K061971 · Nobel Biocare AB · DZE · Mar 6, 2007 · Dental
Device Facts
Record ID
K061971
Device Name
ZIRCONIA IMPLANT
Applicant
Nobel Biocare AB
Product Code
DZE · Dental
Decision Date
Mar 6, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Nobel Biocare's Zirconia Implants are root-form endosseous dental implants intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient esthetics and chewing function. Nobel Biocare's Zirconia Implants are indicated for single or multiple unit restorations in splinted or non-splinted applications. Nobel Biocare Zirconia Implants may be placed immediately and put into immediate function providing that the initial stability requirements detailed in the surgical manual are satisfied.
Device Story
Threaded, root-form dental implant; manufactured from Y2O3 stabilized zirconium oxide; features integrated abutment. Surgically placed in upper or lower jaw bone by dental professionals to support prosthetic devices (artificial teeth). Restores esthetics and chewing function in partially or fully edentulous patients. Supports single or multiple unit restorations in splinted or non-splinted applications. Allows for immediate placement and function if specific stability criteria are met.
Clinical Evidence
No clinical data provided; substantial equivalence based on design and material characteristics.
Technological Characteristics
Material: Y2O3 stabilized zirconium oxide. Design: Threaded, root-form endosseous dental implant with integrated abutment. Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for partially or fully edentulous patients requiring single or multiple unit restorations in the upper or lower jaw. Supports prosthetic devices to restore esthetics and chewing function. May be placed immediately if initial stability requirements are met.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
MAR O 6 2007
## 510(k) Summary of Safety and Effectiveness 1 4
| Submitted by: | Herbert Crane<br>Director Requlatory Affairs |
|----------------------------------------|----------------------------------------------------------------------------------|
| Address: | Nobel Biocare USA LLC<br>22715 Savi Ranch Parkway<br>Yorba Linda, CA 92887 |
| Telephone: | (714) 282-4800, ext. 5074 |
| Facsimile: | (714) 282-9023 |
| Date of Submission: | July 11, 2006 |
| Classification Name: | Endosseous Dental Implant (21 CFR 872.3640) |
| Trade or Proprietary<br>or Model Name: | Zirconia Implant |
| Legally Marketed Device(s): | Replace One Piece Implant (K023952)<br>Branemark System Dental Implant (K022562) |
## Device Description:
Nobel Biocare Zirconia Implants are threaded, root-form dental implants intended for use in the upper and/or lower jaw to support prosthetic devices, such as artificial teeth, in order to restore patient esthetics and chewing function to partially or fully edentulous patients. The Zirconia implant is manufactured from Y20 stabilized zirconium oxide. The Zirconia Implant has an integrated abutment that eliminates the need for attaching an abutment.
## Indications for Use:
Nobel Biocare's Zirconia Implants are root-form endosseous dental implants intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient esthetics and chewing function. Nobel Biocare's Zirconia Implants are indicated for single or multiple unit restorations in splinted applications. Nobel Biocare Zirconia Implant may be placed immediately and put into immediate function providing that the initial stability requirements detailed in the surgical manual are satisfied.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features an abstract image of an eagle with three lines forming its wings and body. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" are arranged in a circular pattern around the eagle. The seal is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Nobel Biocare AB C/O Mr. Herbert Crane Director, Regulatory Affairs Nobel Biocare USA, LLC 22715 Savi Ranch Parkway Yorba Linda, California 92887
MAR 0 6 2007
Re: K061971
Trade/Device Name: Zirconia Implant Regulation Number: 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE Dated: March 1, 2007 Received: March 2, 2007
Dear Mr. Crane:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page 2 - Mr. Crane
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sylette Y. Mcchain, MD.
Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
K661971 510(k) Number (if known):
Device Name: Zirconia Implant
Indications For Use:
Nobel Biocare's Zirconia Implants are root-form endosseous dental implants intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient esthetics and chewing function. Nobel Biocare's Zirconia Implants are indicated for single or multiple unit restorations in splinted or non-splinted applications. Nobel Biocare Zirconia Implants may be placed immediately and put into immediate function providing that the initial stability requirements detailed in the surgical manual are satisfied.