K211551 · Wishbone SA · NPM · Aug 20, 2021 · Dental
Device Facts
Record ID
K211551
Device Name
Wishbone HA
Applicant
Wishbone SA
Product Code
NPM · Dental
Decision Date
Aug 20, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3930
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Filling of infrabony periodontal defects; Filling of periodontal defects in conjunction with products intended for guided tissue regeneration and guided bone regeneration: Filling of defects after root resection, apicectomy and cystectomy; Filling of extraction sockets to enhance preservation of the alveolar ridge; Augmentation or reconstructive treatment of the alveolar ridge; Elevation of the maxillary sinus floor; Filling of peri-implant defects in conjunction with products intended for guided bone regeneration.
Device Story
Wishbone HA is a xenograft biomaterial composed of deproteinized bovine hydroxyapatite; supplied as cancellous and cortical granules (0.25-1.0 mm) in sterile thermoformed blisters. Used by dentists, periodontists, and oral surgeons in clinical settings to fill, augment, or reconstruct periodontal and bony defects of the jaw. Acts as a conductive bone graft matrix to support bone regeneration. Device is single-use and sterilized via gamma irradiation. Clinical benefit includes structural support for bone healing in dental procedures.
Clinical Evidence
Non-clinical bench testing included chemical analysis, phase composition, morphology, porosity, pH, dissolution, and mechanical evaluation per FDA guidance. In-vivo performance was evaluated in a GLP canine mandibular defect model at 4, 12, and 26 weeks. Results at 26 weeks showed biocompatibility and bone regeneration comparable to the predicate device. Biocompatibility testing followed ISO 10993 standards.
Technological Characteristics
Bovine-derived deproteinized hydroxyapatite (≥97%). Granule size: 0.25-1.0 mm. Sterilization: Gamma irradiation (SAL 10^-6). Packaging: PETG blister with Tyvek seal. Standards: ISO 10993 (biocompatibility), ISO 22442 (animal tissue), ISO 11137 (sterilization), ISO 11607 (packaging), and various ASTM standards for packaging integrity (e.g., ASTM F88, F1929, F1980).
Indications for Use
Indicated for adults requiring bone grafting for periodontal defects, extraction socket preservation, alveolar ridge augmentation/reconstruction, maxillary sinus floor elevation, or peri-implant defect filling.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
August 20, 2021
Wishbone SA Emilie Dory Operations Manager 1, Rue de 1'Expansion Flemalle, Liege 4400 BELGIUM
Re: K211551
Trade/Device Name: Wishbone HA Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: Class II Product Code: NPM Dated: June 7, 2021 Received: June 8, 2021
Dear Emilie Dory:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Andrew Steen Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name Wishbone HA
Indications for Use (Describe)
Wishbone HA is intended for the following uses:
· Filling of infrabony periodontal defects;
· Filling of periodontal defects in conjunction with products intended for guided tissue regeneration and guided bone regeneration:
· Filling of defects after root resection, apicectomy and cystectomy;
- · Filling of extraction sockets to enhance preservation of the alveolar ridge;
- · Augmentation or reconstructive treatment of the alveolar ridge;
- · Elevation of the maxillary sinus floor;
· Filling of peri-implant defects in conjunction with products intended for guided bone regeneration.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
## 1. SUBMITTER
| Submitter Name: | Wishbone SA |
|---------------------------------------------------|---------------------------------------------------|
| Submitter Address: | 1, Rue de l'Expansion,<br>4400 Flémalle - Belgium |
| Phone Number: | +32484706172 |
| Contact Person: | Emilie Dory |
| Date Prepared: | 20 August 2021 |
| 2. DEVICE | |
| Device Trade Name: | Wishbone HA |
| Common Name: | Bone graft material |
| Classification Name,<br>Number &<br>Product Code: | Bone grafting material<br>21 CFR 872.3930<br>NPM |
| Class: | II |
| Classification Panel: | Dental |
## 3. PREDICATE DEVICE
K122894, Geistlich Bio-Oss, Geistlich Pharma AG Primary Predicate Device:
## 4. REFERENCE DEVICE
Reference Device: No reference device was used in this submission
## 5. DEVICE DESCRIPTION
Wishbone HA is a xenograft biomaterial composed of deproteinized hydroxyapatite from bovine origin. It is intended to fill, augment, or reconstruct periodontal defects and/or bony defects of the upper or lower jaw.
Wishbone HA is supplied as a mix of cancellous and cortical particles size 0.25 to 1.0 mm) in a single use thermoformed blister, packaged in a secondary thermoformed blister and sterilized by gamma irradiation.
The device is intended to be used in medical procedures, by a qualified physician (academically trained dentists, periodontists and oral surgeons).
{4}------------------------------------------------
# 6. INDICATIONS FOR USE
Wishbone HA is intended for the following uses:
- Filling of infrabony periodontal defects .
- Filling of periodontal defects in conjunction with products intended for guided tissue . regeneration and guided bone regeneration
- Filling of defects after root resection, apicectomy and cystectomy
- . Filling of extraction sockets to enhance preservation of the alveolar ridge
- . Augmentation or reconstructive treatment of the alveolar ridge
- Elevation of the maxillary sinus floor ●
- Filling of peri-implant defects in conjunction with products intended for quided bone . regeneration.
### 7. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICE
| | New Device | Primary Predicate Device |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device name | Wishbone HA | Geistlich Bio-Oss |
| 510(k) Number | K211551 | K122894 |
| Manufacturer | Wishbone SA | Geistlich Pharma AG |
| Regulation Number | 872.3930 | 872.3930 |
| Device<br>Classification Name | Bone grafting material | Bone grafting material |
| Product Code | NPM | NPM |
| Intended Use/<br>Indications for use | • Filling of infrabony periodontal defects;<br>• Filling of periodontal defects in conjunction with products intended for guided tissue regeneration and guided bone regeneration;<br>• Filling of defects after root resection, apicectomy and cystectomy;<br>• Filling of extraction sockets to enhance preservation of the alveolar ridge;<br>• Augmentation or reconstructive treatment of the alveolar ridge<br>• Elevation of the maxillary sinus floor;<br>• Filling of peri-implant defects in conjunction with products intended for guided bone regeneration. | - Augmentation or reconstructive treatment of the alveolar ridge;<br>- Filling of infrabony periodontal defects;<br>- Filling of defects after root resection, apicoectomy, and cystectomy;<br>- Filling of extraction sockets to enhance preservation of the alveolar ridge;<br>- Elevation of the maxillary sinus floor;<br>- Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR); and<br>- Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR). |
| Intended patient<br>population | Adults | Adults |
| Mode of action | Conductive bone graft | Conductive bone graft |
| | | |
| Anatomical sites | Oral, Periodontal | Oral, Periodontal |
| Environment of use | Wishbone HA should only be used<br>by trained dentists, periodontists or<br>oral surgeons. | Geistlich Bio-Oss® should only be<br>used by trained dentists or oral<br>surgeons. |
| Sterilization | Gamma irradiation | Gamma irradiation |
| Reusability | Single use only | Single use only |
| Pyrogenicity | Non-pyrogenic | Non-pyrogenic |
| Packaging | Wishbone HA particles are<br>packaged in a single use<br>thermoformed PETG<br>(Polyethylene Terephthalate<br>Glycol) blister (primary packaging)<br>hermetically sealed with a<br>Tyvek®. This package is placed in<br>a secondary thermoformed. | Double sterile barrier system<br>consisting of a glass vial and an<br>outer blister. |
| Source of materials | Bovine | Bovine |
| Chemical<br>Composition | Calcium: 30 - 40 % (w./w.)<br>Phosphorous: 15 - 20 % (w./w.)<br>Sodium: 0.0 - 1.0 % (w./w.)<br>Magnesium: 0.0 - 1.0 % (w./w.)<br><br>Summed up to 100% by mass: | From the original 510(k):<br>Calcium: 35% - 40% (w./w.)<br>Phosphorous: 13.5% - 18.5%<br>(w./w.)<br><br>Summed up to 100% by mass**: |
| | - Hydroxyapatite $Ca_5(PO_4)_3OH$<br>(CAS® 1306-06-5): ≥97%<br>- $CaCO_3$ (CAS® 471-34-1)<br>- $MgO$ * (CAS® 1309-48-4): ≤2%<br>- $H_2O$ (CAS® 7732-18-5): ≤1%<br>- Organic compounds: 0% | - Hydroxyapatite<br>(Pentacalciumhydroxy-[tri]- phosphate $Ca_5(PO_4)_3(OH)$ )<br>(93.6%),<br>- Calcium carbonate ( $CaO_3$ )<br>(3,4%)<br>- Water (3%)<br>- No organic components (0%) |
| Crystalline phases | Hydroxyapatite > 95% | Hydroxyapatite |
| Dosage form | Mix of cancellous and cortical<br>granules | Cancellous granules or porous<br>block |
| Particle sizes | 0.25 - 1.0 mm | 0.25 - 1.0 mm (small particles)<br>1.0 - 2.0 mm (large particles) |
| Product size (g) | 0.25, 0.5, 1.0, 2.0 | 0.25, 0.5, 2.0, 5.0 (0.25-1.0 mm)<br>0.5, 2.0 (1.0-2.0 mm)<br>For blocks: 1 X 1 X 2 cm (approx.) |
Table 1: Comparison of new device to predicate device
{5}------------------------------------------------
{6}------------------------------------------------
| Biocompatibility testing | Chemical characterization<br>Toxicological evaluation<br>Cytotoxicity<br>Sensitization<br>Intracutaneous Reactivity/Irritation<br>Acute Systemic Toxicity<br>Material Mediated Pyrogenicity<br>Genotoxicity<br>Implantation | Per ISO 10993-1 for the intended use |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|
* Wishbone HA is composed of only bovine bone without any added components. The magnesium oxide (MgO) is not added to Wishbone HA but rather naturally appears during the sintering process in the form of magnesium oxide crystals.
** Data taken from 510(k) summary of Straumann® Cerabone® (K173594) which uses K122894 Geistlich Bio-Oss as the predicate device.
### Equivalences:
The intended use is the same, and the technological characteristics are essentially the same as those of the predicate, K122894, Geistlich Bio-Oss. Both devices are made of deproteinized hydroxyapatite from bovine origin, intended to be implemented as a bone graft matrix in periodontal or bony defects. Both devices are intended to be used by dental health care professionals (e.g. dentists, periodontists and oral surgeons). Both devices are intended for single patient use and sterilized by gamma irradiation.
#### Differences that are demonstrated to be substantially equivalent:
As indicated in Table 1 above, several differences with respect to technological characteristics were identified between Wishbone HA and the primary predicate, namely the chemical composition, crystalline phases, dosage form, specific surface area and the packaging configuration.
Performance testing was conducted to demonstrate substantial equivalence of Wishbone HA to the predicate device. The test results are summarized below.
#### 8. PERFORMANCE DATA
- Non-Clinical Wishbone HA was the subject of the full range of physical and chemical bench testing characterization tests recommended in the FDA's "Class II Special Controls Guidance Documents: Dental Bone Grafting Devices". These bench tests included chemical analysis, phase composition analysis, protein and organic content analysis, morphology analysis, porosity analysis, pH analysis, dissolution analysis and mechanical evaluation.
{7}------------------------------------------------
| Non-Clinical<br>animal<br>performance<br>Testing | A pre-clinical GLP study was performed in canine mandibular defect<br>model system in Beagle dogs to evaluate the in-vivo performance of the<br>Wishbone HA. In this study the local effects of implantation of Wishbone<br>HA in the target locations, as well as bone growth, at subchronic time-<br>points (4, 12 and 26 weeks) were evaluated. |
|--------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | By all parameters assessed at the 26 weeks time-point of the study,<br>biocompatibility and regeneration of defects treated with Wishbone HA<br>showed similar results to defects treated with the predicate device, Bio-<br>Oss®. |
| Biocompatibility | Wishbone HA was the subject of a range of biocompatibility tests in<br>accordance with ISO 10993 series. Test results confirmed that Wishbone<br>HA is biocompatible for the stated intended use. |
| Animal tissue | The methods of sourcing, collection and handling of the bovine-derived<br>bones were assessed to demonstrate compliance with relevant<br>requirements of ISO 22442 series and FDA guidance (2019). Wishbone<br>HA is provided free from organic residues. Protein content is routinely<br>controlled during the production process. The manufacturing process<br>includes several steps intended to inactivate viruses. The results of the |
viral inactivation studies showed that the raw material and device manufacturing processes of the Wishbone HA are capable of achieving at least a 6-log viral reduction of selected model viruses which were tested.
- Sterilization Wishbone HA is provided sterile and is intended for single patient use only. Wishbone HA is sterilized with gamma irradiation to meet a and shelf life minimum sterility assurance level (SAL) of 106. Validation of the sterilization dose was conducted following VDmax method.
Shelf life study supports a shelf life of 36 months for Wishbone HA when stored under the recommended environmental conditions. The shelf life studies confirmed that the packaging maintains the integrity of the device and its sterility throughout the shelf life of the device.
Validation studies for sterilization, packaging and shelf life conform to the following standards:
| Standard<br>reference | Standard title |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| EN ISO 11607-<br>1:2020 | Packaging for terminally sterilized medical devices -<br>Part 1: Requirements for materials, sterile barrier<br>systems and packaging systems |
| EN ISO 11607-<br>2:2020 | Packaging for terminally sterilized medical devices -<br>Part 2: Validation requirements for forming, sealing<br>and assembly processes |
| EN ISO 11137-<br>1:2015/A2:2019 | Sterilization of health care products - Radiation - Part<br>1: Requirements for development, validation and<br>routine control of a sterilization process for medical<br>devices |
| EN ISO 11137-<br>2:2015 | Sterilization of health care products - Radiation - Part<br>2: Establishing the sterilization dose |
| EN ISO 11137-<br>3:2017 | Sterilization of health care products - Radiation - Part<br>3: Part 3: Guidance on dosimetric aspects of<br>development, validation and routine control |
| EN ISO 11737-<br>1:2018 | Sterilization of medical devices - Microbiological<br>methods - Part 1: Determination of a population of<br>microorganisms on products |
| EN ISO 11737-<br>2:2009 | Sterilization of health care products - Microbiological<br>methods - Part 2: Tests of sterility performed in the<br>definition, validation and maintenance of a sterilization<br>process |
| EN 868-5:2009 | Packaging for terminally sterilized medical devices -<br>Part 5: Sealable pouches and reels of porous<br>materials and plastic film construction - Requirements<br>and test methods |
| ASTM F88F88M-<br>15 | Standard Test Method for Seal Strength of Flexible<br>Barrier Materials |
| ASTM<br>F1886F1886M-16 | Standard Test Method for Determining Integrity of<br>Seals for Flexible Packaging by Visual Inspection |
| ASTM F1929-15 | Standard Test Method for Detecting Seal Leaks in<br>Porous Medical Packaging by Dye Penetration |
| ASTM F1980-16 | Standard Guide for Accelerated Aging of Sterile<br>Barrier Systems for Medical Devices |
| ASTM D642-15 | Standard Test Method for Determining Compressive<br>Resistance of Shipping Containers, Components, and<br>Unit Loads |
| ASTM D6344-04<br>(2017) | Standard Test Method for Concentrated Impacts to<br>Transport Packages |
| ASTM D999-08<br>(2015) | Standard Test Methods for Vibration Testing of<br>Shipping Containers |
| ASTM D4169-16 | Standard Practice for Performance Testing of Shipping<br>Containers and Systems |
| ASTM D4332-14 | Standard Practice for Conditioning Containers.<br>Packages, or Packaging Components for Testing |
| ASTM D4728-17 | Standard Test Method for Random Vibration Testing of<br>Shipping Containers |
| ASTM D5276-98<br>(2017) | Standard Test Method for Drop Test of Loaded Containers by Free Fall |
| EN ISO 11140-<br>1:2014 | Sterilization of health care products - Chemical indicators - Part 1: General requirements |
{8}------------------------------------------------
{9}------------------------------------------------
# 9. CONCLUSION
The information discussed above and provided in the 510(k) submission demonstrate that the Wishbone HA is substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.