K083550 · Straumann USA · DZE · Feb 26, 2009 · Dental
Device Facts
Record ID
K083550
Device Name
STRAUMANN DENTAL IMPLANT SYSTEM
Applicant
Straumann USA
Product Code
DZE · Dental
Decision Date
Feb 26, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Straumann® dental implants are suitable for the treatment of oral endosteal implantation in the upper and lower jaw and for the functional and esthetic oral rehabilitation of edentulous and partially dentate patients (unless specific indications and limitations are present, as stated below). Straumann® dental implants can also be used for immediate or early implantation following extraction or loss of natural teeth. Implants can be placed with immediate function on single-tooth and/or multiple tooth applications when good primary stability is achieved and with appropriate occlusal loading, to restore chewing function. The prosthetic restorations used are single crowns, bridges and partial or full dentures, which are connected to the implants by the corresponding elements (abutments). When placing implants in the posterior region, we recommend using only large diameter implants. In cases of fully edentulous patients, 4 or more implants must be used in immediately loaded cases. Specific indications for small diameter (Ø 3.3 mm) implants: Because of their reduced mechanical stability, small diameter implants are only used in cases with a low mechanical load. Placement in the molar region is not recommended.
Device Story
System of endosseous dental implants, abutments, and surgical instruments used for oral rehabilitation. Implants are surgically placed into the upper or lower jaw to support prosthetic restorations (crowns, bridges, dentures). Designed for immediate or early implantation following tooth loss. Requires clinician to achieve primary stability and apply appropriate occlusal loading. Small diameter implants (3.3 mm) are restricted to low mechanical load applications due to reduced stability. Benefits include restoration of chewing function and esthetics for edentulous or partially dentate patients.
Clinical Evidence
No clinical data provided; substantial equivalence is based on design, intended use, and bench testing.
Technological Characteristics
Endosseous root-form dental implants. Materials include TiZr (Titanium-Zirconium) alloy. Features SLActive surface treatment. Available in various diameters including 3.3 mm (Standard, Standard Plus, and Bone Level configurations).
Indications for Use
Indicated for edentulous and partially dentate patients requiring oral endosteal implantation in the upper or lower jaw for functional and esthetic rehabilitation. Suitable for immediate or early implantation post-extraction or tooth loss. Small diameter (3.3 mm) implants are indicated only for low mechanical load cases; molar region placement is contraindicated for small diameter implants.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
K083550
## Section I 510(k) SUMMARY
FEB 2 6 2009
#### 1. Applicant's Name and Address
Straumann US (on behalf of Institut Straumann AG) 60 Minuteman Rd. Andover, MA 01810 978-747-2513 Telephone Number: 978-747-0023 Fax Number: Elaine Alan Contact Person: Regulatory Affairs Specialist Date of Submission: November 26, 2008
#### 2. Name of the Device
Trade Name: Common Name: Classification Name: Requlation Number:
Straumann Dental Implant System Dental implant Implant, Endosseous, Root-Form 21 CFR 872.3640
### 3. Legally Marketed Device to which Equivalence is Claimed (Predicate Device)
Straumann Dental Implant System, K053088 and K062129
#### 4. Description of the Device
The Straumann Dental Implant System is an integrated system of endosseous dental implants, which are designed to support prosthetic devices for partially or fully edentulous patients. The system consists of a variety of dental implants, abutments and surgical and prosthetic parts and instruments.
#### 5. Intended Use of the Device
Straumann® dental implants are suitable for the treatment of oral endosteal implantation in the upper and lower jaw and for the functional and esthetic oral rehabilitation of edentulous and partially dentate patients (unless specific indications and limitations are present, as stated below). Straumann® dental implants can also be used for immediate or early implantation following extraction or loss of natural teeth. Implants can be placed with immediate function on single-tooth and/or multiple tooth applications when good primary stability is achieved and with appropriate occlusal loading, to restore chewing function. The prosthetic restorations used are single crowns, bridges and partial or full dentures, which are connected to the implants by the corresponding elements (abutments). When placing implants in the posterior region, we recommend using only large diameter implants. In cases of fully
510(k) Submission: Modified S/SP/BL 3.3 Implants November 26, 2008
Straumann US Page 68
{1}------------------------------------------------
edentulous patients, 4 or more implants must be used in immediately loaded cases.
Specific indications for small diameter (Ø 3.3 mm) implants: Because of their reduced mechanical stability, small diameter implants are only used in cases with a low mechanical load. Placement in the molar region is not recommended.
#### 6. Technological Characteristics
The modified SLActive Ø3.3mm RN Standard, Standard Plus and Ø3.3mm NC Bone Level TiZr implants are substantially equivalent to the currently marketed devices. The intended use is identical to the predicate devices. The proposed devices have the same basic design and fundamental operating principles to the currently cleared devices.
510(k) Submission: Modified S/SP/BL 3.3 Implants November 26, 2008
Straumann US Page 69
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes on its body, representing service, science, and protection. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# FEB 2 6 2009
Ms. Elaine Alan Regulatory Affairs Specialist Straumann USA 60 Minuteman Road Andover, Massachusetts 01810
Re: K083550
> Trade/Device Name; Modified Dental Implant Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE Dated: November 26, 2008 Received: December 1, 2009
Dear Ms. Alan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
### Page 2 - Ms. Alan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474; For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Clinton, D. Watson Jr.
Ginette Y. Michaud, M.D. Acting Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K08 3550
## Section A INDICATIONS FOR USE STATEMENT
Device Name: Modified Dental Implant
Indications for Use:
Straimann® dental implants are suitable for the treatment of oral endosteal implantation in the upper and lower jaw and for the functional and esthetic oral rehabilitation of edentulous and partially dentate patients (unless specific indications and limitations are present, as stated below). Straumann® dental implants can also be used for immediate or early implantation following extraction or loss of natural teeth. Implants can be placed with immediate function on single-tooth and/or multiple tooth applications when good primary stability is achieved and with appropriate occlusal loading, to restore chewing function. The prosthetic restorations used are single crowns, bridges and partial or full dentures, which are connected to the implants by the corresponding elements (abutments). When placing implants in the posterior region, we recommend using only large diameter implants. In cases of fully edentulous patients, 4 or more implants must be used in immediately loaded cases.
Specific indications for small diameter (Ø 3.3 mm) implants: Because of their reduced mechanical stability, small diameter implants are only used in cases with a low mechanical load. Placement in the molar region is not recommended.
X Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | <img src="data:image/png;base64,iVBORw0KGgoAAAANSUhEUgAAAAUAAAAFCAYAAACNbyblAAAAHElEQVQI12P4//8/w+Y4gImE5f85AQD0HwJ0gO1iogAAAABJRU5ErkJggg=="/> |
|--|-------------------------------------------------------------------------------------------------------------------------------------------------|
| | (Division Sign-Off) |
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices Straumann US
| Page 1 of 1 | Page 5 |
|--------------------------------------------------|--------|
| 510(k) Submission: Modified S/SP/BL 3.3 Implants | |
| November 26, 2008 | |
| 510(k) Number: | K083550 |
|----------------|---------|
|----------------|---------|