K201681 · Institut Straumann AG · NHA · Mar 18, 2021 · Dental
Device Facts
Record ID
K201681
Device Name
Straumann Immediate Temporary Abutments
Applicant
Institut Straumann AG
Product Code
NHA · Dental
Decision Date
Mar 18, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Straumann Immediate Temporary Abutment and associated Plastic Coping are indicated to be placed into Straumann dental implants (BL, BLT or BLX) to provide a support structure for a temporary aesthetic oral rehabilitation of partially edentulous patients with crowns. Temporary components can be used prior to the final components to maintain, stabilize and shape the soft tissue during the healing phase for up to 180 days. The Straumann Immediate Temporary Abutment and associated Plastic Coping must be placed out of occlusion.
Device Story
Device consists of titanium (Ti-6Al-7Nb) abutment and PMMA plastic coping; used with Straumann dental implants (BL, BLT, BLX). Abutment is one-piece design, screwed directly into implant; provides support structure for temporary crowns. Used in dental clinics by clinicians to maintain/shape soft tissue during healing phase (up to 180 days). Must be placed out of occlusion. Abutments provided sterile (gamma irradiation); copings provided non-sterile, requiring manual cleaning/disinfection by user. Output is a stable temporary prosthetic base; aids clinical decision-making by facilitating soft tissue management prior to final restoration.
Clinical Evidence
Bench testing only. Biocompatibility assessed per ISO 10993-1, ISO 10993-5 (cytotoxicity), and ISO 10993-18 (chemical characterization). Sterilization validation performed per ISO 11137 for abutments; high-level disinfectant validation performed for PMMA copings. MRI compatibility established via reference to K190662.
Indicated for partially edentulous patients requiring temporary aesthetic oral rehabilitation with crowns using Straumann dental implants (BL, BLT, or BLX). Components maintain, stabilize, and shape soft tissue during healing for up to 180 days. Contraindicated for use in occlusion.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 18, 2021
Institut Straumann AG % Chanrasmey White Regulatory Affairs Specialist Straumann USA, LLC 60 Minuteman Road Andover, Massachusetts 01801
Re: K201681
Trade/Device Name: Straumann Immediate Temporary Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: February 12, 2021 Received: February 16, 2021
Dear Chanrasmey White:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K201681
Device Name Straumann® Immediate Temporary Abutments
#### Indications for Use (Describe)
The Straumann Immediate Temporary Abutment and associated Plastic Coping are indicated to be placed into Straumann dental implants (BL, BLT or BLX) to provide a support structure for a temporary aesthetic oral rehabilitation of partially edentulous patients with crowns. Temporary components can be used prior to the final components to maintain, stabilize and shape the soft tissue during the healing phase for up to 180 days.
The Straumann Immediate Temporary Abutment and associated Plastic Coping must be placed out of occlusion.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Straumann® Immediate Temporary Abutments
510(k) Summary
#### K201681 510(k) Summary 5
#### Submitter 5.1
60 Minuteman Road
Andover, MA 01810
978-747-0023 Fax Number:
- Contact Person: Chanrasmey White
March 17, 2021 Date of Prepared:
#### 5.2 Device
| Trade Name: | Straumann® Immediate Temporary Abutments |
|----------------------|------------------------------------------|
| Common Name: | Endosseous Dental Implant Abutments |
| Classification Name: | Endosseous Dental Implant Abutments |
| Regulatory Class: | II (21 CFR 872.3630) |
| Product Code: | NHA (21 CFR 872.3630) |
#### 5.3 Predicate Device
Primary Predicate:
- K122192 Straumann Temporary Abutments, PMMA
Reference Predicate:
- K192401 Straumann Screw Retained Abutments
- K181703 BLX Implant System
- K190662 MRI Compatibility for Existing Straumann Dental Implant System
- K041070 Straumann Temporary Coping
{4}------------------------------------------------
### Straumann® Immediate Temporary Abutments
510(k) Summary
#### Device Description 5.4
The Straumann Immediate Temporary Abutment and associated Plastic Copinq consist of a titanium abutment and PMMA coping for the Straumann dental implants BL, BLT and BLX. The abutments are available in a variety of gingival heights and diameters to fit individual patient situations. The PMMA coping is compatible with the subject Immediate Temporary Abutments. The Straumann Immediate Temporary Abutment and associated Plastic Coping are indicated to provide a support structure for a temporary aesthetic oral rehabilitation of partially edentulous patients with crowns. Temporary components can be used prior to the insertion of the final components to maintain, stabilize and shape the soft tissue during the healing phase for up to 180 days.
#### 5.5 Indications for Use
The Straumann Immediate Temporary Abutment and associated Plastic Coping are indicated to be placed into Straumann dental implants (BL, BLT or BLX) to provide a support structure for a temporary aesthetic oral rehabilitation of partially edentulous patients with crowns. Temporary components can be used prior to the insertion of the final components to maintain, stabilize and shape the soft tissue during the healing phase for up to 180 days.
The Straumann Immediate Temporary Abutment and associated Plastic Coping must be placed out of occlusion.
### 5.6 Technological Characteristics
The subject device is equivalent to the primary predicate and reference devices sharing the same material and fundamental operating principles. The differences in indications for use are addressed in performance testing provided in this submission demonstrating equivalent product performance. The subject and the primary predicate devices are intended to support temporary prosthetic reconstructions. The subject devices have a contact duration of up to 180 days compared to the primary predicate and reference devices, biocompatibility testing has been performed to support the 180 duration of use. The technological characteristics of the subject device are compared to the primary predicate and reference devices in Table 2. The subject devices are one-piece abutments opposed to the two-piece temporary abutments cleared per K122192. The one-piece design allows the abutment to be directly screwed into the
{5}------------------------------------------------
## Straumann® Immediate Temporary Abutments
510(k) Summary
implant. The differences in technological differences are addressed in the performance testing provided in the submission which demonstrates equivalency in product performance.
{6}------------------------------------------------
## Straumann® Immediate Temporary Abutments
510(k) Summary
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | REFERENCE DEVICE |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | Subject Device | K122192 | K192401 | K181703 |
| Indications for Use | The Straumann Immediate<br>Temporary Abutment and<br>associated Plastic Coping are<br>indicated to be<br>placed into Straumann dental<br>implants (BL, BLT or BLX) to<br>provide a support structure for a<br>temporary<br>aesthetic oral rehabilitation of<br>partially edentulous patients with<br>crowns. Temporary components<br>can be used prior to the insertion<br>of the final components to<br>maintain, stabilize and shape the<br>soft tissue during the healing<br>phase for up to 180 days.<br>The Straumann Immediate<br>Temporary Abutment and<br>associated Plastic Coping must be<br>placed out of occlusion. | Straumann Temporary Abutments<br>VITA CAD-Temp are indicated for<br>use with Straumann Bone Level<br>and Tissue Level implants for<br>temporary crown and bridge<br>restorations, and to maintain,<br>stabilize and shape the soft tissue<br>during the healing phase for up to<br>six months, and should be placed<br>out of occlusion. | Prosthetic components directly or<br>indirectly connected to the<br>endosseous dental implant are<br>intended for use as an aid in<br>prosthetic rehabilitations.<br>Temporary components can be<br>used prior to the insertion of the<br>final components to maintain,<br>stabilize and shape the soft tissue<br>during the healing phase; they may<br>not be placed into occlusion. Final<br>abutments may be placed into<br>occlusion when the implant is fully<br>osseointegrated.<br>Temporary Abutments have a<br>maximum duration of usage of 180<br>days. | Straumann® BLX Implants<br>Straumann® BLX Implants are<br>suitable for endosteal implantation<br>in the upper and lower jaw and for<br>the functional and esthetic oral<br>rehabilitation of edentulous and<br>partially edentulous patients. BLX<br>Implants can be placed with<br>immediate function on single-tooth.<br>bar and bridge applications when<br>qood primary stability is achieved<br>and with appropriate occlusal<br>loading to restore chewing<br>function. The prosthetic<br>restorations are connected to the<br>implants through the<br>corresponding abutment<br>components.<br>Straumann® BLX SRAs and<br>Anatomic Abutments<br>Prosthetic components directly or<br>indirectly connected to the<br>endosseous dental implant are<br>intended for use as an aid in<br>prosthetic rehabilitations.<br>Temporary components can be<br>used prior to the insertion of the<br>final components to maintain,<br>stabilize and shape the soft tissue<br>during the healing phase; they may<br>not be placed into occlusion. Final<br>abutments may be placed into<br>occlusion when the implant is fully<br>osseointegrated. BLX Temporary<br>Abutments have a maximum<br>duration of usage of 180 days. |
| Material | Abutment:<br>Ti-6Al-7Nb<br>Plastic Coping:<br>Polymethyl Methacrylate (PMMA) | Ti-6Al-7Nb<br>Polymethyl Methacrylate (PMMA) | Ti-6Al-7Nb | Ti-6Al-7Nb |
| Implant Compatibility | BL, BLT, BLX | BL, BLT, TL | BL, BLT | BLX |
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | REFERENCE DEVICE |
| K Number | Subject Device | K122192 | K192401 | K181703 |
| Abutment Diameter | Ø3.8 and 4.5 | Ø5.0, 7.0, and 10.0 mm | Ø3.5 and 4.6 mm | Ø4.6 mm |
| Gingival Heights | BL and BLT - 1, 2, and 3 mm<br>BLX - 1.5, 2.5, and 3.5 mm | 1.0, 2.5, 4.0, and 5.5 mm | 1.5, 2.5, 3.5, 4.5, and 5.5 mm | 1.5, 2.5, 3.5, and 4.5 mm |
| Abutment Angulation | 0° | 0° | 0, 17 and 30° | 0, 17 and 30° |
| Indexing Type/<br>Presence | Non-Engaging | Engaging | Straight - Non-Engaging | Straight - Non-Engaging |
| Integrated Screw<br>Technology | Yes<br>Machined one-piece | No<br>Coupled with a base screw | Yes - The straight abutments are<br>machined one-piece | Yes - The straight abutments are<br>machined one-piece |
| Restoration Type | Cemented | Cemented | Screw-retained | Screw-retained |
| Duration of Use | 180 days | 180 days | unlimited | unlimited |
| Sterilization | Sterile - Gamma Irradiation | Non-Sterile<br>End User - Steam Autoclave | Sterile - Gamma Irradiation | Sterile - Gamma Irradiation |
| Loading | Single Unit | Single and Multi Unit | Single and Multi Unit | Single and Multi Unit |
{7}------------------------------------------------
## Straumann® Immediate Temporary Abutments
### 510(k) Summary
Table 1 – Immediate Temporary Abutment – Comparison Matrix
{8}------------------------------------------------
## Straumann® Immediate Temporary Abutments
### 510(k) Summary
| FEATURE | PROPOSED DEVICE | REFERENCE DEVICE |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| K Number | Subject Device | K041070 |
| Indications for Use | The Straumann Immediate Temporary<br>Abutment and associated Plastic Coping are<br>indicated to be placed into Straumann dental<br>implants (BL, BLT or BLX) to provide a support<br>structure for a temporary aesthetic oral<br>rehabilitation of partially edentulous patients<br>with crowns. Temporary components can be<br>used prior to the insertion of the final<br>components to maintain, stabilize and shape<br>the soft tissue during the healing phase for up<br>to 180 days.<br>The Straumann Immediate Temporary<br>Abutment and associated Plastic Coping must<br>be placed out of occlusion. | Temporary Copings are intended to serve as a<br>base for temporary restorations. |
| Material | Polymethyl methacrylate (PMMA) | Polymethyl methacrylate (PMMA) |
| Restoration Type | Cemented | Cemented |
| Duration of Use | 180 days | 30 days |
| Sterilization | Non-Sterile<br>Cleaned and disinfected | Non-Sterile |
Table 2 – Plastic Coping – Comparison Matrix
{9}------------------------------------------------
### Straumann® Immediate Temporary Abutments
510(k) Summary
#### Performance Data 5.7 -
The following performance data were provided in support of the substantial equivalence determination.
#### Biocompatibility Testing
The subject devices have been assessed for biological safety according to ISO 10993-1:2009 "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" and to the FDA Guidance document "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process', Guidance for Industry and Food and Drug Administration Staff, Document issued on June 16, 2016" for the subject device. The subject abutments are equivalent with regard to nature of body contact, duration of use, material formulation and sterilization methods compared to the primary and reference predicate devices and therefore, no new testing have been performed as biocompatibility is represented in the primary and reference predicate devices.
In support of the subject PMMA plastic copings, Cytotoxicity testing according to ISO 10993-5:2009, "Biological evaluation of medical devices – Part 5: Test for in vitro cytotoxicity" and Chemical Analysis according to ISO 10993-18: 2020 "Chemical characterization of medical device materials within a risk management process" has been performed. In addition to the above testing a toxicology evaluation has been performed in support of the subject PMMA plastic copings. The test and evaluation performed the biocompatibility of the subject PMMA plastic copings.
#### Sterilization
The subject device are single patient devices. The Immediate Temporary Abutments will be delivered in a sterile packaging via Gamma Irradiation and the PMMA plastic coping will be delivered non-sterile. The PMMA plastic coping is to be manually cleaned and disinfected according to the Instructions for Use.
A sterilization validation assessment was performed according to ISO 11137 for the Immediate Temporary Abutments. The assessment concluded the proposed Immediate Temporary
{10}------------------------------------------------
## Straumann® Immediate Temporary Abutments
### 510(k) Summary
Abutment can be adopted into the same sterilization process validated for the Straumann Sterile Prosthetics cleared for reference predicate K192401.
A high-level disinfectant validation was performed on the worst case PMMA device. The worst case was selected based on the most challenging designs for high level disinfection process due to size, surface, complexity and accessibility of the liquid solution. The subject PMMA Plastic Coping is not considered a new worst case and therefore can adopt the validation demonstrating the high-level disinfectant cleaning method can effectively disinfect the Plastic Coping.
The packaging of the Straumann® Immediate Temporary Abutments is identical to the packaging of the reference device K192401. There are no changes to the sterilization method or production process compared to the reference device. The shelf life for the proposed devices is 5 years identical to the reference device.
#### Bench Testing
The subject devices are MRI Conditional and do not introduce a new material or worst-case scenario and is represented under reference device K190662.
#### Conclusion 5.8
The documentation submitted in this premarket notification demonstrates the subject devices are substantially equivalent to the primary predicate and reference devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.