← Product Code [EBW](/submissions/DE/subpart-e%E2%80%94surgical-devices/EBW) · K090727

# IMM DENTAL IRRIGATION TUBING SET (K090727)

_Innovative Medical Manufacturing Co. · EBW · Aug 13, 2009 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/EBW/K090727

## Device Facts

- **Applicant:** Innovative Medical Manufacturing Co.
- **Product Code:** [EBW](/submissions/DE/subpart-e%E2%80%94surgical-devices/EBW.md)
- **Decision Date:** Aug 13, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.4200
- **Device Class:** Class 1
- **Review Panel:** Dental

## Intended Use

The IMM™ Dental Irrigation Tubing Set is intended for providing passage of irrigating fluid from a solution reservoir to a dental handpiece.

## Device Story

The IMM™ Dental Irrigation Tubing Set is a single-use, sterile medical device designed to facilitate the transport of irrigating fluid from a solution reservoir to a dental handpiece. The device consists of inlet spikes, a drip chamber, PVC tubing, clamps, and silicone tubing. It is used in a dental clinical setting to maintain fluid delivery during dental procedures. The device is purely mechanical and does not involve electronic processing, software, or automated decision-making.

## Clinical Evidence

Bench testing only. Performance testing demonstrated compliance with ISO 8536-4 and ISO 8536-9 standards. Biocompatibility testing confirmed the device meets ISO 10993 requirements for limited exposure, tissue/bone/dentin contact, external communicating devices.

## Technological Characteristics

Single-use, sterile device. Materials: PVC plastic tubing, silicone tubing, inlet spikes, drip chamber, clamps, and caps. Standards: ISO 8536-4, ISO 8536-9 (performance), ISO 10993 (biocompatibility). No energy source, software, or electronic components.

## Regulatory Identification

A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.

## Predicate Devices

- W&H Irrigation Tubing Set ([K041124](/device/K041124.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 090727

## AUG 1 3 2009

## Pre-market Notification for IMMTM Dental Irrigation Tubing Set

### 510(k) Summary

#### 1. Submitter's Name:

INNOVATIVE MEDICAL MAUFACTURING COMPANY 107, 181 Lane, Sect. 1 Yong Jane Road Chunan, Miaoli, 350 TAIWAN (ROC)

Contact:

J. P. Lee, General Manager Tel: +886-37-620236 Fax: +886-37-620239

#### 2. Name of Device

Common/Usual Name: Proprietary Name: Classification Name: Product Code: Regulation Number:

Irrigation Tubing Set IMMTM Dental Irrigation Tubing Set Dental handpiece and accessories EBW 21CFR872.4200

#### 3. Predicate Device

| Trade Name                | 510(k) Number | Decision Date |
|---------------------------|---------------|---------------|
| W&H Irrigation Tubing Set | K041124       | 06/07/2004    |

### 4. Device Description

The IMM™ Dental Irrigation Tubing Set is to be used for providing passage of irrigating fluid t from a solution reservoir to a dental hand piece. The device consists of one or more inlet spikes, drip chamber, PVC plastic tubes, clamps, silicone tube, consists of only caps.

## 5. Indication for Use

The IMM™ Dental Irrigation Tubing Set is intended for providing passage of irrigating fluid from a solution reservoir to a dental handpiece.

## 6. Technological Characteristics

The IMM™ Dental Irrigation Tubing Set is a single use device, and is delivered sterile. The device can be readily connected to a dental handpiece in irrigation or fluid delivery.

## 7. Performance Summary

{1}------------------------------------------------

The functional and performance tests demonstrated that IMM™ Dental Irrigation Tubing Set meets specific requirements established in voluntary standards: ISO8536-4 and ISO8536-9. Biocompatibility test indicated that the device meets the requirements per ISO10993 for "limited exposure, tissue/bone/dentin contact, external communicating" devices.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized graphic of an abstract human figure with three flowing lines representing the body and arms. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the graphic.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Innovative Medical Manufacturing Company C/O Joseph J. Chang, Ph.D., P.E. Consultant Innomedtech LLC 7128 Staffordshire Street Houston, Texas 77030

## AUG 1 3 2009

Re: K090727

Trade/Device Name: IMM™ Dental Irrigation Tubing Set Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: II Product Code: EBW Dated: July 21, 2009 Received: July 23, 2009

Dear Dr. Chang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

### Page 2- Dr. Chang

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807,97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Anthony C. m
Susan Purser D.D.S. M.A.

Susan Runner, D.D.S., M. Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

#### Enclosure

{4}------------------------------------------------

#### 1090727 510(k) Number (if known):

Device Name: IMMTM Dental Irrigation Tubing Set

Indications for Use:

The IMM™ Dental Irrigation Tubing Set is intended for providing passage of irrigating fluid from a solution reservoir to a dental handpiece .

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of

Rein Hulsey for MSR
(Division Sign-Off)

Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K090727

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/EBW/K090727](https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/EBW/K090727)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
