K972945 · Aurex Precious Metal Industries (Pty)Ltd · EJS · Nov 26, 1997 · Dental
Device Facts
Record ID
K972945
Device Name
AURENORM 68
Applicant
Aurex Precious Metal Industries (Pty)Ltd
Product Code
EJS · Dental
Decision Date
Nov 26, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3060
Device Class
Class 2
Intended Use
Dental Casting alloy used for fabrication of metallo-ceramic restorations in conjunction with dental ceramics of thermal expansion co-efficient not less than 16.3 x 10 -9K. (25 - 600°C)
Device Story
Aurenorm 68 is a high-gold dental casting alloy; used for fabrication of metallo-ceramic restorations. Device functions as a substrate for dental ceramics; requires ceramic material with thermal expansion coefficient ≥ 16.3 x 10^-6/°C. Operated by dental laboratory technicians/dentists in clinical/lab settings. Output is a cast metal framework; provides structural support for porcelain veneering. Benefits patient by enabling durable, aesthetic dental restorations.
Clinical Evidence
Bench testing only. Performance evaluated via physical and mechanical property testing (melting point, hardness, yield strength, elongation, CTE, density) and compositional analysis compared to predicate. Testing conducted in accordance with ANSI/ADA 5, ISO 1562, and ISO 9693 standards.
Technological Characteristics
High-gold casting alloy. Composition: Au 68.9%, Ag 12.1%, Pt 9.9%, Cu 6.9%, Zn 0.9%, In 0.7%, Ru 0.6%. Physical properties: Melting range 939-977°C; Hardness 160-240 Vickers; CTE 16.4 x 10^-6/°C. Tested per ANSI/ADA 5, ISO 1562, and ISO 9693.
Indications for Use
Indicated for patients requiring metallo-ceramic dental restorations; used in conjunction with dental ceramics having a thermal expansion coefficient of at least 16.3 x 10^-6/°C (25-600°C).
Regulatory Classification
Identification
A noble metal alloy is a device composed primarily of noble metals, such as gold, palladium, platinum, or silver, that is intended for use in the fabrication of cast or porcelain-fused-to-metal crown and bridge restorations.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is FDA's “Class II Special Controls Guidance Document: Dental Noble Metal Alloys.” The devices are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. See § 872.1(e) for availability of guidance information.
Predicate Devices
Jelenko Allround 4
Related Devices
K972949 — AURELUX BIO Y85PF · Aurex Precious Metal Industries (Pty)Ltd · Sep 18, 1997
K972946 — AURELUX 77 · Aurex Precious Metal Industries (Pty)Ltd · Dec 19, 1997
K034049 — HERADOR EC, PF, C, H, NH, G, GG AND MP; HERALOY U AND G; HERABOND N; ALBABOND C, B AND A; BIO HERANORM · Heraeus Kulzer,GmbH · Mar 1, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo for AUREX. The logo consists of a stylized mountain range above the word "AUREX". The mountain range is made up of two overlapping triangles, and the word "AUREX" is written in a simple sans-serif font.
P.O. Box 509 Edenvale 1610 Republic of South Africa Telephone : +27 (11) 609-8640 Telefax : +27 (11) 452-3918
NOV 26 1997
K972945
## 510 (k) Summary
AURENORM 68 Trade name: Dental casting alloy Common name: Gold based alloys and precious metal alloys for clinical use Classification name: Class ification number: EJT
Legally marketed device: Jelenko Allround 4 Description of the device: High gold casting alloy Intended use of the device: Type IV restorations and low fusing porcelain veneering
## Summary of the technological characteristics
Test methods applied: as in ANSI/ADA 5 , ISO 1562 and 9693
Comparison of composition:
| ALLOY | | COMPOSITION (weight%) | | | | | | | |
|-------|-------------|-----------------------|------|-----|-----|-----|-----|-----|---|
| | Name | Au | Ag | Pt | Cu | Zn | In | Ru | I |
| Legal | Allroound 4 | 69 | 13.9 | 9 | 4.5 | 2 | 1.6 | 0 | x |
| New | Aurenorm 68 | 68.9 | 12.1 | 9.9 | 6.9 | 0.9 | 0.7 | 0.6 | 0 |
x is less than 1 %
Comparison of physical and mechanical properties
| ALLOY | | Melting point range (°C) | | Hardness (Vickers 5/30) | | Yield strength (MPa) | | Elongation (%) | | CTE (x10-6/°C ) | Density (g/cm3) |
|-------|-------------|--------------------------|--------|-------------------------|------|----------------------|------|----------------|------|-----------------|-----------------|
| | Name | solid. | liquid | soft | hard | soft | hard | soft | hard | | |
| Legal | Allround 4 | 900 | 970 | | 260 | | 592 | | 10 | 16.9 | 15.8 |
| New | Aurenorm 68 | 939 | 977 | 160 | 240 | 410 | 650 | 14 | 8 | 16.4 | 16.2 |
## Discussion
Both alloys consist of the same major elements. The noble metal content is very high, in excess of 78% and differs only approx. 1 %. The difference in base metal content is even less, i.e. 0.6 %.
## Conclusion
Basically similar composition and physical, mechanical properties. On the basis of compositional similarity it may be assumed that the biological effect is also similar.
Image /page/0/Picture/20 description: The image shows a logo with the letters "TUV" in bold, sans-serif font. Below the letters, there is some text that is difficult to read. The logo is enclosed in a rectangular shape with rounded corners, and there are three lines outlining the shape. The logo appears to be a certification mark or quality seal.
{1}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
J.D. Davis. ·C.E.O. Aurex Precious Metal Industries (PTY) Ltd. P.O. Box 509 Edenvale 1610 Republic of South Africa
Re : K972945 Trade Name: Aurenorm 68 Requlatory Class: II Product Code: EJS September 16, 1997 Dated: Received: September 30, 1997
Dear Mr. Davis:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, qood manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਕੇ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531
Image /page/1/Picture/9 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
{2}------------------------------------------------
Page 2 - Mr. Davis
through 542 of the Act for devices under the Electronic through 342 or the not 101 voisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours
U. Ulatowski
Timothy A Ulatowski Directbr Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
22-5011-1193
Page_1__of_1_
Concession of the consisted on the consisted to the status
510(k) Number (if known): K972945
Device Name:__________________________________________________________________________________________________________________________________________________________________
Indications For Use:
-
-
-
-
-
-
Dental Casting alloy used for fabrication of metallo-ceramic restorations in conjunction with dental ceramics of thermal expansion co-efficient not
less than 16.3 x 10 -9K. (25 - 600°C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) .
> Concurrentse of CDRH, Office of Device Evaluation (ODE) unovi ADDING (Division Sign-Off) Division if Dental Infection Control,
and General Hospial Devices 167794 510(k) Number_
Prescription Use (Per 21 CFR 801.109)
ેન્દ્રનો ભાગના પાકની ખેત
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
・・
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.