K140125 · New Stetic · OIV · Apr 29, 2014 · Dental
Device Facts
Record ID
K140125
Device Name
NEW STETIC DENTAL AMALGAM ALLOY
Applicant
New Stetic
Product Code
OIV · Dental
Decision Date
Apr 29, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3070
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
New Stetic NU Alloy DP 40, NU Alloy DP, NU Alloy DP Active and Micronic Dental Amalgam Alloys are commonly used as a filling material for restoring the morphology and function of posterior teeth (cavities class I and II in molars and premolars), when these have lost their structure due to injuries such as cavities or fractures.
Device Story
Dental amalgam alloy consisting of silver, tin, and copper; zinc-free. Supplied as pre-dosed capsules, tablets, or powder. Intended for use by dental professionals as a restorative filling material for posterior teeth (Class I and II cavities). Operates by mixing alloy with elemental mercury to form a restorative material. Applied in dental clinics to restore tooth structure lost to decay or fracture. Benefits include restoration of tooth morphology and function. No software or electronic components involved.
Clinical Evidence
Bench testing only. Performance testing conducted in accordance with ISO 24234:2004 to verify physical properties (compressive strength, creep, dimensional change) and ensure freedom from contamination. Biocompatibility assessment performed via literature review, concluding no new chemical components or additives compared to predicate devices.
Technological Characteristics
Composition: Silver, tin, copper (zinc-free). Form: Pre-dosed capsules, tablets, or powder. Mixing: Manual or mechanical trituration with mercury. Standards: ISO 24234:2004. No software, electronics, or energy source.
Indications for Use
Indicated for patients requiring restoration of posterior teeth (molars and premolars) with Class I or II cavities or fractures to restore morphology and function.
Regulatory Classification
Identification
Dental amalgam is a device that consists of a combination of elemental mercury, supplied as a liquid in bulk, sachet, or predosed capsule form, and amalgam alloy composed primarily of silver, tin, and copper, supplied as a powder in bulk, tablet, or predosed capsule form, for the direct filling of carious lesions or structural defects in teeth. This device also includes the individual component devices, mercury and amalgam alloy, when intended to be combined with each other to form dental amalgam.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Dental Amalgam, Mercury, and Amalgam Alloy.”*See* § 872.1(e) for the availability of this guidance document.
{1}------------------------------------------------
K140125
# APR 2 9 2014
### 510(k) SUMMARY (as required by 807.92)
| Regulatory Correspondent: | AJW Technology Consultants, Inc.<br>445 Apollo Beach Blvd.<br>Apollo Beach, FL 33572 USA<br>John Obrien<br>Phone: 813-645-2855<br>Fax: 813-645-2856<br>Email: jobrien@ajwtech.com |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter of 510(k): | New Stetic S.A.<br>Carrera 53 Nº 50-09<br>Guarne - Antioquia, Colombia<br>Sandra Maria Montoya<br>Email: smontoya@newstetic.com<br>Phone: +57 4 550 00 00<br>Fax: +57 4 551 31 34 |
| Date of Summary: | December 17, 2013 |
| Trade/Proprietary Name: | New Stetic Dental Amalgam Alloy |
| Common/Usual Name: | Dental Amalgam, Mercury and Amalgam Alloy |
| Classification Name: | Class II |
| Product Code: | OIV |
| Regulation | 21 CFR 872.3070 |
#### Intended Use:
New Stetic NU Alloy DP 40, NU Alloy DP, NU Alloy DP Active and Micronic Dental Amalgam Alloys are commonly used as a filling material for restoring the morphology and function of posterior teeth (cavities class I and II in molars and premolars), when these have lost their structure due to injuries such as cavities or fractures.
{2}------------------------------------------------
#### Device Description:
New Stetic Dental Amalgam Alloys for dental amalgam used as a filling material for oral cavities. They are made of silver, tin and copper and free of zinc. The products are packaged in different forms: pre-dosed capsules, tablets and powder.
Note: Class II Special Controls Guidance Document: Dental Amalgam, Mercury, and Amalgam Alloy - Guidance for Industry and FDA Staff was used
| Predicate Device: | K940675 - Sybraloy |
|-------------------|----------------------------------------------|
| | K820181 - Sybron/Kerr Contour |
| | K820967 - Dispersalloy-Dispersed Phase Alloy |
#### Substantial Equivalence:
The New Stetic Dental Amalgam Alloys are substantially equivalent in intended use and technological characteristics to the Sybron Dental Specialties, Inc "Sybraloy" under K940675, the Sybron Corp "Sybron/Kerr Contour under K820181 and the Johnson & Johnson "Dispersalloy-Dispersed Phase Alloy" under K820967. Any difference that exists between the New Stetic Amalgam and the predicate devices has no negative effect on safety or effectiveness.
{3}------------------------------------------------
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
.
| Product Name<br>Descriptive<br>Information | SYBRALOY | CONTOUR | DISPERSALLOY | NU ALLOY DP 40 | Differences & Remarks |
|------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|---------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) | K940675 | K820181 | K802967 | | |
| INTENDED USE | Used to provide a<br>filling material<br>used in restorative<br>dentistry when<br>mixed with<br>mercury | Intended for use as<br>a filling material for<br>oral cavities. | Restorative material<br>in dental practice. | New Stetic dental<br>amalgam alloys are<br>commonly used as a<br>filling material for<br>restoring the<br>morphology and<br>function of posterior<br>teeth (cavities class l<br>and II in molars and<br>premolars), when<br>these have lost their<br>structure due to<br>injuries such as<br>cavities or fractures. | None.<br>All products are intended<br>to be used as a filling<br>material in restorative<br>dentistry. |
| CHEMICAL COMPOSITION | | | | | |
| Silver | 39.4 - 43.7% | 41% | 70% | 40% | The content of each<br>component meets the<br>requirements of ISO<br>24234:2004. |
| Tin | 28.8 - 32.5% | 31% | 16% | 31.8% | |
| Copper | 25.8 - 29.8% | 28% | 13% | 28.21% | |
| Zinc | 0% | 0% | 1% | 0% | |
| | 1:0,9 - 1:1,0 | | | 1:0.9 | |
| Alloy-Mercury Ratio | (Mercury: 45 -<br>51%) | 1:0,9<br>(Mercury: 47%) | 1:1<br>(Mercury: 50%) | (Mercury: 47.5%) | |
| PHYSICAL PROPERTIES | | | | | |
| Compressive strength (MPa)<br>1h:<br>ISO 24234 Specification: 80<br>MPa min<br>(Results of Regular Set Tablet) | 185 MPa<br>(26832 PSI) | 124,11 MPa<br>(18000 PSI) | 127,5 MPa<br>(18500 PSI) | 87 MPa<br>(12618 PSI)<br>Result of Lot N°:<br>020713B | Although the results are<br>not exactly the same, our<br>tests are carried out<br>according to ISO 24234-<br>2004 and all products are<br>within specifications,<br>which ensure the suitable |
| Compressive strength (MPa)<br>24h:<br>ISO 24234 Specification: 300<br>MPa min<br>(Results of Regular Set Tablet) | 534 MPa<br>(77450 PSI) | 441,26 MPa<br>(64000 PSI) | 427,5 MPa<br>(62000 PSI) | 364 MPa<br>(52794 PSI)<br>Result of Lot Nº:<br>020713B | |
| Maximum Creep (%):<br>ISO 24234 Specification: 2.0%<br>max<br>(Results of Regular Set Tablet) | 0.08% | 0.10% max | 0.86% | 0.23%<br>Result of Lot Nº:<br>020713B | |
| Dimensional Change (%):<br>ISO 24234 Specification:<br>-0.10 to +0.20 %<br>(Results of Regular Set Tablet) | 0.06% | -0.1% | 0.03% | -0.05% | performance of<br>restoration. |
| Particle shape | Spherical | Spherical and<br>Lathe-cut | Spherical and Lathe-<br>cut | Spherical and Lathe-cut | This parameter is not<br>specified by a technical<br>standard. It depends on<br>the characteristics of<br>product. This<br>characteristic does not<br>affect the performance of<br>the restoration. |
| Trituration Time (seconds):<br>High-speed Amalgamator,<br>Product form: Capsules | 7 - 10 seconds | 8 - 10 seconds | 10 - 14 seconds | 11 seconds | This property depends on<br>the characteristics of the<br>product, however, the<br>variations in the<br>amalgamator (machine<br>type, age, line voltage),<br>even of the same brand,<br>can modify the speed and<br>time. It does not affect the<br>safety and effectiveness of<br>the product. |
| Working time (minutes)<br>Regular Set Capsules | 12 min. maximum | 12 min. maximum | Not reported | 8 min. maximum | This parameter is not<br>specified by a technical<br>standard. It depends on the<br>characteristics of the<br>product. This difference does<br>not affect the performance<br>of the restoration. |
| | CAS 7440-22-4 | | | | |
| | CAS 7440-31-5 | | | | |
| | CAS 7440-50-8 | | | | |
K140125 - RTA Response
ക്ഷ
{4}------------------------------------------------
| | Descriptive<br>Information | Product Name | SYBRALOY | CONTOUR | DISPERSALLOY | NU ALLOY DP | Differences & Remarks |
|-----------------------------------------------------------------------------------------------------------------|----------------------------|----------------------|------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|---------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| INTENDED USE | | | Used to provide a<br>filling material used<br>in restorative<br>dentistry when<br>mixed with mercury | Intended for use as<br>a filling material for<br>oral cavities. | Restorative material<br>in dental practice. | New Stetic dental<br>amalgam alloys are<br>commonly used as a<br>filling material for<br>restoring the<br>morphology and<br>function of posterior<br>teeth (cavities class I<br>and II in molars and<br>premolars), when<br>these have lost their<br>structure due to<br>injuries such as<br>cavities or fractures. | None.<br>Both products are intended<br>to be used as a filling<br>material in restorative<br>dentistry. |
| Alloy | CHEMICAL COMPOSITION | | | | | | |
| | | Silver CAS 7440-22-4 | 39.4 – 43.7% | 41% | 70% | 44.85% | |
| | | Tin CAS 7440-31-5 | 28.8 – 32.5% | 31% | 16% | 31.95% | |
| | | Copper CAS 7440-50-8 | 25.8 – 29.8% | 28% | 13% | 23.24% | The content of each<br>component meets the<br>requirements of ISO<br>24234:2004. |
| | | Zinc | 0% | 0% | 1% | 0% | |
| | Alloy-Mercury Ratio | | 1:0,9 – 1:1,0<br>(Mercury: 45 –<br>51%) | 1:0,9<br>(Mercury: 47%) | 1:1,0<br>(Mercury: 50%) | 1:0.98<br>(Mercury: 49,6%) | |
| | | PHYSICAL PROPERTIES | | | | | |
| Compressive strength (MPa)<br>1h<br>ISO 24234 Specification: 80<br>MPa min<br>(Results of Regular Set Tablet) | | | 185 MPa<br>(26832 PSI) | 124,1 MPa<br>(18000 PSI) | 127,5 MPa<br>(18500 PSI) | 182 MPa<br>(26397 PSI)<br>Result of Lot N°: 010913 | Although the results are not<br>exactly the same, the tests<br>are carried out according to<br>ISO 24234-2004 and all<br>products are within<br>specifications, which ensure<br>the suitable performance of<br>restoration. |
| Compressive strength (MPa)<br>24h<br>ISO 24234 Specification: 300<br>MPa min<br>(Results of Regular Set Tablet) | | | 534 MPa<br>(77450 PSI) | 441,26 MPa<br>(64000 PSI) | 427,5 MPa<br>(62000 PSI) | 526 MPa<br>(76290 PSI)<br>Result of Lot N°:<br>010913 | |
| Maximum Creep (%):<br>ISO 24234 Specification: 2.0%<br>max<br>(Results of Regular Set Tablet) | | | 0.08% | 0.10% max | 0.86% max | 0.30%<br>Result of Lot N°:<br>010913 | |
| Dimensional Change (%):<br>ISO 24234 Specification:<br>-0.10 to +0.20 %<br>(Results of Regular Set Tablet) | | | 0.06% | -0.1% | 0.03% | -0.05% | |
| Particle shape | | | Spherical | Spherical and<br>Lathe-cut | Spherical and Lathe-<br>cut | Spherical and Lathe-cut | This parameter is not<br>specified by a technical<br>standard. It depends on the<br>characteristics of product.<br>This characteristic does not<br>affect the performance of the<br>restoration. |
| Trituration Time (seconds)<br>High-speed Amalgamator,<br>Product form: Capsules | | | 7 - 10 seconds | 8 - 10 seconds | 10 - 14 seconds | 12 - 14 seconds | This property depends on<br>the characteristics of the<br>product, however, the<br>variations in the<br>amalgamator (machine type,<br>age, line voltage), even of<br>the same brand, can modify<br>the speed and time. It does<br>not affect the safety and<br>effectiveness of the product. |
| Working time (minutes)<br>Regular Set Capsules | | | 12 min. maximum | 12 min. maximum | Not reported | 8 min. maximum | This parameter is not specified<br>by a technical standard. It<br>depends on the characteristics<br>of the product. This difference<br>does not affect the<br>performance of the<br>restoration. |
-
{5}------------------------------------------------
| | Descriptive<br>Information | Product Name | SYBRALOY | CONTOUR | DISPERSALLOY | NU ALLOY DP<br>ACTIVE | Differences & Remarks | |
|-----------------------------------------------------------------------------------------------------------------|----------------------------|------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|---------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| INTENDED USE | | | Used to provide a<br>filling material used<br>in restorative<br>dentistry when<br>mixed with mercury | Intended for use as<br>a filling material for<br>oral cavities. | Restorative material<br>in dental practice. | New Stetic dental<br>amalgam alloys are<br>commonly used as a<br>filling material for<br>restoring the<br>morphology and<br>function of posterior<br>teeth (cavities class I<br>and II in molars and<br>premolars), when<br>these have lost their<br>structure due to<br>injuries such as<br>cavities or fractures. | None.<br>Both products are intended to<br>be used as a filling material in<br>restorative dentistry. | |
| | CHEMICAL COMPOSITION | | | | | | | |
| | | Silver<br>CAS 7440-22-4 | 39.4 - 43.7% | 41% | 70% | 44.74% | | |
| Alloy | | Tin<br>CAS 7440-31-5 | 28.8 - 32.5% | 31% | 16% | 31.54% | The content of each | |
| | | Copper<br>CAS 7440-50-8 | 25.8 - 29.8% | 28% | 13% | 23.96% | component meets the | |
| | | Zinc | 0% | 0% | 1% | 0% | requirements of ISO | |
| | Alloy-Mercury Ratio | | 1:0,9 - 1:1,0<br>(Mercury: 45 - 51%) | 1:0.9<br>(Mercury: 47%) | 1:1,0<br>(Mercury: 50%) | 1:0.98<br>(Mercury: 49.6%) | 24234:2004. | |
| | | PHYSICAL PROPERTIES | | | | | | |
| | | Compressive strength (MPa) 1h<br>ISO 24234 Specification: 80 MPa min<br>(Results of Regular Set Tablet) | | 185 MPa<br>(26832 PSI) | 124,11 MPa<br>(18000 PSI) | 127,5 MPa<br>(18500 PSI) | 159 MPa<br>(23061 PSI)<br>Result of Lot Nº: 010713 | |
| | | Compressive strength (MPa) 24h<br>ISO 24234 Specification: 300 MPa min<br>(Results of Regular Set Tablet) | | 534 MPa<br>(77450 PSI) | 441,26 MPa<br>(64000 PSI) | 427,5 MPa<br>(62000 PSI) | 500 MPa<br>(72519 PSI)<br>Result of Lot Nº:<br>010713 | Although the results are not<br>exactly the same, the tests<br>are carried out according to<br>ISO 24234-2004 and all<br>products are within<br>specifications, which ensure<br>the suitable performance of<br>restoration. |
| | | Maximum Creep (%):<br>ISO 24234 Specification: 2.0% max<br>(Results of Regular Set Tablet) | | 0.08% | 0.10% max | 0.86% max | 0.18%<br>Result of Lot Nº:<br>010713 | |
| | | Dimensional Change (%):<br>ISO 24234 Specification:<br>-0.10 to +0.20 %<br>(Results of Regular Set Tablet) | | 0.06% | -0.1% | 0.03% | -0.06% | |
| | | Particle shape | Spherical | Spherical and<br>Lathe-cut | Spherical and Lathe<br>cut | Spherical and Lathe cut | This parameter is not<br>specified by a technical<br>standard. It depends on the<br>characteristics of product.<br>This characteristic does not<br>affect the performance of the<br>restoration. | |
| | | Trituration Time (seconds)<br>High-speed Amalgamator,<br>Product form: Capsules | 7 - 10 seconds | 8 - 10 seconds | 10 - 14 seconds | 12 seconds | This property depends on the<br>characteristics of the product,<br>however, the variations in the<br>amalgamator (machine type,<br>age, line voltage), even of the<br>same brand, can modify the<br>speed and time. It does not<br>affect the safety and<br>effectiveness of the product. | |
| | | Working time (minutes)<br>Regular Set Capsules | 12 min. maximum | 12 min. maximum | Not reported | 8 min. maximum | This parameter is not specified<br>by a technical standard. It<br>depends on the characteristics<br>of the product. This difference<br>does not affect the performance<br>of the restoration. | |
| Product Name<br>Descriptive<br>Information | | | SYBRALOY | CONTOUR | DISPERSALLOY | MICRONIC | Differences & Remarks | |
| INTENDED USE | | | Used to provide a<br>filling material used<br>in restorative<br>dentistry when<br>mixed with mercury | Intended for use as<br>a filling material for<br>oral cavities. | Restorative material<br>in dental practice. | New Stetic dental<br>amalgam alloys are<br>commonly used as a<br>filling material for<br>restoring the<br>morphology and<br>function of posterior<br>teeth (cavities class I<br>and II in molars and<br>premolars), when these<br>have lost their structure<br>due to injuries such as<br>cavities or fractures. | None.<br>Both products are intended to<br>be used as a filling material<br>in restorative dentistry. | |
| | | CHEMICAL COMPOSITION | | | | | | |
| | Silver | CAS 7440-22-4 | 39.4 - 43.7% | 41% | 70% | 70.14% | | |
| Alloy | Tin | CAS 7440-31-5 | 28.8 - 32.5% | 31% | 16% | 26.89% | The content of each | |
| | Copper | CAS 7440-50-8 | 25.8 - 29.8% | 28% | 13% | 3.86% | component meets the | |
| | Zinc | | 0% | 0% | 1% | 0% | requirements of ISO | |
| | Alloy-Mercury Ratio | | 1:0,9 - 1:1,0<br>(Mercury: 45 -<br>51%) | 1:0,9<br>(Mercury: 47%) | 1:1<br>(Mercury: 50%) | 1:1.1 - 1:1.2<br>(Mercury: 52.4% -<br>54.5%) | 24234:2004. | |
| | | PHYSICAL PROPERTIES | | | | | | |
| Compressive strength (MPa) 1h<br>ISO 24234 Specification: 80<br>MPa min<br>(Results of Regular Set Tablet) | | | 185 MPa<br>(26832 PSI) | 124,11 MPa<br>(18000 PSI) | 127,5 MPa<br>(18500 PSI) | 183 MPa<br>(26542 PSI)<br>Result of Lot Nº: 010913M | Although the results are not | |
| Compressive strength (MPa)<br>24h<br>ISO 24234 Specification: 300<br>MPa min<br>(Results of Regular Set Tablet) | | | 534 MPa<br>(77450 PSI) | 441,26 MPa<br>(64000 PSI) | 427,5 MPa<br>(62000 PSI) | 398 MPa<br>(57725 PSI)<br>Result of Lot Nº:<br>010913M | exactly the same, the tests<br>are carried out according to<br>ISO 24234-2004 and all<br>products are within<br>specifications, which ensure | |
| Maximum Creep (%):<br>ISO 24234 Specification: 2.0%<br>max<br>(Results of Regular Set Tablet) | | | 0.08% | 0.10% max | 0.86% max | 0.20%<br>Result of Lot Nº:<br>010913M | the suitable performance of<br>restoration. | |
| Dimensional Change (%):<br>ISO 24234 Specification:<br>-0.10 to +0.20 %<br>(Results of Regular Set Tablet) | | | 0.06% | -0.1% | 0.03% | -0.03% | | |
| Particle shape | | | Spherical | Spherical and<br>Lathe-cut…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.