K242897 · Clemde SA DE CV · EBI · Jun 3, 2025 · Dental
Device Facts
Record ID
K242897
Device Name
Partial Flex
Applicant
Clemde SA DE CV
Product Code
EBI · Dental
Decision Date
Jun 3, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3760
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Partial Flex® is indicated for the fabrication of partial or full removable dentures, as well occasional splints or night guards
Device Story
Partial Flex is a polypropylene resin material supplied as pellets in aluminum tubes or plastic bags. It is intended for use by dental professionals to fabricate partial or full removable dentures, splints, and night guards via injection molding. The material is designed as a biocompatible, hypoallergenic alternative to traditional acrylic denture resins, offering improved impact resistance and fracture toughness. The healthcare provider processes the resin into the final prosthetic form; the resulting device provides structural support and functional restoration for the patient's oral cavity. The device is used in a dental laboratory or clinical setting.
Clinical Evidence
Bench testing only. Testing included biocompatibility (Intracutaneous Injection, Kligman Maximization, and L929 MEM Elution tests per ISO 10993) and physical performance testing per ISO 20795 (including flexural strength, flexural modulus, fracture work, sorption, and solubility). All tests passed acceptance criteria.
Technological Characteristics
Polypropylene resin material; supplied as pellets. Formulated for injection molding. Biocompatible and hypoallergenic. Tested per ISO 20795 (dental base polymers) and ISO 10993 (biocompatibility).
Indications for Use
Indicated for patients requiring partial or full removable dentures, or occasional splints or night guards.
Regulatory Classification
Identification
A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.
K063626 — DURAFLEX · Cosmetic Dental Materials, Inc. · Feb 8, 2007
K061501 — FLEXO DENTAL RESIN MATERIAL · Aci, Inc. · Jul 26, 2006
K123396 — I.FLEX BY TCS · Thermoplastic Comfort Systems, Inc. · Jan 10, 2013
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 3, 2025
Clemde Sa De Cv
Cabrera Francisco
General Manager
Mexico City, 07800
MEXICO
Re: K242897
Trade/Device Name: Partial Flex
Regulation Number: 21 CFR 872.3760
Regulation Name: Denture Relining, Repairing, Or Rebasing Resin.
Regulatory Class: Class II
Product Code: EBI, MQC
Dated: April 4, 2025
Received: May 1, 2025
Dear Cabrera Francisco:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K242897 - Cabrera Francisco
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K242897 - Cabrera Francisco
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
Submission Number (if known)
K242897
Device Name
PARTIAL FLEX
Indications for Use (Describe)
Partial Flex® is indicated for the fabrication of partial or full removable dentures, as well occasional splints or night guards
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
Partial Flex® 510(K) Premarket Notification
CLEMDE
# K 242897
# Section 6 - 510(k) Summary
## 1. Submitter
**APPLICANT:** Av la Fortuna #136, Industrial, México City 07800
Tel.: (+52) 55 3537 7103
Internet: www.clemde.com
**CONTACT PERSON:** Mariana Correa
Tel.: 956 414 1295
E-mail: apcoconsultores@gmail.com
## 2. Device Name
Trade Names: Partial Flex
Common Name: Dental Resin
Classification Name: EBI Denture Relining, Repairing, or Rebasing Resin
## 3. Predicate Device
Partial Flex is substantially equivalent to the following predicate devices already cleared by the FDA:
| Predicate Device | 510(k) Number | 510(k) Holder |
| --- | --- | --- |
| TCS UNBREAKABLE | K053060 | THERMOPLASTIC CONFORT SYSTEM |
## 4. Description of the Device
### 4.1. Partial Flex
Partial Flex® dental resin is a polymer base material used for the fabrication of partial removable dentures and prosthesis. Partial Flex is made from a resin of polypropylene that is designed and manufactured for injection molding. Partial Flex® is available in two shades as a granular material (so-called pellets) contained in an aluminum tube with 27 g or in a plastic bag of 250 g. In addition to high biocompatibility, Partial Flex® material features hypoallergenic properties, no risk of discoloration and the dental prosthesis made are unbreakable, which differentiates the from acrylic materials.
001_510K_Summary_v1
{5}
Partial Flex® 510(K) Premarket Notification
CLEMDE
## 4.2. Partial Flex Plastic bag
The resin material is pigmented in one color and without pigment presentation:
300116 Light pink
300118 Natural color (without pigment)
## 4.3. Partial Flex Aluminum tube
The resin material is pigmented in one color and without pigment presentation:
300073 Light pink
300079 Natural color (without pigment)
## 5. Indications for Use
### 5.1. Partial Flex
Partial Flex® is indicated for fabrication of partial or full removable dentures, as well occasional splints or night guards.
## 6. Technological Characteristics
| | NEW DEVICE | PREDICATE DEVICE |
| --- | --- | --- |
| 510(k) Submitter/ Holder | CLEMDE | THERMOPLASTI CONFORT SYSTEM |
| Trade Name | Partial Flex | TCS UNBREAKABLE |
| 510(k) Number | K242897 | K053060 |
| Material | Resin of polypropylene | polyamide |
| Shape | Pellets | Granulate - resin |
| Indications for use | Partial Flex® is indicated for fabrication of partial or full removable dentures, as well occasional splints or night guards. | Fabrication and repair of removable dental prosthetic devices, such as full and partial dentures, orthodontic devices, occlusal splints, and guards. |
| Shape | Pellet | granular resin |
001_510K_Summary_v1
{6}
Partial Flex® 510(K) Premarket Notification
CLEMDE
# Color stability
The material complies with the requirements
| Translucency | The material complies with the requirements | - |
| --- | --- | --- |
| Freedom from porosity | The material complies with the requirements | - |
| Ultimate Flexural Strength | The material complies with the requirements | Physical testing meets standard for dental base polymers |
| Flexural modulus | The material complies with the requirements | Physical testing meets standard for dental base polymers |
| Maximum stress intensity factor for materials with improved impact resistance | The material complies with the requirements | Physical testing meets standard for dental base polymers |
| Total fracture work | The material complies with the requirements | Physical testing meets standard for dental base polymers |
| Sorption | The material complies with the requirements | Physical testing meets standard for dental base polymers |
| Solubility | The material complies with the requirements | Physical testing meets standard for dental base polymers |
001_510K_Summary_v1
{7}
Partial Flex® 510(K) Premarket Notification
CLEMDE
# 7. Nonclinical Performance Testing
Partial Flex® has been tested in the following test modes:
- Bench testing per Iso 20795.
- Biocompatibility:
- INTRACUTANEOUS INJECTION TEST - per Iso 10993 passed the acceptance criteria
- KLIGMAN MAXIMIZATION TEST - per Iso 10993 passed the acceptance criteria
- L929 MEM ELUTION TEST - per Iso 10993 passed the acceptance criteria
The results of this nonclinical testing show that the PARTIAL FLEX is sufficient for its intended to used and is substantially equivalent to the legally marketed predicate device.
# 8. Substantial Equivalence Summary / Conclusion
Conclusion Based on available 510(k) information provided here in, Partial Flex is considered substantial equivalent to the predicate devices in terms of indications for use, material, technology, design, and performance specifications.
001_510K_Summary_v1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.