Any-Com Bulk
K242713 · Mediclus Co., Ltd. · EBF · Feb 7, 2025 · Dental
Device Facts
| Record ID | K242713 |
| Device Name | Any-Com Bulk |
| Applicant | Mediclus Co., Ltd. |
| Product Code | EBF · Dental |
| Decision Date | Feb 7, 2025 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 872.3690 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
Base under Class I and Class II direct restorations Liner under direct restorative materials Restoration of small cavities Class III and V restorations Core build-ups
Device Story
Any-Com Bulk is a light-cured, radiopaque, flowable resin composite restorative material. Used by dentists in clinical settings for anterior and posterior restorations. Applied as a base-liner or restorative material in cavities up to 4mm deep; cured in a single increment using a dental curing light. After curing, the material is finished and polished with conventional instruments. Provides structural support and aesthetic tooth-matching via five available shades (A1, A2, A3, TL, Blue). Benefits include simplified placement technique and restoration of tooth structure.
Clinical Evidence
Bench testing only. Performance evaluated per ISO 4049, including appearance, weight, packaging, sensitivity to ambient light, curing depth (3.5 mm), flexural strength (100 MPa), water sorption (30 µg/mm³), solubility (0.6 µg/mm³), radiopacity (>2.9 mm), and color stability. Biocompatibility testing conducted per ISO 10993 series (cytotoxicity, sensitization, irritation/intracutaneous reactivity, acute systemic toxicity).
Technological Characteristics
Light-cured, flowable resin composite. Composition: Bis-GMA, UDMA, TEDMA, Ba-Glass, Camphorquinone. Inorganic filler loading ~66% by weight; particle size 0.1–2.0 µm. Form factor: Syringe delivery. Complies with ISO 4049 for dental restorative materials. Biocompatible per ISO 10993.
Indications for Use
Indicated for patients requiring dental restorations, specifically as a base or liner for Class I and II direct restorations, restoration of small cavities, Class III and V restorations, and core build-ups.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
Predicate Devices
- 3M Filtek Bulk Fill Flowable Restorative (K120453)
Related Devices
- K203598 — Omnichroma Flow Bulk · Tokuyama Dental Corporation · Feb 3, 2021
- K161353 — ESTELITE BULK FILL Flow · Tokuyama Dental Corporation · Apr 3, 2017
- K152471 — CAPO BULK FILL · Schuetz Dental GmbH · Aug 4, 2016
- K220605 — Venus Bulk Flow ONE · Kulzer, LLC · Apr 21, 2022
- K250823 — Any-Com Flow · Mediclus Co., Ltd. · Apr 14, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
February 7, 2025
Mediclus Co., Ltd. % Priscilla Chung Official Correspondent LK Consulting Group USA, Inc. 18881 Von Karman Ave STE 160 Irvine, California 92612
Re: K242713
Trade/Device Name: Any-Com Bulk Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF Dated: January 8, 2025 Received: January 8, 2025
Dear Priscilla Chung:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bobak
Shirmohammadi -S
For Michael E. Adjodha, M.ChE., RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
#### K242713
Device Name
Any-Com Bulk
Indications for Use (Describe)
- Base under Class I and Class II direct restorations
- Liner under direct restorative materials
- Restoration of small cavities
- Class III and V restorations
- Core build-ups
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary (K242713)
This summary of 510(K) information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: 1/7/2025
## 1. Submitter
MEDICLUS Co., Ltd. No. 1210, 134, Gongdan-ro, Heungdeok-gu, Cheongju-si. Chungcheongbuk-do. Republic of Korea TEL : +82(43)211-2877 FAX : +82(43)211-2866
## 2. U.S Agent/Contact Person
Priscilla Chung LK Consulting Group USA, Inc. 18881 Von Karman Ave STE 160, Irvine CA 92612 Fax: 714-409-3357 Phone: 714.202.5789 Email: juhee.c(@lkconsultinggroup.com
## 3. Device
- Trade Name: Any-Com Bulk
- Common Name: Tooth Shade Resin
- Classification Name: Tooth Shade Resin Material ●
- Product Code: EBF ●
- Classification regulation: 21 CFR 872.3690
### 4. Predicate Device:
3M Filtek Bulk Fill Flowable Restorative by 3M ESPE (K120453) (Primary predicate)
### 5. Description:
Any-Com Bulk is a light cured. radiopaque, flowable restorative material for both posterior and anterior restorations. It is designed to be used as a base-liner for Class I and II restorations. The product allows a technique in which a cavity up to 4mm in depth can be filled and cured in a single increment.
The product offers 5 shades options available. Users can select shades such as A 1, A2, A3, TL and Blue, based on the desired brightness, yellowness, and darkness that match the
{5}------------------------------------------------
patient's natural teeth.
After the appropriate tooth pretreatment process, fill the appropriate amount of product to applying area. Once light cured, the composite can be finished and polished using conventional finishing and polishing instruments.
#### 6. Indication for use:
- Base under Class I and Class II direct restorations
- Liner under direct restorative materials
- " Restoration of small cavities
- . Class III and V restorations
- . Core build-ups
#### 7. Basis for Substantial Equivalence
| | Subject Device | Predicate Device | Equivalence<br>evaluation | |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| 510K# | K242713 | K120453 | | |
| Manufacturer | MEDICLUS Co., Ltd.<br>Any-Com Bulk | 3M ESPE | | |
| Product Name | Image: Any-Com Bulk | 3M Filtek Bulk Fill Flowable<br>Restorative<br>Image: 3M Filtek Bulk Fill Flowable Restorative | | |
| 510k# | - | K120453 | | |
| Product Code | EBF | EBF | | |
| Material | Bis-GMA<br>UDMA<br>TEDMA<br>Ba-Glass<br>Camphorquinone<br>• Inorganic filler loading is<br>approximately 66% by weight<br>• particles size : 0.1~2.0 μm | Bis-GMA<br>UDMA<br>BisEMA<br>Procrylat resins<br>• Inorganic filler loading is<br>approximately 64.5% by<br>weight<br>• Particle size : 0.1 to 5.0 μm | Similar | |
| Curing type | Light Curing | Light Curing | Same | |
| Indications for<br>Use Statement | • Base under Class I and Class II<br>direct restorations<br>• Liner under direct restorative<br>materials<br>• Restoration of small cavities<br>• Class III and V restorations<br>• Core build-ups | • Base under Class I and 11<br>direct restorations<br>• Liner under direct<br>restorative materials<br>• Pit and fissure sealant<br>• Restoration of minimally<br>invasive cavity<br>preparations (including | Same | |
| | | | small, non stress-bearing<br>occlusal restorations) Class<br>III and V restorations<br>Undercut blockout<br>• Repair of small enamel<br>defects Repair of small<br>defects in esthetic indirect<br>restorations Repair of resin<br>and acrylic temporary<br>materials<br>• As a core build-up where<br>at least half the coronal<br>tooth structure is<br>remaining to provide<br>structural support for the<br>crown | |
| | Standard | ISO 4049 | ISO 4049 | |
| Technological<br>Characteristics | Curing Depth | 3.5 mm | 4 mm | |
| | Flexural<br>strength | 100 Mpa | 126.5 Mpa | |
| | Water<br>Absorption | 30 $\mu g/mm^3$ | - | Same |
| | Solubility | 0.6 $\mu g/mm^3$ | - | |
| | Radiopacity | over-2.9 mm | 2.4 mm | |
| | Color and<br>color stability | All samples have<br>a matching shade<br>guide | - | |
| | Sensitivity to<br>ambient light | All samples are<br>physically<br>homogeneous | - | |
| Biocompatibility | Using ISO 10993 series<br>Biocompatible | | Using ISO 10993 series<br>Biocompatible | |
| | • Delivery System: Syringe | | • Delivery System: Syringe | |
| | • Weight: 2g | | • Weight: 2g | |
| Delivery method | • Units per Pack: 1, 2, 3 | | • Units per Pack: 2 | |
| | • Accessories: Tip | | • Accessories: Tip | |
| Shade | A1, A2, A3, Blue, TL | | A1, A2, A3, Universal | Same |
| Shelf-Life | 3 years | | 3 years | Same |
#### 7.1. Comparison Chart
510(k) summary
{6}------------------------------------------------
# 7.2. Comparison Chart
The subject device has the same indications for use and the technological characteristics as the predicate device. The minor raw materials are different between the devices but the performance and the biocompatibility test results show that it does not raise a concern in safety and effectiveness.
{7}------------------------------------------------
## 8. Non-Clinical Testing
- Performance Tests including Appearance, Weight, Packaging, Sensitivity to Ambient ● Light, Curing Depth, Flexural Strength, Water Sorption/Solubility, Color and Color stability, Radiopacity in accordance with ISO 4049
- . Biocompatibility Tests
| No | Test Item | Test Standard |
|----|-----------------------------------------|----------------------------------------|
| 1 | Cytotoxicity | ISO 10993-5 |
| 2 | Sensitization | ISO 10993-10 |
| 3 | Irritation or Intracutaneous Reactivity | ISO 10993-10 |
| 4 | Acute Systemic Toxicity | ISO 10993-11 |
| 5 | Material-Mediated Pyrogenicity | |
| 6 | Subacute/Subchronic Toxicity | |
| 7 | Genotoxicity | |
| 8 | Implantation | BIOLOGICAL SAFETY<br>ASSESSMENT REPORT |
| 9 | Chronic Toxicity | |
| 10 | Carcinogenicity | |
### 9. Conclusion
The subject device and the predicate device have the same intended use and have the same technological characteristics. Based on the similarities and the test results, we conclude that the subject device is substantially equivalent to the predicate device.