The DSM Biomedical Dental Bone Graft Plus is indicated for: - Augmentation or reconstructive treatment of the alveolar ridge - Filling of infrabony periodontal defects - Filling of defects after root resection, apicoectomy, and cystectomy - Filling of extraction sockets to enhance preservation of the alveolar ridge - Elevation of the maxillary sinus floor - Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR) - Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR)
Device Story
Non-pyrogenic porous bone mineral and collagen matrix; composed of anorganic porcine bone granules and bovine collagen formed into small cylinders. Used in periodontal, oral, and maxillofacial surgery to facilitate bone formation. Supplied sterile via gamma irradiation for single-use. Clinician implants device into bone defects; acts as osteoconductive scaffold to support natural bone regeneration. Benefits patient by enhancing bone volume and structural integrity in oral sites.
Clinical Evidence
Bench testing included material characterization, stability, and packaging validation (ISO 11607-1). Biocompatibility testing per ISO 10993-1 (cytotoxicity, sensitization, irritation, systemic toxicity, genotoxicity, implantation) and pyrogenicity (USP <85>) confirmed safety. Animal testing in a canine intrabony defect model demonstrated bone healing performance equivalent to the predicate device.
Technological Characteristics
Osteoconductive hydroxyapatite bone mineral (porcine-derived) with bovine collagen. Form: cylinder. Dimensions: 100mg, 250mg, 500mg. Porosity: 73.42-77.26%. Sterilization: Gamma irradiation (ISO 11137-1/2). Biocompatibility: ISO 10993-1. Animal source compliance: ISO 22442-2/3.
Indications for Use
Indicated for human oral and periodontal surgery requiring alveolar ridge augmentation, infrabony defect filling, root resection/apicoectomy/cystectomy defect filling, extraction socket preservation, maxillary sinus floor elevation, and GTR/GBR-assisted periodontal or peri-implant defect filling.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
September 10, 2020
DSM Biomedical Susan Pileggi Senior Regulatory Specialist 735 Pennsylvania Drive Exton, Pennsylvania 19431
Re: K193212
Trade/Device Name: DSM Biomedical Dental Bone Graft Plus Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: Class II Product Code: NPM Dated: August 17, 2020 Received: August 19, 2020
Dear Susan Pileggi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-
combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K193212
Device Name
DSM Biomedical Dental Bone Graft Plus
Indications for Use (Describe)
The DSM Biomedical Dental Bone Graft Plus is indicated for:
- Augmentation or reconstructive treatment of the alveolar ridge
- Filling of infrabony periodontal defects
- Filling of defects after root resection, apicoectomy, and cystectomy
- Filling of extraction sockets to enhance preservation of the alveolar ridge
- Elevation of the maxillary sinus floor
- Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)
- Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR)
| Type of Use (Select one or both, as applicable) | <table><tr><td><input checked="true" type="checkbox"/> Commercial Use (including CBD extraction) </td><td><input type="checkbox"/> Personal Use (including CBD extraction) </td></tr></table> | <input checked="true" type="checkbox"/> Commercial Use (including CBD extraction) | <input type="checkbox"/> Personal Use (including CBD extraction) |
|-----------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|------------------------------------------------------------------|
| <input checked="true" type="checkbox"/> Commercial Use (including CBD extraction) | <input type="checkbox"/> Personal Use (including CBD extraction) | | |
X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for DSM, a global science-based company. The logo consists of a colorful abstract graphic on the left and the letters "DSM" in bold, blue font on the right. The graphic is made up of overlapping shapes in various colors, including blue, green, yellow, and purple, creating a sense of movement and dynamism.
# K193212
# 510(k) Summary
# I. SUBMITTER
| Submitter: | DSM Biomedical<br>735 Pennsylvania Drive<br>Exton, PA 19341 |
|-----------------|-------------------------------------------------------------|
| Phone: | 484-713-2100 |
| Contact Person: | Susan Pileggi |
| Date Prepared: | September 10, 2020 |
### II. DEVICE
| Name of Device: | DSM Biomedical Dental Bone Graft Plus |
|----------------------|------------------------------------------|
| Common/Usual Name: | Bone Grafting Material |
| Classification Name: | Bone Grafting Material (21 CFR 872.3930) |
| Device Class: | II |
| Device Code: | NPM |
# III. PREDICATE DEVICE
Substantial equivalence is claimed to the following device:
- Bio-OSS® Collagen, K122894 ●
The following device is referenced within the submission:
- DSM Biomedical Dental Bone Graft, K170245 ●
# IV. DEVICE DESCRIPTION
The DSM Biomedical Dental Bone Graft Plus is a non-pyrogenic porous bone mineral and collagen matrix for use in periodontal, oral, and maxillofacial surgery. The DSM Biomedical Dental Bone Graft Plus is composed of anorganic porcine bone granules combined with bovine collagen to form small cylinders. The device is provided in sizes ranging from 100 - 500mg. DSM Dental Bone Graft Plus is supplied sterile by gamma irradiation and is for single use only.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the DSM company logo. The logo consists of a colorful abstract design on the left and the letters "DSM" in blue on the right. The abstract design is made up of overlapping shapes in various colors, including blue, green, orange, and purple. The letters "DSM" are in a bold, sans-serif font.
### V. Indications For USE:
DSM Biomedical Dental Bone Graft Plus is indicated for:
- . Augmentation or reconstructive treatment of the alveolar ridge
- Filling of infrabony periodontal defects .
- Filling of defects after root resection, apicoectomy, and cystectomy
- Filling of extraction sockets to enhance preservation of the alveolar ridge ●
- Elevation of the maxillary sinus floor
- . Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)
- Filling of peri-implant defects in conjunction with products intended for Guided Bone . Regeneration (GBR)
# VII. COMPARISON OF TECHNOLOGICAL CHARACTERSTICS WITH THE PREDICATE DEVICE
The DSM Biomedical Dental Bone Graft Plus is substantially equivalent in terms of indications for use, material composition, technological characteristics, and performance characteristics to the predicate device, Bio-OSS® Collagen [Geistlich-Pharma] K122894. The subject device is substantially equivalent in terms of indications for use, technological characteristics and performance characteristics to the reference device, DSM Biomedical Dental Bone Graft K170245, but differs in material composition as the subject device contains collagen in addition to the anorganic bone material of the reference device. The devices are compared in the Table 1.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for DSM, a global science-based company. The logo consists of a colorful abstract graphic on the left and the letters "DSM" in blue on the right. The graphic is made up of overlapping shapes in various colors, including blue, green, orange, and purple. The letters "DSM" are in a bold, sans-serif font.
| | Table | | 1: |
|--|-------|--|----|
|--|-------|--|----|
| | DSM Biomedical Dental<br>Bone Graft Plus | Bio-OSS® Collagen | DSM Biomedical Dental<br>Bone Graft |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | (K193212: Subject Device) | (K122894: Predicate<br>Device) | (K170245: Reference<br>Device) |
| Indications for<br>Use | DSM Biomedical Dental Bone<br>Graft Plus is indicated for<br>augmentation or<br>reconstructive treatment of<br>the alveolar ridge; filling of<br>infrabony periodontal<br>defects; filling of defects<br>after root resection,<br>apicoectomy, and<br>cystectomy; filling of<br>extraction sockets to<br>enhance preservation of the<br>alveolar ridge; elevation of<br>the maxillary sinus floor;<br>filling of periodontal defects<br>in conjunction with products<br>intended for Guided Tissue<br>Regeneration (GTR) and<br>Guided Bone Regeneration<br>(GBR); filling of peri-implant<br>defects in conjunction with<br>products intended for<br>Guided Bone Regeneration<br>(GBR). | Augmentation or<br>reconstructive treatment of<br>the alveolar ridge; filling of<br>periodontal defects; filling<br>of defects after root<br>resection, apicoectomy, and<br>cystectomy; filling of<br>extraction sockets to<br>enhance preservation of the<br>alveolar ridge; elevation of<br>the maxillary sinus floor;<br>filling of periodontal defects<br>in conjunction with products<br>intended for Guided Tissue<br>Regeneration (GTR) and<br>Guided Bone Regeneration<br>(GBR); filling of peri-implant<br>defects in conjunction with<br>products intended for<br>Guided Bone Regeneration<br>(GBR). | The DSM Biomedical Dental<br>Bone Graft is indicated for<br>use in augmentation or<br>reconstructive treatment of<br>the alveolar ridge; filling of<br>infrabony periodontal<br>defects; filling of defects<br>after root resection,<br>apicoectomy, and<br>cystectomy; filling of<br>extraction sockets to<br>enhance preservation of the<br>alveolar ridge; elevation of<br>the maxillary sinus floor;<br>filling of periodontal defects<br>in conjunction with products<br>intended for Guided Tissue<br>Regeneration (GTR) and<br>Guided Bone Regeneration<br>(GBR); filling of peri-implant<br>defects in conjunction with<br>products intended for<br>Guided Bone Regeneration<br>(GBR). |
| Target<br>Population | Human, oral, periodontal | Human, oral, periodontal | Human, oral, periodontal |
| Material<br>Composition | Porcine-derived<br>osteoconductive<br>hydroxyapatite bone<br>mineral with bovine<br>collagen. | Bovine-derived<br>osteoconductive<br>hydroxyapatite bone<br>mineral with porcine<br>collagen | Porcine-derived<br>osteoconductive<br>hydroxyapatite bone<br>mineral |
| Form | Cylinder | Block | Granules |
| Dimensions | 100mg, 250mg, and 500mg | 100mg, 250mg, and 500mg | 0.5cc, 1.0cc, 2.5cc |
| Cumulative Pore<br>Volume | 1.6272 - 1.8329 mL/g | 0.9524 - 1.1751 mL/g | 1.3494mL/g |
| Characteristic | DSM Biomedical Dental<br>Bone Graft Plus<br>(K193212: Subject Device) | Bio-OSS® Collagen<br>(K122894: Predicate<br>Device) | DSM Biomedical Dental<br>Bone Graft<br>(K170245: Reference<br>Device) |
| Pore Size<br>Distribution | 0.003 - 360.86μm | 0.003 - 346.995μm | 0.003 - 645μm |
| Pore Size Range | Primarily macropores | Primarily macropores | Primarily macropores |
| Porosity | 73.42 - 77.26% | 53.47 - 61.34% | 65.97 - 68.74% |
| Package | Double sterile barrier<br>consisting of an inner and<br>outer tray | Double sterile barrier<br>consisting of an inner and<br>outer tray | Granules supplied in a<br>single use sterilized vial. |
| Performance | Bone formation | Bone formation | Bone formation |
| Biocompatible | Yes (ISO 10993-1) | Yes | Yes (ISO 10993-1) |
| Reusable | Single Use Device | Single Use Device | Single Use Device |
| Sterility | Sterile, SAL 10-6<br>Gamma Irradiation<br>(ISO 11137-1) | Sterile, SAL 10-6<br>Gamma Irradiation | Sterile, SAL 10-6<br>Gamma Irradiation<br>(ISO 11137-1) |
| Pyrogencity | Non-pyrogenic<br>(USP <85> and ISO 10993-<br>11) | Non-pyrogenic | Non-pyrogenic<br>(USP <85> and ISO 10993-<br>11) |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for DSM, a global science-based company. The logo consists of a colorful, abstract swirl design on the left, with the letters "DSM" in bold, blue font on the right. The swirl design incorporates various colors, including blue, green, orange, and purple, creating a dynamic and visually appealing element.
DSM Biomedical Dental Bone Graft Plus and the predicate device have identical indications for use and are for use in the same target population. They are provided in the same sizes. Both are sterilized by gamma irradiation and supplied in a double tray configuration. The materials are shown to be an osteoconductive hydroxyapatite bone mineral with a collagen component to aid in bone formation. The DSM Biomedical Dental Bone Graft Plus is derived from porcine bone and bovine collagen while the predicate device is derived from bovine bone and porcine collagen. This does not raise questions of substantial equivalence, as chemical and physical characteristics have been shown to be substantially equivalent in benchtop and animal testing.
### VII. NON-CLINICAL TESTING DATA:
The following performance data was provided in support of the substantial equivalence determination.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for DSM, a global science-based company. The logo consists of a colorful abstract graphic on the left and the letters "DSM" in bold, blue font on the right. The graphic is composed of overlapping, curved shapes in shades of blue, green, orange, and purple, creating a dynamic and modern design.
### Mechanical and Physical Testing
Material characterization of the DSM Biomedical Dental Bone Graft Plus was completed in accordance with the FDA Guidance Document, Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices. Bench testing was completed to verify that the device can be hydrated and trimmed prior to implantation.
The packaging of DSM Dental Bone Graft Plus was designed to meet the requirements of ISO 11607-1. The product is gamma sterilized and has been validated per ISO 11137-1 and ISO 11137-2. Stability testing on the device included evaluations of appearance, mass, collagen characterization, and simulated use. The stability assessments verify that the passage of time did not affect the device performance, collagen characterization or package integrity.
# Biocompatibility Testing
Biocompatibility testing was completed in accordance with the requirements of ISO 10993-1 for a permanent implant device with tissue/bone contact. Biocompatibility testing included the following biological effects:
- . Cytotoxicity per ISO 10993-5
- Sensitization per ISO 10993-10 ●
- Irritation per ISO 10993-10
- Acute Systemic Toxicity per ISO 10993-11 ●
- Genotoxicity per ISO 10993-3 ●
- Subacute Systemic Toxicity per ISO 10993-11
- Chronic Systemic Toxicity per ISO 10993-6 .
- . Implantation per ISO 10993-6
Pyrogenicity testing per USP <85> and Rabbit Material Mediated Pyrogenicity testing per ISO 10993-11 have been completed to verify that the device is non-pyrogenic.
The porcine bone and bovine collagen are sourced in accordance with ISO 22442-2 and per the FDA Guidance Document, Medical Devices Containing Materials Derived from Animal Sources (Except for In Vitro Diagnostic Devices).
Viral inactivation studies per ISO 22442-3 and a residual chemical assessment per ISO 10993-17 were conducted.
Results indicate that the device's biocompatibility profile is acceptable.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for DSM, a global science-based company. The logo consists of a colorful abstract graphic on the left and the letters "DSM" in bold blue font on the right. The graphic is composed of overlapping shapes in various colors, including blue, green, orange, and purple, creating a sense of dynamism and innovation. The overall design is modern and professional, reflecting the company's focus on science and technology.
# Animal Testing
A canine intrabony defect animal study was performed to evaluate bone healing following treatment with the DSM Biomedical Dental Bone Graft Plus compared to the predicate device.
# VIII. CONCLUSIONS
Pursuant to section 510(k), DSM Biomedical Dental Bone Graft Plus is substantially equivalent to the predicate device Bio-OSS® Collagen regarding indication for use, material, technological characteristics, including principles of operation, and performance characteristics as shown in an anatomically relevant animal model.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.