The Dentis s-Clean Abutment Mini is an endosseous dental implant that is indicated for surgical placement in the upper and lower jaw arches, to provide a root form means for single or multiple-units prosthetic appliance attachment to restore a patient's chewing function. Implants can be placed with a conventional two stage surgical process with an option for transmucosal healing or they can be placed in a single stage surgical process for immediate loading when good primary stability has been achieved and with appropriate occlusal loading.
Device Story
The Dentis s-Clean Abutment Mini is a dental implant abutment system used to connect prosthetic appliances to endosseous dental implants. The device is used by dental clinicians in a clinical setting. It acts as an interface between the implant fixture and the final prosthesis, supporting chewing function. The system includes various abutment types (O-Ring, TiN-coated Sub-Octa, and CCM UCLA) that are screw-retained to the implant fixture. The clinician selects the appropriate abutment based on the clinical requirement (e.g., overdenture retention or fixed restoration). The device is provided non-sterile for end-user sterilization. By providing a stable connection to the implant, the device enables the restoration of dental function for the patient.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including biocompatibility (ISO 10993-1), end-user sterilization validation (ANSI/AAMI ST79, ISO 17665, ISO 11737, ISO 11138), and MRI safety assessment based on scientific rationale and published literature.
Technological Characteristics
Materials: Ti-6Al-4V ELI (ASTM F136) and Cobalt-28Chromium-6Molybdenum Alloy (ASTM F1537). Features TiN coating on specific models. Connection: Screw-retained. Dimensions: Various diameters (4.0mm–4.8mm) and lengths. Sterilization: End-user sterilization required. No angled abutments; fatigue testing not required per FDA guidance.
Indications for Use
Indicated for patients requiring single or multiple-unit prosthetic restoration in the upper or lower jaw arches. Suitable for conventional two-stage surgical processes with transmucosal healing or single-stage processes with immediate loading, provided primary stability and appropriate occlusal loading are achieved.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Dentis Co., Ltd. % April Lee Consultant Withus Group Inc 106 Superior Irvine, California 92620
Re: K220440
Trade/Device Name: Dentis s-Clean Abutment Mini Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: May 14, 2022 Received: May 17, 2022
Dear April Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew I Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K220440
Device Name Dentis s-Clean Abutment Mini
#### Indications for Use (Describe)
The Dentis s-Clean Abutment Mini is an endosseous dental implant that is indicated for surgical placement in the upper and lower jaw arches, to provide a root form means for single or multiple-units prosthetic appliance attachment to restore a patient's chewing function. Implants can be placed with a conventional two stage surgical process with an option for transmucosal healing or they can be placed in a single stage surgical process for immediate loading when good primary stability has been achieved and with appropriate occlusal loading.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| <div> <span style="font-family: DejaVu Sans, sans-serif;">✓</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span style="font-family: DejaVu Sans, sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
#### Submitter
Dentis Co., Ltd. Gyu Ri Kim 99, Seongseoseo-ro, Dalseo-gu Daegu, 42718 Korea Email: kgr1026@dentis.co.kr Tel. +82-53-589-3541 Fax. +82-53-289-7922
## Official Correspondent
Withus Group Inc. April Lee 106 Superior, Irvine, CA 92620 USA Email: withus6664@gmail.com Phone: 1-909-274-9971 Fax: 1-909-460-8122
### Device Information
- . Trade Name: Dentis s-Clean Abutment Mini
- . Common Name: Endosseous dental implant abutment
- Classification Name: Endosseous dental implant abutment ●
- . Product Code: NHA
- Panel: Dental
- Regulation Number: 872.3630
- Device Class: Class II
- Date prepared: 06/15/2022 ●
### Predicate Devices:
Primary Predicate K171027, Dentis Dental Implant System by Dentis Co., Ltd
Reference Device K161244, s-Clean OneQ-SL Narrow Implant System by Dentis Co., Ltd. K210080, Dentis s-Clean s-Line Mini by Dentis Co., Ltd.
#### Indication for Use:
The Dentis s-Clean Abutment Mini is an endosseous dental implant that is indicated for surgical placement in the upper and lower jaw arches, to provide a root form means for single or multiple-units prosthetic appliance attachment to restore a patient's chewing function. Implants can be placed with a conventional two stage surgical process with an option for transmucosal healing or they can be placed in a single stage surgical process for immediate loading when good primary stability has been achieved and with appropriate occlusal loading.
#### Device Description:
Dentis s-Clean Abutment Mini is composed of Abutments are made of Ti-6AI-4V ELI according to ASTM F136 and Cobalt-28Chromium-6Molybdenum Alloy ASTM F1537. The s-Clean O-Ring Abutment s-Line and s-Clean TiN Half Coating Sub-Octa Abutment is screw assembled design but s-Clean CCM UCLA Abutment s-Line is screw separation design.
The s-Clean TiN Half Coating Sub-Octa Abutment is always to be combined with a casting abutment (e.g., Gold Cylinder) or burn-out plastic sleeve for single-unit loading. The minimum post height of s
{4}------------------------------------------------
Clean TiN Half Coating Sub-Octa Abutment when combined with casting abutment (e.g., Gold Cylinder) or burn-out plastic sleeve for single-unit loading is 4mm. The minimum post height of s-Clean CCM UCLA Abutment s-Line for single-unit loading is 4mm.
The Subject device is compatible with implants cleared in K210080. Compatible fixture diameters are Ø5.8. 6.8 and Ø7.8 and lengths are 7.5. 9.5, 11.4 and 11.5mm.
Compatible fixture, K210080, has dual screw-joint (structured with Upper and Lower screw joint) and a hex anti-rotation design connection. As explained in K210080, there are two-screw-joint features, 1.6M screw (Lower part) or 2.0M Screw (Upper Part) for diverse screw connection with other abutments. s-Clean O-Ring Abutment s-Line and s-Clean TiN Half Coating Sub-Octa Abutment have Upper M2.0 Screw and s-Clean CCM UCLA Abutment s-Line has Lower M1.6 screw. When dividing screw assemble design, screw assembled one-piece abutment is connected with upper screw and screw separated twopiece abutment is connected with lower screw.
| No. | Device Name | Diameter X Total Length | Angulation |
|-----|--------------------------------------------|----------------------------------------------------------------|------------|
| 1 | s-Clean O-Ring Abutment s-Line | $\varnothing$ 4.5 X 10.71 and 12.71mm | 0° |
| 2 | s-Clean TiN Half Coating Sub-Octa Abutment | $\varnothing$ 4.8 X 8.16, 8.66, 9.66, 10.66, 11.66 and 12.66mm | 0° |
| 3 | s-Clean CCM UCLA Abutment s-Line | $\varnothing$ 4.0 X 1, 3mm (Gingival Height) | 0° |
The dimensions of abutments are as following:
#### The Abutments have below featured:
| Name | Uses | Surface | Connection |
|-----------------------------------------------|-----------------------------------------------------------------------------------------------------|-------------|--------------|
| s-Clean O-Ring Abutment<br>s-Line | It is used to retain an overdenture prosthetics. | N/A | Screw retain |
| s-Clean TiN Half Coating<br>Sub-Octa Abutment | This Abutment is connected with fixture and it<br>supports prosthesis which restore tooth function. | TiN Coating | Screw retain |
| s-Clean CCM UCLA Abutment<br>s-Line | This Abutment is connected with fixture and it<br>supports prosthesis which restore tooth function. | N/A | Screw retain |
### Materials:
- . s-Clean O-Ring Abutment s-Line and s-Clean TiN Half Coating Sub-Octa Abutment is fabricated from Ti-6A1-4V ELI (Conforming to ASTM Standard F136).
- S-Clean CCM UCLA Abutment s-Line is fabricated from Cobalt-28Chromium-6Molybdenum . Alloy (Conforming to ASTM F1537)
- CCM UCLA s-Line Abutment Screw is fabricated from Ti-6A1-4V ELI (Conforming to ASTM Standard F136).
{5}------------------------------------------------
## Summaries of Technology Characteristics
The subject device is substantially equivalent to the current cleared devices. They are substantially equivalent in intended use, material and product spec as diameter and length. Comparison demonstrating Substantial Equivalence follows:
## 1. s-Clean O-Ring Abutment s-Line
| | Subject Device | Primary Predicate |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant | Dentis Co., Ltd. | Dentis Co., Ltd. |
| Trade Name | Dentis s-Clean Abutment Mini | Dentis Dental Implant System |
| 510(k) No. | | K171027 |
| Classification<br>Name | Endosseous Dental Implant | Endosseous Dental Implant |
| Product Code | NHA | DZE, NHA |
| Description | Image: Dentis s-Clean Abutment Mini | Image: Dentis Dental Implant System |
| Material | Ti-6Al-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM F136) |
| Diameter (mm) | $\text{\O}4.5$ | $\text{\O}3.4$ and $\text{\O}4.5$ |
| Length (mm) | 10.71 and 12.71 | 10.1, 11.6 and 13.6 |
| Sterile | End User Sterilization | End User Sterilization |
| Indications For<br>Use | The Dentis s-Clean Abutment Mini is an endosseous<br>dental implant that is indicated for surgical<br>placement in the upper and lower jaw arches, to<br>provide a root form means for single or multiple-<br>units prosthetic appliance attachment to restore a<br>patient's chewing function. Implants can be placed<br>with a conventional two stage surgical process with<br>an option for transmucosal healing or they can be<br>placed in a single stage surgical process for<br>immediate loading when good primary stability has<br>been achieved and with appropriate occlusal loading. | The Dentis Dental Implant System is an<br>endosseous dental implant that is indicated for<br>surgical placement in the upper and lower jaw<br>arches, to provide a root form means for single or<br>multiple-units prosthetic appliance attachment to<br>restore a patient's chewing function. Implants<br>can be placed with a conventional two stage<br>surgical process with an option for transmucosal<br>healing or they can be placed in a single stage<br>surgical process for immediate loading when<br>good primary stability has been achieved and<br>with appropriate occlusal loading. |
| Substantial<br>Equivalence<br>Comparison | The subject device has same indications for Use, fundamental scientific technology, principle of<br>operation, technology, functions, and materials and has similar design with the primary predicate.<br>The difference between the subject and primary predicate is the Dimensions of the device and the<br>compatible implant body. The subject device's diameter( $\text{\O}2.88$ ) with length (9.11mm) is smaller size<br>than the primary predicate but doesn't affect device's fundamental functions and safety; therefore, it is<br>substantial equivalent. | |
{6}------------------------------------------------
| | Subject Device | Primary Predicate | | Subject Device | Primary Predicate Device | Reference Device |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant | Dentis Co., Ltd. | Dentis Co., Ltd. | Applicant | Dentis Co., Ltd. | Dentis Co., Ltd. | Dentis Co., Ltd. |
| Trade Name | Dentis s-Clean Abutment Mini | Dentis Dental Implant System | Trade Name | Dentis s-Clean Abutment Mini | Dentis Dental Implant System | s-Clean OneQ-SL Narrow Implant<br>System |
| 510(k) No. | - | K171027 | 510(k) No. | - | K171027 | K161244 |
| Classification<br>Name | Endosseous Dental Implant | Endosseous Dental Implant | Classification<br>Name | Endosseous Dental Implant | Endosseous Dental Implant | Endosseous Dental Implant |
| Product Code | NHA | DZE, NHA | Product Code | NHA | DZE, NHA | DZE, NHA |
| Description | Image: Dentis s-Clean Abutment Mini | Image: Dentis Dental Implant System | Class | Class II | Class II | Class II |
| Material | Ti-6Al-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM F136) | Description | | | |
| Diameter (mm) | Ø4.8 | Ø4.8 | Material | Cobalt-28Chromium-6Molybdenum<br>Alloy<br>(ASTM F1537) | Cobalt-28Chromium-<br>6Molybdenum Alloy<br>(ASTM F1537) | Cobalt-28Chromium-6Molybdenum<br>Alloy<br>(ASTM F1537) |
| Length (mm) | 8.16, 8.66, 9.66, 10.66, 11.66 and 12.66 | 9.15, 9.65, 10.65, 11.65, 12.65 and 13.65 | Diameter (mm) | $Ø4.0$ | $Ø4.5$ | $Ø4.0$ |
| Surface Treatment | TiN Coating | TiN Coating | Gingival Height | 1mm and 3mm | 1mm and 3mm | 1mm |
| Sterile | End User Sterilization | End User Sterilization | Sterile | End User Sterilization | End User Sterilization | End User Sterilization |
| Indications For<br>Use | The Dentis s-Clean Abutment Mini is an endosseous<br>dental implant that is indicated for surgical placement<br>in the upper and lower jaw arches, to provide a root<br>form means for single or multiple-units prosthetic<br>appliance attachment to restore a patient's chewing<br>function. Implants can be placed with a conventional<br>two stage surgical process with an option for<br>transmucosal healing or they can be placed in a single<br>stage surgical process for immediate loading when<br>good primary stability has been achieved and with<br>appropriate occlusal loading. | The Dentis Dental Implant System is an<br>endosseous dental implant that is indicated for<br>surgical placement in the upper and lower jaw<br>arches, to provide a root form means for single or<br>multiple-units prosthetic appliance attachment to<br>restore a patient's chewing function. Implants can<br>be placed with a conventional two stage surgical<br>process with an option for transmucosal healing or<br>they can be placed in a single stage surgical<br>process for immediate loading when good primary<br>stability has been achieved and with appropriate<br>occlusal loading. | Indications For<br>Use/<br>Intended Use | The Dentis s-Clean Abutment Mini<br>is an endosseous dental implant that<br>is indicated for surgical placement<br>in the upper and lower jaw arches,<br>to provide a root form means for<br>single or multiple-units prosthetic<br>appliance attachment to restore a<br>patient's chewing function.<br>Implants can be placed with a<br>conventional two stage surgical<br>process with an option for<br>transmucosal healing or they can be<br>placed in a single stage surgical<br>process for immediate loading when<br>good primary stability has been<br>achieved and with appropriate<br>occlusal loading. | The Dentis Dental Implant<br>System is an endosseous dental<br>implant that is indicated for<br>surgical placement in the upper<br>and lower jaw arches, to provide<br>a root form means for single or<br>multiple-units prosthetic<br>appliance attachment to restore<br>a patient's chewing function.<br>Implants can be placed with a<br>conventional two stage surgical<br>process with an option for<br>transmucosal healing or they<br>can be placed in a single stage<br>surgical process for immediate<br>loading when good primary<br>stability has been achieved and<br>with appropriate occlusal<br>loading. | The s-Clean OneQ-SL Narrow<br>Implant System is indicated for<br>use in partially or fully edentulous<br>mandibles and maxillae, in support<br>of single or multiple-unit<br>restorations including; cemented,<br>retained, screw retained, or<br>overdenture restorations, and<br>terminal or intermediate abutment<br>support for fixed bridgework. This<br>system is dedicated for one and<br>two stage surgical procedures and<br>not dedicated for immediate<br>loading. This system is intended<br>for delayed loading. Only, s-Clean<br>OneQ-SL Narrow fixture (3.0, 3.3,<br>3.7mm) is limited to replacement<br>of maxillary lateral incisors and<br>mandibular incisors. |
| Substantial<br>Equivalence<br>Comparison | The subject device has same indications for Use, fundamental scientific technology, principle of operation,<br>technology, functions, and materials and has similar design with the primary predicate.<br>The difference between the subject and primary predicate is the lengths of the device and the compatible<br>implant body. The subject device has shorter size than the primary predicate but doesn't affect device's<br>fundamental functions and safety; therefore, it is substantial equivalent. | | Substantial<br>Equivalence<br>Comparison | The subject device has same indications for Use, fundamental scientific technology, principle of operation,<br>technology, functions, and materials and has similar design with the primary predicate.<br>The difference between the subject and primary predicate is the diameter of the device and the compatible<br>implant body. To support this discrepancy, K161244 was added; therefore, it is substantial equivalent. | | |
# 2. s-Clean TiN Half Coating Sub-Octa Abutment
{7}------------------------------------------------
# 3. s-Clean CCM UCLA Abutment s-Line
{8}------------------------------------------------
## Non-Clinical Testing
Below tests were performed for predicate devices and leveraged for the subject device:
- Biocompatibility testing according to ISO 10993-1:2009 for Ti-6A1-4V material abutments . referenced in K210080, for Cobalt-28Chrominum-6Molybdenum material abutment referenced in K161244 and for Ti-6Al-4V with TiN Coating abutments referenced in K171027
- End User Sterilization Validation Test Report according to ANSI/AAMI ST79, ISO 17665-1, ● ISO 17665-2. ISO 11737-1, ISO 11737-2, and ISO 11138-1 referenced in K111364 for Ti-6Al-4V material abutments non-coating and TiN-Coating.
The results of the above tests have met the criteria of the standards and demonstrated the substantial equivalence with the predicate device.
Non-clinical test data was conducted in accordance with FDA Guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments".
Biocompatibility testing for Ti-6AI-4V wth TiN Coating material abutments was conducted on the predicate device, K171027 and can be leveraged for the subject device because both devices have same material and manufacturing process which is substantially equivalent.
Fatigue testing was not performed for the subject device according to the ISO 14801:2016 because this subject device system does not include any angled abutments. Based on the FDA Guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments", the fatigue testing is not necessary if the system does not consist of angled abutments. Also, the surface modification for the subject device was not tested or provided because the surface modification information was provided in the predicate devices and can be leveraged for the subject device.
End User Sterilization Validation Test Report for Cobalt-27Chrominum-6Molybdenum was conducted on the device that is not cleared by FDA. The test article is the largest surface area among the devices that have the same manufacturing process and material.
Clinical testing was not necessary to establish substantial equivalency of the device.
- MR Environment Condition -
Non-clinical worst-case MRI review was performed to evaluate the Dentis s-Clean Abutment Mini in the MRI environment using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan. "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795), based on the entire system including all variations (all compatible implant bodies, dental abutments, and fixation screws) and material composition. Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque
{9}------------------------------------------------
### Conclusion
Dentis s-Clean Abutment Mini constitutes a substantially equivalent medical device, meeting all the declared requirements of its intended use. This system has the same intended use and fundamental scientific technology as its predicate de…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.