric<br>micromotor<br>drive                                                                                                                                         | as drilling into<br>the tooth canal,<br>and general<br>dentistry, such<br>as removing<br>carious material<br>from the dentin.<br>Electric<br>micromotor<br>drive | surgery and<br>endodontics for<br>treatment of<br>dental hard<br>tissue and<br>mechanical<br>rotating root<br>canal<br>preparation<br>Electric<br>micromotor<br>drive                                                                                                                                          | endodontics for<br>treatment of<br>dental hard<br>tissue and<br>mechanical<br>rotating root<br>canal<br>preparation<br>Electric<br>micromotor<br>drive                                                                                                     |
|------------------------|------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number          | K141482                                                                                              | K123608                                                                                              | K123695                                                                                                | K092758                                                                                                      | Package<br>contents                      | Power cord,<br>Control unit,<br>Irrigation tube,<br>Y-tube, Tube<br>holder, Coolant<br>hanger, Foot<br>control,<br>handpiece<br>stand, E-type<br>motor&Motor<br>cord, Foot<br>hanger, Contra<br>angle,<br>E-type spray<br>nozzle,<br>External spray<br>nozzle, Nozzle<br>clamp, Internal<br>spray nozzle | Control Unit,<br>E-type Motor &<br>Motor Cord, E-<br>type<br>Handpiece, foot<br>switch,<br>Handpiece<br>stand, Power<br>Cord,<br>Autoclaving<br>Plugs, Manual    | Hanger, Foot<br>controller,<br>Control box,<br>Power cord,<br>Motor stand,<br>Irrigation tube,<br>Internal spray<br>nozzle, Motor<br>cap for<br>autoclave,<br>Angle<br>handpiece,<br>BLDC motor,<br>Tube holder,<br>Tube clamp, Y-<br>tube, Foot<br>hanger, Foot<br>hanger joint<br>cap, Instruction<br>Manual | Hanger, Foot<br>controller,<br>Control box,<br>Power cord,<br>Motor stand,<br>Irrigation tube,<br>Internal spray<br>nozzle, Motor<br>cap for<br>autoclave,<br>Angle<br>handpiece,<br>BLDC motor,<br>Tube holder,<br>Tube clamp, Y-<br>tube, Foot<br>hanger |
| Device Name            | Ki-20                                                                                                | Endo E Class                                                                                         | TRAUS SIP10                                                                                            | X-CUBE                                                                                                       | Patient<br>Contacting<br>Part's Material | Stainless Steel<br>& BSBM                                                                                                                                                                                                                                                                                | Stainless Steel<br>& BSBM                                                                                                                                        | Stainless Steel<br>& BSBM                                                                                                                                                                                                                                                                                      | Stainless Steel<br>& BSBM                                                                                                                                                                                                                                  |
| Common Name            | Dental<br>Handpiece and<br>accessories                                                               | Dental<br>Handpiece and<br>accessories                                                               | Dental<br>Handpiece and<br>accessories                                                                 | Surgical motor<br>unit for<br>implantology<br>and maxilla<br>surgery                                         | Handpiece<br>chuck                       | Push button &<br>handle type                                                                                                                                                                                                                                                                             | Push button &<br>handle type                                                                                                                                     | Push button &<br>handle type                                                                                                                                                                                                                                                                                   | Push button &<br>handle type                                                                                                                                                                                                                               |
| Manufacturer           | Saeyang<br>Microtech Co.,<br>Ltd.                                                                    | Saeyang<br>Microtech Co.,<br>Ltd.                                                                    | Saeshin<br>Precision Co.,<br>Ltd.                                                                      | Saeshin<br>Precision Co.,<br>Ltd.                                                                            | Coolant Motor                            | DC motor                                                                                                                                                                                                                                                                                                 | No                                                                                                                                                               | DC motor                                                                                                                                                                                                                                                                                                       | DC motor                                                                                                                                                                                                                                                   |
| Intended Use           | For use in a<br>wide range of<br>dental<br>procedures<br>including;<br>endodontic<br>surgeries, such | For use in a<br>wide range of<br>dental<br>procedures<br>including;<br>endodontic<br>surgeries, such | The TRAUS<br>SIP 10 is<br>intended for use<br>in dental<br>surgery,<br>implantology,<br>maxilla-facial | The X-CUBE is<br>intended for use<br>in dental<br>surgery,<br>implantology,<br>maxilla-facial<br>surgery and | Shank length &<br>Type                   | Shank length :<br>11mm(Type 1<br>Shank)                                                                                                                                                                                                                                                                  | Shank length :<br>11mm(Type 1<br>Shank)                                                                                                                          | Shank length :<br>11mm(Type 1<br>Shank)                                                                                                                                                                                                                                                                        | Shank length :<br>11mm(Type 1<br>Shank)                                                                                                                                                                                                                    |
| Operation<br>Principle | Speed control,<br>Torque control,<br>Program<br>Memory                                               | Speed control,<br>Torque control,<br>Program<br>Memory                                               | Speed control,<br>Torque control,<br>Program<br>Memory                                                 | Speed control,<br>Torque control,<br>Program<br>Memory                                                       |                                          |                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                            |

{5}------------------------------------------------

{6}------------------------------------------------

| Motor speed<br>range                                                                       | 600~40,000rpm                                      | 20~17500rpm                                        | 0~40,000 rpm                                       | 600~50,000<br>rpm                                  |
|--------------------------------------------------------------------------------------------|----------------------------------------------------|----------------------------------------------------|----------------------------------------------------|----------------------------------------------------|
| Handpiece rpm<br>range                                                                     | 20 ~ 80,000 rpm                                    | 20~17,500rpm                                       | 10~200,000rpm                                      | 10~200,000rpm                                      |
| Torque setting<br>range applied to<br>the motor in<br>Ncm                                  | 5~65Ncm                                            | 0.1~9.9Ncm                                         | 5~65Ncm                                            | 5~65Ncm                                            |
| Allows<br>reciprocating<br>drive<br>(forward/reverse<br>cycling)                           | Yes                                                | Yes                                                | Yes                                                | Yes                                                |
| Allows selection<br>of gear ratios for<br>different geared<br>E-type<br>handpieces         | 1:1, 1:2,<br>20:1, 32:1                            | 1:1, 4:1, 6:1,<br>8:1, 10:1, 16:1,<br>20:1, 64:1   | 1:5, 1:4, 1:1,<br>16:1, 20:1,<br>27:1, 32:1, 64:1  | 1:5, 1:4, 1:1,<br>16:1, 20:1,<br>27:1, 32:1, 64:1  |
| Allows selection<br>of forward or<br>Auto reverse<br>drive rotation                        | Yes                                                | Yes                                                | Yes                                                | Yes                                                |
| Allows selection<br>of Auto stop                                                           | Yes                                                | Yes                                                | Yes                                                | Yes                                                |
| Allows use of a<br>foot switch<br>control to<br>operate the<br>attached<br>handpiece motor | Yes<br>Electronic foot<br>control                  | Yes<br>Electronic foot<br>control                  | Yes<br>Electronic foot<br>control                  | Yes<br>Electronic foot<br>control                  |
| Allows the user<br>to define their<br>own presets for<br>speed and<br>torque               | Yes                                                | Yes                                                | Yes                                                | Yes                                                |
| Allows<br>programmable<br>doctor's choice                                                  | Yes                                                | Yes                                                | Yes                                                | Yes                                                |
|                                                                                            | AC100V~120V,<br>50/60Hz<br>AC220V~240V,<br>50/60Hz | AC100V~120V,<br>50/60Hz<br>AC220V~240V,<br>50/60Hz | AC100V~120V,<br>50/60Hz<br>AC220V~240V,<br>50/60Hz | AC100V~120V,<br>50/60Hz<br>AC220V~240V,<br>50/60Hz |
| Input<br>voltage(charger)                                                                  |                                                    |                                                    |                                                    |                                                    |
| Handpiece<br>Coupling type                                                                 | E-type                                             | E-type                                             | E-type                                             | E-type                                             |
| Lubricant                                                                                  | NSK PANA<br>SPRAY PLUS                             | NSK PANA<br>SPRAY PLUS                             | Information not<br>available                       | Information not<br>available                       |

{7}------------------------------------------------

| (K131014) | (K131014) |
|-----------|-----------|
|-----------|-----------|

### 10. Performance Tests (Non-clinical)

Non-clinical bench tests were performed as followings:

- ISO 3964:2014 Dental Handpieces Coupling dimensions ●
- ISO 7494-1:2004 Dentistry Dental units -Part 1: General requirements and test methods ●
- ISO 14457:2012 Dentistry Handpieces and motors ●
- IEC 60601-1, IEC 60601-1-2, IEC61000-3-2, IEC61000-3-3: Electrical safety and EMC ●

Along with the above tests, the following tests were performed:

- Sterilization validation in accordance with ISO 11135-1, ISO 11138-1, ISO 11138-3, ISO ● 11607-1, ISO 17664, ISO 17665-1, and ISO 17665-2
- Software validation in accordance with the FDA Guidance, "General Principles of ● Software Validation; Final Guidance for Industry and FDA Staff"
- Usability test EN60601-1-6, EN 62366 and IEC 62366
- . Temperature rise testing

None of the testing demonstrated any design characteristics that violated the requirements of the standards or resulted in any safety hazard.

### 11. Conclusions:

Based on the information provided in this premarket notification, Saeyang Microtech Co., Ltd. concludes that the Ki-20 is substantially equivalent to the predicate device as described herein in.

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/EBW/K141482](https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/EBW/K141482)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
