SonicFill 2

K143209 · Sybron Dental Specialties · EBF · Feb 5, 2015 · Dental

Device Facts

Record IDK143209
Device NameSonicFill 2
ApplicantSybron Dental Specialties
Product CodeEBF · Dental
Decision DateFeb 5, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3690
Device ClassClass 2
AttributesTherapeutic

Intended Use

SonicFill 2 is indicated for direct placement in all cavity classes in anterior and posterior teeth. Additional indications include: base/liner material, repair of enamel defects, repair of provisionals, repair of porcelain restorations, minor occlusal build-ups, pit and fissure sealant, luting of composite/ceramic veneers, core build-ups and incisal abrasions.

Device Story

SonicFill 2 is a light-cured, nano-hybrid, resin-based dental restorative material. Used with SonicFill 2010 Handpiece (K091091) for sonic activation during delivery. Sonic energy reduces material viscosity, allowing close adaptation to cavity walls; deactivation restores higher viscosity for sculpting. Enables bulk filling of cavities up to 5mm deep in single increment. Used by licensed dental professionals in clinical settings. Benefits include reduced shrinkage stress, high depth of cure, and efficient placement.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing including mechanical strength, polishability, water sorption/solubility, flexural strength, depth of cure, light sensitivity, radiopacity, volumetric shrinkage, shrinkage stress, color stability, and biocompatibility per ISO 10993 and ISO 4049 standards.

Technological Characteristics

Nano-hybrid, light-cured, resin-based dental restorative. Sonic-activated delivery via SonicFill 2010 Handpiece. Complies with ISO 4049 (dentistry polymer-based filling/restorative/luting materials) and ISO 10993 (biocompatibility).

Indications for Use

Indicated for direct placement in all cavity classes in anterior and posterior teeth; base/liner material; repair of enamel defects, provisionals, and porcelain restorations; minor occlusal build-ups; pit and fissure sealant; luting of composite/ceramic veneers; core build-ups; and incisal abrasions.

Regulatory Classification

Identification

Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of three human profiles facing right, arranged in a way that they appear to be interconnected. The profiles are rendered in black and have a flowing, abstract design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the profile graphic. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 5, 2015 Sybron Dental Specialties Ms. Kerri Casino Regulatory Affairs Manager 1717 West Collins Avenue Orange, California 92687 Re: K143209 Trade/Device Name: SonicFill 2 Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth shade resin material Regulatory Class: II Product Code: EBF, EBC Dated: November 7, 2014 Received: November 10, 2014 Dear Ms. Casino: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Tina Kiang -S for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K143209 Device Name SonicFill 2 ### Indications for Use (Describe) SonicFill 2 is indicated for direct placement in all cavity classes in anterior teeth. Additional indications include: base/liner material, repair of enamel defects, repair of povisionals, repair of porcelain restorations, minor occlusal build-ups, pit and fissure sealant, luting of composite/ceramic veneers, core build-ups and incisal abrasions. Type of Use (Select one or both, as applicable) | <span></span> Prescription Use (Part 21 CER 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) | |------------------------------------------------------------|-----------------------------------------------------------| |------------------------------------------------------------|-----------------------------------------------------------| ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) SUMMARY for SonicFill 2 Image /page/3/Picture/1 description: The image shows the logo for Sybron Dental Specialties. The logo features the letters 'sds' in a bold, sans-serif font, with a curved line above the letters. Below the letters, the words 'Sybron Dental Specialties' are written in a smaller, sans-serif font. The color of the text and curved line is a dark blue. ### SonicFill 2 - 1. Submitter Information: Sybron Dental Specialties 1717 W. Collins Ave. Orange CA, 92687 | Contact Person: | Kerri Casino | |-------------------|--------------| | Telephone Number: | 714-516-7634 | | Fax Number: | 714-516-7472 | January 16, 2015 Date Prepared: - 2. Device Name: - Proprietary Name: SonicFill 2 . ● - Tooth Shade Resin Material Classification Name: - 872.3690 CFR Number: ● - Device Class: ● II - . Product Code: EBF, EBC - 3. Predicate Device: SonicFill 2 is substantially equivalent to the legally marketed device Metamorphosis (K091023) cleared on May 21, 2009, product code EBF, EBC. - 4. Description of Device: SonicFill 2 is a light-cured, low-shrink, resin-based, dental restorative designed for direct placement. SonicFill 2 is used in combination with the SonicFill 2010 Handpiece (K091091), which activates the restorative sonically for delivery. Upon delivery the restorative drops in viscosity, allowing the composite to closely adapt to the cavity walls. When the cavity is filled and the SonicFill 2010 Handpiece (K091091) is deactivated, the restorative resin returns to its original viscosity. The composite in the cavity is then shaped and sculpted to the contours of the tooth. SonicFill 2 is designed to have a high depth of cure and low shrinkage stress allowing a cavity up to 5mm in depth to be filled and cured in a single bulk increment. - న. Statement of Intended Use: SonicFill 2 is indicated for direct placement in all cavity classes in anterior and posterior teeth. Additional indications include: base/liner material, repair of enamel defects, repair of provisionals, repair of porcelain restorations, minor occlusal build-ups, pit and fissure sealant, luting of composite/ceramic veneers, core build-ups and incisal abrasions. {4}------------------------------------------------ #### 6. Identification of Risk Analysis Method Risk analysis was performed on SonicFill 2 utilizing a process based on ISO 14971:2007. The results of the risk analysis performed on SonicFill 2 concluded that all device design controls and process controls will be able to mitigate known potential failures and effects. In addition, performance testing and biocompatibility testing were performed to mitigate other potential risks. #### 7. Description of Safety and Substantial Equivalence: ### Technological Characteristics The technological characteristics of SonicFill 2 are very similar to those of the predicate, Metamorphosis (K091023). Both composites are sonically activated using the SonicFill SonicFill 2010 Handpiece (K091091) during delivery. Both SonicFill 2 and Metamorphosis (K091023) composites behave in the same manner in terms of viscosity dropping during activation while adapting to the cavity, and viscosity increasing after activation ends. Both SonicFill 2 and Metamorphosis (K091023) use the same SonicFill 2010 Handpiece (K091091) and the same unidose delivery system. ## Non-Clinical Performance Data Non-clinical performance data included testing for mechanical strength, polishability, water sorption and solubility, flexural strength, depth of cure, light sensitivity, radiopacity, volumetric shrinkage, shrinkage stress, and color stability. Working Time testing, stability testing and biocompatibility testing were also performed. The data analyzed from the various tests substantiate that SonicFill 2 is as safe and effective as the predicate Metamorphosis (K091023). The following standards were utilized for the nonclinical performance testing of SonicFill 2: - . Guidance for Industry and FDA Staff: Dental Composite Resin Devices -Premarket Notification [510(k)] Submissions, November 27, 1998 - ISO 10993-1: 2009 Biological evaluation of medical devices ● - ISO 10993-3:2003 Biological Evaluation of Medical Devices- Part 3: Tests for . Genotoxicity, Carcinogenicity, and Reproductive Toxicity - . ISO 10993-5:2009 Biological Evaluation of Medical Devices- Part 5: Tests for in Vitro Cytotoxicity - . ISO 10993-6:2007 Biological Evaluation of Medical Devices- Part 6: Tests for Local Effects after Implantation - . ISO 10993-10:2010 Biological Evaluation of Medical Devices- Part 10: Tests for Irritation and Skin Sensitization - ISO 14971:2007 Risk Management - ISO 4049:2009 Dentistry- Polymer based Filling, Restorative, and Luting . Materials. {5}------------------------------------------------ | Element | Predicate Device-<br>Metamorphosis | SonicFill 2 | |-----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) | Metamorphosis (K091023) | To be assigned | | Trade Name | SonicFill, Sonic-Activated Bulk Fill<br>Composite | SonicFill 2, Sonic-Activated Bulk<br>Fill Composite | | Target Users | Licensed dental professionals | Licensed dental professionals | | Device Description | Metamorphosis, a nano-hybrid<br>composite, is a light-cured, resin-<br>based, dental restorative designed for<br>direct placement. | SonicFill2, a nano-hybrid<br>composite, is a light cured, resin-<br>based dental restorative designed<br>for direct placement. | | | Metamorphosis is packaged into tips,<br>which are custom-designed, single-<br>dose and made from plastic. The<br>single-dose tips are dispensed using<br>the SonicFill 2010 Handpiece<br>(K091091). | SonicFill 2 uses the same delivery<br>system as the predicate device<br>Metamorphosis, which are<br>custom-designed, single dose and<br>made from plastic. The single-<br>dose tips are dispensed using the<br>same SonicFill 2010 Handpiece<br>(K091091) as the predicate device. | | Common Name | Dental Composite Restorative<br>Material | Dental Composite Restorative<br>Material | | Classification Name | Tooth Shade Resin Material, per<br>CFR § 872.3690 and<br>Pit and Fissure Sealant and<br>Conditioner, per CFR § 872.3765 | Tooth Shade Resin Material, per<br>CFR § 872.3690 | | Class | II | II | | Product Code | EBF/ EBC | EBF/EBC | | Storage | Ambient Temperature | Ambient Temperature | | Curing Mechanism | Photo initiation | Photo initiation | | Material<br>Compatibility | Biocompatibility meets requirements | Biocompatibility meets<br>requirements | | Shelf Life | 24 months based on real time data | 24 months based on accelerated<br>data | | UV Color Stability<br>per ISO 4049 | Pass | Pass | | Volumetric Shrinkage | Pass | Equivalent to Predicate | | Shrinkage Stress | Pass | Equivalent to Predicate | | Depth of Cure | Pass | Pass | | Light Sensitivity per<br>ISO 4049 | Pass | Pass | | Flexural Strength per<br>ISO 4049 | Pass | Pass | | Radiopacity per ISO<br>4049 | Pass | Pass | | Translucency | Pass | Pass | | Working Time | Pass | Pass | | Water Sorption/<br>Solubility per ISO<br>4049 | Pass | Pass | | Polishability | N/A | Pass | # Table 5.1: Predicate and Proposed Device Comparison Table {6}------------------------------------------------ ## Clinical Performance Data Clinical performance testing has not been performed for SonicFill 2. ## Conclusion as to Substantial Equivalence SonicFill 2 has been tested for its strength, depth of cure, shrinkage, polishability, color It also has been verified as being substantially equivalent in stability, and safety. performance, safety, and effectiveness to the predicate Metamorphosis (K091023).
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...