IMTEC BioBarrier is a temporarily implantable material intended to be used as a space maintaining barrier over bone. It is intended for use in the oral cavity. The material is conformable to a variety of shapes as required for specific anatomical limitations. The IMTEC BioBarrier is additionally indicated for: - simultaneous use of GBR-membrane (BioBarrier) and implants: - augmentation around implants placed in immediate extraction sockets; - augmentation around implants placed in delayed extraction sockets; - localized ridge augmentation for later implantation. Because of its elasticity, the IMTEC BioBarrier should be used in combination with space-making bone graft materials, e.g., autogenous bone or bone substitutes.
Device Story
IMTEC BioBarrier is a non-absorbable, 100% polytetrafluoroethylene (PTFE) membrane; acts as a passive mechanical barrier to prevent epithelial and gingival connective tissue migration into osseous cavities. Used in oral surgery; placed by dental clinicians over bone sites to facilitate guided bone regeneration (GBR). Requires use with space-making bone graft materials (e.g., autogenous bone or substitutes) due to material elasticity. Provides mechanical support for bone repair while maintaining tissue compliance. Benefits include improved bone regeneration outcomes in dental implant procedures. Device is sterile, available in porous (5u) or non-porous full-density forms, and can be trimmed to fit anatomical requirements.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Constructed of 100% polytetrafluoroethylene (PTFE). Available in porous (5 micron) or non-porous (full density) forms. Passive mechanical barrier; non-absorbable. Supplied sterile. Dimensions are variable/customizable by clinician.
Indications for Use
Indicated for patients requiring guided bone regeneration in the oral cavity, including augmentation around dental implants in immediate or delayed extraction sockets and localized ridge augmentation. Contraindicated in patients with local infection, vascular impairment, uncontrolled diabetes, chronic high-dose steroid use, clotting disorders, anticoagulant therapy, metabolic bone disease, or other systemic disorders affecting bone/wound healing. Not for use to stabilize failed implants.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
K974752
MAR - 3 1998
#### 510 (k) Summary
# SUBMITTED BY:
M. K. Patterson, Jr. PhD Sr Vice President Regulatory Affairs IMTEC Corporation 2401 North Commerce Ardmore, Oklahoma 73401 (405) 223-4456
F.D.A Registration Number: 1645158 Owner / Operator Number: 9003407
Date Submitted: December 20 ,1997
# CLASSIFICATION/COMMON OR USUAL NAME/ DEVICE NAME:
Classification Name: AUGMENTATION MEMBRANE, 76LYC. Common/ Usual Name: GUIDED TISSUE MEMBRANE, PTFE. Proprietary Name: IMTEC/ BIOBARRIER MEMBRANE (K950306 / K972240)
### PREDICATE DEVICE:
Osteohealth, Bio-gide, K960724
# DEVICE DESCRIPTION:
IMTEC/BIOBARRIER is a non-absorbable guided tissue membrane composed of 100% Polytetrafluoroethylene. It is provided in various dimensions and can be cut to a preferred shape or size The device will be marketed as a sterile device. It is available in either Porous (5 u) or Non-Porous, "full density" PTFE.
{1}------------------------------------------------
# INDICATIONS FOR USE:
IMTEC BioBarrier is a temporarily implantable material intended to be used as a space maintaining barrier over bone. It is intended for use in the oral cavity. The material is conformable to a variety of shapes as required for specific anatomical limitations.
The IMTEC BioBarrier is additionally indicated for:
- simultaneous use of GBR-membrane (BioBarrier) and implants: -
- augmentation around implants placed in immediate extraction sockets; -
- augmentation around implants placed in delayed extraction sockets; -
- localized ridge augmentation for later implantation. 1
Because of its elasticity, the IMTEC BioBarrier should be used in combination with space-making bone graft materials, e.g., autogenous bone or bone substitutes.
## PRINCIPLES OF OPERATION:
IMTEC/ BIOBARRIER is manufactured from biocompatable,100% PTFE. It has been demonstrated to serve as a temporary mechanical barrier to undesirable tissue growth into an osseous cavity. It is designed to be a passive barrier which reduces the migration of epithelial and gingival connective tissue into a bony repair site. The material is firm enough to provide sufficient mechanical support over a site while maintaining adequately supple characteristics during placement such that tissue compliance related problems are reduced.
# CONTRAINDICATIONS:
It is critical that when used in dental implant repair applications that the implant is stable. It may NOT be used to stabilize a failed implant. Contraindications customary to the use of bone grafts and membrane techniques should be observed. These include, but are not limited to, current local infection, vascular impairment, uncontrolled diabetes, chronic high dose of steroids, clotting disorders, current anticoagulant therapy, etabolic bone disease, and other metabolic or systemic disorders which will affect bone or wound healing.
{2}------------------------------------------------
# COMPLICATIONS:
Possible complications with any oral reconstructive surgery include infection, closure perforation, abcess formation, bone loss, pain, soft tissue irregularities, and additional complications associated with anesthesia and dental surgery. Specific to this surgery is augmentation material perforation or exfoliation. Depending on the type and severity of the complications the removal of the membrane may be indicated.
# MATERIALS OF CONSTRUCTION:
IMTEC/ BIOBARRIER is constructed of 100% polytetrafluoroethylene (PTFE).
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS:
The uses of IMTEC/BIOBARRIER are similar to the currently marketed predicate devices which are used for Guided Bone Regeneration (GBR) in conjunction with endosseous dental implant placement and repair.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three human profiles incorporated into its design. The eagle is positioned in the center of a circular border, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" encircling it.
Public Health Service
OCT 1 0 2007
Food and Drug Administration 9200 Corporate Boulevard Rockville, Maryland 20850
Mr. M.K. Patterson, Jr. Ph.D. Sr. Vice President Regulatory Affairs IMTEC Corporation 2401 North Commerce Ardmore, Oklahoma 73401
Re: K974752
Trade Name: Imtec/Biobarrier Membrane Regulation Number: 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: 2 Product Code: NPK Dated: December 19, 1997 Received: December 19, 1997
Dear Dr. Patterson:
This letter corrects our substantially equivalent letter of March 3, 1998.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting (1111), is may be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
### Page 2 -
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Sattie Y. Michael Davis.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Image /page/4/Picture/7 description: The image is a circular logo with the text "FDA Centennial" in the center. The logo also includes the years "1906-2006" at the top. The text around the circle reads "Celebrating 100 Years of Service & Public Health Protection". There are also stars included in the logo.
Protecting and Promoting Public Health
{5}------------------------------------------------
# 1974753
### 510(k) Number (if known):
#### IMTEC BIOBARRIER MEMBRANE Device Name:
### Indications For Use:
IMTEC BioBarrier is a temporarily implantable material intended to be used as a space maintaining barrier over bone. It is intended for use in the oral cavity. The material is conformable to a variety of shapes as required for specific anatomical limitations.
The IMTEC BioBarrier is additionally indicated for:
- simultancous use of GBR-membrane (BioBarrier) and implants:
- augmentation around implants placed in immediate extraction sockets;
- augmentation around implants placed in delayed extraction sockets;
- localized ridge augmentation for later implantation.
Because of its clasticity, the IMTEC BioBarrier should be used in combination with space-making bone graft malerials, e.g., autogenous bone or bone substitutes.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF YEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susen Runner
(Division Sign-Off)
Division of Dental, Infection Control,
and General Hospital Devices
510(k) Number 1974752
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.