K220253 · Dio Corporation · NHA · Aug 18, 2023 · Dental
Device Facts
Record ID
K220253
Device Name
Eco Abutment, Multiunit Abutment
Applicant
Dio Corporation
Product Code
NHA · Dental
Decision Date
Aug 18, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Eco Abutment is a premanufactured prosthetic component directly connected to an endosseous implant and it is intended for use in prosthetic rehabilitation. The Multiunit Abutment is a premanufactured prosthetic component directly connected to an endosseous implant and it is intended for use in prosthetic rehabilitation.
Device Story
Eco Abutment and Multiunit Abutment are premanufactured prosthetic components for dental implants. Eco Abutment is a two-piece system (base abutment and post) available in various diameters and angulations (0°, 6°, 12°, 18°). Multiunit Abutment is available in straight (0°) and angled (20°, 30°) configurations. Both are fabricated from titanium alloy (ASTM F136). Devices are provided non-sterile and require steam sterilization by the clinician before use. They are used by dentists to connect prosthetic restorations (crowns, bridges, overdentures) to endosseous implants. The output is a stable mechanical interface between the implant and the final dental restoration, facilitating patient chewing function and aesthetic rehabilitation. Clinical decision-making is supported by the device's ability to accommodate different anatomical requirements through varied cuff heights, diameters, and angulations.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including fatigue testing per ISO 14801:2016, sterilization validation per ISO 17665, and biocompatibility evaluation per ISO 10993 series.
Indicated for prosthetic rehabilitation in patients requiring single or multiple unit restorations (crowns, bridges, bars, overdentures) supported by endosseous dental implants in the mandible or maxilla.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
K203252 — Multi-unit Abutments for CONELOG · BioHorizons Implant Systems, Inc. · Mar 16, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
DIO Corporation % Peter Kang General Manager DIO USA 2729 Bristol St. Costa Mesa, California 92626
Re: K220253
Trade/Device Name: Eco Abutment, Multiunit Abutment Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: July 19, 2023 Received: July 19, 2023
Dear Peter Kang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K220253
Device Name Eco Abutment, Multiunit Abutment
## Indications for Use (Describe)
Indications for Use for Eco Abutment
The Eco Abutment is a premanufactured prosthetic component directly connected to an endosseous implant and it is intended for use in prosthetic rehabilitation.
Indications for Use for Multiunit Abutment
The Multiunit Abutment is a premanufactured prosthetic component directly connected to an endosseous implant and it is intended for use in prosthetic rehabilitation.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for DIO Implant Solution. The logo consists of the letters "DIO" in a bold, sans-serif font, with the "D" and "I" in gray and the "O" in pink. A small pink circle is positioned above and to the right of the "O". Below the letters, the words "Implant Solution" are written in a smaller, sans-serif font.
## 510(k) Summary - K220253
#### Submitter
DIO Corporation Han-Seul, Kim 66, Centum Seo-ro, Haeundae-gu Busan, 48058 Republic of Korea Email: khs@dio.co.kr Phone: +82-51-745-7875
Contact/US agent DIO USA Peter Kang 2729 Bristol St., Costa Mesa, California USA Email: peter@dioimplantusa.com Phone: 1-714-8804141
## 1. Device Information
Trade Name: Eco Abutment, Multiunit Abutment Common Name: Dental Implant Abutment Classification Name: Endosseous dental implant abutment Product Code: NHA Regulation Number: 21 CFR 872.3630 Device Class: Class II Date prepared: August/18/2023
## 2. Predicate Device
## Primary Predicate
K161655 - On1 Concept
## Reference Device
K182194 - UV Active Implant System K173975 – UF(II) Wide Fixture K122519 – DIO UF HSA INTERNAL SUB-MERGED IMPLANT SYSTEM K170608 – UF(II) Implant System K080559 – DIO SM IMPLANT SYSTEM
## 3. Device Description
The Eco Abutment is compatible with following Implant system. The Eco abutments cannot be used with implant diameters less than Ø3.8mm.
| Manufacturer | Compatible Implant System | 510(k) Number | Implant diameter (mm) | Connection type |
|-----------------|-----------------------------|---------------|---------------------------------|----------------------|
| DIO Corporation | UV Active Implant System | K182194 | 3.8/4.0/4.5/5.0/5.5/5.9/6.4 | Internal Hex/Non-hex |
| | UF(II) Wide Fixture | K173975 | 5.9/6.4/6.9 | Internal Hex/Non-hex |
| | UF Submerged Implant system | K122519 | 3.8/4.0/4.5/5.0/5.5/5.9/6.4/6.9 | Internal Hex/Non-hex |
| | UF(II) Implant System | K170608 | 3.8/4.0/4.5/5.0/5.5 | Internal Hex/Non-hex |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for DIO Implant Solution. The logo consists of the letters "DIO" in a bold, sans-serif font. The "DI" portion of the logo is in gray, while the "O" is in pink with a pink dot above it. Below the letters, the words "Implant Solution" are written in a smaller, sans-serif font, also in gray.
The Multiunit abutment is compatible with following Implant system. The Multiunit abutments cannot be used with implant diameters less than Ø4.5mm.
| Manufacturer | Compatible Implant<br>System | 510(k)<br>Number | Implant diameter<br>(mm) | Connection<br>type |
|--------------------------------|------------------------------|-----------------------------|--------------------------|-------------------------|
| DIO<br>Corporation | UV Active Implant<br>System | K182194 | 4.5/5.0/<br>5.5/5.9/6.4 | Internal<br>Hex/Non-hex |
| UF(II) Wide Fixture | K173975 | 5.9/6.4/6.9 | Internal<br>Hex/Non-hex | |
| UF Submerged<br>Implant system | K122519 | 4.5/5.0/<br>5.5/5.9/6.4/6.9 | Internal<br>Hex/Non-hex | |
| UF(II) Implant System | K170608 | 4.5/5.0/<br>5.5 | Internal<br>Hex/Non-hex | |
Eco Abutment is a two-piece abutment. It consists of a base abutment which is used with temporary post, cemented post, angled post and healing cap. The base abutment is first secured to the dental implant with a base screw and the post is secured to the base abutment with post screw. The Eco Abutment has 4.5, 4.8, 5.5 and 6.5 of diameter and consists of 0, 6, 12 and 18 degree.
The Multiunit Abutment is intended for use in conjunction with the fixture in partially or fully edentulous mandibles and maxillae, in support of single unit loading (i.e., crown) or multi-unit loaded restorations (i.e., bridge, bars, overdentures). It is consists of Multiunit Straight Abutment, Multiunit Angled Abutment and Temporary Cylinder. The Multiunit Abutment has 4.8mm of diameter and consists of two kind of design that has three angles. The Multiunit Straight abutment has 0 degree and the Multiunit Angled abutment has 20 or 30 degree
The Eco Abutment and Multiunit Abutment are made from titanium alloy conforming to ASTM F136. It is provided non-sterile and is steam sterilized before use.
Non-Hex connection abutments are intended for multi-unit restorations only.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for DIO Implant Solution. The logo consists of the word "DIO" in a bold, sans-serif font, with the "O" in pink and the rest of the letters in gray. There is a pink circle above and to the right of the "O". Below the word "DIO" are the words "Implant Solution" in a smaller, gray, sans-serif font.
## 4. Indications for Use
#### <Eco Abutment>
The Eco Abutment is a premanufactured prosthetic component directly connected to an endosseous implant and it is intended for use in prosthetic rehabilitation
<Multiunit Abutment>
The Multiunit Abutment is a premanufactured prosthetic component directly connected to an endosseous implant and it is intended for use in prosthetic rehabilitation
## 5. Comparison of Technological Characteristics with the Predicate Device.
| | Subject Device | Primary Predicate Device | Reference Device | Remark | |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------|--|
| Applicant | DIO Corporation | Nobel Biocare AB | DIO Corporation | - | |
| Trade Name | Eco Abutment | On1 Concept | UV Active Implant system -<br>Multi-unit Abutment | - | |
| 510(k) No. | K220253 | K161655 | K182194 | - | |
| Classification<br>Name | Endosseous Dental Implant,<br>Abutment (872.3630) | Endosseous Dental Implant,<br>Abutment (872.3630) | Endosseous Dental Implant<br>(872.3640) | - | |
| Product Code | NHA | NHA | DZE, NHA | - | |
| Class | II | II | II | - | |
| Indications<br>For Use | The Eco Abutment is a<br>premanufactured prosthetic<br>component directly connected to<br>an endosseous implant and it is<br>intended for use in prosthetic<br>rehabilitation. | The On1™ device is a<br>premanufactured prosthetic<br>component directly<br>connected to an endosseous<br>implant and it is intended<br>for use in prosthetic<br>rehabilitation. | The UV Active Implant<br>System is indicated for<br>surgical placement in the<br>upper and lower jaw arches,<br>to provide a root form<br>means for single or multiple<br>units' prosthetic attachment<br>to restore a patient's<br>chewing function.<br><br>The narrow (⌀3.0, ⌀3.3)<br>implant is limited to the<br>replacement of maxillary<br>lateral incisors and<br>mandibular incisors. It is<br>intended for delayed<br>loading.<br><br>The Regular (⌀3.8 ~ ⌀5.5)<br>implants can be placed with<br>a conventional two stage<br>surgical process with an<br>option for transmucosal<br>healing or they can be<br>placed in a single stage<br>surgical process for<br>immediate loading when<br>good primary stability is<br>achieved with appropriate<br>occlusal loading.<br><br>The Wide (⌀6.0 ~ ⌀6.4) | Same | |
| | Subject Device | Primary Predicate Device | Reference Device | Remark | |
| | | | a conventional two stage<br>surgical process with an<br>option for transmucosal<br>healing and are indicated<br>for the molar region with<br>delayed loading | | |
| Material | Eco Base Abutment, Eco Cemented<br>post, Eco Angled post, Temporary<br>Post<br>: Ti-6Al-4V ELI (ASTM F136) | 0n1 Base, Temporary<br>Abutment, On1 Healing Cap,<br>Esthetic Abutment Titanium,<br>Clinical Screws and Prosthetic<br>Screws - Titanium vanadium<br>alloy (ASTM F1472, ASTM<br>F136)<br>On1 Esthetic Abutment<br>Zirconia – Y-TZP zirconium<br>oxide (ISO 6872, ISO 13356) | Ti-6Al-4V ELI (ASTM F136) | Same | |
| Design | - Post (Eco Cemented post, Eco<br>Angled post, Eco Temporary post):<br>Divided into Hex/Non-Hex<br>connection<br>-Eco Base Abutment: Divided into<br>Hex/Non-Hex connection | 2 piece (base placed either<br>at time of implant<br>placement or with final<br>abutment)<br>Abutment shape fixed | Multiunit Abutment:<br>Divided into Hex/Non-Hex<br>Connection | Similar;<br>Except for its<br>shape | |
| Diameters<br>(mm) | 4.5/4.8/5.5/6.5 | At base 4.8/5.3/6.5 | 4.8 | Different but<br>similar range | |
| Length (mm) | Combined base abutment and post<br>height<br>Cemented post -<br>5.5/6.5/7.0/7.5/8.0/8.5/9.0/9.5/<br>10.0/10.5/11.0/11.5/12.5<br>Angled post -<br>9.5/10.5/11.5/12.5/13.5<br>Temporary post -<br>5.5/6.5/7.0/7.5/8.0/8.5/9.0/9.5/<br>10.0/10.5/11.0/11.5/12.5/13.5/1<br>4.5 | Combined base and post<br>height.<br>Temporary Abut - 8.3 /9.0<br>Esthetic Abut Ti - 8.2/9.0<br>Esthetic Abut Zi - 8.2/9.0 | 10.3/11.3/12.3/13.3 | Similar;<br>Longer range<br>than<br>predicate | |
| Gingival<br>heights (mm) | 1.5/2.5/3.5/4.5/5.5 | 2.0/3.5 | 1.5/2.5/3.5/4.5/5.5 | Same; | |
| Angle | Eco Base abutment & Eco<br>Cemented post: 0°<br>Eco Base abutment & Eco Angled<br>post: 6°/12°/18° | No abutment angulation | 0 °/ 20 ° / 30° | Similar;<br>Within range<br>of reference<br>device. | |
| Sterile | Steam Sterilization by user<br>(Delivered non sterile) | Steam Sterilization by user<br>(Delivered non sterile) | Steam Sterilization by user<br>(Delivered non sterile) | Same | |
| Type of<br>Retention | Screw-retained or cemented<br>retained | Screw-retained or cemented<br>retained | Screw-retained or cemented<br>retained | Same | |
| Substantial Equivalence Discussion | | | | | |
| | Subject Device | Primary Predicate Device | Reference Device | Remark | |
| Similarities | | | | | |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for DIO Implant Solution. The logo has the letters "DIO" in a bold, sans-serif font. The letters "DI" are in gray, while the "O" is in pink and has a small pink circle above it. Below the letters, the words "Implant Solution" are written in a smaller, gray font.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for DIO Implant Solution. The logo consists of the letters "DIO" in a bold, sans-serif font. The "DI" is in gray, while the "O" is in pink and has a small pink circle above and to the right of it. Below the letters, in a smaller font, are the words "Implant Solution".
The subject device is a two-piece abutment. It consists of Eco Base abutment and 3 types of post (Cemented, Angled and Temporary).
The subject device has the same indications for use compared to the predicate device (K161655) and design (except for its shape), identical material composition and manufacturing methods by the same manufacturer compared to the reference device (K182194). The Indications for Use Statement has minor differences; however has the same intended use as being used with Endosseous dental implant to support a single or multiple prosthetic devices.
#### Differences
The diameter and lengths are different compared to the predicate device; however, the 4.5~6.5mm for the subject device and 4.8, 5.3 and 6.5mm for the predicate devices is comparable.
The total length of 5.5 to 14.5 mm for the subject device is an extended range compared to the predicate devices. To account for these differences, fatigue testing was conducted in accordance with ISO 14801 to verify that subject device is substantially equivalent to the predicate devices. The results of fatigue testing were comparable between the predicate and subject device.
#### Discussion
The subject device is similar in fundamental scientific technology to the predicate device (K161655) in that they have been designed in compliance with FDA's Class II special controls guidance document root-form Endosseous dental implants and Endosseous dental implant abutment. There are slight mechanical differences between subject devices which was evaluated through comparative performance testing. The documentation submitted in the premarket notification demonstrates that the subject device is substantially equivalent to the predicate devices.
Therefore, we conclude that the Eco Base abutment is substantially equivalent to the predicate device.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for DIO Implant Solution. The logo consists of the letters "DIO" in a stylized font, with "DI" in gray and "O" in pink with a pink dot above it. Below the letters, the words "Implant Solution" are written in a smaller, gray font.
## 2) Multiunit Abutment
| | Subject Device | Reference Device | Remark |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|
| Applicant | DIO Corporation | DIO Corporation | - |
| Trade Name | Multiunit Abutment | UV Active Implant system – Multi-unit Abutment | - |
| 510(k) No. | K220253 | K182194 | - |
| Classification Name | Endosseous Dental Implant, Abutment (872.3630) | Endosseous Dental Implant (872.3640) | - |
| Product Code | NHA | DZE, NHA | - |
| Class | II | II | - |
| Indications For Use | The Multiunit Abutment is a premanufactured prosthetic component directly connected to an endosseous implant and it is intended for use in prosthetic rehabilitation. | The UV Active Implant System is indicated for surgical placement in the upper and lower jaw arches, to provide a root form means for single or multiple units' prosthetic attachment to restore a patient's chewing function.<br><br>The narrow (Ø3.0, Ø3.3) implant is limited to the replacement of maxillary lateral incisors and mandibular incisors. It is intended for delayed loading.<br><br>The Regular (Ø3.8 ~ Ø5.5) implants can be placed with a conventional two stage surgical process with an option for transmucosal healing or they can be placed in a single stage surgical process for immediate loading when good primary stability is achieved with appropriate occlusal loading.<br><br>The Wide (Ø6.0 ~ Ø6.4) implants can be placed with a conventional two stage surgical process with an option for transmucosal healing and are indicated for the molar region with delayed loading. | Similar – The reference device include both implant bodies and abutments. |
| Material | Ti-6Al-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM F136) | Same |
| Design | Multiunit Straight<br>Image: Multiunit Straight<br><br>Multiunit Angled<br>Image: Multiunit Angled Hex<br>Image: Multiunit Angled Non-Hex | Multiunit Straight<br>Image: Multiunit Straight<br><br>Multiunit Angled<br>Image: Multiunit Angled Hex<br>Image: Multiunit Angled Non-Hex | Same |
| Diameters (mm) | 4.8 | 4.8 | Same |
| Cuff size (mm) | 2.5/3.0/3.5/4.5/5.5/6.5/7.5 | 1.5/2.5/3.5/4.5/5.5 | Similar; Subject device has longer range compared to reference device |
| Connected | Regular type | Regular type | Same |
| | Subject Device | Reference Device | Remark |
| fixture type | | | |
| Angle(°) | 0/20/30 | 0/20/30 | Same |
| Sterile | Steam Sterilization by user (Delivered non<br>sterile) | Steam Sterilization by user (Delivered non<br>sterile) | Same |
| Type of<br>Retention | Screw-retained or cement retained | Screw-retained or cement retained | Same |
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for DIO Implant Solution. The logo consists of the letters "DIO" in a bold, sans-serif font. The "DI" portion of the logo is in gray, while the "O" is in pink and has a small pink circle above and to the right of it. Below the logo, in a smaller font, are the words "Implant Solution".
#### Substantial Equivalence Discussion
#### Similarities
The subject device has the same indications for use compared to the predicate device (K161655) and design, and manufacturing methods and by the same manufacturer as the reference device (K182194). The Indications for Use Statement has minor differences; however has the same intended use as being used with Endosseous dental implant to support a single or multiple prosthetic devices.
#### Differences
The subject device is substantially equivalent to the indication for use, connected fixture type, design and angle as compared to the predicate devices, but includes additional specifications in order to cover different clinical situations. The subject device has additional cuff sizes as compared to the predicate devices. To account for these differences, fatigue testing was conducted in accordance with ISO 14081 to verify substantial equivalence with the predicate device. The results of the fatigue testing were comparable between the predicate and subject device.
#### Discussion
There are slight mechanical differences between the subject devices. These differences were evaluated through comparative performance testing. The documentation submitted it the premarket notification demonstrates that the subject device is substantially equivalent to the marketed devices.
Therefore, we conclude that the Multiunit abutment is substantially equivalent to the predicate device.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for DIO Implant Solution. The logo consists of the letters "DIO" in a bold, sans-serif font. The "DI" is in gray, while the "O" is in pink with a small pink circle above and to the right of it. Below the letters, in a smaller font, are the words "Implant Solution" in gray.
## 6. Summary of Non-clinical Testing
The results of the non-clinical testing demonstrate that the results have met the criteria of the standards, and the subject device is substantially equivalent to the predicate device.
## Sterilization Validation
The subject device is provided non-sterile. Sterilization has been performed in accordance with ISO 17665-1 and ISO 17665-2 for steam sterilization. Test results have demonstrated that the SAL of 10-6 was achieved and all testing requirements were met. Also, Sterilization Validation was conducted according to FDA Guidance "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling".
## Biocompatibility
Biocompatibility tests were conducted in accordance with FDA Guidance Document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" and ISO 10993-10, ISO 10993-10, ISO 10993-23, ISO 10993-11, ISO 10093-3, and ISO 10993-6. The Abutments have the identical nature of body contact, contact duration, material formulation, manufacturing processes, and sterilization methods compared to the reference devices, K182194 and K080559. No new issues of biocompatibility are raised for the subject devices.
## Mechanical Properties
Fatigue testing was performed on the subject device including both the eco abutments and multiunit abutments in accordance with ISO 14801:2016 Dentistry-Implants-Dynamic fatigue test for Endosseous Dental Implants. The worst case scenario was chosen based on the FDA guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments".
Comparative fatigue testing was also performed on comparative abutments cleared in K182194.
## MR Environment Condition
Non-clinical worst-case MRI review was performed to evaluate the metallic. Eco Abutment and Multiunit Abutment in the MRI environment using scientific rationale and published literature (e.g., Woods, Terry 0., Jana G. Delfino, and Sunder Rajan. 'Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices.' Journal of Testing and Evaluation 49.2 (2019): 783-795), based on the entire system including all variations (all compatible implant bodied, dental abutments and, fixation screws) and material composition. Rationale addressed parameters per the FDA Guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque.
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for DIO Implant Solution. The logo consists of the letters "DIO" in a bold, sans-serif font, with "DIO" in gray and the "O" in pink. Above the "O" is a small pink circle. Below the letters is the text "Implant Solution" in a smaller, gray, sans-serif font.
## 7. Summary of Clinical Testing
No clinical studies are submitted.
## 8. Conclusions
The Eco Abutment and Multiunit Abutment constitute a substantially equivalent medical device, meeting all the declared requirements of its intended use. These abutments have the same intended use and fundamental scientific technology as its predicate device. Therefore, Eco Abutment and Multiunit Abutment and its predicates are substantially equivalent.
End of Document
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.