TRI-PHASIX CHROMATIC ALGINATE
Device Facts
| Record ID | K962741 |
|---|---|
| Device Name | TRI-PHASIX CHROMATIC ALGINATE |
| Applicant | Parkell, Inc. |
| Product Code | ELW · Dental |
| Decision Date | Sep 6, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 872.3660 |
| Device Class | Class 2 |
Intended Use
The product is intended for use as impression material as defined in §872.3660 of CFR 21, "to reproduce the structure of a patient's teeth and gums ... to provide models for study and for production of restorative prosthetic devices," etc.
Device Story
Vanilla-flavored chromatic alginate impression material; changes color (violet to pink to white) to indicate mixing completion, tray loading readiness, and impressioning readiness. User-adjustable water-to-powder ratios allow for heavy, medium, or light body consistencies. Used by dental professionals in clinical settings to capture dental anatomy; output is a physical impression used to create gypsum models for study or prosthetic restoration. Benefits include visual cues for optimal handling and 50-micron detail reproduction.
Clinical Evidence
Bench testing only. Complies with ISO 1563/1990. Reproduces 50-micron detail. Mixing time 45 seconds; total working time 2 minutes 25 seconds.
Technological Characteristics
Chromatic alginate impression material; vanilla flavored. Complies with ISO 1563/1990. Compatible with dental model gypsum. Reproduces 50-micron detail.
Indications for Use
Indicated for dental patients requiring impressions of teeth and gums for study models or restorative prosthetic device fabrication.
Regulatory Classification
Identification
Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.
Special Controls
*Classification.* Class II (Special Controls).
Predicate Devices
- Chromaclone Color Changing Alginate
- Orthalgenat Impression Material
- Algidenco
- Algiprint Impress Material
- Dental Alginate Impression Material
- Proof Alginate Impression Material
- Cutter-Jel Alginate Impression Material
Related Devices
- K981970 — ALGETRAL, CAVEX CHROMATIC, CA 37 SUPERIOR PINK, CA 37 FAST SET, CAVEX RAINBOW, CAVEX IMPRESSIONAL NORMAL SETTING, CAVEXI · Cavex Holland BV · Sep 2, 1998
- K980965 — SPECTRA FORM CHROMATIC ALGINATE IMPRESSION MATERIAL · Cavex Holland BV · May 1, 1998
- K160441 — Identic and KromaFaze Alginate Dental Impression Materials · Kerr Corporation · Jun 30, 2016
- K061267 — SIMULATE, COLOR CHANGE ALGINATE · Cavex Holland BV · Jul 10, 2006
- K140074 — HYGEDENT ALGINATE IMPRESSION MATERIAL · Hygedent, Inc. · May 21, 2014