K221301 · Terrats Medical SL · NHA · Mar 9, 2023 · Dental
Device Facts
Record ID
K221301
Device Name
DESS Dental Smart Solutions
Applicant
Terrats Medical SL
Product Code
NHA · Dental
Decision Date
Mar 9, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations. All digitally designed custom abutments for use with DESS Bases or Pre-milled Blanks are to be sent to a Terrats Medical validated milling center for manufacture, or to be designed and manufactured according to the digital dentistry workflow integrates multiple components: scan files from intra-oral and lab (desktop) scanners, CAM software, ceramic material, milling machine, and associated tooling and accessories.
Device Story
DESS Dental Smart Solutions abutments provide prosthetic support for dental implants. System includes Base Abutments and Pre-milled (Blank) Abutments. Input: digital scan files from intra-oral or lab scanners. Workflow: CAD software designs custom abutment/superstructure; CAM software generates milling instructions; milling machine fabricates components from ceramic (VITA YZ ST/XT) or titanium/CoCr blanks. Used in dental clinics/labs by clinicians/technicians. Output: patient-specific custom abutment or zirconia superstructure. Clinical benefit: functional and esthetic rehabilitation of edentulous maxilla/mandible. Software locks design parameters (e.g., wall thickness, gingival height) to ensure structural integrity. Provider uses output to restore chewing function.
Clinical Evidence
No clinical data included. Evidence consists of bench testing: sterilization validation (ISO 17665-1/2, ISO 14937), biocompatibility (ISO 10993-5/12), fatigue testing, and reverse engineering of OEM implant connections. Software verification confirmed design parameter restrictions and locking of CAM zones to prevent user modification of connection geometry.
Technological Characteristics
Materials: Ti-6Al-4V ELI (ASTM F136), Co-Cr-Mo (ASTM F1537), and Y-TZP Zirconia (ISO 6872). Components: Base Abutments, Pre-milled Blanks, screws. Surface: SelectGrip®, DLC coating, anodization. Connectivity: Digital workflow integration (CAD/CAM). Sterilization: Non-sterile. Software: Locked design library/parameters.
Indications for Use
Indicated for patients requiring prosthetic restoration supported by endosseous dental implants in the maxillary or mandibular arch. No specific age or gender restrictions. Contraindications include conditions precluding dental implant surgery.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Terrats Medical SL % Melissa Burbage Senior Regulatory Specialist PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K221301
Trade/Device Name: DESS Dental Smart Solutions Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA, PNP Dated: February 8, 2023 Received: February 8, 2023
Dear Melissa Burbage:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221301
Device Name
#### DESS Dental Smart Solutions
Indications for Use (Describe)
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
All digitally designed custom abutments for use with DESS Bases or Pre-milled Blanks are to be sent to a Terrats Medical validated milling center for manufacture, or to be designed and manufactured according to the digital dentistry workflow integrates multiple components: scan files from intra-oral and lab (desktop) scanners, CAM software, ceramic material, milling machine, and associated tooling and accessories.
| Implant System Compatibility | Implant Diameter (mm) | Implant Platform Name |
|------------------------------------------|------------------------------------|-----------------------|
| Ankylos C/X | 3.5, 4.5, 5.5 | 3.5, 4.5, 5.5 |
| | 3.0 | 3.0 |
| Astra Tech EV | 3.6 | 3.6 |
| | 4.2 | 4.2 |
| | 4.8 | 4.8 |
| | 5.4 | 5.4 |
| Astra Tech OsseoSpeed™ | 3.0 | 3.0 |
| | 3.5/4.0 | 3.5/4.0 |
| | 4.5/5.0 | 4.5/5.0 |
| BioHorizons Internal | 3.0, 3.4, 3.8 | 3.0 |
| | 3.8, 4.6 | 3.5 |
| | 4.6, 5.8 | 4.5 |
| | 5.8 | 5.7 |
| Biomet 3i Certain® | 3.25 | 3.4 |
| | 4.0 | 4.1 |
| | 5.0 | 5.0 |
| Biomet 3i OSSEOTITE® | 3.25 | 3.4 |
| | 3.75, 4.0 | 4.1 |
| | 5.0 | 5.0 |
| Camlog | 3.3 | 3.3 |
| | 3.8 | 3.8 |
| | 4.3 | 4.3 |
| | 5.0 | 5.0 |
| Friadent XiVE® | 3.4 | 3.4 |
| | 3.8 | 3.8 |
| | 4.5 | 4.5 |
| | 5.5 | 5.5 |
| MegaGen AnyRidge | 3.5, 4.0, 4.5, 5.0, 5.5 | 3.5 |
| Neodent Grand Morse | 3.5, 3.75, 4.0, 4.3, 5.0, 6.0, 7.0 | Grand Morse (GM) |
| NobelActive® | 3.0 | 3.0 |
| | 3.5 | NP |
| NobelReplace/ NobelParallel Conical | 4.3, 5.0 | RP |
| | 5.5 | WP |
| NobelReplace® Trilobe | 3.5 | NP |
| | 4.3 | RP |
| | 5.0 | WP |
| | 6.0 | 6.0 |
| Nobel Brånemark System® | 3.3 | NP |
| | 3.75, 4.0 | RP |
| | 5.0 | WP |
| Osstem TS | 3.5 | Mini |
| | 4.0, 4.5, 5.0, 6.0, 7.0 | Regular |
| Straumann® BLX | 3.5, 3.75, 4.0, 4.5 | RB |
| | 5.0, 5.5, 6.5 | WB |
| Straumann® Bone Level | 3.3 | NC |
| | 4.1, 4.8 | RC |
| | 3.3 | NNC |
| Straumann® Tissue Level | 3.3, 4.1, 4.8 | RN |
| Zimmer Screw-Vent® / Tapered Screw-Vent® | 3.3, 3.7, 4.1 | 3.5 |
| | 4.7 | 4.5 |
| | 6.0 | 5.7 |
{3}------------------------------------------------
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of2
PSC Publishing Services (301) 443-6740 EF
{4}------------------------------------------------
#### 510(k) Summary
# Terrats Medical SL DESS® Dental Smart Solutions
# March 9, 2023
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | Terrats Medical SL<br>Carrer Mogoda 75-99<br>Barberà del Vallès 08210<br>Barcelona, Spain |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Telephone: +34 935 646 006<br>Fax: +34 935 647 317 |
| Official Contact | Roger Terrats, CEO |
| Representative/Consultant | Melissa Burbage<br>Kevin A. Thomas, PhD; Floyd G. Larson, MS, MBA<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130 |
Telephone: +1 858-792-1235 Fax: +1 858-792-1236 Email: mburbage@paxmed.com; kthomas@paxmed.com; flarson@paxmed.com
#### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | DESS Dental Smart Solutions |
|------------------------|------------------------------------|
| Common Name | Dental implant abutment |
| Regulation Number | 21 CFR 872.3630 |
| Regulation Name | Endosseous dental implant abutment |
| Regulatory Class | Class II |
| Product Code | NHA, PNP |
| Classification Panel | Dental Products Panel |
| Reviewing Division | DHT1B: Division of Dental Devices |
#### PREDICATE DEVICE INFORMATION
Primary Predicate Device K222288, DESS Dental Smart Solutions, Terrats Medical SL
Reference Devices K201860, Elos Accurate® Hybrid Base™, Elos Medtech Pinol A/S K180703, VITA YZ ST and VITA YZ XT, VITA Zahnfabrik H.Rauter GmbH. Co. K130436, Multilink Hybrid Abutment Cement, Ivoclar Vivadent AG K151455, 3Shape Abutment Designer Software, 3Shape A/S
{5}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
All digitally designed custom abutments for use with DESS Bases or Pre-milled Blank are to be sent to a Terrats Medical validated milling center for manufacture, or to be designed and manufactured according to the digital dentistry workflow. The digital dentistry workflow integrates multiple components: scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine, and associated tooling and accessories. .. . . .
| Compatible Implant Systems | | |
|------------------------------------------|------------------------------------|-----------------------|
| Implant System Compatibility | Implant Diameter (mm) | Implant Platform Name |
| Ankylos C/X | 3.5, 4.5, 5.5 | 3.5, 4.5, 5.5 |
| | 3.0 | 3.0 |
| | 3.6 | 3.6 |
| Astra Tech EV | 4.2 | 4.2 |
| | 4.8 | 4.8 |
| | 5.4 | 5.4 |
| | 3.0 | 3.0 |
| | 3.5/4.0 | 3.5/4.0 |
| Astra Tech OsseoSpeedTM | | |
| | 4.5/5.0 | 4.5/5.0 |
| | 3.0, 3.4, 3.8 | 3.0 |
| BioHorizons Internal | 3.8, 4.6 | 3.5 |
| | 4.6, 5.8 | 4.5 |
| | 5.8 | 5.7 |
| | 3.25 | 3.4 |
| Biomet 3i Certain® | 4.0 | 4.1 |
| | 5.0 | 5.0 |
| | 3.25 | 3.4 |
| Biomet 3i OSSEOTITE® | 3.75, 4.0 | 4.1 |
| | 5.0 | 5.0 |
| | 3.3 | 3.3 |
| | 3.8 | 3.8 |
| Camlog | 4.3 | 4.3 |
| | 5.0 | 5.0 |
| | 3.4 | 3.4 |
| | 3.8 | 3.8 |
| Friadent XiVE® | 4.5 | 4.5 |
| | 5.5 | 5.5 |
| MegaGen AnyRidge | 3.5, 4.0, 4.5, 5.0, 5.5 | 3.5 |
| Neodent Grand Morse | 3.5, 3.75, 4.0, 4.3, 5.0, 6.0, 7.0 | Grand Morse (GM) |
| | 3.0 | 3.0 |
| NobelActive® | 3.5 | NP |
| NobelReplace/ NobelParallel Conical | 4.3, 5.0 | RP |
| | 5.5 | WP |
| | 3.5 | NP |
| | 4.3 | RP |
| NobelReplace® Trilobe | 5.0 | WP |
| | 6.0 | 6.0 |
| | 3.3 | NP |
| Nobel Brånemark System® | 3.75, 4.0 | RP |
| | 5.0 | WP |
| | 3.5 | Mini |
| Osstem TS | 4.0, 4.5, 5.0, 6.0, 7.0 | Regular |
| | 3.5, 3.75, 4.0, 4.5 | RB |
| Straumann® BLX | 5.0, 5.5, 6.5 | WB |
| | 3.3 | NC |
| Straumann® Bone Level | 4.1, 4.8 | RC |
| | 3.3 | NNC |
| Straumann® Tissue Level | 3.3, 4.1, 4.8 | RN |
| | 4.8 | WN |
| | 3.3, 3.7, 4.1 | 3.5 |
| Zimmer Screw-Vent® / Tapered Screw-Vent® | 4.7 | 4.5 |
| | 6.0 | 5.7 |
{6}------------------------------------------------
# SUBJECT DEVICE DESCRIPTION
The purpose of this submission is to expand the DESS Dental Smart Solutions abutment system cleared under K22288 to add the ability for the subject device Base Abutments and Pre-milled (Blank) Abutments to be manufactured via a digital dentistry workflow by using scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories. There are no changes to the abutment design or implant compatibilities, however, there is a new material for the zircomia superstructure. All part numbers have been cleared within previous submissions for manufacturing at a validated milling center. The purpose of this submission is to allow manufacturing via digital dentistry workflow and to add a new zirconia material for the superstructures of the two-piece abutment.
The subject device DESS Dental Smart Solutions abutments provide a range of prosthetic solutions for dental implant restoration. DESS abutments are offered in a variety of connection types to enable compatibility with currently marketed dental implants. All abutments are provided non-sterile, and each abutment is supplied with the appropriate abutment screw (if applicable) for attachment to the corresponding implant.
Subject device Base Abutments are designed for fabrication of a patient-specific CAD/CAM zirconia superstructure on which a crown may be placed. They are two-piece abutments for which the second part (or top half) is the ceramic superstructure. They also may be used for support of a crown directly on the abutment.
All patient-specific custom abutment fabrication for Base Abutments and Pre-milled (Blank) Abutments is by prescription on the order of the clinician. The subject device Pre-milled (Blank) Abutments and all zirconia superstructures for use with the subject device Ti Base Interface, DESS Aurum Base, ELLIPTIBase, and DESS C-Base will be manufactured using a digital dentistry workflow by using scan files from intra-oral and lab (desktop) scanners, CAD software, ceramic material, milling machine and associated tooling and accessories.
The digital dentistry workflow uses scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories. The digital workflow includes the following products (not subject devices of this submission):
- . Ceramic material: VITA YZ ST and VITA YZ XT (K180703)
- Cement: Ivoclar Vivadent Multilink Hybrid Abutment Cement (K130436)
- Intraoral Scanner: 3Shape TRIOS A/S Series Intraoral Scanner (510(k) exempt under 21 CFR 872.3661)
- Desktop scanner: 3Shape D900 Dental Lab Scanner (510(k) exempt under 21 CFR 872.3661)
- Abutment design software: 3Shape Abutment Designer Software (K151455)
- . Milling machine: VHF R5 by vhf camfacture AG with DentalCAM and DentalCNC 7 software
The design parameters for the CAD/CAM zirconia superstructure to be used on Base Abutments: Minimum wall thickness – 0.4 mm Minimum post height for single-unit loading Ti Base Interface - 4.2 mm DESS Aurum Base – 4.0 mm ELLIPTIBase - 4.0 mm C-Base - 4.7 mm CoCr Base - 4.5 mm Minimum gingival height - 0.5 mm Maximum gingival height – 6.0 mm
All zirconia superstructures are for straight abutments only.
{7}------------------------------------------------
The design parameters for the CAD/CAM Pre-Milled Blank custom abutments: Minimum wall thickness - 0.4 mm Minimum post height - 4.0 mm Maximum gingival height - 6.0 mm Minimum gingival height - 0.3 mm All Pre-Milled Blank are for straight abutments only.
For the CAD/CAM Pre-milled Blanks that are compatible with Astra Tech EV (except for 3.0 mm implants), Astra Tech OsseoSpeed, Biomet 3i Certain, Nobel Active/NobelParallel Conical (except for 3.0 mm implants), NobelReplace Trilobe. Nobel Brånemark, Straumann Bone Level, and Zimmer Screw-Vent/Tapered Screw-Vent, the following design parameters may be used:
Minimum wall thickness - 0.45 mm Minimum post height - 4.0 mm Maximum gingival height - 6.0 mm Minimum gingival height - 0.3 mm Maximum angulation of the final abutment - 30°
# Screws
DESS Dental Smart Solutions screws are designed to attach the abutment to the implant or the prosthesis to the abutment. The screws were cleared in K170588, K173908, K191986, K203464, K212577, and K212628. There have been no changes to the screws and there are no new subject device screws. Screws are available with and without a DLC (Diamond-like Carbon) coating.
{8}------------------------------------------------
| Compatible Implant Systems | DESS Abutment System | Ti Base Interface, Engaging | Ti Base Interface, Non-engaging | DESS Aurum Base, Engaging | DESS Aurum Base, Non-engaging | ELLIPTIBase, Engaging | DESS C-Base | CoCr Base | Pre-milled Blank Ti, Engaging | Screws |
|----------------------------------------|------------------------|-----------------------------|---------------------------------|---------------------------|-------------------------------|-----------------------|--------------------|--------------------|-------------------------------|--------|
| Ankylos C/X | Internal Ank | 3.5, 4.5, 5.5 | 3.5, 4.5, 5.5 | | | | 3.5, 4.5, 5.5 | 3.5, 4.5, 5.5 | 3.5, 4.5, 5.5 | X |
| Astra Tech EV | Conic EVO | 3.0, 3.6, 4.2, 4.8, 5.4 | 3.0, 3.6, 4.2, 4.8, 5.4 | 3.6, 4.2, 4.8 | 3.6, 4.2, 4.8 | 3.0 | 3.6, 4.2, 4.8 | 3.6, 4.2, 4.8 | 3.0, 3.6, 4.2, 4.8, 5.4 | X |
| Astra Tech OsseoSpeed | Internal Hex Conic | 3.0, 3.5/4.0, 4.5/5.0 | 3.0, 3.5/4.0, 4.5/5.0 | 3.5/4.0, 4.5/5.0 | 3.5/4.0, 4.5/5.0 | | 3.5/4.0, 4.5/5.0 | | 3.0, 3.5/4.0, 4.5/5.0 | X |
| BioHorizons Internal | BH Internal | 3.0, 3.5, 4.5, 5.7 | 3.0, 3.5, 4.5, 5.7 | 3.5, 4.5, 5.7 | 3.5, 4.5, 5.7 | 3.0 | | 3.0, 3.5, 4.5, 5.7 | 3.5, 4.5, 5.7 | X |
| Biomet 3i Certain | Internal Hex Click | 3.4, 4.1, 5.0 | 3.4, 4.1, 5.0 | 3.4, 4.1, 5.0 | 3.4, 4.1, 5.0 | | 3.4, 4.1, 5.0 | 3.4, 4.1, 5.0 | 3.4, 4.1, 5.0 | X |
| Biomet 3i OSSEOTITE | External Hex USA | 3.4, 4.1, 5.0 | 3.4, 4.1, 5.0 | 3.4, 4.1, 5.0 | 3.4, 4.1, 5.0 | | 3.4, 4.1, 5.0 | | 3.4, 4.1, 5.0 | X |
| Camlog | Internal CAM | 3.3, 3.8, 4.3, 5.0 | 3.3, 3.8, 4.3, 5.0 | 3.8, 4.3 | 3.8, 4.3 | | | 3.8, 4.3 | 3.3, 3.8, 4.3, 5.0 | X |
| Frident XiVE | Internal Hex FD | 3.4, 3.8, 4.5, 5.5 | 3.4, 3.8, 4.5, 5.5 | 3.4, 3.8, 4.5 | 3.4, 3.8, 4.5 | | 3.4, 3.8, 4.5, 5.5 | 3.8, 4.5, 5.5 | 3.8, 4.5, 5.5 | X |
| Neodent Grand Morse | Neo GM | Grand Morse | Grand Morse | Grand Morse | Grand Morse | | | | Grand Morse | X |
| NobelActive, NobelParallel Conical | Active Hex | 3.0, NP, RP, WP | 3.0, NP, RP, WP | NP, RP | NP, RP | 3.0 | NP, RP, WP | NP, RP | 3.0, NP, RP, WP | X |
| NobelReplace Trilobe | Tri-lobe | NP, RP, WP, 6.0 | NP, RP, WP, 6.0 | NP, RP, WP | NP, RP, WP | | NP, RP, WP, 6.0 | | NP, RP, WP, 6.0 | X |
| Nobel Brånemark System | External Hex Universal | NP, RP, WP | NP, RP, WP | NP, RP | NP, RP | | NP, RP | | NP, RP, WP | X |
| Osstem TS | Conic OSS | Mini, Regular | Mini, Regular | Mini, Regular | Mini, Regular | | Mini, Regular | | Mini, Regular | X |
| Straumann BLX | Conical BLX | RB/WB, WB | RB/WB, WB | RB/WB, WB | RB/WB, WB | | | | RB/WB, WB | X |
| Straumann Bone Level | Conical BL | NC, RC | NC, RC | NC, RC | NC, RC | NC | NC, RC | NC, RC | NC, RC | X |
| Straumann Tissue Level | Octagon | NNC, RN, WN | NNC, RN, WN | RN, WN | RN, WN | NNC | RN, WN | RN, WN | NNC, RN, WN | X |
| Zimmer Screw Vent / Tapered Screw Vent | Internal Hex USA | 3.5, 4.5, 5.7 | 3.5, 4.5, 5.7 | 3.5, 4.5, 5.7 | 3.5, 4.5, 5.7 | 3.5 | 3.5, 4.5, 5.7 | 3.5, 4.5, 5.7 | 3.5, 4.5, 5.7 | X |
# Summary of Subject Device Components and Implant Platforms for Compatible Implant Systems
{9}------------------------------------------------
# MATERIAL COMPOSITION
All subject device abutments and screws are made of titanium alloy conforming to ASTM F136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401) or from Co-Cr-Mo alloy conforming to ASTM F1537 Standard Specification for Wrought Cobalt-28Chromium-6Molybdenum Alloys for Surgical Implants (UNS R31537, UNS R31538, and UNS R31539). These materials are the same as those used for devices cleared in K222288.
The new material recommended for zirconia superstructures on Ti Base Interface, DESS Aurum Base and ELLIPTIBase, C-Base, and CoCr Base is VITA YZ ST and VITA YZ XT, conforming to ISO 6872 Dentistry - Ceramic Materials and cleared in K180703. This material is equivalent to those cleared in K22288.
The cement recommended in labeling for bonding of superstructures is Multilink Hybrid Abutment Cement from Ivoclar Vivadent AG, cleared under K130436. This material is the same as that used for devices cleared in K222288.
All subject device components are manufactured from the same or similar materials, are treated with the same surface treatments (SelectGrip® surface, DLC coating and anodization), and are manufactured in the same or similar facilities using the same manufacturing processes cleared in K170588, K173908, K191986, K203464, K212577, and K212628. The facilities may be different in that the CAD/CAM portion of the abutment may be manufactured in dental laboratories; however, the equipment and procedures are intended to be the same as those used by the validated milling centers.
# PERFORMANCE DATA
Non-clinical data submitted to demonstrate substantial equivalence included the following: Sterilization validation according to ISO 17665-1, ISO 17665-2 and ISO 14937; biocompatibility testing according to ISO 10993-5 and ISO 10993-12 for titanium; fatigue testing and reverse engineering analysis of OEM implant bodies, OEM abutments, and OEM abutment screws to confirm compatibility; and MR Conditional labeling were leveraged from K222288. Biocompatibility testing according to ISO 10993-5 and ISO 10993-12 was included for the new zirconia material used for zirconia superstructures, tested alone and cemented to Ti Base abutments and for patient-specific abutments fabricated from Pre-milled Blanks using the new manufacturing (point of care) process. Software verification included testing of restrictions that prevent design of components outside of the stated design parameters. In addition, the abutment design library was validated to demonstrate that the established design limitations and specifications are locked and cannot be modified by the user. To address the potential risk of damage to the implant-abutment connection geometry during the milling of the patient-matched portions of the abutment blanks, validation testing of the CAM restriction zones was conducted, including verification to show avoidance of damage or modifications of the connection geometry, and locking of restriction zones from user editing in the CAM software. No clinical data were included in this submission.
# EQUIVALENCE TO MARKETED DEVICES
Subject device abutments are substantially equivalent in intended use to the primary predicate device K222288 and reference devices K201860, K180703, K130436, and K151455. All are intended for use with endosseous dental implants to provide functional and esthetic rehabilitation of the edentulous maxilla and mandible. The Indications for Use Statement (IFUS) for the subject device is substantially equivalent to that of the primary predicate device K222288 for compatible OEM implants and manufacturing using a validated milling center.
{10}------------------------------------------------
The Indications for Use Statement (IFUS) for the subject device is substantially equivalent to that of the reference device K201860 for manufacturing using a digital dentistry workflow. The digital dentistry workflow integrates multiple components: scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine, and associated tooling and accessories. The differences between the subject device IFUS and primary predicate device K222288 do not raise different questions of substantial equivalence, as demonstrated by manufacturing software validation and workflow validation.
All subject device abutments (base and blanks) and screws are equivalent in design, materials and technological characteristics to those of the primary predicate device K222288. There are no changes to the abutment designs or implant compatibilities. However, there is a new Y-TZP material for the zirconia superstructure. All part numbers have been cleared within previous submissions for manufacturing via a validated milling center. The purpose of this submission is to allow manufacturing via digital dentistry workflow.
# CONCLUSION
The subject device, the primary predicate device, the additional predicate devices, and reference device have the same intended use, have similar technological characteristics, and are made of the same or similar materials. The subject device, the primary predicate device and reference devices encompass the same range of physical dimensions, manufactured by similar methods, are packaged in similar materials, and are to be sterilized using similar methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
{11}------------------------------------------------
# Device Comparison Table
| Comparison | Subject Device | Predicate Devices | Reference Device |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | DESS Dental Smart Solutions<br>Terrats Medical SL | K222288<br>DESS Dental Smart Solutions<br>Terrats Medical SL | K201860<br>Elos Accurate® Hybrid Base™<br>Elos Medtech Pinol A/S |
| Product Code | NHA, PNP | NHA | NHA, PNP |
| Reason for<br>predicate/reference | n/a | Abutment design and OEM<br>compatibility | Digital dentistry workflow |
| Intended Use | Support of a prosthesis to restore<br>chewing function | Support of a prosthesis to restore<br>chewing function | Support of a prosthesis to restore<br>chewing function |
| Indications | DESS Dental Smart Solutions<br>abutments are intended for dental<br>prosthetic restorations. DESS<br>Dental Smart Solutions abutments<br>are used as an interface between a<br>dental implant or dental abutment<br>and a dental restoration and will be<br>attached to the implant or abutment<br>using a prosthetic screw and<br>attached to the dental restoration by<br>cementing.<br>All digitally designed custom<br>abutments for use with Base<br>abutment or Pre-milled Blank are<br>to be sent to a Terrats Medical<br>validated milling center for<br>manufacture, or to be designed and<br>manufactured according to the<br>digital dentistry workflow. The<br>digital dentistry workflow<br>integrates multiple components:<br>scan files from intra-oral and lab<br>(desktop) scanners, CAD software,<br>CAM software, ceramic material,<br>milling machine, and associated<br>tooling and accessories.<br>For complete Indications for Use<br>statement on OEM Compatibility<br>see Section 4. | DESS Dental Smart Solutions<br>abutments are intended to be used<br>in conjunction with endosseous<br>dental implants in the maxillary or<br>mandibular arch to provide support<br>for prosthetic restorations.<br>All digitally designed custom<br>abutments for use with DESS<br>Bases or Blanks are to be sent to a<br>Terrats Medical validated milling<br>center for manufacture.<br>For complete Indications for Use<br>statement on OEM Compatibility<br>see 510(k) Summaries for<br>K170588, K173908, K191986,<br>K203464, K212577, K212628<br>in Section 12. | The Elos Accurate® Hybrid<br>Base™ is intended for attaching to<br>dental implants in order to provide<br>basis for single or multiple tooth<br>prosthetic restorations. The Hybrid<br>Base™ is used as an interface<br>between a dental implant and a<br>zirconia superstructure and will be<br>attached to the implant using a<br>prosthetic screw and attached to the<br>zirconia superstructure by<br>cementing.<br>The Elos Accurate® Hybrid<br>Base™ is compatible with the<br>implant systems listed in Table 1:<br>The zirconia superstructures for use<br>with the Elos Accurate® Hybrid<br>Base™ are only intended to be<br>designed and manufactured<br>according to digital dentistry<br>workflow. The workflow system<br>integrates multiple components of<br>the digital dentistry workflow: scan<br>files from Intra-Oral Scanners,<br>CAD software, CAM software,<br>ceramic material, milling machine<br>and associated tooling and<br>accessories.<br>For complete Indications for Use<br>statement on OEM Compatibility<br>see 510(k) Summary for K151789<br>in Section 12. |
| Design | | | |
| Abutment Designs | CAD/CAM Bases, CAD/CAM<br>Blanks | Healing, Temporary, Straight,<br>Multi-unit, Locator-type,<br>CAD/CAM Bases, CAD/CAM<br>Blanks | CAD/CAM Bases |
| Prosthesis<br>Attachment | Cement-retained<br>Screw-retained | Cement-retained<br>Screw-retained | Screw Retained Abutment<br>Cement Retained Superstructure |
| Restoration | Single-unit,<br>Multi-unit | Single-unit,<br>Multi-unit | Single-unit,<br>Multi-unit |
| Abutment/Implant<br>Platform Ø, mm | 2.52 - 6.5 | 2.52 - 6.5 | 3.0-6.0 |
| Abutment Angle | 30° max* for Blanks<br>0° for TiBases | 30° max* for Blanks<br>30° max* for TiBases | 20° max for TiBases |
| Material | | | |
| Abutment Material | Ti-6Al-4V ELI | Ti-6Al-4V ELI<br>Co-Cr-Mo Alloy | n/a |
| Comparison | Subject Device | Predicate Devices | Reference Device |
| | DESS Dental Smart Solutions<br>Terrats Medical SL…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.