MERIDIAN

K131753 · Ab Ardent · EBF · Nov 14, 2013 · Dental

Device Facts

Record IDK131753
Device NameMERIDIAN
ApplicantAb Ardent
Product CodeEBF · Dental
Decision DateNov 14, 2013
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 872.3690
Device ClassClass 2
AttributesTherapeutic, Pediatric

Intended Use

Meridian is a direct filling material for restoration of Class I to V cavities and for restoration of deciduous teeth.

Device Story

Meridian is a light-curing nanohybrid dental composite material used by dentists for direct cavity restorations. It functions as a restorative filling material for Class I through V cavities and deciduous teeth. The device is applied by the clinician to the prepared tooth cavity and cured using a light source. It serves to restore tooth structure and function. No complex processing, software, or remote workflows are involved.

Clinical Evidence

Bench testing only. The device was evaluated according to ISO 4049:2009 for water absorption, water solubility, radiopacity, flexural strength, depth of cure, modulus of elasticity, and compressive strength. Biocompatibility was evaluated per ISO 10993. Results demonstrate substantial equivalence to the predicate device.

Technological Characteristics

Nanohybrid light-curing dental composite resin. Tested per ISO 4049:2009 (physical properties) and ISO 10993 (biocompatibility).

Indications for Use

Indicated for the restoration of Class I to V cavities and the restoration of deciduous teeth in patients requiring dental restorative procedures.

Regulatory Classification

Identification

Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K131753 ## 510(K) SUMMA Meridian | Contact: | Donna Marie Hartnett | |----------------------|----------------------------------------------------------------------| | Company: | Ardent, Inc. 175 Pineview Drive, Amherst, NY 14228<br>(716) 691-0010 | | Date Prepared: | June 13, 2013 | | Proprietary Name: | Meridian | | Classification Name: | Material, Tooth Shade, Resin (872.3690) | | Predicate Devices: | Tetric EvoCeram (K042819) | Meridian is a light curing universal dental composite material of the Device Description: latest nanohybrid type having an optimized chemical composition. The predicate device to which Meridian has been compared is Tetric EvoCeram (K042819). For this application. Tetric EvoCeram has been compared to its predicate with regard to chemical composition, performance data and indications for use. The comparison shows that Meridian is substantially equivalent to the predicate device. Intended Use: Meridian is a direct filling material for restoration of Class I to V cavities and for restoration of deciduous teeth. Technological Characteristics: The device design, i.e. delivery form, and intended use of Meridian and the predicate device are the same. The composition of the subject device has been modified from the predicate, however, there are no ingredients in the subject device which pose any new issues of safety and effectiveness. Guidance Document: The submission was prepared in accordance with FDA Guidance document entitled Dental Composite Resin Devices - Premarket Notification (510K) Submissions dated October 26, 2005. Testing Summary: The device was tested in accordance with ISO 4049:2009 for Polymer based dental restorative materials for Water Absorption, Water solubility, Radiopacity, Flexural Strength, Depth of Cure. Modulus of Elasticity, and Compressive Strength and the results from testing demonstrates that Meridian is substantially equivalent to the predicate device. Biocompatibility testing and evaluation was also carried out according to ISO 10993. Conclusion: Meridian is substantially equivalent to the predicate device. {1}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three wing-like shapes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol. Public Health Service Food and Drug Administration 0903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 14, 2013 AB Ardent C/O Ms. Donna Marie Hartnett Director Quality Assurance/Regulatory Affairs 175 Pineview Drive AMHERST NY 14228 Re: K131753 Trade/Device Name: MERIDIAN Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: II Product Code: EBF Dated: September 3, 2013 Received: September 5, 2013 #### Dear Ms Hartnett: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ ### Page 2 - Ms. Hartnett Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. # Mary S. Runner -S Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K134753 Device Name: MERIDIAN Indications For Use: Meridian is a direct filling material for restoration of Class I to V cavities and for restoration of deciduous teeth. Prescription Use _____________________________________________________________________________________________________________________________________________________________ Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) > Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 > > Andrew I. Steen -S 2013.11.14 08:21:40 -05'00'
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...