ENDOSEAL is a root canal filling material composed of natural pure cement, zirconium dioxide, and citric acid anhydrous. It is supplied as a powder to be mixed with water, forming a putty that hardens in the oral environment. Used by dental professionals for root canal filling and perforation repair, the material provides sealing and biocompatibility. It is applied directly to the root canal or perforation site to facilitate healing and structural integrity.
Clinical Evidence
Bench testing only. Testing included ISO 6876 (visual, packaging, setting time, solubility, radiopacity), ISO 7405 (cytotoxicity), ISO 10993-10 (skin sensitization, LLNA), ISO 10993-11 (acute systemic toxicity), and additional bench testing for capacity and pH. All results met preset criteria.
Technological Characteristics
Composition: Natural Pure Cement, Zirconium dioxide, Citric acid anhydrous. Form: Powder/water mixture, putty consistency. Hardening: Self-setting in oral environment. Standards: ISO 6876, ISO 7405, ISO 10993-10, ISO 10993-11.
Indications for Use
Indicated for the repair of root perforations and as a root canal filling material in patients requiring endodontic treatment.
Regulatory Classification
Identification
A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.
Predicate Devices
ENDOCEM MTA (Mineral Trioxide Aggregate) (K112078)
Submission Summary (Full Text)
{0}------------------------------------------------
KI33054
MAY 1 2 2014
# 510(k) Summary
This summary of 510(k)-safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: 02/26/2014
# Submission Applicant / Submitter: 1.
MARUCHI 2-208, 42-10, Taejanggongdan-gil, Wonju-si. Gangwon-do, 220-962, Republic of Korea
Tel: +82-33-734-0330 Fax: +82-33-746-2804
#### 2. Submission Correspondent:
Priscilla Chung LK Consulting Group USA, Inc. 2651 E Chapman Ave Ste 110, Fullerton CA 92833 Phone: 714-202-5789 Fax: 714-409-3357 Email: juhee.c@lkconsultinggroup.com
# 3. Device:
| Proprietary Name: | ENDOSEAL |
|------------------------------|---------------------------|
| Common Name: | Root Filling Material |
| Classification Name: | Root Canal Filling Resin |
| Classification: | Class II, 21 CFR 872.3820 |
| Classification Product Code: | KIF |
### 4. Predicate Device:
ENDOCEM MTA (Mineral Trioxide Aggregate) (K112078) by MARUCHI
### 5. Device Description:
The raw materials of the ENDOSEAL are Natural Pure Cement, Zirconium dioxide, and Citric acid anhydrous, and the device has been showing good sealing ability and biocompatibility. It is prepared as a mixture of powder and water, and it is used in a putty form which gradually hardens in the oral environment.
1
{1}------------------------------------------------
## દ. Intended Use:
- . Repair of perforation
- . Root canal filling
# Performance Data(Non-Clinical): 7.
The following properties were tested based on the referenced standard. All the test results met the preset test criteria.
- ISO 6876 Visual, Packaging, Setting time, Solubility and Radiopacity
- ISO 7405 Cytotoxicity
- ISO 10993-10 Skin sensitization, Local Lymph Node Assay, LLNA
- ISO 10993-11 Acute systemic toxicity
- · Other bench testing Capacity and pH (Saline Extracts)
# 8. Substantial Equivalence
ENDOSEAL has similar physical and biocompatible properties, and demonstrates comparable performance specifications to ENDOCEM MTA (Mineral Trioxide Aggregate). In addition. ENDOSEAL has a comparable delivery system to ENDOCEM MTA. In comparison, ENDOSEAL contains Zirconium dioxide and Citric acid anhydrous, but Bismuth Trioxide, which is in ENDOCEM MTA, is not present in ENDOSEAL. The results of bench and biocompatibility testing performed demonstrate that this difference does not raise any new questions as to safety and effectiveness. Therefore, it is concluded that ENDOSEAL is substantially equivalent in safety and
effectiveness to the predicate device.
### 9. Conclusion:
Based on the testing results, MARUCHI concludes that the ENDOSEAL is substantially equivalent in safety and effectiveness to the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in all capital letters. The text is in a bold, sans-serif font. The words are arranged on a single line and are centered horizontally. The text appears to be part of a document or letterhead.
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
# May 12, 2014
MARUCHI C/O Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Incorporated 2651 East Chapman Avenue, Suite 110 Fullerton, CA 92833
Re: K133054
Trade/Device Name: ENDOSEAL Regulation Number: 21 CFR 872.3820 Regulation Name: Root Canal Filling Resin Regulatory Class: II Product Code: KIF Dated: February 26, 2014 Received: March 4, 2014
Dear Ms. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
{3}------------------------------------------------
Page 2 - Ms. Chung
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part
807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mary S. Runner -S
Erin I. Keith, M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K133054
Device Name ENDOSEAL
Indications for Use (Describe)
- · Repair of perforation
· Root canal filling
Type of Use (Select one or both, as applicable)
[x] Prescription Use (Part 21 CFR 801 Subpart D)
O Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR 1. Banks 2017 - 12:10 PM 11. 12. 12. 12. 12. 12. 12. 12. 12. 12. 12. 12. 12. 12. 12. 12. 12. 12. 12. 12. 12. 12. 12. 12. 12. 12. 12. 12. 12. 12. 12. 12. 12. 12. 12. 12. 12. 1
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/4/Picture/13 description: The image shows the name "Sheena" followed by the date "2014.05.12 22:22:21". There is a symbol that looks like a triangle in the middle of the date. The image also contains the text "-04'00'".
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.