TENDER FIBER

K063170 · Micerium S.P.A. · EBI · Jan 12, 2007 · Dental

Device Facts

Record IDK063170
Device NameTENDER FIBER
ApplicantMicerium S.P.A.
Product CodeEBI · Dental
Decision DateJan 12, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3760
Device ClassClass 2
AttributesTherapeutic

Intended Use

Tender fiber is a semi-manufactured product made of fibers used to reinforce composite and acrylic prosthodontic appliances and to repair or reinforce denture and orthodontic appliances. It can be use as well to immobilize teeth for periodontal or orthodontic treatments. These products are for use only by trained and experienced dentists and dental technicians and have become well-known in this area of dentistry, therefore they do not require prescription (Per 21 CFR 801 Subpart D)

Device Story

Tender Fiber consists of silanized glass fibers impregnated with light-curing resin. Used by dentists and dental technicians in clinical or laboratory settings to reinforce prosthodontic, denture, and orthodontic appliances, or to splint/immobilize teeth. The material is applied to the appliance or tooth structure and cured using external dental halogen, plasma, or LED light systems. The light triggers initiators within the resin, causing polymerization and hardening of the fiber reinforcement. This process provides structural support to dental appliances or stabilizes mobile teeth, potentially improving clinical outcomes and appliance durability.

Clinical Evidence

No clinical data provided. Evidence consists of bench testing and cytotoxicity testing performed in accordance with ISO 10993-5 and ISO 10993-12 standards.

Technological Characteristics

Silanized glass fibers impregnated with light-curing resin. Polymerization initiated by dental halogen, plasma, or LED light systems. Complies with ISO 4049:2000 (Polymer-based filling, restorative and luting materials) and ISO 13485:2003 (Quality management).

Indications for Use

Indicated for reinforcement of composite and acrylic prosthodontic appliances, repair or reinforcement of denture and orthodontic appliances, and immobilization of teeth for periodontal or orthodontic treatments in patients requiring such dental interventions.

Regulatory Classification

Identification

A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K063170 JAN 1 2 2007 ## CONFIDENTIAL ### 510(k) Summary Submitter: Mr. Eugenio Miceli, QA Manager, Micerium SpA, Via Marconi, 83, 16030 Avegno (GE), Italy. Phone: 39 0185 7885 880. - I. Classification Name and Number: Resin, denture, relining, rebasing - 872.3760 (Product Code 76EBI) . Class: II - II. Common/Usual Name: Fiber reinforcement material - III. Proprietary Name: Tender Fiber Ortho, Tender Fiber Due, Tender Fiber Quattro. - IV. Registration No .: Foreign, in process XI. - V. Compliance with Performance Standards: "ISO 4049:2000 Dentistry - Polymer - Based Filling, Restorative and Luting materials". "ISO 13485:2003 Medical devices -- Quality management systems -- Requirements for regulatory purposes" For the Citotoxicity tests: "ISO 10993 -1:2003 -- Evaluation and testing" "ISO 10993 -5:1999--Tests for in vitro citotoxicity" "ISO 10993-12:2002-- Sample preparation and reference materials" For risk analysis "ISO 14971:2000/Amd 1:2003 - MEDICAL DEVICES - Application of risk management to medical devices." VI. Premarket Notification truthful and accurate statement VII. Description of the Device: > Tender Fiber are silanized glass fibers, impregnated with light curing resin used to reinforce prosthodontic appliances and to repair or reinforce denture and orthodontic appliances. - Labels and Labeling: Draft labels of the Tender Fiber and instructions for use are VIII. provided. {1}------------------------------------------------ ### CONFIDENTIAL #### IX. Substantial Equivalence: The Tender fiber are substantially equivalent to: - Everstick by Stick Tech Ltd., 510 (k) no .: K011788 । - Everstick Perio by Stick Tech Ltd., 510 (k) no .: K030072 - - Everstick Ortho by Stick Tech Lid., 510 (k) no .: K021126 ı - Stick by Stick Tech Ltd., 510(k) no .: K003333 ・ - Ribbond (TM) by Ribbond Inc, 510(k) no .: K913040 ı 1. These products have the same intended use as predicate devices. 2. Technological characteristics: Ribbond an ultra high modulus polyethylene fiber ribbon with no impregnation, while Stick is a polymer pre-impregnated unidirectional continuous glass fiber. Everstick, Everstick Perio, Everstick Ortho and Tender fiber have the same technological characteristics of Stick but they are also impregnated with light curing resin. Curing is made with help of dental halogen, plasma or LED light systems. The curing process is triggered by initiators, which initiate the polymerization. These initiators are activated by certain wavelengths of the different light systems and are similar or exactly the same in all listed materials. Test results indicate that there are no hazards presented with the use of Tender Fiber as compared with the predicate devices. The different technological characteristics of Tender Fiber (practically the same of Everstick, Everstick Perio, Everstick Ortho) compared with Stick and Ribbond does not raise new questions of safety and effectiveness and demonstrates that the device is safe and effective as the predicate devices. #### IX.1 Risk to Health. Potential adverse affects and complications common to glass fibers impregnated with a light curing resin include: - Allergies or hypersensivities - No adhesion or incomplete adhesion to teeth - incorrect or incomplete curing Cytotoxicity tests appears in appendix II.1 - X. Indications for Use. Tender fiber is a semi-manufactured product made of fibers used to reinforce composite and acrylic prosthodontic appliances and to repair or reinforce denture and orthodontic appliances. It can be use as well to immobilize teeth for periodontal or orthodontic treatments. (End of Summary) {2}------------------------------------------------ Image /page/2/Picture/1 description: The image is a circular seal for the Department of Health & Human Services, USA. The seal features the department's emblem, which is a stylized caduceus with a human figure in the center. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged around the emblem in a circular fashion. The emblem is black and the background is white. ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Eugenio Miceli Quality Assurance Manager Micerium S.P.A. Via Marconi 83 Avegno, Italy 16030 JAN 1 2 2007 Re: K063170 Trade/Device Name: Tender Fiber Regulation Number: 872.3760 Regulation Name: Denture Relining, Repairing or Rebasing Resin Regulatory Class: II Product Code: EBI Dated: October 12, 2006 Received: October 18, 2006 Dear Mr. Miceli: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Miceli Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies, You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Smitte y. Michou m.D. Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### CONFIDENTIAL # X. Indications for Use [Separate Page] 510(k) Number: (not assigned) 106 3170 Device Name: Tender Fiber Indications for Use: Tender fiber is a semi-manufactured product made of fibers used to reinforce composite and acrylic prosthodontic appliances and to repair or reinforce denture and rethodontic appliances. It can be use as well to immobilize teeth for periodontic or orthodontic treatments. These products are for use only by trained and experienced dentists and dental technicians and have become well-known in this area of dentistry, therefore they do not require prescription (Per 21 CFR 801 Subpart D) Susan Quarre 100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000 (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED ### Concurrence of CDRH, Office of Device Evaluation (ODE) File: tender fiber_510K.doc pag. 11/ 15
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...