KETAC-MOLAR APLICAP
Device Facts
| Record ID | K960954 |
|---|---|
| Device Name | KETAC-MOLAR APLICAP |
| Applicant | Espe GmbH & Co. Kg. |
| Product Code | EMA · Dental |
| Decision Date | May 16, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 872.3275 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
Ketac-Molar® Aplicap® is a glass ionomer cement material indicated for the following uses: 1. linings for Class I and II cavities filled with composite; 2. core build-ups; 3. fillings in deciduous teeth; 4. fillings in Class I cavities located in non-occlusal load bearing areas; 5. fillings in Class V cavities if the aesthetics are not of primary importance; and 6. temporary fillings in Class I and II cavities.
Device Story
Glass ionomer cement material; modification of Ketac-Bond® Aplicap® and Ketac-Silver® Aplicap®; designed for improved compressive strength. Used by dental professionals in clinical settings for restorative procedures; provides structural support and filling material for various cavity classes and core build-ups. Applied via Aplicap® delivery system; hardens to provide dental restoration. Benefits include enhanced mechanical properties compared to predecessors while maintaining established glass ionomer clinical utility.
Clinical Evidence
Bench testing only. Comparison of physical and technical characteristics and water-soluble fluoride content against predicate devices. Compliance with relevant DIN and ISO standards for dental cement confirmed.
Technological Characteristics
Glass ionomer cement material; improved formulation for higher compressive strength. Meets relevant DIN and ISO standards for dental cement. Delivered via Aplicap® system.
Indications for Use
Indicated for dental patients requiring cavity linings (Class I/II), core build-ups, deciduous tooth fillings, non-occlusal Class I cavity fillings, Class V fillings (where aesthetics are secondary), and temporary fillings for Class I/II cavities.
Regulatory Classification
Identification
Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.