A clinical case series was used to demonstrate defect healing and new bone formation of sufficient quality for dental implant placement, supporting the device's biocompatibility and performance.
Clinical case series; Bone graft; Dental implant
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Clinical case series; Clinical case series
Patients requiring dental bone grafting/implant placement
Not applicable for this study
Defect healing and formation of new bone of sufficient quality to obtain dental implant placement
Indications for Use
OCS-B Collagen® is recommended for: Augmentation or reconstructive treatment of the alveolar ridge. Filling of infrabony periodontal defects. Filling of defects after root resection, apicoectomy, and cystectomy Filling of extraction sockets to enhance preservation of the alveolar ridge Elevation of maxillary sinus floor Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR) Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR)
Device Story
OCS-B Collagen is a bone grafting material composed of purified bovine cancellous bone mineral granules and 10% porcine-derived collagen. It is provided in block or cylindrical forms. Used by dental professionals in oral and maxillofacial surgery to fill intra-osseous cavities and defects. The device acts as an osteoconductive scaffold, facilitating new bone formation and mineralization by osteoblasts. It is intended to be used in conjunction with GTR or GBR membranes where applicable. The device is sterilized via gamma irradiation. Clinical benefit includes successful defect healing and preparation of bone quality sufficient for dental implant placement.
Clinical Evidence
Evidence includes chemical/physical characterization, animal studies demonstrating new bone growth and resorption rates comparable to the predicate, and a clinical case series showing successful defect healing and bone quality sufficient for dental implant placement. Biocompatibility testing performed per ISO 10993, including cytotoxicity, hemolysis, sensitization, genotoxicity, and implantation studies.
Technological Characteristics
Material: Anorganic bovine bone mineral granules and porcine-derived type I collagen. Morphology: Trabecular, interconnecting macro/micro pores. Dimensions: 0.2-1.0mm or 1.0-2.0mm granules; various block/cylinder sizes. Sterilization: Gamma irradiation (SAL 10^-6). Standards: ASTM F1581-99, ISO 10993, ISO 11137, ASTM F1980-99. Non-pyrogenic.
Indications for Use
Indicated for human oral and periodontal bone grafting, including alveolar ridge augmentation, filling of infrabony periodontal defects, defects post-root resection/apicoectomy/cystectomy, extraction sockets, maxillary sinus floor elevation, and peri-implant defects, often in conjunction with GTR/GBR products.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are simple and abstract, with only the basic outline of the face visible.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 31, 2015
NIBEC CO., LTD c/o Mr. Daniel Nam PATS CORP 904 E. Windsor Road #102 Glendale, CA 91203
Re: K142040
Trade/Device Name: OCS-B Collagen® Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: NPM Dated: June 30, 2015 Received: July 2, 2015
Dear Mr. Nam:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142040
Device Name
OCS-B Collagen®
#### Indications for Use (Describe)
OCS-B Collagen ® is recommended for:
- Augmentation or reconstructive treatment of the alveolar ridge.
- Filling of infrabony periodontal defects.
- Filling of defects after root resection, apicoectomy, and cystectomy
- Filling of extraction sockets to enhance preservation of the alveolar ridge
- Elevation of maxillary sinus floor
- Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)
- Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR)
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K142040 510(k) Summary
#### Sponsor/Applicant
NIBEC Co., Ltd. Iwol electricity-electronic Agro-industrial Complex, 116, Bamdi-gil, Iwol-myeon, Jincheon-gun, Chungcheongbuk-do, Korea Phone: 82-10-2889-8590 Fax: 82-2-744-8732 Contact: Dr. Park, Yoon-Jeong
Date Prepared : July 29, 2015
## Device Name and Identification
Proprietary Name: OCS-B Collagen® Bone Grafting Material Plus Collagen Common/Usual Name: Anorganic Bovine Bone Grafting Material Classification Name: Bone Filling Material Bone grafting material, animal source (NPM)
### Predicate devices
Bio-Oss Collagen® (K092428) (Primary Predicate) Manufactured by: Geistlich Pharma AG Bahnhofstrasse 40 CH-6110 Wolhusen Switzerland
OCS-B® (K113246) (Reference Predicate) Manufactured by: NIBEC Co., Ltd. lwol electricity-electronic Agro-industrial Complex, 116, Bamdi-gil, Iwol-myeon, Jincheon-gun, Chungcheongbuk-do, Korea
### Device Category/Class
Device Class: Regulation Number: Product Code:
Class II 21 C.F.R. 872.3930 NPM
### Intended use
OCS-B Collagen®is recommended for: Filling of large oral and maxillofacial intra-osseous cavities
### Indications for use
OCS-B Collagen® is recommended for:
{4}------------------------------------------------
- Augmentation or reconstructive treatment of the alveolar ridge.
- Filling of infrabony periodontal defects.
- -Filling of defects after root resection, apicoectomy, and cystectomy
- Filling of extraction sockets to enhance preservation of the alveolar ridge -
- Elevation of maxillary sinus floor -
- -Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)
- -Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR)
### Device Description
OCS-B Collagen® is a combination of purified cancellous bone mineral granules (OCS-B") and 10% collagen in a block form in a blister and cylinderic form in a syringe and blister. It is sterilized by yirradiation.
### Basis for Substantial Equivalence
OCS-B Collagen® and Bio-Oss® Collagen have a similar physical and chemical structure. Both are porous, biocompatible bone grafts that facilitate the formation and mineralization of new bone by the osteoblast. Both products have same component ratio of bone mineral granules and collagen. Both products have same source of bone (bovine bone), and collagen (porcine skin). The bone granules were treated through a specific process for removal of organic compound. The collagen was purified through specific process for removal of other proteins and organic compounds. Therefore, OCS-B Collagen® is substantially equivalent to Bio-Oss® Collagen.
The following is a table comparing OCS-B Collagen® and OCS-B® a Bone graft cleared for GBR.
#### Table 1: Substantial Equivalence Comparison
| | OCS-B Collagen® | Primary Predicate<br>Bio-Oss® Collagen*<br>(K092428) | | Reference Predicate<br>OCS-B®**<br>(K113246) | | Cylinderic<br>5.6X45 | 500mg | | | | |
|----------------------|----------------------------------------------------------------------------|--------------------------------------------------------------|---------------|--------------------------------------------------------------|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|------------------------------------------------------------------------|----------------------------------|----------------------------------|----------------------------------|
| ITEM | | | | | Material | Anorganic derived<br>osteoconductive hydroxyapatite<br>(pre-sterilized product of OCS-<br>B® (K113246), Collagen | Anorganic derived<br>osteoconductive<br>hydroxyapatite,<br>Collagen | Anorganic derived<br>osteoconductive<br>hydroxyapatite bone<br>mineral | | | |
| Intended<br>Use | Used as an adjective therapy in<br>restoring bony defects | Used as an adjective<br>therapy in restoring<br>bony defects | | Used as an adjective<br>therapy in restoring<br>bony defects | Source<br>of bone | Bovine bone | Bovine bone | Bovine bone | | | |
| Target<br>population | Human oral, periodontal | Human oral, periodontal | | Human oral,<br>periodontal | Source<br>of collagen | Porcine skin | Porcine skin | - | | | |
| Dosage<br>form | a block form in a blister<br>a cylinderic form in a syringe<br>and blister | a block form in a blister | | Granules contained in<br>single use container | Physical<br>Morphology | Trabecular, interconnecting<br>macro and micro pores | Trabecular, interconnecting<br>macro and micro pores | Trabecular,<br>interconnecting macro<br>and micro pores | | | |
| | 0.2mm to 1.0mm or 1.0mm to<br>2.0mm granules | 0.25mm to 1.0mm or<br>1.0mm to 2.0mm granules | | 0.2mm to 1.0mm or<br>1.0mm to 2.0mm<br>granules | Biocompatible | Biocompatible, as demonstrated<br>by :<br>- Cytotoxicity testing<br>- Hemolysis testing<br>- Acute systemic injection testing<br>- Intracutaneous reactivity<br>testing<br>- Skin sensitization testing<br>(LLNA)<br>- Genotoxicity testing<br>(Micronucleus test)<br>- Oral Mucosa Irritation<br>- Genotoxicity testing (AMES<br>test)<br>- Implantation testing<br>- Genotoxicity<br>testing(Chromosome aberration<br>test)<br>- Preclinical quality testing<br>- Clinical case series | Biocompatible<br>(as demonstrated<br>in published literature) | Biocompatible (as<br>demonstrated in<br>published literature) | | | |
| Dimension | Shape | Dimension<br>(mm) | Weight<br>(g) | Shape | Dimension<br>(mm) | Weight<br>(g) | | Performance | Bone formation | Bone formation | Bone formation |
| | Block | 6X6X3 | 50mg | | | | | Compatibility<br>with other<br>devices | Can be used with GTR<br>membrane | Can be used with GTR<br>membrane | Can be used with<br>GTR membrane |
| | Block | 6X6X6 | 100mg | Block | 6X6X6 | 100mg | | | | | |
| | Block | 7X8X9 | 250mg | Block | 7X8X9 | 250mg | | | | | |
| | Block | 9X10X11 | 500mg | Block | 9X10X11 | 500mg | | | | | |
| | Cylinderic | 4.6X40 | 250mg | | | | | | | | |
*= Primary Predicate (K092428)
**=Reference Predicate(K113246)
{5}------------------------------------------------
{6}------------------------------------------------
| Sterilization<br>Process | Sterile by Gamma Irradiation | Sterile by Gamma<br>Irradiation | Sterile by Gamma<br>Irradiation |
|--------------------------|------------------------------|---------------------------------|---------------------------------|
| Biocom-<br>patibility | Biocompatible | Biocompatible | Biocompatible |
| Anatomical<br>sites | Oral, Periodontal | Oral, Periodontal | Oral, Periodontal |
| Non-<br>Pyrogenic | Yes | Yes | Yes |
| Shelf-Life | 36 Months | 36 Months | 36 Months |
## Brief Summary of Data Submitted
The Sponsor evaluated the performance characteristics of OCS-B Collagen® and Bio-Oss® Collagen with a thorough chemical and physical characterization. The physical and chemical characteristics of the products were found to be comparable. Further, in several animal studies, both products were found to grow new bone and be subsequently resorbed at similar rates. Finally, in a clinical case series, use of OCS-B Collagen resulted in defect healing and formation of new bone of sufficient quality to obtain dental implant placement.
OCS-B Collagen was the subject of the full range of biocompatibility tests recommended in the FDA's "Class II Special Controls Guidance Document: Dental Bone Grafting Devices" and in accordance with ISO 10993. Test results confirmed equivalence to predicates. OCS-B® (K113246) was manufactured to remove organic impurities and product specifications have been established to limit protein content. Throughout the risk analysis for each production step, for example, cleaning validation, the removal of organic solvent, the risk control was conducted during the manufacturing process. In addition, the TSE inactivation as well as virus inactivation study result confirmed product suitability. Collagen is purified from porcine skin by a standardized controlled manufacturing process. The type I collagen has been purified from veterinary certified porcine skin. OCS-B Collagen® is composed of cancellous bone mineral granules (pre-sterilized OCS-B®) and 10% collagen. The dosage form is a blister or a cylinderic form in a syringe and blister. Further, the product is sterilized to achieve a sterility assurance level SAL 1 x 10 °. Further, OCS-B Collagen® is sterilized to achieve a sterility assurance level SAL 1 x 10 °.
Specifically, OCS-B Collagen® was evaluated and tested in accordance with the following performance standards:
- · ISO 10993 "Biological Evaluation of Medical Devices"
- · ASTM F 1581-99 "Standard Specification for Composition of Anorganic Bone for Surgical Implants" (1999)
- · FDA's "Medical Device Materials Derived from Animal Sources" (1998)
- · ISO 11137 "Sterilization of Healthcare Products Radiation"
- · ASTM F 1980-99 "Standard Guide for Accelerated Aging of Sterile Medical device Packages"
{7}------------------------------------------------
Based on the information presented herein, it has been demonstrated that OCS-B Collagen® is substantially equivalent to Bio-Oss® Collagen, and suitable for the proposed indications for use.
## Conclusion
OCS-B Collagen® has the same indications for use, same intended use, same sterilization process, same shelf-life, biocompatibility and similar technologies as its primary predicate, Bio-Oss® Collagen. The subject device and predicate device encompass the same range of physical dimensions. They share similar physical and chemical structure. Both are porous, biocompatible bone grafts that facilitate the formation and mineralization of new bone by the osteoblast. Both products have same component ratio of bone mineral granules and collagen.
Both products have same source of bone (bovine bone), and collagen (porcine skin). The bone granules were treated through specific process for removal of organic compound. The collagen was purified through specific process for removal of other proteins and organic compounds.
OCS-B Collagen® presents the same types of potential risks to consumers as the predicate device Bio-Oss Collagen, and has controlled these risks in a similar manner. Accordingly, OCS-B Collagen® is deemed suitable for its intended uses. Biocompatibility tests and additional performance testing demonstrate that OCS-B Collagen® meets the requirements of those standards.
The nonclinical, animal and clinical data included in this submission support the substantial equivalence of OCS-B Collagen® to the predicate devices listed in this submission. Literature and post market experience show that the device is substantially equivalent. Comparison with the predicate device shows that the device has similar specifications and performance. Therefore, it is concluded that OCS-B Collagen® is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.