← Product Code [DZJ](/submissions/DE/subpart-e%E2%80%94surgical-devices/DZJ) · K223024

# MedCAD AccuPlan System (K223024)

_Medcad · DZJ · Mar 7, 2023 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/DZJ/K223024

## Device Facts

- **Applicant:** Medcad
- **Product Code:** [DZJ](/submissions/DE/subpart-e%E2%80%94surgical-devices/DZJ.md)
- **Decision Date:** Mar 7, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.4120
- **Device Class:** Class 2
- **Review Panel:** Dental

## Intended Use

The MedCAD® AccuPlan® System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the MedCAD® AccuPlan® System and the result is an output data file that may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces physical outputs including anatomical models, surgical guides, and dental splints for use in maxillofacial surgery. The surgical guides and dental splints are intended to be used for the maxillofacial bone in maxillofacial surgery. The MedCAD® AccuPlan® System is also intended as a pre-operative software tool for simulating / evaluating surgical treatment options.

## Device Story

System processes 3D medical scan data (CT, CBCT) or physical model scans to create patient-specific anatomical models, dental splints, and surgical guides for maxillofacial surgery. Operated by trained MedCAD employees using COTS software; physician provides input/feedback on digital models during planning. Outputs support surgical simulation and execution. Addition of titanium cutting/drilling guides allows for bone cutting/drilling; polymer guides used for marking/positioning. Physician reviews all outputs prior to finalization. Benefits include improved surgical planning and patient-specific anatomical accuracy.

## Clinical Evidence

Bench testing only. Wear debris testing on titanium surgical guides showed debris levels below literature-reported safety thresholds. Fit and form validation confirmed manufactured devices aligned with 3D models and fit representative anatomical defects.

## Technological Characteristics

Materials: Biocompatible polymers and Ti-6Al-4V ELI Titanium Alloy (ASTM F136). Additive manufacturing. Steam sterilization (ISO 17665) to SAL 1x10^-6. Software-based image segmentation and planning tool. Connectivity: Input from medical scanners (CT/CBCT) and physical model scans.

## Regulatory Identification

A bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert a wire, pin, or screw. The device includes the manual bone drill and wire driver, powered bone drill, rotary bone cutting handpiece, and AC-powered bone saw.

## Predicate Devices

- MedCAD® AccuPlan® System ([K192282](/device/K192282.md))

## Reference Devices

- KLS Martin Individual Patient Solutions (IPS) Planning System ([K181241](/device/K181241.md))
- MedCAD® AccuShape® Titanium Patient-Specific Cranial Plate ([K220357](/device/K220357.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the agency's name on the right. The symbol on the left is a stylized depiction of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. The words "FOOD & DRUG" are larger than the other words.

March 7, 2023

MedCAD % Linda Braddon CEO Secure BioMed Evaluations 7828 Hickory Flat Highway Suite 120 Woodstock, Georgia 30188

Re: K223024

Trade/Device Name: MedCAD® AccuPlan® System Regulation Number: 21 CFR 872.4120 Regulation Name: Bone cutting instrument and accessories Regulatory Class: Class II Product Code: DZJ, LLZ Dated: February 6, 2023 Received: February 7, 2023

Dear Linda Braddon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

## Bobak Shirmohammadi -S

For Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K223024

Device Name MedCAD® AccuPlan® System

#### Indications for Use (Describe)

The MedCAD® AccuPlan® System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the MedCAD® AccuPlan® System and the result is an output data file that may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces physical outputs including anatomical models, surgical guides, and dental splints for use in maxillofacial surgery. The surgical guides and dental splints are intended to be used for the maxillofacial bone in maxillofacial surgery. The MedCAD® AccuPlan® System is also intended as a pre-operative software tool for simulating / evaluating surgical treatment options.

Type of Use (Select one or both, as applicable)

| <span style="font-family: sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for MedCAD. The logo is composed of the word "Med" in a serif font, followed by "CAD" in a sans-serif font. The "CAD" portion of the logo is in a teal color and is slightly offset from the "Med" portion.

#### K223024 510(k) Summary MedCAD® AccuPlan® System

#### March 7, 2023

#### Sponsor

MedCAD 501 S 2nd Ave, Suite A-1000 Dallas, TX 75226 (214) 453-8864 x305

#### Contacts

Secure BioMed Evaluations Linda Braddon, Ph.D. Justin Gracyalny, MSE 7828 Hickory Flat Highway Suite 120 Woodstock, GA 30188 770-837-2681 Regulatory(@SecureBME.com

#### Name of Device and Classification Name

Device Name: MedCAD® AccuPlan® System Regulation Name: Bone Cutting Instruments and Accessories Regulation Number: 872.4120 Product Code: Primary - DZJ; Secondary - LLZ Classification Panel: Dental

#### Predicate Device(s)

MedCAD® AccuPlan® System (K192282)

#### Reference Device(s)

KLS Martin Individual Patient Solutions (IPS) Planning System (K181241) MedCAD® AccuShape® Titanium Patient-Specific Cranial Plate (K220357)

#### Indication for Use:

The MedCAD® AccuPlan® System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the MedCAD® AccuPlan® System and the result is an output data file that may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces physical outputs including anatomical models, surgical guides, and dental splints for use in maxillofacial surgery. The surgical guides and dental splints are intended to be used for the maxillofacial bone in maxillofacial surgery. The MedCAD® AccuPlan® System is also intended as a pre-operative software tool for simulating /evaluating surgical treatment options.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for MedCAD. The word "Med" is in a serif font and is black. The word "CAD" is in a sans-serif font, is white, and is on a blue-green background. There is a registered trademark symbol in the upper right corner of the blue-green background.

#### Device Description

The MedCAD® AccuPlan® System is a collection of software and associated additive manufacturing equipment intended to provide a variety of outputs to support orthognathic or reconstructive surgery. The system uses electronic medical images of the patient's anatomy or stone castings made from the patient anatomy with input from the physician, to manipulate original patient images for planning and executing surgery. The patient specific outputs from the system includes anatomical models, dental splints, surgical guides, and patient-specific case reports.

Following the MedCAD® Quality System and specific Work Instructions, trained employees utilize Commercial Off-The-Shelf (COTS) software to manipulate 3-D medical scan images which can include Computed Tomography (CT), Cone Beam CT (CBCT), and/or 3-D scan images from patient physical models (stone models of the patient's teeth) to create patient-specific physical and digital outputs. The process requires clinical input and review from the physician during planning and prior to delivery of the final outputs. While the process and dataflow vary somewhat based on the requirements of a given patient and physician, the following description outlines the functions of key sub-components of the system, and how they interact to produce the defined system outputs. It should be noted that the system is operated only by trained MedCAD employees, and the physician does not directly input information. The physician provides input for model manipulation and interactive feedback through viewing of digital models of system outputs that are modified by the engineer during the planning session.

The MedCAD® AccuPlan® System is made up of 4 individual pieces of software for the design and various manufacturing equipment integrated to provide a range of anatomical models (physical and digital), dental splints, surgical guides, and patient-specific planning reports for reconstructive surgery in the maxillofacial region.

The MedCAD® AccuPlan® System requires an input 3-D image file from medical imaging systems (i.e. - CT) and/or implant file. This input is then used, with support from the prescribing physician to provide the following potential outputs to support reconstructive surgery. Each system output is designed with physician input and reviewed by the physician prior to finalization. All outputs are used only with direct physician involvement to reduce the criticality of the outputs.

System outputs include:

- Anatomical Models ●
- Surgical Guides
- Dental Splints ●
- Patient-Specific Case Reports .

The purpose of this submission was to add titanium cutting / drilling guides to the family of available patient specific outputs. Cutting and drilling instruments can only be used with titanium cutting / drilling guides. Polymer guides are to be used for marking and positioning of anatomy only.

#### Sterilization Validation

Sterlization validation was conducted in accordance with international standard ISO 17665 and FDA guidance document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling." to a sterility Assurance Level (SAL) of 1x10". All test method acceptance criteria were met.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for MedCAD. The text "Med" is in a large, bold, black serif font. To the right of "Med" is "CAD" in a smaller, white sans-serif font, set against a teal, rectangular background. A registered trademark symbol is in the upper right corner of the teal rectangle.

#### Biocompatibility Validation

Biocompatibility validation was conducted in accordance with international standard ISO 10993-1 and FDA guidance document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The results of the testing adequately address biocompatibility for the output devices and their intended use.

#### Performance Testing

Performance testing for the MedCAD® AccuPlan® System is summarized in the table below:

| Test                       | Test Method Summary                                                                                                                                                                                                                                                                          | Results                                                                                                                      |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Wear Debris<br>Testing     | Cutting / drilling instruments were used on a<br>worst-case titanium surgical guide. The quantity<br>of wear debris generation was characterized.                                                                                                                                            | PASS<br>The wear debris<br>generated by the subject<br>device is less than that<br>reported in the literature<br>to be safe. |
| Fit and Form<br>Validation | Subject devices from historical cases were<br>manufactured. The manufactured devices were<br>optically scanned to verify alignment with the<br>3D model. Evaluation of fit was also validated<br>by fitting the guide over the corresponding<br>defect in a representative anatomical model. | PASS<br>All samples met the<br>predetermined<br>acceptance criteria.                                                         |

#### Substantial Equivalence

MedCAD® AccuPlan® System is substantially equivalent to the identified predicate based on indications for use, principles of operation, technological characteristics, inputs, and outputs. Minor differences in the surgical planning and manufacturing processes are verified and validated in the performance data in accordance with the intended use.

The input is images from medical scanners, dental models, and/or implants files. Physical outputs include dental splints and surgical guides. Biocompatible materials are used in the subject devices.

All devices are intended to aid in maxillofacial surgeries. These systems are intended to be utilized by trained employees with the approval by the physician.

{6}------------------------------------------------

# Med CAD

|                                                             | COMPARISON OF TECHNOLOGICAL CHARACTERISITICS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|-------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                             | MedCAD® AccuPlan® System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | MedCAD® AccuPlan® System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| 510(k) Number                                               | K223024                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | K192282                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Indications for Use                                         | The MedCAD® AccuPlan® System is<br>intended for use as a software system<br>and image segmentation system for the<br>transfer of imaging information from a<br>medical scanner such as a CT based<br>system. The input data file is processed<br>by the MedCAD® AccuPlan® System<br>and the result is an output data file that<br>may then be provided as digital models<br>or used as input to a rapid prototyping<br>portion of the system that produces<br>physical outputs including anatomical<br>models, surgical guides, and dental<br>splints for use in maxillofacial surgery.<br>The surgical guides and dental splints<br>are intended to be used for the<br>maxillofacial bone in maxillofacial<br>surgery. The MedCAD® AccuPlan®<br>System is also intended as a pre-<br>operative software tool for simulating /<br>evaluating surgical treatment options. | The MedCAD® AccuPlan® System is<br>intended for use as a software system<br>and image segmentation system for the<br>transfer of imaging information from a<br>medical scanner such as a CT based<br>system. The input data file is processed<br>by the MedCAD® AccuPlan® System<br>and the result is an output data file that<br>may then be provided as digital models<br>or used as input to a rapid prototyping<br>portion of the system that produces<br>physical outputs including anatomical<br>models, surgical guides, and dental<br>splints for use in maxillofacial surgery.<br>The surgical guides and dental splints<br>are intended to be used for the<br>maxillofacial bone in maxillofacial<br>surgery. The MedCAD® AccuPlan®<br>System is also intended as a pre-<br>operative software tool for simulating /<br>evaluating surgical treatment options. |
| Preoperative software                                       | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Additive manufacturing<br>of splints, guides, and<br>models | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Data inputs                                                 | Images from medical scanners                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Images from medical scanners                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Data outputs                                                | Output for dental splints, surgical<br>guides, and anatomical models                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Output for dental splints, surgical<br>guides, and anatomical models                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Physical outputs                                            | Dental splints, surgical guides,<br>anatomical models, and patient-specific<br>case reports                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Dental splints, surgical guides,<br>anatomical models, and patient-specific<br>case reports                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Materials                                                   | Biocompatible polymers and Ti-6Al-4V<br>ELI Titanium Alloy per ASTM F136                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Biocompatible polymers                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Sterilization                                               | Provided non-sterile and is steam<br>sterilized by the end-user                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Provided non-sterile and is steam<br>sterilized by the end-user                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Manufacturing Method                                        | Additive Manufacturing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Additive Manufacturing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Patient Contact                                             | Surgical Guides: Implant - Tissue /<br>Bone; Limited (< 24 hours)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Surgical Guides: Implant - Tissue /<br>Bone; Limited (< 24 hours)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                                             | Splint - Intermediate: Surface<br>Contacting - Mucosal Membrane;<br>Limited (< 24 hours)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Splint - Intermediate: Surface<br>Contacting - Mucosal Membrane;<br>Limited (< 24 hours)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                                             | Splint — Final: Externally<br>Communicating - Tissue / Bone /<br>Dentin: Limited (< 24 hours)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Splint — Final: Externally<br>Communicating - Tissue / Bone /<br>Dentin: Limited (< 24 hours)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |

{7}------------------------------------------------

#### Comparison of Technological Characteristics with the Predicate Device

MedCAD® AccuPlan® system is substantially equivalent to and it is safe and effective as its predicate device (MedCAD® AccuPlan® system - K192282).

#### Similarities to Predicate

The subject device is identical to that cleared in K192282 with the exception that the subject device also includes titanium cutting and drilling surgical guides. The anatomical models, splints, and positioning surgical guides are identical to those cleared in K192282.

The MedCAD® AccuPlan® System has the same intended use and similar technological characteristics as the identified predicate device. The system employs similar fundamental technologies as the identified predicates including software image transfer, manipulation, and surgical planning. The principals of operation and technological characteristics are either identical or substantially equivalent to the predicate. The system has similar technological characteristics including:

- System inputs: Images from medical scanners (ex: CT), dental models, and/or implant files (.STL)
- System outputs: Physical and/ or Digital outputs such as patient-specific anatomical models, ● marking and positioning guides, and splints
- Materials: Biocompatible polymers and Ti-6A1-4V ELI Titanium Alloy per ASTM F136 ●
- Sterility assurance level of 1x10-6 ●

The intended use of the subject device and the predicate both provide tools and accessories (software for image manipulation, anatomical models, splints, and guides) for use in reconstructive surgery, Additionally, both the subject and the predicate device are intended to be used by trained personnel with active support from the surgeon.

#### Differences to Predicate

The subject device is identical to that cleared in K192282 with the exception that the subject device also includes titanium cutting and drilling surgical guides. The inclusion of these components within the subject device system is similar to those cleared in the reference device (K181241). Safety and performance testing demonstrate that the addition of these components does not raise new questions for substantial equivalence.

#### Conclusion

The MedCAD® AccuPlan® System is substantially equivalent to its predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/DZJ/K223024](https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/DZJ/K223024)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
