← Product Code [NPM](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NPM) · K231672

# S1 (K231672)

_Medpark Co., Ltd. · NPM · Oct 13, 2023 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NPM/K231672

## Device Facts

- **Applicant:** Medpark Co., Ltd.
- **Product Code:** [NPM](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NPM.md)
- **Decision Date:** Oct 13, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3930
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

Augmentation or reconstructive treatment of the alveolar ridge; Filling of infrabony periodontal defects; Filling of defects after root resection, apicocectomy, and cystectomy; Filling of extraction sockets to enhance preservation of the alveolar ridge; Elevation of the maxillary sinus floor; Filling of periodontal defects in conjuncts intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR); and Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR).

## Device Story

S1 is a bone grafting material consisting of bovine-derived mineralized hydroxyapatite (HAp) mixed with hydroxypropyl methylcellulose (HPMC). Supplied as sterile granules (powder or chip form) in vials for single-use. Used by clinicians in maxillofacial surgery, implantology, and periodontology to fill and augment bony voids. The porous matrix supports bone formation. The device is implanted into the surgical site; it acts as a scaffold for new bone growth and is gradually resorbed over 24+ months. Clinical benefit includes preservation of alveolar ridge and successful bone regeneration in dental defects. The addition of HPMC improves handling compared to pure HAp grafts.

## Clinical Evidence

No human clinical data provided. Evidence consists of bench testing (ISO 13779-3, ASTM F1581-08) and an animal study (beagle mandibular defect model). Animal study compared S1 to Bio-Oss® at 4, 8, and 12 weeks using radiographic, Micro CT, histology, and histomorphometry analyses, demonstrating equivalent bone formation performance. Biocompatibility testing performed per ISO 10993-1 and FDA guidance for dental bone grafting devices. Sterilization validated to SAL 10^-6.

## Technological Characteristics

Material: Bovine cancellous bone-derived hydroxyapatite (HAp) and hydroxypropyl methylcellulose (HPMC). Form: Granules (0.2-1.0mm or 1.0-2.0mm). Sterilization: Gamma irradiation. Standards: ISO 13779-3 (crystallinity, Ca/P ratio), ASTM F1581-08 (heavy metals), ISO 10993 (biocompatibility), ISO 22442 (animal tissue derivatives).

## Regulatory Identification

Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.

## Special Controls

*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Geistlich Bio-Oss® ([K122894](/device/K122894.md))

## Reference Devices

- MBCP Gel™ ([K060732](/device/K060732.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a seal with an abstract image of a human figure, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in three lines, with "FDA" in a larger font size than the rest of the text. The logo is clean and professional, and it is easily recognizable as the symbol of the FDA.

October 13, 2023

MedPark Co., Ltd. Woojong Kim Official Correspondent Busan Venture Tower 606-ho, 22, Mora-ro Sasang-gu, Busan KOREA, SOUTH

Re: K231672

Trade/Device Name: S1 Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: Class II Product Code: NPM Dated: July 04, 2023 Received: July 17, 2023

Dear Woojong Kim:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Sherrill Lathrop Blitzer

for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices

{2}------------------------------------------------

Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K231672

Device Name

ਟ I

#### Indications for Use (Describe)

S1 is intended for the following uses:

· Augmentation or reconstructive treatment of the alveolar ridge;

- · Filling of infrabony periodontal defects
- · Filling of defects after root resection, apicocectomy, and cystectomy
- · Filling of extraction sockets to enhance preservation of the alveolar ridge
- · Elevation of the maxillary sinus floor

· Filling of periodontal defects in conjuncts intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR); and

· Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR).

| Type of Use (Select one or both, as applicable)                                                                                                                                         |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <input checked="true" type="checkbox"/> Research involving NIH-funded Stem Cells   </div> <div> <input type="checkbox"/> Grantee-owned or held Stem Cells for future use   </div> |

> Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

| Image: MedPark logo<br>Your Reliable Partner | <b>510(k) Submission Files</b> | Revision No.<br>0                |
|----------------------------------------------|--------------------------------|----------------------------------|
|                                              |                                | Revision Date<br>October-13-2023 |
|                                              | S1                             | Page<br>1 / 7                    |

# K231672

# Submitter

MedPark Co., Ltd. Woojong Kim 24, Nakdong-daero 1570 beon-gil, Buk-gu, Busan Republic of Korea Email: qm@medpark.net Phone: +82-51-301-8777 Fax: +82-50-5877-7778

# Official Correspondent

MedPark Co., Ltd. Woojong Kim Busan Venture tower 606-ho, 22, Mora-ro, Sasang-gu, Busan, Republic of Korea Email: qm@medpark.net Phone: +82-51-301-8777 Fax: +82-50-5877-7778

## Device Information

- Trade Name: S1
- Common Name: Bone Grafting Material
- Classification Name: Bone Grafting Material, Animal Source
- Device Panel: Dental
- Regulation Number: 21 CFR 872.3930
- Product Code: NPM
- Device Class: Class II
- Date Prepared: O October/13/2023

## Primary Predicate

The subject device is substantially equivalent to the following predicate device

- · K122894, Geistlich Bio-Oss® manufactured by Geistlich Pharma AG
#### Reference Device

- K060732, MBCP Gel™ manufactured by BIOMATLANTE
#### Indication for Use

S1 is intended for the following uses:

- Augmentation or reconstructive treatment of the alveolar ridge

{5}------------------------------------------------

| Image: MedPark logo<br>Your Reliable Partner | <b>510(k) Submission Files</b> | Revision No.  | 0               |
|----------------------------------------------|--------------------------------|---------------|-----------------|
| S1                                           |                                | Revision Date | October-13-2023 |
|                                              |                                | Page          | 2 / 7           |

- . Filling of infrabony periodontal defects
- Filling of defects after root resection, apicocectomy, and cystectomy
- Filling of extraction sockets to enhance preservation of the alveolar ridge ●
- Elevation of the maxillary sinus floor
- · Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)
- Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR)

## Device Description

S1 is intended for filling and augmentation of bony voids and gaps in maxillofacial surgery, implantology, and periodontology according to the intended use of the product.

S1 is serving as a matrix consisting of interconnected macro- and micropores. The material is porous and has inner surface area.

S1 is a mixture of Hydroxyapatite (HAp) and hydroxypropyl methylcellulose (HPMC). Hydroxyapatite (HAp) is made from bovine cancellous bone and is mineralized hydroxyapatite.

The bone particle size is 0.2 ~ 1.0mm for the powder type and 1.0 ~ 2.0 mm for the chip type. S1 is packaged in vials and it is supplied sterile by gamma irradiation and is for single use only.

{6}------------------------------------------------

| Image: MedPark logo<br>Your Reliable Partner | <b>510(k) Submission Files</b> | Revision No.  | 0               |
|----------------------------------------------|--------------------------------|---------------|-----------------|
| S1                                           |                                | Revision Date | October-13-2023 |
|                                              |                                | Page          | 3 / 7           |

# Summary of Technological Characteristics

| Descriptive Information                      | Subject Device          | Primary Predicate                                                                                               | Reference Predicate                                           | Remark                                                                                                                                                                       |                                                                                                    |
|----------------------------------------------|-------------------------|-----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| K number                                     |                         | Unknown                                                                                                         | K122894                                                       | K060732                                                                                                                                                                      |                                                                                                    |
| Manufacturer                                 |                         | MedPark Co., Ltd.                                                                                               | Geistlich Pharma Ag                                           | BIOMATLANTE                                                                                                                                                                  |                                                                                                    |
| Model (Device name)<br>/ Type (Product name) |                         | S1                                                                                                              | Bio-Oss®                                                      | MBCP Gel™                                                                                                                                                                    |                                                                                                    |
| Product Code                                 |                         | NPM                                                                                                             | NPM                                                           | LYC                                                                                                                                                                          | Same                                                                                               |
| Regulation number                            |                         | 21CFR872.3930                                                                                                   | 21CFR872.3930                                                 | 21CFR872.3930                                                                                                                                                                | Same                                                                                               |
| Intended Use                                 |                         | Intended for filling of bone<br>defects and bone augmentation                                                   | Intended for filling of bone<br>defects and bone augmentation | Intended for use as bone<br>grafting material to fill,<br>augment or reconstruct osseous<br>bone defects in particular in<br>periodontal oral/maxillofacial<br>applications. | Same                                                                                               |
| Target Population                            |                         | Adults ( => 22 years old )                                                                                      | Adults                                                        | Adults                                                                                                                                                                       | Same                                                                                               |
| Anatomical site                              |                         | Oral, periodontal                                                                                               | Oral, periodontal                                             | Oral, periodontal                                                                                                                                                            | Same                                                                                               |
| Device<br>Design                             | Source<br>Bone          | Bovine bone<br>- Cancellous                                                                                     | Bovine bone<br>- Cancellous                                   | Biphasic Calcium Phosphate<br>(60% Hydroxyapatite and 40%<br>beta-Tricalcium Phosphate)                                                                                      | Similar                                                                                            |
|                                              | Particle size<br>range  | 0.2 - 1.0 mm<br>1.0 - 2.0 mm                                                                                    | 0.25 - 1.0 mm<br>1.0 - 2.0 mm                                 | 0.5 - 1.0 mm                                                                                                                                                                 | Similar                                                                                            |
|                                              | Form                    | Granules                                                                                                        | Granules                                                      | Granules (Putty type)                                                                                                                                                        | Same                                                                                               |
|                                              | Reusable                | Single Use only                                                                                                 | Single Use only                                               | Single Use only                                                                                                                                                              | Same                                                                                               |
|                                              | Sterilization           | Gamma Irradiation                                                                                               | Gamma Irradiation                                             | Not identified                                                                                                                                                               | Same                                                                                               |
| Composition<br>of Materials                  | Chemical<br>composition | Hydroxyapatite with excipient<br>vehicle of pharmaceutical grade<br>quality (hydroxypropyl-<br>methylcellulose) | Hydroxyapatite                                                | Biphasic Calcium Phosphate<br>with excipient vehicle of<br>pharmaceutical grade quality<br>(hydroxypropyl-<br>methylcellulose)                                               | Different<br>(Biocompatibility,<br>animal<br>performance, and<br>bench testing was<br>performed to |

{7}------------------------------------------------

| MedPark<br>Your Reliable Partner | 510(k) Submission Files | Revision No.  | 0               |
|----------------------------------|-------------------------|---------------|-----------------|
|                                  | S1                      | Revision Date | October-13-2023 |
|                                  |                         | Page          | 4 / 7           |

|                        |                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | demonstrate<br>substantial<br>equivalence.)                               |
|------------------------|-----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|
| Physical<br>Properties | Phase<br>purity | Ca5(PO4)3(OH) (≥95%)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Ca5(PO4)3(OH) (≥95%)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Not identified                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same                                                                      |
| Shelf-life             |                 | 2 years                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 3 years                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Not identified                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | The Validation of<br>Gamma<br>Sterilization<br>(Report No.<br>210107-E-2) |
| Performance            |                 | Bone formation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Bone formation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Bone formation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same                                                                      |
| Biocompatibility       |                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                      |
| Indication for use     |                 | S1 is intended for the following<br>uses:<br>• Augmentation or<br>reconstructive treatment of the<br>alveolar ridge;<br>• Filling of infrabony<br>periodontal defects;<br>• Filling of defects after root<br>resection, apicoectomy, and<br>cystectomy;<br>• Filling of extraction sockets<br>to enhance preservation of the<br>alveolar ridge;<br>• Elevation of the maxillary<br>sinus floor;<br>• Filling of periodontal<br>defects in conjunction with<br>products intended for Guided<br>Tissue Regeneration (GTR) and<br>Guided Bone Regeneration<br>(GBR); and<br>• Filling of peri-implant | Bio-OSS® is intended for the<br>following uses:<br>• Augmentation or<br>reconstructive treatment of the<br>alveolar ridge;<br>• Filling of infrabony<br>periodontal defects;<br>• Filling of defects after root<br>resection, apicoectomy, and<br>cystectomy;<br>• Filling of extraction sockets<br>to enhance preservation of the<br>alveolar ridge;<br>• Elevation of the maxillary<br>sinus floor;<br>• Filling of periodontal<br>defects in conjunction with<br>products intended for Guided<br>Tissue Regeneration (GTR) and<br>Guided Bone Regeneration<br>(GBR); and<br>• Filling of peri-implant | MBCP Gel™ is intended for<br>use as a bone grafting material<br>to fill, augment or reconstruct<br>osseous bone defects in<br>aprticular in periodontal or<br>oral/maxillofacial application.<br>Theses defects may be<br>surgically created osseous<br>defects or osseous defects<br>created from traumatic injury to<br>the bone.<br>MBCP Gel™ can generally be<br>used for bone filling in closed<br>cavities.<br>MBCP Gel™ can be used with<br>autogenous bone grafting<br>materials.<br>Typical uses include but are not<br>limited to:<br>• Periodontal/Infrabony defects<br>• Ridge augmentation | Same                                                                      |

{8}------------------------------------------------

| Image: MedPark logo   | <b>510(k) Submission Files</b> | Revision No.  | 0               |
|-----------------------|--------------------------------|---------------|-----------------|
| Your Reliable Partner | S1                             | Revision Date | October-13-2023 |
|                       |                                | Page          | 5 / 7           |

| defects in conjunction with<br>products intended for Guided<br>Bone Regeneration (GBR). | defects in conjunction with<br>products intended for Guided<br>Bone Regeneration (GBR). | • Extractions sites (implant<br>preparation/placement)<br>• Sinus lifts<br>• Cystic cavities |
|-----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|

# Similarity and Equivalence Discussion

Device characteristics with the Primary predicate device, suceal shape, sizes, structure, findamental technologies and applied production method are equivalent.

- The new device has the same intended use (as discussed above).
- . The main material of both the primary predicate device and S1 is the Hydroxyapatite (HAp).
- Through the Biocompatibility Test and Animal Performance Test, it was verformance of S1 are equivalent . to the Primary Predicate device.

## Difference Discussion

The difference compared to the primary predicate device(Bio-Oss® K122894) is that S1 is provided as a mixture of Hydroxyapatite (HAp) and hydroxypropyl methylcellulose (HPMC).

The hydroxypropyl methylcellulose (HPMC) used as an additive is pharmaceutical grade, and was also used in Reference Predicate Device.

Through the Biocompatibility Test and Animal Test, it was verformance of S1 were equivalent to that of the Primary Predicate device.

## Non-clinical testing data

- The subject device was evaluated its substantial equvalence through physical tests according to ISO 13779-3:2018 as the below table.

{9}------------------------------------------------

|                       |                         | Revision No.  | 0               |
|-----------------------|-------------------------|---------------|-----------------|
| Image: MedPark logo   | 510(k) Submission Files | Revision Date | October-13-2023 |
| Your Reliable Partner | S1                      | Page          | 6 / 7           |

| Test Item                       | Methods/Standards             | Criteria                                                                                                                                                    | Result                                                                   |
|---------------------------------|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|
| Crystallinity Ratio             | ISO 13779-3                   | The crystallinity ratio<br>should be more than 95%.                                                                                                         | 100 %                                                                    |
| Ca/P ratio                      | ISO 13779-3                   | Ratio of calcium (Ca) and<br>phosphate (P) must be<br>$1.66\pm0.1$ .                                                                                        | 1.66                                                                     |
| Hydroxyapatite(HAp)<br>Contents | ISO 13779-3                   | The content of<br>Hydroxyapatite(HAp)<br>should be more than 97<br>wt%.                                                                                     | 100 wt%                                                                  |
| Heavy metals                    | ASTM F1581-08,<br>ISO 13779-3 | As $\leq$ 3 mg/kg<br>Pb $\leq$ 30 mg/kg<br>Hg $\leq$ 5 mg/kg<br>Cd $\leq$ 5 mg/kg<br>(The total contents of<br>heavy metal shall be less<br>than 50 mg/kg.) | As: Not Detected<br>Pb: 0.8 mg/kg<br>Hg: 0.012 mg/kg<br>Cd: Not Detected |

- Sterilization Validation Test according to ISO 11137-1:2006 , ISO 11137-2:2013, ISO ● 11137-3:2017
- Packaging and shelf-life testing according to ASTM F1980-16, ASTM F1980, ASTM F88-15, ASTM F1140, ASTM F2096, ASTM D4169-22, ISO 11607-1:2019, ISO 11607-2:2019, ASTM F1929-15, ISO 13779-3:2018, ISO 11737-2:2019
- Biocompatibility testing according to ISO 10993-1:2018, ISO 10993-3:2014. ISO 10993-4:2017, ISO 10993-5:2009, ISO 10993-6:2016, ISO 10993-10:2010 and ISO 10993-11:2017, ISO 10993-12:2012, ISO 10993-23:2021, USP (151)
- Medical devices utilizing animal tissues and their derivatives according to ISO 22442-1:2020, ISO 22442-2:2020
- Virus Inactivation Testing according to ISO 22442-3:2007

S1 was the subject of the full range of biocompatibility test recommended in the FDA's "Class II Special Controls Guidance Documents: Dental Bone Grafting Devices" and in accordance with ISO 10993-1. Test results confirmed product safety. In addition, virus inactivation study was conducted. Further, the product is sterilized to achieve a sterility assurance level SAL 1 X10-6.

The results of these studies confirm the substantial equivalence of S1 to its predicate devices.

{10}------------------------------------------------

| Image: MedPark logo          | 510(k) Submission Files | Revision No.  | 0               |
|------------------------------|-------------------------|---------------|-----------------|
| Image: Your Reliable Partner |                         | Revision Date | October-13-2023 |
|                              | S1                      | Page          | 7 / 7           |

# Animal Test

The performance of the beagle mandibular defect model was compared to the performance of the predicate device, Bio-Oss®. Radiographic, Micro CT, Histology and Histomorphometry analyses were conducted following implantation at 4, 8, and 12 weeks for the subject device, predicate device and negative control.

The predicate device Bio-Oss® and the subject device S1 were compared in the preclinical study, and the results demonstrated substantially equivalent performances under intended uses.

Eventually, the majority of the device will be resorbed by 24 months or more, and was found to be similar to the predicate device Bio-Oss®.

# Conclusion

S1 constitutes a substantially equivalent medical device, meeting all the declared requirements of its intended use. This system has the same intended use and fundamental scientific technology as its predicate devices. The conclusions drawn from the nonclinical tests demonstrate that the device is substantially equivalent to its predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NPM/K231672](https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NPM/K231672)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
