← Product Code [NHA](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA) · K252384

# DESS® Dental Smart Solutions (K252384)

_Terrats Medical SL · NHA · Oct 23, 2025 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA/K252384

## Device Facts

- **Applicant:** Terrats Medical SL
- **Product Code:** [NHA](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA.md)
- **Decision Date:** Oct 23, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3630
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

DESS® Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations. All digitally designed custom abutments for use with DESS® Pre-Milled Blank abutments are to be sent to a Terrats Medical validated milling center for manufacture.

## Device Story

DESS® Dental Smart Solutions are dental implant abutments and screws used to support prosthetic restorations. The system includes Pre-Milled Blank Abutments for CAD-CAM custom fabrication, and Multi-Unit Abutments (straight, 17°, 30°) for screw-retained restorations. Inputs include patient-specific digital designs for custom abutments, which must be manufactured at a Terrats Medical validated milling center. The device is used in a clinical setting by dental professionals. The abutments attach to endosseous dental implants to provide a stable platform for crowns or bridges. Benefits include functional and esthetic rehabilitation of edentulous areas. The device is provided either sterile or non-sterile for clinical use.

## Clinical Evidence

No clinical data included. Evidence consists of bench testing: mechanical fatigue testing per ISO 14801, bacterial endotoxin testing (LAL) per ANSI/AAMI ST72, and MR environment safety analysis per FDA guidance. Sterilization validation (gamma and moist heat) referenced from previous submissions.

## Technological Characteristics

Materials: Ti-6Al-4V alloy (ASTM F136, ISO 5832-3). Surface treatments: Anodization, Zirconium Nitride (ZrN) coating, Diamond-like Carbon (DLC) coating. Design: Pre-Milled Blanks (max 10/14mm diameter), Multi-Unit Abutments (0°, 17°, 30°). Sterilization: Gamma irradiation or moist heat. Connectivity: None (mechanical interface).

## Regulatory Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

## Predicate Devices

- DESS Dental Smart Solutions ([K242340](/device/K242340.md))

## Reference Devices

- DESS Dental Smart Solutions ([K240208](/device/K240208.md))
- DESS Dental Smart Solutions ([K222288](/device/K222288.md))
- DESS Dental Smart Solutions ([K230143](/device/K230143.md))
- DESS Dental Smart Solutions ([K170588](/device/K170588.md))
- DESS Dental Smart Solutions ([K191986](/device/K191986.md))
- DESS Dental Implants ([K212538](/device/K212538.md))
- DESS Dental Smart Solutions ([K212628](/device/K212628.md))
- Astra Tech Implant System Plus ([K111287](/device/K111287.md))
- Astra Tech Implant System ([K101732](/device/K101732.md))
- BioHorizons The Prodigy System™ Endosseous Implants ([K042429](/device/K042429.md))
- BioHorizons Tapered Internal Implant System ([K071638](/device/K071638.md))
- BioHorizons Laser-Lok 3.0 Implant System ([K093321](/device/K093321.md))
- 3i OSSEOTITE® Certain® Dental Implants ([K063341](/device/K063341.md))
- Neodent Implant System – GM Line ([K163194](/device/K163194.md))
- Neodent Implant System – GM Line ([K180536](/device/K180536.md))
- Neodent Implant System – GM Helix Implants 7.0 ([K201225](/device/K201225.md))
- NobelActive® ([K142260](/device/K142260.md))
- NobelReplace Hexagonal Implants ([K073142](/device/K073142.md))
- NobelParallel™ Conical Connection ([K173418](/device/K173418.md))
- Straumann® BLX Implant System ([K173961](/device/K173961.md))
- Straumann® BLX Line Extension ([K181703](/device/K181703.md))
- Straumann BLX Ø3.5 mm Implants ([K191256](/device/K191256.md))
- Straumann BLX Implant System ([K210855](/device/K210855.md))
- BLX WB Ø5.0, Ø5.5 and Ø6.5 mm Implants ([K212533](/device/K212533.md))
- Straumann® Bone Level Tapered Implants ([K140878](/device/K140878.md))
- Screw Vent Implant; Tapered Screw Vent Implant ([K013227](/device/K013227.md))
- Tapered Screw-Vent Implant ([K072589](/device/K072589.md))
- Spectra Dental Implant System ([K061319](/device/K061319.md))
- Legacy2, Legacy3, Legacy4, simplyLegacy2, simplyLegacy3 dental implants ([K192221](/device/K192221.md))
- InterActive / SwishPlus 2 Implant System ([K130572](/device/K130572.md))
- Simply Iconic™ Implants ([K201553](/device/K201553.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

Terrats Medical SL
% Rebecca Kattan
Regulatory Specialist
PaxMed International, LLC
1925 Palomar Oaks Way
Suite 210
Carlsbad, California 92008

October 23, 2025

Re: K252384
Trade/Device Name: DESS® Dental Smart Solutions
Regulation Number: 21 CFR 872.3630
Regulation Name: Endosseous Dental Implant Abutment
Regulatory Class: Class II
Product Code: NHA
Dated: July 30, 2025
Received: July 31, 2025

Dear Rebecca Kattan:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K252384 - Rebecca Kattan
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

{2}

K252384 - Rebecca Kattan
Page 3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Andrew I. Steen -S

Andrew I. Steen
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K252384

Device Name
DESS® Dental Smart Solutions

Indications for Use (Describe)

DESS® Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.

All digitally designed custom abutments for use with DESS® Pre-Milled Blank abutments are to be sent to a Terrats Medical validated milling center for manufacture.

Compatible Implant Systems for Pre-Milled Abutments

|  Compatible Implant System (Connection) | Implant Body Diameter, mm | Implant Platform, mm  |
| --- | --- | --- |
|  Astra Tech EV (Internal Taper) | 3.0 | 3.0  |
|   |  3.6 | 3.6  |
|   |  4.2 | 4.2  |
|  BioHorizons Internal (Internal Hex) | 3.0, 3.4, 3.8 | 3.0  |
|   |  3.8, 4.2, 4.6 | 3.5  |
|   |  4.6, 5.2, 5.8 | 4.5  |
|  Simply Iconic™ (Internal Conical) | 3.2, 3.7, 4.2, 4.7 | 3.0  |
|   |  4.7, 5.2, 5.7 | 3.4  |
|  InterActive (Internal Conical) | 3.2, 3.7 | 3.0  |
|   |  4.3, 5.0 | 3.4  |
|  Legacy™1 (Internal Hex) | 3.7, 4.2 | 3.5  |
|   |  4.7 | 4.5  |
|  Legacy™2, simplyLegacy™2, Legacy™3, simplyLegacy™3, Legacy™4 (Internal Hex) | 3.2 | 3.0  |
|   |  3.7, 4.2 | 3.5  |
|   |  4.7, 5.2 | 4.5  |
|  Neodent GM (Morse taper) | 3.5, 3.75, 4.0, 4.3, 5.0, 6.0, 7.0 | GM  |
|  Nobel Biocare NobelActive® (Conical Connection) | 3.0 | 3.0  |
|   |  3.5 | NP (3.5)  |
|   |  4.3, 5.0 | RP (3.9)  |
|  NobelReplace Conical Connection (Conical Connection) | 3.5 | NP (3.5)  |
|   |  4.3, 5.0 | RP (3.9)  |
|  NobelParallel Conical Connection (Conical Connection) | 3.75 | NP (3.5)  |
|   |  4.3, 5.0 | RP (3.9)  |
|  Osstem® TS Hiossen® ET (Internal Taper) | 3.5 | Mini  |
|   |  4.0, 4.5, 5.0, 5.5, 6.0, 7.0 | Regular  |
|  Straumann Bone Level (CrossFit® Morse Taper) | 3.3 | NC  |
|   |  4.1, 4.8 | RC  |
|  Straumann BLX (TorcFit™ Internal Hexalobular) | 3.5, 3.75, 4.0, 4.5 | RB  |
|   |  5.0, 5.5, 6.5 | WB  |
|  Zimmer Screw-Vent® Tapered Screw-Vent® (Internal Hex) | 3.3, 3.7, 4.1 | 3.5  |
|   |  4.7 | 4.5  |

Page 18 of 19

{4}

Compatible Implant Systems for Multi-Unit Abutments

|  Compatible Implant System (Connection) | Implant Body Diameter, mm | Implant Platform, mm  |
| --- | --- | --- |
|  Biomet 3i OSSEOTITE®
(External Hex) | 3.25 | 3.4  |
|   |  3.75, 4.0 | 4.1  |
|  InterActive
(Internal Conical) | 3.2, 3.7 | 3.0  |
|   |  4.3, 5.0 | 3.4  |
|  Simply Iconic™
(Internal Conical) | 3.2, 3.7, 4.2, 4.7 | 3.0  |
|   |  4.7, 5.2, 5.7 | 3.4  |
|  Legacy™1
(Internal Hex) | 3.7, 4.2 | 3.5  |
|   |  4.7 | 4.5  |
|  Legacy™2, simplyLegacy™2,
Legacy™3, simplyLegacy™3,
Legacy™4
(Internal Hex) | 3.2 | 3.0  |
|   |  3.7, 4.2 | 3.5  |
|   |  4.7, 5.2 | 4.5  |
|  Neodent GM
(Morse taper) | 3.5, 3.75, 4.0, 4.3, 5.0, 6.0, 7.0 | GM  |
|  Nobel Biocare NobelActive®
(Conical Connection) | 3.5 | NP (3.5)  |
|   |  4.3, 5.0 | RP (3.9)  |
|  NobelReplace Conical Connection
(Conical Connection) | 3.5 | NP (3.5)  |
|   |  4.3, 5.0 | RP (3.9)  |
|  NobelParallel Conical Connection
(Conical Connection) | 3.75 | NP (3.5)  |
|   |  4.3, 5.0 | RP (3.9)  |
|  Straumann Bone Level
(CrossFit® Morse Taper) | 3.3 | NC  |
|   |  4.1, 4.8 | RC  |

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

"DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740

{5}

510(k) Summary
Page 1 of 8
DESS® Dental Smart Solutions
K252384

510(k) Summary
K252384
Terrats Medical SL
DESS® Dental Smart Solutions
October 8, 2025

ADMINISTRATIVE INFORMATION

Manufacturer Name
Terrats Medical SL
Carrer Mogoda, 75-99 Barberà del Vallès
08210 Barcelona, Spain
Telephone
+34 935 646 006
Official Contact
Roger Terrats, CEO

Representative/Consultant
Rebecca E. Kattan, PhD
Kevin A. Thomas, PhD
Floyd G. Larson, MS, MBA
PaxMed International, LLC
1925 Palomar Oaks Way, Suite 210
Carlsbad, CA 92008
Telephone
+1 858-792-1235
Email
rkattan@paxmed.com
kthomas@paxmed.com
flarson@paxmed.com

DEVICE NAME AND CLASSIFICATION

Trade/Proprietary Name
DESS® Dental Smart Solutions
Common Name
Dental implant abutment

Regulation Number
21 CFR 872.3630

Regulation Name
Endosseous dental implant abutment

Regulatory Class
Class II

Product Code
NHA

Classification Panel
Dental

Reviewing Office
Office of Health Technology 1
(Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices)

Reviewing Division
Division of Dental and ENT Devices

PREDICATE AND REFERENCE DEVICE INFORMATION

Primary Predicate Device
K242340, DESS Dental Smart Solutions, Terrats Medical SL

Reference Devices
K240208, DESS Dental Smart Solutions, Terrats Medical SL
K222288, DESS Dental Smart Solutions, Terrats Medical SL
K230143, DESS Dental Smart Solutions, Terrats Medical SL
K170588, DESS Dental Smart Solutions, Terrats Medical SL
K191986, DESS Dental Smart Solutions, Terrats Medical SL
K212538, DESS Dental Implants, Terrats Medical SL
K212628, DESS Dental Smart Solutions, Terrats Medical SL

{6}

510(k) Summary
Page 2 of 8
DESS® Dental Smart Solutions
K252384

Reference Devices for OEM implant body clearances
K111287, Astra Tech Implant System Plus, Astra Tech AB
K101732, Astra Tech Implant System, Astra Tech AB
K042429, BioHorizons The Prodigy System™ Endosseous Implants, BioHorizons Implant Systems, Inc.
K071638, BioHorizons Tapered Internal Implant System, BioHorizons Implant Systems, Inc.
K093321, BioHorizons Laser-Lok 3.0 Implant System, BioHorizons Implant Systems, Inc.
K063341, 3i OSSEOTITE® Certain® Dental Implants, Implant Innovations, Inc.
K163194, Neodent Implant System – GM Line, JJGC Indústria e Comércio de Materiais Dentários SA
K180536, Neodent Implant System – GM Line, JJGC Indústria e Comércio de Materiais Dentários S.A.
K201225, Neodent Implant System – GM Helix Implants 7.0, JJGC Indústria e Comércio de Materiais Dentários S.A.
K142260, NobelActive®, Nobel Biocare AB
K073142, NobelReplace Hexagonal Implants, Nobel Biocare AB
K173418, NobelParallel™ Conical Connection, Nobel Biocare AB
K173961, Straumann® BLX Implant System, Institut Straumann AG
K181703, Straumann® BLX Line Extension – Implants, SRAs and Anatomic Abutments, Institut Straumann AG
K191256, Straumann BLX Ø3.5 mm Implants, Institut Straumann AG
K210855, Straumann BLX Implant System, Institut Straumann AG
K212533, BLX WB Ø5.0 (L18), Ø5.5 and Ø6.5 mm (L14 and L16) Implants, Institut Straumann AG
K140878, Straumann® Bone Level Tapered Implants, Straumann USA, LLC
K013227, Screw Vent Implant; Tapered Screw Vent Implant, Sulzer Dental, Inc.
K072589, Tapered Screw-Vent Implant, 4.1mmD, Zimmer Dental, Inc.
K061319, Spectra Dental Implant System, Implant Direct, LLC
K192221, Legacy2, Legacy3, Legacy4, simplyLegacy2, simplyLegacy3 dental implants; Legacy2, Legacy3, Legacy4 fixture-mounts, Implant Direct Sybron Manufacturing, LLC
K130572, InterActive / SwishPlus 2 Implant System, Implant Direct Sybron Manufacturing LLC
K201553, Simply Iconic™ Implants, Implant Direct Sybron Manufacturing LLC

## INDICATIONS FOR USE STATEMENT

DESS® Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.

All digitally designed custom abutments for use with DESS® Pre-Milled Blank abutments are to be sent to a Terrats Medical validated milling center for manufacture.

Compatible Implant Systems for Pre-Milled Abutments

|  Compatible Implant System (Connection) | Implant Body Diameter, mm | Implant Platform, mm  |
| --- | --- | --- |
|  Astra Tech EV (Internal Taper) | 3.0 | 3.0  |
|   |  3.6 | 3.6  |
|   |  4.2 | 4.2  |
|  BioHorizons Internal (Internal Hex) | 3.0, 3.4, 3.8 | 3.0  |
|   |  3.8, 4.2, 4.6 | 3.5  |
|   |  4.6, 5.2, 5.8 | 4.5  |

{7}

510(k) Summary
Page 3 of 8
DESS® Dental Smart Solutions
K252384

|  Compatible Implant System (Connection) | Implant Body Diameter, mm | Implant Platform, mm  |
| --- | --- | --- |
|  Simply Iconic™
(Internal Conical) | 3.2, 3.7, 4.2, 4.7 | 3.0  |
|   |  4.7, 5.2, 5.7 | 3.4  |
|  InterActive
(Internal Conical) | 3.2, 3.7 | 3.0  |
|   |  4.3, 5.0 | 3.4  |
|  Legacy™1
(Internal Hex) | 3.7, 4.2 | 3.5  |
|   |  4.7 | 4.5  |
|  Legacy™2, simplyLegacy™2, Legacy™3, simplyLegacy™3, Legacy™4
(Internal Hex) | 3.2 | 3.0  |
|   |  3.7, 4.2 | 3.5  |
|   |  4.7, 5.2 | 4.5  |
|  Neodent GM
(Morse taper) | 3.5, 3.75, 4.0, 4.3, 5.0, 6.0, 7.0 | GM  |
|  Nobel Biocare NobelActive®
(Conical Connection) | 3.0 | 3.0  |
|   |  3.5 | NP (3.5)  |
|   |  4.3, 5.0 | RP (3.9)  |
|  NobelReplace Conical Connection
(Conical Connection) | 3.5 | NP (3.5)  |
|   |  4.3, 5.0 | RP (3.9)  |
|  NobelParallel Conical Connection
(Conical Connection) | 3.75 | NP (3.5)  |
|   |  4.3, 5.0 | RP (3.9)  |
|  Osstem® TS
Hiossen® ET
(Internal Taper) | 3.5 | Mini  |
|   |  4.0, 4.5, 5.0, 5.5, 6.0, 7.0 | Regular  |
|  Straumann Bone Level
(CrossFit® Morse Taper) | 3.3 | NC  |
|   |  4.1, 4.8 | RC  |
|  Straumann BLX
(TorcFit™ Internal Hexalobular) | 3.5, 3.75, 4.0, 4.5 | RB  |
|   |  5.0, 5.5, 6.5 | WB  |
|  Zimmer Screw-Vent®
Tapered Screw-Vent®
(Internal Hex) | 3.3, 3.7, 4.1 | 3.5  |
|   |  4.7 | 4.5  |

## Compatible Implant Systems for Multi-Unit Abutments

|  Compatible Implant System (Connection) | Implant Body Diameter, mm | Implant Platform, mm  |
| --- | --- | --- |
|  Biomet 3i OSSEOTITE®
(External Hex) | 3.25 | 3.4  |
|   |  3.75, 4.0 | 4.1  |
|  InterActive
(Internal Conical) | 3.2, 3.7 | 3.0  |
|   |  4.3, 5.0 | 3.4  |
|  Simply Iconic™
(Internal Conical) | 3.2, 3.7, 4.2, 4.7 | 3.0  |
|   |  4.7, 5.2, 5.7 | 3.4  |
|  Legacy™1
(Internal Hex) | 3.7, 4.2 | 3.5  |
|   |  4.7 | 4.5  |
|  Legacy™2, simplyLegacy™2, Legacy™3, simplyLegacy™3, Legacy™4
(Internal Hex) | 3.2 | 3.0  |
|   |  3.7, 4.2 | 3.5  |
|   |  4.7, 5.2 | 4.5  |
|  Neodent GM | 3.5, 3.75, 4.0, 4.3, 5.0, 6.0, 7.0 | GM  |

{8}

510(k) Summary
Page 4 of 8
DESS® Dental Smart Solutions
K252384

|  Compatible Implant System (Connection) | Implant Body Diameter, mm | Implant Platform, mm  |
| --- | --- | --- |
|  (Morse taper) |  |   |
|  Nobel Biocare NobelActive® (Conical Connection) | 3.5 | NP (3.5)  |
|   |  4.3, 5.0 | RP (3.9)  |
|  NobelReplace Conical Connection (Conical Connection) | 3.5 | NP (3.5)  |
|   |  4.3, 5.0 | RP (3.9)  |
|  NobelParallel Conical Connection (Conical Connection) | 3.75 | NP (3.5)  |
|   |  4.3, 5.0 | RP (3.9)  |
|  Straumann Bone Level (CrossFit® Morse Taper) | 3.3 | NC  |
|   |  4.1, 4.8 | RC  |

# SUBJECT DEVICE DESCRIPTION

The purpose of this submission is to add abutments to the DESS Dental Smart Solutions system, which includes abutments cleared previously in K242340 and K240208. The subject device abutment designs include Pre-Milled Blank Abutments, Multi-Unit Abutments (straight and angled), and abutment screws.

Pre-Milled Blank Abutments are designed for custom abutment fabrication by a CAD-CAM process. All patient-specific custom abutment fabrication is by prescription on the order of the clinician. The Pre-Milled Blank Abutments have a maximum (before milling) diameter of 10 mm or 14 mm and a solid cylindrical design with an engaging or non-engaging implant connection. They are available in versions with a straight screw channel and with an angled screw channel for esthetic considerations. The Pre-Milled Blank Abutments are manufactured from titanium alloy (Ti 6Al-4V).

The design parameters for the CAD-CAM fabrication of a custom abutment from the Pre-Milled Blank Abutment are:

- Minimum wall thickness – 0.45 mm
- Minimum cementable post height for single-unit restoration – 4.0 mm
(minimum cementable post height for single-unit restoration is defined as the height above the restorative margin)
- Minimum gingival height – 0.5 mm
- Maximum gingival height – 6.0 mm
- Pre-Milled Blanks are for straight abutments only

Multi-Unit Abutments: Straight and Angled are designed for attachment of multi-unit screw-retained restorations and are provided in three (3) designs, straight (0°), angled 17°, and angled 30°. The designs of the subject Multi-Unit Abutments are similar to the designs of Multi-Unit Abutments cleared in K242340 and K240208. All Multi-Unit Abutments are manufactured from titanium alloy (Ti-6Al-4V).

The subject device straight Multi-Unit Abutments have a non-engaging, threaded design that attaches directly to the implant. The subject device straight (0°) Multi-Unit Abutments are included in this submission are similar in design to straight Multi-Unit Abutments cleared in K242340, K230143, and K222288. Subject device straight Multi-Unit Abutments are provided with a prosthetic platform diameter of 4.8 mm, and with a gingival height ranging from 1 mm to 5.5 mm.

The subject device angled Multi-Unit Abutments are provided only in an engaging design that requires an abutment screw. The subject device Multi-Unit Abutments angled 17° and 30° are provided with a prosthetic platform diameter of 4.8 mm, and with a gingival height ranging from 2.5 to 4.5 mm.

{9}

510(k) Summary
Page 5 of 8
DESS® Dental Smart Solutions
K252384

Select straight and angled Multi-Unit Abutments have an added coating of zirconium nitride (ZrN). This coating is identical to the ZrN coating used on abutments with ZrN cleared in K242340.

DESS® Dental Smart Solutions Screws are designed to attach the abutment to the implant or the prosthesis to the abutment. There are a total of seventeen (17) subject device screws compatible with the subject device components or previously cleared components. The new screws have designs that are similar to those of screws cleared in K242340 and K240208. Screws are made of titanium alloy (Ti-6Al-4V). Select subject device screws are available with DLC (Diamond-like Carbon) coating and are identical to that used on screws cleared in K240208.

## OEM Implant Compatibility

To ensure that the subject abutments and screws are designed to fit the corresponding OEM implants, the dimensions and tolerances of subject abutments and screws have been established by reverse engineering dimensional analysis of OEM implant bodies, OEM abutments, and OEM abutment screws or by a cooperative arrangement with the OEM implant manufacturer. The subject device abutments have the same OEM implant compatibilities as the components cleared in K170588, K191986, K212628, and K230143, except for the 3.2 mm diameter implants of the Legacy™2, simplyLegacy™2, Legacy™3, simplyLegacy™3 and Legacy™4 implant lines from Implant Direct Sybron Manufacturing, LLC, for which a new Letter of Authorization has been provided.

## Materials

The subject device abutments and the subject device screws are manufactured from Ti-6Al-4V alloy conforming to ASTM F136 and ISO 5832-3. Select abutments are colored by an anodization process that is identical to that used on abutments cleared in K242340. Select subject device straight and angled Multi-Unit abutments have a Zirconium Nitride (ZrN) coating used to increase the surface hardness and reduce wear on the surface. This coating is identical to the ZrN coating used on abutments with ZrN cleared in K242340. Select subject device screws are available with DLC (Diamond-like Carbon) coating and is identical to that used on screws cleared in K240208.

## PERFORMANCE DATA

Non-clinical data submitted or referenced to demonstrate substantial equivalence included:

- provided in this submission was non-clinical analysis to evaluate the subject devices (including all abutments and abutment screws) in the MR environment using scientific rationale and published literature (TO Woods, JG Delfino, and S Rajan, "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices," Journal of Testing and Evaluation, Volume 49, No. 2, 2021, pp. 783-795); the analysis addressed parameters per the FDA guidance Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment (issued October 2023) including magnetically induced displacement force and torque;
- referenced from K212538 was gamma irradiation sterilization validation (for the subject device implants) to a sterility assurance level of 10⁻⁶ by selecting and substantiating a 25 kGy dose using method VDmax²⁵, according to ISO 11137-1 and ISO 11137-2; analysis showed that the subject device abutments do not create a new worst case for gamma sterilization;
- provided in this submission was bacterial endotoxin testing (Limulus amebocyte lysate, LAL) according to ANSI/AAMI ST72 to demonstrate sterile product met a limit of &lt; 20 EU/device;
- referenced from K170588 was moist heat sterilization for subject device abutments provided non-sterile to the end user, validated to a sterility assurance level of 10⁻⁶ by the overkill method according to ANSI/AAMI/ISO 17665-1 and ANSI/AAMI/ISO TIR 17665-2; analysis showed that the subject device abutments do not create a new worst case for moist heat sterilization;
- referenced from K242340 and K240208 was biocompatibility of the subject device components;
- referenced from K170588, K191986, K212628, and K230143 was compatibility of the subject abutments and the OEM implants listed in the Indications for Use Statement;

{10}

510(k) Summary
Page 6 of 8
DESS® Dental Smart Solutions
K252384

- provided in this submission was a new Letter of Authorization from the OEM implant manufacturer for 3.2 mm diameter implants from Implant Direct Sybron Manufacturing, LLC; and
- provided in this submission was mechanical testing conducted according to ISO 14801 to support the performance of the subject device abutments in conjunction with the compatible OEM implants.

No clinical data were included in this submission.

## EQUIVALENCE TO MARKETED DEVICES

The subject device is substantially equivalent in intended use to the primary predicate device K242340 and the reference device K240208. All are intended to provide functional and esthetic rehabilitation of the edentulous maxilla and mandible.

The subject device IFUS and the IFUS for the primary predicate device K242340 both contain identical language regarding the use of abutments and the requirement to send Pre-Milled Blank abutments to the Terrats Medical validated milling center for manufacture. The IFUS for the subject device differs slightly from that of the primary predicate device K242340 and reference device K240208 in terms of the lists of the specific OEM implant compatibilities. These differences do not change the intended use, do not raise different questions of safety or effectiveness, and do not impact substantial equivalence.

The primary predicate device K242340 and the reference device K240208, are in support of substantial equivalence of the subject device designs, material, manufacturing, and biocompatibility.

## CONCLUSION

The subject device, the primary predicate device, and the reference devices have the same intended use, have similar technological characteristics, and are made of identical materials. The subject device, the primary predicate, and the reference devices encompass the same range of physical dimensions, are packaged in similar materials, and are sterilized using similar methods. Any differences in the technological characteristics between the subject device, the predicate device, and reference device do not raise different questions of safety or effectiveness. The data included in this submission demonstrate substantial equivalence to the predicate device and reference devices listed above.

The basis for the belief of Terrats Medical SL that the subject device is substantially equivalent to the predicate devices is summarized in the following Table of Substantial Equivalence.

{11}

510(k) Summary

Page 7 of 8

DESS® Dental Smart Solutions

K252384

Table of Substantial Equivalence

|  Comparison | Subject Device | Primary Predicate Device | Reference Device  |
| --- | --- | --- | --- |
|   |  DESS Dental Smart Solutions Terrats Medical SL | K242340 DESS Dental Smart Solutions Terrats Medical SL | K240208 DESS Dental Smart Solutions Terrats Medical SL  |
|  Product Code | NHA | NHA | NHA  |
|  Reason for Predicate Device /Reference Device | n/a | Abutment designs, OEM compatibilities, materials, manufacturing, biocompatibility | Abutment designs, OEM compatibilities, materials, manufacturing, biocompatibility  |
|  Intended Use | Functional and esthetic rehabilitation of the edentulous mandible or maxilla | Functional and esthetic rehabilitation of the edentulous mandible or maxilla | Functional and esthetic rehabilitation of the edentulous mandible or maxilla  |
|  Indications for Use Statement | DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations. All digitally designed custom abutments for use with DESS Pre-Milled Blank abutments are to be sent to a Terrats Medical validated milling center for manufacture. The complete Indications for Use Statement with OEM implant compatibilities is provided in this 510(k) Summary. | DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations. All digitally designed custom abutments for use with DESS Ti Base abutments or Pre-milled Blank Abutments are to be sent to a Terrats Medical validated milling center for manufacture. The complete Indications for Use Statement with OEM implant compatibilities is provided in the 510(k) Summary for K242340. | DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations. All digitally designed custom abutments for use with DESS Ti Base abutments or Pre-Milled Blank abutments are to be sent to a Terrats Medical validated milling center for manufacture. The complete Indications for Use Statement with OEM implant compatibilities is provided in the 510(k) Summary for K240208.  |
|  Designs | Pre-Milled Blank Abutments | Pre-Milled Blank Abutments | Pre-Milled Blank Abutments  |
|   |  Multi-Unit Abutments, Straight | Multi-Unit Abutments, Straight | Multi-Unit Abutments, Straight  |
|   |  Multi-Unit Abutments, Angled 17° and 30° | Multi-Unit Abutments, Angled 17° and 30° | Multi-Unit Abutments, Angled 17° and 30°  |
|   |  Abutment Screws | Abutment Screws | Abutment Screws  |
|  Prosthesis Attachment | Cement-retained Screw-retained | Cement-retained Screw-retained | Cement-retained Screw-retained  |
|  Restoration | Single-unit Multi-Unit | Single-unit Multi-Unit | Single-unit Multi-Unit  |
|  Prosthetic Platform Ø, mm | Pre-Milled: 10 (max) and 14 (max) Multi-Unit: 4.8 | Pre-Milled: 10 (max) and 14 (max) Multi-Unit: 4.8 | Pre-Milled: 10 (max) and 14 (max) Multi-Unit: 4.8  |

{12}

510(k) Summary
Page 8 of 8
DESS® Dental Smart Solutions
K252384

| Comparison | Subject Device | Primary Predicate Device | Reference Device |
| --- | --- | --- | --- |
| Comparison | DESS Dental Smart Solutions
Terrats Medical SL | K242340
DESS Dental Smart Solutions
Terrats Medical SL | K240208
DESS Dental Smart Solutions
Terrats Medical SL |
| Abutment Angle | Pre-Milled: Straight (0°)
Multi-Unit: 0°, 17°, 30° | Pre-Milled: 30°
Multi-Unit: 0°, 17°, 30° | Pre-Milled: Straight (0°)
Multi-Unit: 0°, 17°, 30° |
| Gingival Height, mm | Pre-Milled Blank: 6 (max)
Multi-Unit: 1.0 - 5.5 | Pre-Milled Blank: 6 (max)
Multi-Unit: 0.8 - 5.5 | Pre-Milled Blank: 6 (max)
Multi-Unit: 0.8 - 5.5 |
| Materials |  |  |  |
| Abutment Material | Ti-6Al-4V alloy | Ti-6Al-4V alloy | Ti-6Al-4V alloy
Co-Cr-Mo alloy |
| Abutment Surface | Anodization
Zirconium nitride (ZrN) | Anodization
SelectGrip® surface
Zirconium nitride (ZrN) | Anodization
SelectGrip® surface
Zirconium nitride (ZrN) |
| Screw Material | Ti-6Al-4V alloy
DLC coating | Ti-6Al-4V alloy | Ti-6Al-4V alloy
DLC coating |
| How Provided |  |  |  |
| Sterilization | Sterile by gamma irradiation, and
Non-sterile | Sterile by gamma irradiation, and
Non-sterile | Sterile by gamma irradiation, and
Non-sterile |
| Usage – All Components | Single patient, single use | Single patient, single use | Single patient, single use |

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA/K252384](https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA/K252384)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
