← Product Code [NHA](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA) · K242686

# Neodent Implant System (K242686)

_Jjgc Indústria E Comércia DE Materiais Dentários S.A. · NHA · Dec 4, 2024 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA/K242686

## Device Facts

- **Applicant:** Jjgc Indústria E Comércia DE Materiais Dentários S.A.
- **Product Code:** [NHA](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA.md)
- **Decision Date:** Dec 4, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3630
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

NeoConvert Solution: Instruments and components used to perform the procedure of converting removable prostheses into temporary fixed prostheses on GM line implants installed in the maxilla and/or mandible. Direct Screw to MUA: The Direct Screw to MUA is suitable for attaching the multiple restoration milled from a CAD/CAM library to the prosthetic abutment (mini or micro abutments). GM Attachment TiN 30°: This product is used to stabilize implant-mucosa supported removable full prostheses (retained on the implant and supported on the mucosa), on implants installed in the maxilla or mandibula. It can be used in immediate or conventional rehabilitation procedures.

## Device Story

Neodent Implant System additions include NeoConvert Solution, Direct Screw to MUA, and GM Attachment TiN 30°. NeoConvert converts removable temporary prostheses into screw-retained fixed temporary prostheses. Direct Screw to MUA fixes digitally designed multi-unit milled restorations directly onto mini/micro abutments, bypassing passive seating cylinders. GM Attachment TiN 30° stabilizes implant-mucosa-supported removable full prostheses. Devices are used in dental clinics by clinicians; components are compatible with existing Neodent implants. Personalized copings for Direct Screw to MUA are manufactured at a Straumann-validated milling center. These components provide additional options for multi-unit rehabilitation, improving chewing function and patient aesthetics.

## Clinical Evidence

Bench testing only. Dynamic fatigue testing per ISO 14801 and torsion testing demonstrated performance equal or superior to predicates. Biocompatibility evaluated per ISO 10993-1, -5, and -18. Sterilization validated per ISO 11135 and ISO 17665. No clinical data presented.

## Technological Characteristics

Materials: Titanium alloy (ASTM F136). Components: NeoConvert (screw, coping, distal bar), Direct Screw to MUA, GM Attachment TiN 30°. Sterilization: Ethylene Oxide (SAL 1x10^-6) or moist heat (steam) for non-sterile components. Connectivity: None (mechanical prosthetic components). Manufacturing: CAD/CAM library-based milling for personalized copings.

## Regulatory Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

## Predicate Devices

- Neodent Implant System ([K163194](/device/K163194.md))

## Reference Devices

- Neodent Implant System ([K101945](/device/K101945.md))
- Nuvo IF Implant System ([K193234](/device/K193234.md))
- Straumann SC Variobase Abutments ([K162890](/device/K162890.md))
- Neodent Implant System - GM Line ([K180536](/device/K180536.md))
- GM Titanium Base for Bridge ([K192229](/device/K192229.md))
- NobelProcera Zirconia Implant Bridge ([K202452](/device/K202452.md))
- NUVO CF Implant System ([K203309](/device/K203309.md))
- Mini Abutment 60° ([K203542](/device/K203542.md))
- GM Narrow Implant System ([K220251](/device/K220251.md))
- Helix Short Implant System ([K223638](/device/K223638.md))
- Neodent Implant System – GM Line ([K173902](/device/K173902.md))
- MRI Compatibility for Existing Neodent Implant System ([K182620](/device/K182620.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 4, 2024

JJGC Indústria e Comércia de Materiais Dentários S.A. % Jennifer Jackson Sr. Director, Regulatory Affairs and Quality Straumann USA. LLC 60 Minuteman Road Andover, Massachusetts 01810

Re: K242686

Trade/Device Name: Neodent Implant System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: September 6, 2024 Received: September 6, 2024

Dear Jennifer Jackson:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,
Andrew I. Steen -S

Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

### Indications for Use

510(k) Number (if known) K242686

Device Name Neodent Implant System

### Indications for Use (Describe)

NeoConvert Solution:

The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. The NeoConvert Solution may be used with single-stage or two-stage procedures, for temporary multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.

#### Direct Screw to MUA:

The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. The Direct Screw to MUA may be used with single-stage or two-stage procedures, for screw-retained multi-unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. All digitally designed personalized copings to be used with the Direct to Multi Unit Abutment Screw are intended to be sent to Straumann for manufacture at a validated milling center.

#### GM Attachment TiN 30°:

The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. The GM Attachment TiN 30° may be used with single-stage or two-stage procedures, for multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------|---------------------------------------------|
|----------------------------------------------|---------------------------------------------|

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

### Neodent Implant System

510(k) Summary

# 510(k) Summary

### Submitter's Contact Information

| Submitter:             | Straumann USA, LLC<br>60 Minuteman Road<br>Andover, MA 01810, USA<br>Registration No.: 1222315 Owner/Operator No.: 9005052                                                                                             |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                        | On the behalf of:                                                                                                                                                                                                      |
|                        | JJGC Indústria e Comércio de Materiais Dentários S.A (dba Neodent)<br>Av. Juscelino Kubitschek de Oliveira, 3291<br>Curitiba, Paraná, Brazil 81270-200<br>Registration No.: 3008261720<br>Owner/Operator No.: 10031702 |
| Contact Person:        | Jennifer M. Jackson, MS, RAC<br>Sr. Director, Regulatory Affairs and Quality<br>Phone Number: +1 978 747-2509<br>Fax Number: +1 978 747-0023<br>Email: jennifer.jackson@straumann.com                                  |
| Prepared By &          | Leticia Milani                                                                                                                                                                                                         |
| Alternate Contact:     | Regulatory Affairs Analyst<br>JJGC Indústria e Comércio de Materiais Dentários SA                                                                                                                                      |
| Date Prepared:         | December 2, 2024                                                                                                                                                                                                       |
| Name of the Device     |                                                                                                                                                                                                                        |
| Trade Names:           | Neodent Implant System                                                                                                                                                                                                 |
| Common Name:           | Endosseous dental implant abutment                                                                                                                                                                                     |
| Classification Name:   | Endosseous dental implant abutment                                                                                                                                                                                     |
| Regulation Number:     | 21 CFR 872.3630                                                                                                                                                                                                        |
| Device Classification: | II                                                                                                                                                                                                                     |
| Product Code(s):       | NHA                                                                                                                                                                                                                    |
| Classification Panel:  | Dental                                                                                                                                                                                                                 |

{5}------------------------------------------------

### Neodent Implant System

510(k) Summary

#### Predicate Device(s)

#### NeoConvert Solution:

Primary Predicate:

K163194 Neodent Implant System (JJGC Indústria e Comércio de Materiais Dentários S.A)

Reference Devices:

| K101945 | Neodent Implant System (JJGC Indústria e Comércio de Materiais Dentários S.A. |
|---------|-------------------------------------------------------------------------------|
|---------|-------------------------------------------------------------------------------|

K193234 Nuvo IF Implant System (JJGC Indústria e Comércio de Materiais Dentários S.A)

#### Direct Screw to MUA:

Primary Predicate:

K163194 Neodent Implant System (JJGC Indústria e Comércio de Materiais Dentários S.A)

#### Reference Devices:

| K162890 | Straumann SC Variobase Abutments (Straumann USA, LLC)                                   |
|---------|-----------------------------------------------------------------------------------------|
| K180536 | Neodent Implant System - GM Line (JJGC Indústria e Comércio de Materiais Dentários S.A) |
| K192229 | GM Titanium Base for Bridge (JJGC Indústria e Comércio de Materiais Dentários S.A)      |
| K202452 | NobelProcera Zirconia Implant Bridge (Nobel Biocare AB)                                 |
| K203309 | NUVO CF Implant System (JJGC Indústria e Comércio de Materiais Dentários S.A)           |
| K203542 | Mini Abutment 60°(JJGC Indústria e Comércio de Materiais Dentários S.A)                 |
| K220251 | GM Narrow Implant System (JJGC Indústria e Comércio de Materiais Dentários S.A)         |
| K223638 | Helix Short Implant System (JJGC Indústria e Comércio de Materiais Dentários S.A)       |

#### GM Attachment TiN 30°:

Primary Predicate:

- K163194 Neodent Implant System (JJGC Indústria e Comércio de Materiais Dentários S.A)
#### Reference Devices:

Neodent Implant System – GM Line (JJGC Indústria e Comércio de Materiais Dentários S.A) K173902

- K192229 GM Titanium Base for Bridge (JJGC Indústria e Comércio de Materiais Dentários S.A)
#### MRI:

K182620 MRI Compatibility for Existing Neodent Implant System

{6}------------------------------------------------

### Neodent Implant System

510(k) Summary

### Device Description

This premarket notification includes the following additions to the Neodent Implant System:

- . NeoConvert Solution
- . Direct Screw to MUA
- GM Attachment TiN 30° .

The Neodent Multi-Unit Abutments proposed on this submission are similar to devices already cleared in previous submissions of Neodent Implant System, according to predicate devices described above. This submission intends to expand the portfolio of Neodent Abutments with these new solutions for multi-unit restorations, to provide more treatment options to customers.

The NeoConvert Solution is a technique comprised by some abutments (not subject of this submission) used for converting removable temporary prosthesis into a screw-retained fixed temporary prosthesis.

The Direct Screw to MUA is a prosthetic component designed to fix planned and digitally designed multi-unit milled restorations directly onto mini or micro abutments, eliminating the need for using the passive seating cylinder to restore chewing function.

The GM Attachment TiN 30° is a prosthetic component used to stabilize implant-mucosa-supported removable full prosthesis (retained on the implant and supported on the mucosa), on implants installed in the maxilla or mandible.

Although all these three new solutions are compatible with the same implants and components of Neodent Implant System and used for multiple unit rehabilitation, they are not used in conjunction with each other.

#### Intended Use

NeoConvert Solution:

Instruments and components used to perform the procedure of converting removable prostheses into temporary fixed prostheses on GM line implants installed in the maxilla and/or mandible.

#### Direct Screw to MUA:

The Direct Screw to MUA is suitable for attaching the multiple restoration milled from a CAD/CAM library to the prosthetic abutment (mini or micro abutments).

{7}------------------------------------------------

### Neodent Implant System

510(k) Summary

### GM Attachment TiN 30°:

This product is used to stabilize implant-mucosa supported removable full prostheses (retained on the implant and supported on the mucosa), on implants installed in the maxilla or mandibula. It can be used in immediate or conventional rehabilitation procedures.

### Indications for Use

### NeoConvert Solution:

The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. The NeoConvert Solution may be used with single-stage or two-stage procedures, for temporary multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.

### Direct Screw to MUA:

The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. The Direct Screw to MUA may be used with single-stage or two-stage procedures, for screw-retained multi-unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. All digitally designed personalized copings to be used with the Direct to Multi Unit Abutment Screw are intended to be sent to Straumann for manufacture at a validated milling center.

### GM Attachment TiN 30°:

The Neodent Implant System is intended to be surqically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. The GM Attachment TiN 30° may be used with single-stage or two-stage procedures, for multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.

{8}------------------------------------------------

## Neodent Implant System

510(k) Summary

### Technological Characteristics

The technological characteristics of the subject devices are compared to the primary predicate and reference devices in the following tables:

| FEATURE                  | PROPOSED DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                                                               | PRIMARY PREDICATE DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                            | REFERENCE DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number                 | K242686<br>Neodent Implant System                                                                                                                                                                                                                                                                                                                                                                                                                                             | K163194<br>Neodent Implant System                                                                                                                                                                                                                                                                                                                                                                                                                   | K101945<br>Neodent Implant System                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | K193234<br>Nuvo IF Implant System                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Indications<br>for Use   | The Neodent Implant System is intended<br>to be surgically placed in the bone of the<br>upper or lower jaw to provide support for<br>prosthetic devices, such as artificial teeth,<br>to restore chewing function.<br>The<br>NeoConvert Solution may be used with<br>single-stage or two-stage procedures, for<br>temporary multiple unit restorations, and<br>may be loaded immediately when good<br>primary stability is achieved and with<br>appropriate occlusal loading. | The Neodent Implant System is intended to<br>be surgically placed in the bone of the upper<br>or lower jaw to provide support for prosthetic<br>devices, such as artificial teeth, to restore<br>chewing function. It may be used with single-<br>stage or two-stage procedures, for single or<br>multiple unit restorations, and may be loaded<br>immediately when good primary stability is<br>achieved and with appropriate occlusal<br>loading. | The Neodent Implant System is intended to be<br>surgically placed in the bone of the upper or<br>lower jaw to provide support for prosthetic<br>devices, such as artificial teeth, to restore<br>chewing function. It may be used with single-<br>stage or two-stage procedures, for single or<br>multiple unit restorations, and may be loaded<br>immediately when good primary stability is<br>achieved and with appropriate occlusal loading.<br>Multiple tooth applications may be rigidly<br>splinted. | The Implant System is intended to be<br>surgically placed in the maxilla or mandible to<br>provide support for prosthetic devices such<br>as artificial teeth in order to restore chewing<br>function. It may be used with single-stage or<br>two-stage procedures, for single or multiple<br>unit restorations, and may be loaded<br>immediately when good primary stability is<br>achieved and with appropriate occlusal<br>loading. Multiple tooth applications may be<br>rigidly splinted. |
| Material                 | Titanium alloy, according to ASTM F136.                                                                                                                                                                                                                                                                                                                                                                                                                                       | Titanium alloy, according to ASTM F136.                                                                                                                                                                                                                                                                                                                                                                                                             | Titanium alloy, according to ASTM F136.                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Titanium alloy, according to ASTM F136.                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Overall<br>dimensions    | NeoConvert Screw:<br>● Laying diameter: 1.4 mm<br>● Height: 3.7 mm<br><br>NeoConvert Coping:<br>● Laying diameter: 4.8 mm<br>● Height: 5.0 and 6.0 mm<br><br>NeoConvert Distal Bar:<br>● Laying diameter: 4.8 mm<br>● Height: 5 mm                                                                                                                                                                                                                                            | Mini Conical Abutment Coping Screw:<br>● Laying diameter: 1.4 mm<br>● Height: 3.7 mm                                                                                                                                                                                                                                                                                                                                                                | Mini Conical Abutment Coping:<br>● Laying diameter: 4.8 mm<br>● Height: 12 mm                                                                                                                                                                                                                                                                                                                                                                                                                               | Nuvo Distal Bar:<br>● Laying diameter: 5.2 mm<br>● Height: 10 mm                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Commercially<br>Supplied | NeoConvert Coping and Screw:<br>One cavity blister with five units.<br><br>NeoConvert Distal Bar:<br>Two cavity blister with one unit and one<br>screw.                                                                                                                                                                                                                                                                                                                       | Mini Conical Abutment Coping Screw:<br>One cavity blister with one unit.                                                                                                                                                                                                                                                                                                                                                                            | Mini Conical Abutment Coping:<br>Two cavity blister with one unit and one screw.                                                                                                                                                                                                                                                                                                                                                                                                                            | Nuvo Distal Bar:<br>Two cavities blister with one unit and one<br>screw.                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Single use               | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| FEATURE                  | PROPOSED DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                                                               | PRIMARY PREDICATE DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                            | REFERENCE DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| K Number                 | K242686<br>Neodent Implant System                                                                                                                                                                                                                                                                                                                                                                                                                                             | K163194<br>Neodent Implant System                                                                                                                                                                                                                                                                                                                                                                                                                   | K101945<br>Neodent Implant System<br>K193234<br>Nuvo IF Implant System                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Sterilization<br>Method  | Provided sterile via Ethylene Oxide. The<br>coping and screw, if opened and not<br>used, must be terminally sterilized by<br>user via moist heat with cycle parameters<br>validated to a SAL of 1x10-6.                                                                                                                                                                                                                                                                       | Provided sterile via Ethylene Oxide.                                                                                                                                                                                                                                                                                                                                                                                                                | Provided sterile via Ethylene Oxide. Terminally<br>sterilized by user via moist heat with cycle<br>parameters validated to a SAL of 1x10-6.<br>Provided sterile via Ethylene Oxide.                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |

#### Table 1 – Table of Substantial Equivalence NeoConvert

{9}------------------------------------------------

# Neodent Implant System

510(k) Summary

{10}------------------------------------------------

# Neodent Implant System

### 510(k) Summary

### Table 2 – Table of Substantial Equivalence Direct Screw to MUA (Zygomatic Implants)

| FEATURE                  | PROPOSED DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | PRIMARY PREDICATE DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                            | REFERENCE DEVICE                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                         |  |  |  |  |  |  |  |  |  |  |  |  |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|
| K Number                 | K242686<br>Neodent Implant System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | K163194<br>Neodent Implant System                                                                                                                                                                                                                                                                                                                                                                                                                   | K202452<br>NobelProcera Zirconia Implant Bridge                                                                                                                                                                                     | K203542<br>Mini Abutment 60°                                                                                                                                                                                                                                                                                                                                                                            |  |  |  |  |  |  |  |  |  |  |  |  |
| Indications<br>for Use   | The Neodent Implant System is intended to<br>be surgically placed in the bone of the<br>upper or lower jaw to provide support for<br>prosthetic devices, such as artificial teeth,<br>to restore chewing function. The Direct<br>Screw to MUA may be used with single-<br>stage or two-stage procedures, for screw-<br>retained multi-unit restorations, and may be<br>loaded immediately when good primary<br>stability is achieved and with appropriate<br>occlusal loading. All digitally designed<br>personalized copings to be used with the<br>Direct to Multi Unit Abutment Screw are<br>intended to be sent to Straumann for<br>manufacture at a validated milling center. | The Neodent Implant System is intended to<br>be surgically placed in the bone of the upper<br>or lower jaw to provide support for prosthetic<br>devices, such as artificial teeth, to restore<br>chewing function. It may be used with single-<br>stage or two-stage procedures, for single or<br>multiple unit restorations, and may be loaded<br>immediately when good primary stability is<br>achieved and with appropriate occlusal<br>loading. | The NobelProcera Zirconia Implant Bridge<br>are indicated for use as a bridge anatomically<br>shaped and/or framework in the treatment of<br>partially or totally edentulous jaws for the<br>purpose of restoring chewing function. | The Mini Conical Abutments are indicated for<br>use with Zygomatic Implants, in case of<br>severe jaw resorption, in order to restore<br>patient aesthetics and chewing function. It<br>may be used with single-stage or two-stage<br>procedures, for multiple unit restorations, and<br>may be loaded immediately when good<br>primary stability is achieved and with<br>appropriate occlusal loading. |  |  |  |  |  |  |  |  |  |  |  |  |
| Material                 | Titanium alloy, according to ASTM F136.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Titanium alloy, according to ASTM F136.                                                                                                                                                                                                                                                                                                                                                                                                             | Titanium alloy, according to ASTM F136.                                                                                                                                                                                             | Titanium alloy, according to ASTM F136.                                                                                                                                                                                                                                                                                                                                                                 |  |  |  |  |  |  |  |  |  |  |  |  |
| Overall<br>dimensions    | Diameter of the screw head: 2.2 mm<br>Height of screw head: 1.7 mm<br>Total height: 4.1 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Diameter of the screw head: 1.9 mm<br>Diameter of the external thread: 1.4 mm<br>Height of screw head: 1.7 mm<br>Total height: 3.7 mm                                                                                                                                                                                                                                                                                                               | Diameter of the screw head: 2.0 mm<br>Total height: 4.0 mm                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                         |  |  |  |  |  |  |  |  |  |  |  |  |
| Commercially<br>Supplied | One cavity blister with five units.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | One cavity blister with one units.                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                         |  |  |  |  |  |  |  |  |  |  |  |  |
| Single Use               | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                     |  |  |  |  |  |  |  |  |  |  |  |  |
| Sterilization<br>Method  | Provided non-sterile. Must be sterilized<br>before installation via moist heat (steam),<br>using either gravity or vacuum, with an<br>exposure time of 3 minutes at 132°C<br>(270°F).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Provided non-sterile. Must be sterilized<br>before installation via moist heat (steam),<br>using either gravity or vacuum, at 132°C<br>(270°F) or 134°C (273°F).                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                         |  |  |  |  |  |  |  |  |  |  |  |  |
| Top Half<br>Material     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                     | Material Minimum thickness   N!ce® HT 0.4 mm   N!ce® LT 0.4 mm   N!ce® XT 0.4 mm   Ticon® 0.4 mm   Coron® 0.4 mm                                                                                                                                                                                                                                                                                        |  |  |  |  |  |  |  |  |  |  |  |  |

{11}------------------------------------------------

# Neodent Implant System

### 510(k) Summary

### Table 3 – Table of Substantial Equivalence Direct Screw to MUA (GM Implants)

| FEATURE                   | PROPOSED DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | PRIMARY PREDICATE DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                           | REFERENCE PREDICATE DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                              |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number                  | K242686<br>Neodent Implant System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | K163194<br>Neodent Implant System                                                                                                                                                                                                                                                                                                                                                                                                                  | K192229<br>GM Titanium Base for Bridge<br>K203309<br>NUVO CF Implant System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                              |
| Indications<br>for Use    | The Neodent Implant System is intended to<br>be surgically placed in the bone of the upper<br>or lower jaw to provide support for prosthetic<br>devices, such as artificial teeth, to restore<br>chewing function. The Direct Screw to MUA<br>may be used with single-stage or two-stage<br>procedures, for screw-retained multi-unit<br>restorations, and may be loaded immediately<br>when good primary stability is achieved and<br>with appropriate occlusal loading. All digitally<br>designed personalized copings to be used<br>with the Direct to Multi Unit Abutment Screw<br>are intended to be sent to Straumann for<br>manufacture at a validated milling center. | The Neodent Implant System is intended to<br>be surgically placed in the bone of the<br>upper or lower jaw to provide support for<br>prosthetic devices, such as artificial teeth,<br>to restore chewing function. It may be used<br>with single-stage or two-stage procedures,<br>for single or multiple unit restorations, and<br>may be loaded immediately when good<br>primary stability is achieved and with<br>appropriate occlusal loading. | Titanium Base Abutment is a titanium base<br>placed onto Neodent dental implants to provide<br>support for customized prosthetic restorations.<br>It is used with a coping and crown, or crown<br>alone, and is indicated for cement-retained<br>single or multi-unit restorations or screw-<br>retained single restorations. All digitally<br>designed copings and/or crowns to be used<br>with the Neodent Titanium Base Abutment<br>System are intended to be sent to Straumann<br>for manufacture at a validated milling center.<br>The GM Titanium Base for Bridge is indicated<br>for cement or screw-retained multi-unit<br>restorations. | The CF Titanium Base for Bridge is a<br>titanium abutment placed onto dental<br>implants to provide support for customized<br>prosthetic restorations (multiple structures).<br>It is indicated for multi-unit restorations,<br>cement-retained or screw-retained in<br>aesthetic areas on implants installed in the<br>maxilla or mandible. |
| Material                  | Titanium alloy, according to ASTM F136.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Titanium alloy, according to ASTM F136.                                                                                                                                                                                                                                                                                                                                                                                                            | Titanium alloy, according to ASTM F136.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Titanium alloy, according to ASTM F136.                                                                                                                                                                                                                                                                                                      |
| Overall<br>dimensions     | Diameter of the screw head: 2.2 mm<br>Height of screw head: 1.7 mm<br>Total height: 4.1 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Diameter of the screw head: 1.9 mm<br>Diameter of the external thread: 1.4 mm<br>Height of screw head: 1.7 mm<br>Total height: 3.7 mm                                                                                                                                                                                                                                                                                                              | Diameter: 3.5mm<br>Cementable area height 4.0mm<br>Gingival Height: 0.8 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Diameter: 3.5 mm<br>Cementable area height: 4.5 mm<br>Gingival height: 0.5 mm                                                                                                                                                                                                                                                                |
| Commercially<br>Supplied  | One cavity blister with five units.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | One cavity blister with one units.                                                                                                                                                                                                                                                                                                                                                                                                                 | Two cavity blister with one unit and one screw.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Two cavity blister with one unit and one<br>screw.                                                                                                                                                                                                                                                                                           |
| Single Use                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Yes                                                                                                                                                                                                                                                                                                                                          |
| Sterilization<br>Method   | Provided non-sterile. Must be sterilized<br>before installation via moist heat (steam),<br>using either gravity or vacuum, with an<br>exposure time of 3 minutes at 132°C (270°F).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Provided sterile by Ethylene Oxide. Must be<br>sterilized before installation via moist heat<br>(steam), using either gravity or vacuum, at<br>132°C (270°F).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Provided sterile by Ethylene Oxide. Must be<br>sterilized before installation via moist heat<br>(steam), using either gravity or vacuum, at<br>132°C (270°F).                                                                                                                                                                                |
| Top Half<br>Material      | Material                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Minimum thickness                                                                                                                                                                                                                                                                                                                                                                                                                                  | Material…

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA/K242686](https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA/K242686)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
