BRANEMARK SYSTEMS MKII SELK-TAPPING FIXTURE
Device Facts
| Record ID | K962130 |
|---|---|
| Device Name | BRANEMARK SYSTEMS MKII SELK-TAPPING FIXTURE |
| Applicant | Nobelpharma USA, Inc. |
| Product Code | DZE · Dental |
| Decision Date | Jul 22, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 872.3640 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The Nobelpharma BRÄNEMARK SYSTEM - MK II Self-Tapping Fixtures are intended to be used in edentulous and partially edentulous patients to restore chewing functions.
Device Story
Endosseous dental implant; Mk II Self-Tapping Fixture; implanted in upper/lower jaw bone; supports prosthetic devices to restore chewing function. Modification to 5.0mm and 5.5mm diameter fixtures: increased internal thread diameter for abutment screws; increased hexagonal dimension for abutment seating; 5.5mm fixture includes minor flange (0.4mm) and apex (0.3mm) diameter increases. Used by dental clinicians in surgical/prosthetic procedures.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Endosseous dental implant; titanium material (implied by system lineage); threaded design; varying diameters (3.75mm to 5.5mm). Modifications: increased internal thread diameter, increased hex dimension, minor flange/apex diameter adjustments.
Indications for Use
Indicated for edentulous and partially edentulous patients requiring restoration of chewing function via endosseous dental implants.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Predicate Devices
- Mk II Self-Tapping Fixture (K945398)