GEISTLICH BIO-OSS, GEISTLICH BIO-OSS COLLAGEN, GEISTLICH COMBI-KIT COLLAGEN, GEISTLICH PERIO SYSTEM COMBI PACK, GEISTLIC

K122894 · Geistlich Pharma AG · NPM · Feb 15, 2013 · Dental

Device Facts

Record IDK122894
Device NameGEISTLICH BIO-OSS, GEISTLICH BIO-OSS COLLAGEN, GEISTLICH COMBI-KIT COLLAGEN, GEISTLICH PERIO SYSTEM COMBI PACK, GEISTLIC
ApplicantGeistlich Pharma AG
Product CodeNPM · Dental
Decision DateFeb 15, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3930
Device ClassClass 2
AttributesTherapeutic

Intended Use

Geistlich Bio-Oss® is intended for the following uses: - Augmentation or reconstructive treatment of the alveolar ridge; - Filling of infrabony periodontal defects; - Filling of defects after root resection, apicoectomy, and cystectomy; - Filling of extraction sockets to enhance preservation of the alveolar ridge; - Elevation of the maxillary sinus floor; - Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR); and - Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR).

Device Story

Geistlich Bio-Oss is a bovine-derived, porous bone mineral matrix used as a bone grafting material. It features interconnected macro- and micropores, providing a hydrophilic scaffold for bone regeneration. The device is supplied as granules or blocks, sometimes combined with 10% porcine collagen to improve handling and defect adaptation. Used by clinicians in maxillofacial, implant, and periodontal surgeries to fill bony voids, gaps, or defects. The material acts as a physical matrix to support bone growth; the collagen component is resorbed over time. It is provided sterile for single-use application.

Clinical Evidence

Bench testing only. No clinical data presented. Equivalence supported by comparative analytical testing (XRD, FTIR, Mercury Intrusion Porosimetry) between standard and alternative source materials.

Technological Characteristics

Bovine-derived crystalline hydroxyapatite bone mineral matrix. Highly porous with interconnected macro- and micropores. Geistlich Bio-Oss Collagen includes 10% porcine collagen. Sterilized via gamma-irradiation. Supplied in double sterile barrier packaging (vials/blisters).

Indications for Use

Indicated for patients requiring bone grafting in maxillofacial surgery, implantology, and periodontology for alveolar ridge augmentation, periodontal defect filling, root resection/cystectomy defect filling, extraction socket preservation, maxillary sinus floor elevation, and GTR/GBR procedures.

Regulatory Classification

Identification

Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.

Special Controls

*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### 510(k) Summary FEB 1 5 2013 # GEISTLICH BIO-OSS® #### SPONSOR Geistlich Pharma AG Bahnhofstrasse 40 CH-6110 Wolhusen Switzerland Contact Person: E-mail: Phone: Fax: Date Prepared: Daniel Kracov, Arnold & Porter, LLP Counsel to Geistlich Pharma AG Daniel.Kracov@aporter.com 202-942-5120 202-942-5999 February 13, 2013 #### DEVICE NAME | Proprietary Name: | Geistlich Bio-Oss® | |----------------------|---------------------------------------------| | Common/Usual Names: | Bone Grafting Material | | Classification Name: | Bone grafting material, animal source (NPM) | # PREDICATE DEVICES Geistlich Bio-Oss® and Geistlich Bio-Oss® Blocks (K112572, K033815, K970321, K952618) ### DEVICE DESCRIPTION Geistlich Bio-Oss® spongiosa (cancellous) granules and Geistlich Bio-Oss® spongiosa (cancellous) Block (referred as Geistlich Bio-Oss®) are intended for filling and augmentation of bony voids and gaps in maxillofacial surgery, implantology, and periodontology according to the intended use of the product. Geistlich Bio-Oss® is serving as a matrix consisting of interconnected macro- and micropores. The material is highly porous and has a large inner surface area. Geistlich Bio-Oss® is a biocompatible bone mineral matrix and is manufactured from bovine bone in a validated multistage purification process to remove the organic components. Due to the interconnected macro and microporous system the device is hydrophilic and easv to moisten. {1}------------------------------------------------ Geistlich Bio-Oss® is packed in a double sterile barrier system consisting of a glass vial and an outer blister. Geistlich Bio-Oss is sterilized according to validated processes by y-irradiation. It is available in the following sizes and amounts: | Product | Weight | Particle Size | |----------------------------------------------------|--------|---------------| | Geistlich Bio-Oss® spongiosa (cancellous) granules | 0.25 g | 0.25 - 1.0 mm | | Geistlich Bio-Oss® spongiosa (cancellous) granules | 0.5 g | 0.25 - 1.0 mm | | Geistlich Bio-Oss® spongiosa (cancellous) granules | 2.0 g | 0.25 - 1.0 mm | | Geistlich Bio-Oss® spongiosa (cancellous) granules | 5.0 g | 0.25 - 1.0 mm | | Geistlich Bio-Oss® spongiosa (cancellous) granules | 0.5 g | 1.0 - 2.0 mm | | Geistlich Bio-Oss® spongiosa (cancellous) granules | 2.0 g | 1.0 - 2.0 mm | Geistlich Bio-Oss® spongiosa (cancellous) block 1 x 1 x 2 cm (approx.) ### INTENDED USE Geistlich Bio-Oss® is intended for the following uses: - Augmentation or reconstructive treatment of the alveolar ridge; - Filling of infrabony periodontal defects; - Filling of defects after root resection, apicoectomy, and cystectomy; - Filling of extraction sockets to enhance preservation of the alveolar ridge; - Elevation of the maxillary sinus floor; - Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR); and - Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR). ### SUMMARY OF TECHNOLOGICAL CHARACTERISTICS Appearance: white granulate or white, porous block Moisture: less than 5% (w./w.) Calcium: 35%-40% (w.lw.) Phosphorous: 13.5% - 18.5% (w./w.) #### SUMMARY OF DATA TO SUPPORT SUBSTANTIAL EQUIVALENCE Geistlich Bio-Oss® from standard manufacturing lots was compared with Geistlich Bio-Oss® manufactured with the raw material from the alternative geographic source by means of X-ray diffraction analysis, Fourier Transform Infrared Spectroscopy and Mercury Intrusion Porosimetry. The X-ray spectra of Geistlich Bio-Oss® manufactured with bovine bone from the alternative geographic source did not reveal any differences and confirmed that Geistlich Bio-Oss® is crystalline hydroxyapatite with no detectable impurities. {2}------------------------------------------------ Fourier Transform Infrared Spectroscopy data was generated to determine the molecule structure characteristics of Geistlich Bio-Oss®. The data from regular manufacturing lots were equal to Geistlich Bio-Oss® manufactured from bovine bone of the alternative geographic source. Mercury Intrusion Porosimetry revealed that porosity, specific pore surface area, bulk density and bimodal pore size distribution demonstrates that the micro and macro pores of Geistlich Bio-Oss® from regular manufacturing lots are identical to Geistlich Bio-Oss® manufactured from bovine bone of the alternative geographic source. # CONCLUSION Results of the bench tests support that Geistlich Bio-Oss® is Substantially Equivalent to the identified predicate devices. {3}------------------------------------------------ ### 510(k) Summary # GEISTLICH BIO-OSS COLLAGEN® ### SPONSOR Geistlich Pharma AG Bahnhofstrasse 40 CH-6110 Wolhusen Switzerland | Contact Person: | Daniel Kracov, Arnold & Porter, LLP<br>Counsel to Geistlich Pharma AG | |-----------------|-----------------------------------------------------------------------| | E-mail: | Daniel.Kracov@aporter.com | | Phone: | 202-942-5120 | | Fax: | 202-942-5999 | | Date Prepared: | February 13, 2013 | #### DEVICE NAME | Proprietary Name: | Geistlich Bio-Oss Collagen® | |----------------------|---------------------------------------------| | Common/Usual Names: | Bone Grafting Material Plus Collagen | | Classification Name: | Bone grafting material, animal source (NPM) | ### PREDICATE DEVICES Geistlich Bio-Oss Collagen® (K112572, K092428, K033815, K974399) ### DEVICE DESCRIPTION GBOC Geistlich Bio-Oss Collagen® is a combination of purified spongiosa (cancellous) bone mineral granules (Geistlich Bio-Oss®) and 10% collagen fibers in a block form. The collagen facilitates handling and application of Geistlich Bio-Oss Collagen® and acts to hold the Bio-Oss particles at the desired place. The consistency of this material readily allows to take the shape of the defect. The collagen component is resorbed after application. Geistlich Bio-Oss Collagen® is intended for filling and augmentation of bony voids and gaps in maxillofacial surgery, implantology, and periodontology according to the intended use of the product. Geistlich Bio-Oss Collagen® is serving as a matrix consisting of interconnected macroand micropores. The material is highly porous and has a large inner surface area. 60123676v1 {4}------------------------------------------------ Geistlich Bio-Oss Collagen®is a biocompatible bone mineral matrix and is manufactured from boyine bone and collagen from connective tissue of pigs certified for human consumption according to a controlled and validated multistage purification process. From the bovine bone all organic components are removed. Due to the interconnected macro and microporous system the device is hydrophilic and easy to moisten. Geistlich Bio-Oss Collagen® is packed in a double sterile barrier system consisting of an inner and an outer blister. Geistlich Bio-Oss® is sterilized according to validated processes by y-irradiation. It is available as part of two convenience kits and different sizes: Single Products: - 1 block 100 mg Geistlich Bio-Oss® spongiosa (cancellous) granules and 10% collagen . - 1 block 250 mg Geistlich Bio-Oss® spongiosa (cancellous) granules and 10% collagen . - 1block 500 mg Geistlich Bio-Oss® spongiosa (cancellous) granules and 10% collagen . Convenience Kits: - Geistlich Combi-Kit Collagen® . - 1 block 100 mg Geistlich Bio-Oss® spongiosa (cancellous) granules and 10% collagen, and 1 Geistlich Bio-Gide® membrane, 16 x 22 mm - Geistlich Perio System Combi-Pack . 1 block 100 mg Geistlich Bio-Oss® spongiosa (cancellous) granules and 10% collagen, and 1 Geistlich Bio-Gide® Perio membrane, 16 x 22 mm #### INTENDED USE Geistlich Bio-Oss Collagen® is intended for the following uses: - Augmentation or reconstructive treatment of the alveolar ridge; - Filling of periodontal defects; - Filling of defects after root resection, apicoectomy, and cystectomy: - Filling of extraction sockets to enhance preservation of the alveolar ridge: - Elevation of the maxillary sinus floor; - Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR); and - Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR). # SUMMARY OF TECHNOLOGICAL CHARACTERISTICS Appearance: white sponge-like, hard pieces of customized size 60123676v1 {5}------------------------------------------------ Water: less than 8% (w./w.) Calcium: 38%-42% (w./w.) Phosphorous: 12.5% - 17.5% (w./w.) # SUMMARY OF DATA TO SUPPORT SUBSTANTIAL EQUIVALENCE Geistlich Bio-Oss® from standard manufacturing lots was compared with Geistlich Bio-Oss manufactured with the raw material from the alternative geographic source by means of X-ray diffraction analysis, Fourier Transform Infrared Spectroscopy and Mercury Intrusion Porosimetry. The X-ray spectra of Geistlich Bio-Oss® manufactured from bovine bone of the alternative geographic source did not reveal any differences and confirmed that Geistlich Bio-Oss® is crystalline hydroxyapatite with no detectable impurities. Fourier Transform Infrared Spectroscopy data was generated to determine the crystal structure, crystal size and molecule structure characteristics of Geistlich Bio-Oss® The data from reqular manufacturing lots were equal to Geistlich Bio-Oss® manufactured from bovine bone of the alternative geographic source. Mercury Intrusion Porosimetry revealed that porosity, specific pore surface area, bulk density and bimodal pore size of Geistlich Bio-Oss® from regular manufacturing lots are identical to Geistlich Bio-Oss® manufactured from bovine bone of the alternative geographic source. #### CONCLUSION Geistlich Bio-Oss Collagen® is Substantially Equivalent to the identified predicate devices. {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a bird-like figure with three curved lines representing wings or streams. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 15, 2013 Geistlich Pharma AG C/O Mr. Daniel A. Kracov Arnold & Porter, Limited Liability Partnership 555 Twelfth Street, North West WASHINGTON DC 20004-1206 Re: K122894 Trade/Device Name: Geistlich Bio-Oss® and Geistlich Bio-Oss Collagen® Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: NPM Dated: January 31, 2013 Received: February 1, 2013 Dear Mr. Kracov: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {7}------------------------------------------------ Page 2 - Mr. Kracov Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Mary S. Runner Susan Runner, DDS, MA 2013.02.21 13:27:34 -05'00' Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ # Indications for Use K122894 510(k) Number (if known): (not yet known) Device Name: Geistlich Bio-Oss® Indications For Use: Geistlich Bio-Oss® is intended for the following uses: - Augmentation or reconstructive treatment of the alveolar ridge; - Filling of infrabony periodontal defects; - Filling of defects after root resection, apicoectomy, and cystectomy; - Filling of extraction sockets to enhance preservation of the alveolar ridge; - Elevation of the maxillary sinus floor; - Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR); and - Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR). AND/OR Prescription Use × CFR 801 Subpart D) (21 CFR 801 Subpart C) Over-The-Counter Use_ (Part 21 (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------# [2013:02.15 09:45:55 -05'00' ADDAIA Page 1 of _ 2 (Division Sign-Off) (Division Sign-On) Division of Anesthesiology, General Hospital Division Control, Dental Devices 510(k) Number; ---
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...