Sirona Dental CAD/CAM System with CEREC Chairside Software
Applicant
Dentsply Sirona
Product Code
NHA · Dental
Decision Date
Aug 16, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
The Sirona Dental CAD/CAM System with CEREC Chairside Software is intended for use in partially or fully edentulous mandibles and maxillae in support of single or multiple-unit cetained restorations. For the AT TX 3.0 S, BH 3.0 S, SSO 3.5 L, and SBL 3.3 L titanium bases, the indication is restricted to the replacement of single lateral incisors in the maxilla and lateral and central incisors in the mandible. The system consists of three major parts: TiBase, inCoris mesostructure and CAD/CAM software. Specifically, the inCoris mesostructure and TiBase components make up a two- piece abutment which is used in conjunction with endosseous dental implants to restore the function and aesthetics in the oral cavity. The inCoris mesostructure may also be used in conjunction with the Camlog Titanium base CAD/CAM (types K2244.XXXX) (K083496) in the Camlog Implant System. The CAD/CAM software is intended to design and fabricate the inCoris mesostructure. The inCoris mesostructure and TiBase two-piece abutment is compatible with the following implant systems:
Device Story
System captures 3D digital topography of teeth/impressions via intraoral scanner (CEREC Omnicam) and acquisition units (CEREC AC/AC Connect); software (CEREC Chairside) designs patient-specific two-piece abutments (TiBase + inCoris ZI mesostructure). Milling unit (CEREC MCXL) fabricates mesostructure from zirconium blocks. Used in dental clinics by clinicians to restore function/aesthetics in edentulous patients. Output is a custom abutment for endosseous implants. Benefits include chairside fabrication of restorative components.
Clinical Evidence
No clinical data included. Substantial equivalence supported by bench testing: geometric compatibility verification, dynamic fatigue testing per ISO 14801, and validated steam sterilization parameters per ISO 17665-1 and ANSI/AAMI ST79.
Technological Characteristics
System components: TiBase (titanium alloy), inCoris ZI (zirconium oxide ceramic), CEREC AC/Omnicam scanner, CEREC MCXL milling unit. Connectivity: Standalone/networked CAD/CAM workflow. Sterilization: Moist heat (steam) per ISO 17665-1. Software: CAD/CAM design and fabrication control.
Indications for Use
Indicated for partially or fully edentulous patients requiring single or multiple-unit cement-retained restorations. Specific TiBase variants (AT TX 3.0 S, BH 3.0 S, SSO 3.5 L, SBL 3.3 L) restricted to single lateral incisors (maxilla) and lateral/central incisors (mandible).
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 16, 2020
Dentsply Sirona Karl Nittinger Vice President, Corporate Regulatory Affairs 221 West Philadelphia St., Suite 60W York, Pennsylvania 17401
Re: K193408
Trade/Device Name: Sirona Dental CAD/CAM System with CEREC Chairside Software Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA, PNP Dated: July 16, 2020 Received: July 17, 2020
Dear Karl Nittinger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Srinivas Nandkumar, Ph.D. for Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K193408
Device Name
Sirona Dental CAD/CAM System with CEREC Chairside Software
Indications for Use (Describe)
The Sirona Dental CAD/CAM System with CEREC Chairside Software is intended for use in partially or fully edentulous mandibles and maxillae in support of single or multiple-unit cetained restorations. For the AT TX 3.0 S, BH 3.0 S, SSO 3.5 L, and SBL 3.3 L titanium bases, the indication is restricted to the replacement of single lateral incisors in the maxilla and lateral and central incisors in the mandible. The system consists of three major parts: TiBase, inCoris mesostructure and CAD/CAM software. Specifically, the inCoris mesostructure and TiBase components make up a two- piece abutment which is used in conjunction with endosseous dental implants to restore the function and aesthetics in the oral cavity. The inCoris mesostructure may also be used in conjunction with the Camlog Titanium base CAD/CAM (types K2244.XXXX) (K083496) in the Camlog Implant System. The CAD/CAM software is intended to design and fabricate the inCoris mesostructure. The inCoris mesostructure and TiBase two-piece abutment is compatible with the following implant systems:
| Manufacturer | Name of Implant System | Implant Size | |
|--------------------------------------|----------------------------------------------------------|-----------------|---------------------|
| | | Platform | Diameter |
| Nobel Biocare | Replace | NP | 3.5 |
| | | RP | 4.3 |
| | | WP | 5.0 |
| | | 6.0 | 6.0 |
| Nobel Biocare | Active | NP | 3.5 |
| | | RP | 4.3/5.0 |
| | Branemark | NP | 3.3 |
| | | RP | 3.75/4.0 |
| Straumann | Synocta | NN (3.5mm) | 3.3 |
| | | RN (4.8mm) | 3.3/4.1/4.8 |
| | | WN (6.5mm) | 4.8 |
| | Bone Level | NC (3.3mm) | 3.3 |
| | | RC (4.1mm/4.8mm | 4.1/4.8 |
| Dentsply Sirona Implants | Osseospeed | 3.5/4.0 | 3.5 S / 4.0 S |
| | | 4.5/5.0 | 4.5/5.0/5.0 S |
| | Xive | 3.4 | 3.4 |
| | | 3.8 | 3.8 |
| | | 4.5 | 4.5 |
| | | 5.5 | 5.5 |
| Manufacturer | Name of Implant System | Implant Size | |
| | | Platform | Diameter |
| | Osseospeed EV | 3.6 | 3.6 |
| | | 4.2 | 4.2 |
| | | 4.8 | 4.8 |
| Dentsply Sirona Implants | | 5.4 | 5.4 |
| | Ankylos | C/X | A, B, C, D |
| | Osseospeed TX | 3.0 | 3.0 |
| | | 3.5/4.0 | 3.5/4.0 |
| | | 4.5/5.0 | 4.5/5.0 |
| | Osseotite | 3.4 | 3.25 |
| | | | 3.75 |
| | | 4.1 | 4.1 |
| | | | 3/4 |
| | | 5.0 | 5.0 |
| | | | 4/5 |
| | Certain | 3.4 | 3.25 |
| Biomet 3i | | | 4/3 |
| | | | 3/4/3 |
| | | 4.1 | 4.0 |
| | | | 4/5/4 |
| | | | 5/4 |
| | | 5.0 | 5.0 |
| | | | 4/5 |
| | Tapered Screw-Vent | 3.5 | 3.7/4.1 |
| Zimmer | | 4.5 | 4.7 |
| | | 5.7 | 6 |
| | Thommen Medical Implants | 3.5 | 3.5 |
| Thommen Medical | | 4 | 4 |
| | | 4.5 | 4.5 |
| | | 5 | 5 |
| | | 6 | 6 |
| | Osstem TS Implant System | Mini | 3.5 |
| Osstem / Hiossen | Hiossen Implant System | Regular | 4.0/4.5/5.0/6.0/7.0 |
| Manufacturer | Name of Implant System | Implant Size | |
| | | Platform | Diameter |
| | Tapered 3.0,Tapered plus | 3.0 | 3.0/3.4/3.8 |
| | Tapered internal | 3.0 | 3.0 |
| | Tapered plus | | 4.6 |
| | Tapered internal, Tapered<br>internal tissue level | 3.5 | 3.0/3.8 |
| | Internal dental implant | | 3.5 |
| | Single stage dental implants | | 3.5/4.0 |
| BioHorizons<br>(Internal Connection) | Tapered Plus | | 5.8 |
| | Tapered internal, Tapered<br>internal tissue level | 4.5 | 4.6 |
| | Internal dental implant | | 4.0 |
| | Single stage dental implants | | 4.0/5.0 |
| | Tapered internal, Tapered<br>internal tissue level | 5.7 | 5.8 |
| | Internal dental implant,<br>Single stage dental implants | | 5.0/6.0 |
{3}------------------------------------------------
{4}------------------------------------------------
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 3of 3
PSC Publishing Services (301) 443-6740 ER
{5}------------------------------------------------
Dentsply Sirona 221 West Philadelphia Street Suite 60W York, PA 17401
Image /page/5/Picture/1 description: The image contains the logo for Dentsply Sirona. The logo consists of a stylized, curved shape on the left, followed by the text "Dentsply" on the first line and "Sirona" on the second line. The text is in a simple, sans-serif font and is aligned to the right of the curved shape.
## 510(k) SUMMARY
## for
## Sirona Dental CAD/CAM System with CEREC Chairside Software (K193408)
- 1. Submitter Information:
Dentsply Sirona 221 West Philadelphia Street Suite 60W York, PA 17401
| Contact Person: | Karl Nittinger |
|-------------------|-----------------------------------|
| Telephone Number: | 717-849-4424 |
| Fax Number: | 717-849-4343 |
| Email: | karl.nittinger@dentsplysirona.com |
11 August 2020 Date Prepared:
- 2. Device Name:
| • | Proprietary Name: | Sirona Dental CAD/CAM System with<br>CEREC Chairside Software |
|---|-------------------------|----------------------------------------------------------------|
| • | Classification Name: | Endosseous dental implant abutment. |
| • | CFR Number: | 21 CFR 872.3630 |
| • | Device Class: | Class II |
| • | Primary Product Code: | NHA (Abutment, Implant, Dental Endosseous) |
| • | Secondary Product Code: | PNP (Dental Abutment Design Software for Dental<br>Laboratory) |
#### 3. Predicate Device:
The predicate device that has been identified relating to the substantial equivalence of the Sirona Dental CAD/CAM System with CEREC Chairside Software is:
| Primary Predicate Device Name | 510(k) | Company Name |
|--------------------------------------------|---------|-------------------------------------------------|
| Sirona Dental CAD/CAM System | K181520 | Sirona Dental Systems GmbH<br>(Dentsply Sirona) |
| Reference Device Name | | |
| Osseospeed TX implant (3.5, 4.0, 4.5, 5.0) | K080156 | Astra Tech AB<br>(Dentsply Sirona) |
| Osseospeed TX Implant (3.0) | K080396 | Astra Tech AB<br>(Dentsply Sirona) |
{6}------------------------------------------------
#### 4. Description of Device:
The Sirona Dental CAD/CAM System with CEREC Chairside Software which is the subject of this premarket notification is a modification to the Sirona Dental CAD/CAM System as previously cleared under K181520. The modified Sirona Dental CAD/CAM System with CEREC Chairside Software that is the subject of this premarket notification includes a line extension to the existing offerings. These additional TiBase variants facilitate compatibility with currently marketed dental implant systems.
The modified Sirona Dental CAD/CAM System with CEREC Chairside Software which is the subject of this premarket notification consists of:
- CEREC SW "chairside" CAD/CAM software
- CEREC AC digital acquisition unit
- CEREC AC Connect digital acquisition unit ●
- CEREC Omnicam 3D digital intraoral scanner ●
- CEREC MCXL product family of CAM milling units
- Additional Sirona TiBase titanium base components (line extension subject to this submission) compatible with Dentsply Sirona Osseospeed TX 3.0, 3.5, 4.0, 4.5, and 5.0 implants.
- inCoris ZI zirconium mesostructure blocks ●
As subject to this premarket notification, the Sirona Dental CAD/CAM System with CEREC Chairside Software is utilized to digitally acquire and record the topographical characteristics of teeth, dental impressions, or physical stone models in order to facilitate the computer aided design (CAD) and computer aided manufacturing (CAM) of two-piece "CAD/CAM" abutments. The patient-specific two-piece abutments consist of prefabricated "TiBase" components which are designed with interface geometry to facilitate compatibility and connection with currently marketed dental implant system.
As subject to this premarket notification, the subject Sirona Dental CAD/CAM System with CEREC Chairside Software is modified to include a line extension to the existing TiBase offerings by introducing TiBases which are compatible with the Dentsply Sirona Osseospeed TX 3.0, 3.5, 4.0, 4.5, and 5.0 dental implants.
The CEREC SW CAD/CAM software is utilized to drive the specified acquisition unit hardware to acquire the intraoral dental scans and to design the mesostructure component of the CAD/CAM abutments. Following the completion of the design, the CEREC SW drives the CAM fabrication of the mesostructure component in the "chairside" workflow by utilizing the CEREC MCXL milling equipment and the defined zirconium/zirconia block materials.
The Sirona Dental CAD/CAM System with CEREC Chairside Software is intended for the design and fabrication of two-piece, CAD/CAM dental abutments.
{7}------------------------------------------------
#### 5. Indications for Use:
The Sirona Dental CAD/CAM System with CEREC Chairside Software is intended for use in partially or fully edentulous mandibles and maxillae in support of single or multiple-unit cement retained restorations. For the AT TX 3.0 S, BH 3.0 S, SSO 3.5 L, and SBL 3.3 L titanium bases, the indication is restricted to the replacement of single lateral incisors in the maxilla and lateral and central incisors in the mandible. The system consists of three major parts: TiBase, inCoris mesostructure and CAD/CAM software. Specifically, the inCoris mesostructure and TiBase components make up a two- piece abutment which is used in conjunction with endosseous dental implants to restore the function and aesthetics in the oral cavity. The inCoris mesostructure may also be used in conjunction with the Camlog Titanium base CAD/CAM (types K2244.xxxx) (K083496) in the Camlog Implant System. The CAD/CAM software is intended to design and fabricate the inCoris mesostructure. The inCoris mesostructure and TiBase two- piece abutment is compatible with the following implant systems:
| Manufacturer | Name of Implant System | Implant Size | |
|--------------------------------------|----------------------------------------------------------|-----------------|---------------------|
| | | Platform | Diameter |
| Nobel Biocare | Replace | NP | 3.5 |
| | | RP | 4.3 |
| | | WP | 5.0 |
| | | 6.0 | 6.0 |
| Nobel Biocare | Active | NP | 3.5 |
| | | RP | 4.3/5.0 |
| | Branemark | NP | 3.3 |
| | | RP | 3.75/4.0 |
| Straumann | Synocta | NN (3.5mm) | 3.3 |
| | | RN (4.8mm) | 3.3/4.1/4.8 |
| | | WN (6.5mm) | 4.8 |
| | Bone Level | NC (3.3mm) | 3.3 |
| | | RC (4.1mm/4.8mm | 4.1/4.8 |
| Dentsply Sirona Implants | Osseospeed | 3.5/4.0 | 3.5 S / 4.0 S |
| | | 4.5/5.0 | 4.5/5.0/5.0 S |
| | Xive | 3.4 | 3.4 |
| | | 3.8 | 3.8 |
| | | 4.5 | 4.5 |
| | | 5.5 | 5.5 |
| Manufacturer | Name of Implant System | Implant Size | |
| | | Platform | Diameter |
| Dentsply Sirona Implants | Osseospeed EV | 3.6 | 3.6 |
| | | 4.2 | 4.2 |
| | | 4.8 | 4.8 |
| | | 5.4 | 5.4 |
| | Ankylos | C/X | A, B, C, D |
| | | 3.0 | 3.0 |
| | Osseospeed TX | 3.5/4.0 | 3.5/4.0 |
| | | 4.5/5.0 | 4.5/5.0 |
| Biomet 3i | Osseotite | 3.4 | 3.25 |
| | | | 3.75 |
| | | 4.1 | 4.1 |
| | | | 3/4 |
| | | 5.0 | 5.0 |
| | | | 5.0 |
| | Certain | | 3.25 |
| | | 3.4 | 4/3 |
| | | | 3/4/3 |
| | | | 4.0 |
| | | 4.1 | 4/5/4 |
| | | | 5/4 |
| | | 5.0 | 5.0 |
| | | | 4/5 |
| Zimmer | Tapered Screw-Vent | 3.5 | 3.7/4.1 |
| | | 4.5 | 4.7 |
| | | 5.7 | 6 |
| Thommen Medical | Thommen Medical Implants | 3.5 | 3.5 |
| | | 4 | 4 |
| | | 4.5 | 4.5 |
| | | 5 | 5 |
| | | 6 | 6 |
| Osstem / Hiossen | Osstem TS Implant System | Mini | 3.5 |
| | Hiossen Implant System | Regular | 4.0/4.5/5.0/6.0/7.0 |
| Manufacturer | Name of Implant System | Implant Size | |
| | | Platform | Diameter |
| BioHorizons<br>(Internal Connection) | Tapered 3.0,Tapered plus | 3.0 | 3.0/3.4/3.8 |
| | Tapered internal | | 3.0 |
| | Tapered plus | | 4.6 |
| | Tapered internal, Tapered<br>internal tissue level | 3.5 | 3.0/3.8 |
| | Internal dental implant | | 3.5 |
| | Single stage dental implants | | 3.5/4.0 |
| BioHorizons<br>(Internal Connection) | Tapered Plus | | 5.8 |
| BioHorizons<br>(Internal Connection) | Tapered internal, Tapered<br>internal tissue level | 4.5 | 4.6 |
| | Internal dental implant | | 4.0 |
| | Single stage dental implants | | 4.0/5.0 |
| BioHorizons<br>(Internal Connection) | Tapered internal, Tapered<br>internal tissue level | | 5.8 |
| | Internal dental implant,<br>Single stage dental implants | 5.7 | 5.0/6.0 |
{8}------------------------------------------------
{9}------------------------------------------------
#### 6. Substantial Equivalence:
The modified Sirona Dental CAD/CAM System with CEREC Chairside Software has the same intended use as the predicate Sirona Dental CAD/CAM System cleared under premarket notification K181520. Both the modified Sirona Dental CAD/CAM System with CEREC Chairside Software and the predicate device are intended as optical impression systems for the 3D digital acquisition of the topography of teeth for use in the design and manufacturing of two-piece "CAD/CAM" dental abutments. As such, the subject Sirona Dental CAD/CAM System with CEREC Chairside Software and the predicate device cleared under premarket notification K181520 are regulated under 21 CFR 872.3630.
Osseospeed TX implant (K080156 and K080396) is listed as a reference device for the purpose of showing clearance of the additional implant system that the modified Sirona Dental CAD/CAM System with CEREC Chairside Software is compatible with. The previously cleared reference device, Osseospeed TX implant (K080156 and K080396) is not changed in any aspect (e.g. diameter) under the scope of this submission.
The modified Sirona Dental CAD/CAM System with CEREC Chairside Software which is the subject of this premarket notification and the predicate cleared under premarket notification K181520 include the same scanning, acquisition, and milling equipment, and utilize the same TiBase and inCoris ZI zirconia mesostructure materials for the design and fabrication of two-piece, CAD/CAM dental abutments.
The primary differences between the modified Sirona Dental CAD/CAM System with CEREC Chairside Software and the predicate device cleared in K181520 is the line extension of the available variants of the existing Sirona TiBase titanium base component offerings to facilitate compatibility with an additional dental implant system (Osseospeed TX).
{10}------------------------------------------------
Summary comparison of the intended use, indications for use, and design of the modified Sirona Dental CAD/CAM System with CEREC Chairside Software and the predicate, Sirona Dental CAD/CAM System (K181520), is presented in Tables 6.1 and 6.2.
Table 6.1: Indications for Use
| Modified Device | Predicate Device |
|---------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| Sirona Dental CAD/CAM System with | Sirona Dental CAD/CAM System |
| CEREC Chairside Software | (K181520) |
| | |
| The Sirona Dental CAD/CAM System with CEREC Chairside Software is | The Sirona Dental CAD/CAM System is intended for use in partially or |
| intended for use in partially or fully edentulous mandibles and maxillae in | fully edentulous mandibles and maxillae in support of single or multiple- |
| support of single or multiple-unit cement retained restorations. For the AT | unit cement retained restorations. For the BH 3.0 S, SSO 3.5 L and SBL |
| TX 3.0 S, BH 3.0 S, SSO 3.5 L and SBL 3.3 L titanium bases, the indication | 3.3 L titanium bases, the indication is restricted to the replacement of |
| is restricted to the replacement of single lateral incisors in the maxilla and | single lateral incisors in the maxilla and lateral and central incisors in the |
| lateral and central incisors in the mandible. The system consists of three | mandible. The system consists of three major parts: TiBase, inCoris |
| major parts: TiBase, inCoris mesostructure, and CAD/CAM software. | mesostructure, and CAD/CAM software. Specifically, the inCoris |
| Specifically, the inCoris mesostructure and TiBase components make up a | mesostructure and TiBase components make up a two-piece abutment |
| two-piece abutment which is used in conjunction with endosseous dental | which is used in conjunction with endosseous dental implants to restore |
| implants to restore the function and aesthetics in the oral cavity. The inCoris | the function and aesthetics in the oral cavity. The inCoris mesostructures |
| mesostructure may also be used in conjunction with the Camlog Titanium | may also be used in conjunction with the Camlog Titanium base |
| base CAD/CAM (types K2244.xxxx) (K083496) in the Camlog Implant | CAD/CAM (types K2244.xxxx) (K083496) in the Camlog Implant |
| System. The CAD/CAM software is intended to design and fabricate the | System. The CAD/CAM software is intended to design and fabricate the |
| inCoris mesostructure. The inCoris mesostructure and TiBase two-piece | inCoris mesostructure. The inCoris mesostructure and TiBase two-piece |
| abutment is compatible with the following implant systems: | abutment is compatible with the following implant systems: |
{11}------------------------------------------------
| Manufacturer | Name of<br>Implant<br>System | Implant Sizes | |
|--------------------------------|----------------------------------------------------------------------------|------------------------------------------------------|-----------------------------------------------------------|
| | | Platform | Diameter |
| Nobel Biocare | Replace | NP<br>TP<br>WP<br>6.0 | 3.5<br>4.3<br>5.0<br>6.0 |
| Nobel Biocare | Active | NP<br>RP | 3.5<br>4.3/5.0 |
| Branemark | | NP<br>RP | 3.3<br>3.75/4.0 |
| Straumann | Synocta | NN (3.5mm)<br>RN (4.8mm)<br>WN (6.5mm)<br>NC (3.3mm) | 3.3<br>3.3/4.1/4.8<br>4.8<br>3.3 |
| Straumann | Bone Level | RC<br>(4.1mm/4.8mm) | 4.1/4.8 |
| | Osseospeed | 3.5/4.0<br>4.5/5.0 | 3.5 S / 4.0 S<br>4.5/5.0/5.0 S |
| | Xive | 3.4<br>3.8<br>4.5<br>5.5 | 3.4<br>3.8<br>4.5<br>5.5 |
| Dentsply<br>Sirona<br>Implants | Osseospeed<br>EV | 3.6<br>4.2<br>4.8<br>5.4 | 3.6<br>4.2<br>4.8<br>5.4 |
| | Ankylos | C/X | A, B, C, D |
| | Osseospeed<br>TX | 3.0<br>3.5/4.0<br>4.5/5.0 | 3.0<br>3.5/4.0<br>4.5/5.0 |
| | Osseotite | 3.4<br>4.1<br>5.0 | 3.25<br>3.75<br>4.1<br>3/4<br>5.0<br>4/5 |
| Biomet 3i | Certain | 3.4<br>4.1<br>5.0 | 3.25<br>4/3<br>3/4/3<br>4.0<br>4/5/4<br>5/4<br>5.0<br>4/5 |
| Zimmer | Tapered<br>Screw-Vent | 3.5<br>4.5<br>5.7 | 3.7/4.1<br>4.7<br>6 |
| Thommen<br>Medical | SPI<br>ELEMENT,<br>SPI<br>ELEMENT<br>INICELL,<br>SPI<br>CONTACT<br>INICELL | 4<br>4.5<br>5<br>6 | 4<br>4.5<br>5<br>6 |
{12}------------------------------------------------
| | CONTACT<br>INICELL | | | | | Osstem TS<br>Implant | Mini | 3.5 |
|-----------------------------------------|----------------------------------------------------------------------|----------|--|---------------------------|-----------------------------------------|-------------------------------------------------------------|----------|---------------------------|
| Manufacturer | Name of<br>Implant<br>System | Platform | | Implant Sizes<br>Diameter | Osstem /<br>Hiossen | System-<br>Hiossen<br>Implant<br>System | Regular | 4.0/4.5/5.0/6.0/7.0 |
| | Osstem TS<br>Implant<br>System- | Mini | | 3.5 | Manufacturer | Name of<br>Implant | Platform | Implant Sizes<br>Diameter |
| Osstem /<br>Hiossen | Hiossen<br>Implant<br>System | Regular | | 4.0/4.5/5.0/6.0/7.0 | | System<br>Tapered<br>3.0,<br>Tapered | | 3.0/3.4/3.8 |
| | Tapered<br>3.0,<br>Tapered | | | 3.0/3.4/3.8 | | plus<br>Tapered<br>Internal | 3.0 | 3.0 |
| | plus<br>Tapered | 3.0 | | 3.0 | | Tapered<br>plus | | 4.6 |
| | Internal<br>Tapered<br>plus<br>Tapered<br>internal, | | | 4.6 | | Tapered<br>internal,<br>Tapered<br>internal<br>tissue level | 3.5 | 3.0/3.8 |
| | Tapered<br>internal<br>tissue level | 3.5 | | 3.0/3.8 | | Internal<br>dental<br>implant | | 3.5 |
| | Internal<br>dental<br>implant | | | 3.5 | | Single stage<br>dental<br>implants | | 3.5/4.0 |
| | Single stage<br>dental | | | 3.5/4.0 | BioHorizons<br>(Internal<br>Connection) | Tapered<br>plus | 4.5 | 5.8 |
| BioHorizons<br>(Internal<br>Connection) | implants<br>Tapered<br>plus<br>Tapered | 4.5 | | 5.8 | | Tapered<br>internal,<br>Tapered<br>internal | | 4.6 |
| | internal,<br>Tapered<br>internal<br>tissue level | | | 4.6 | | tissue level<br>Internal<br>dental<br>implant | | 4.0 |
| | Internal<br>dental<br>implant | | | 4.0 | | Single stage<br>dental<br>implants | | 4.0/5.0 |
| | Single stage<br>dental<br>implants<br>Tapered | | | 4.0/5.0 | | Tapered<br>internal,<br>Tapered<br>internal | 5.7 | 5.8 |
| | internal,<br>Tapered<br>internal<br>tissue level | 5.7 | | 5.8 | | tissue level<br>Internal<br>dental<br>implant, | | |
| | Internal<br>dental<br>implant,<br>Single stage<br>dental<br>implants | | | 5.0/6.0 | | Single stage<br>dental<br>implants | | 5.0/6.0 |
{13}------------------------------------------------
Table 6.2: Design
| Modified Device<br>Sirona Dental CAD/CAM System with<br>CEREC Chairside Software | Predicate Device<br>Sirona Dental CAD/CAM System<br>(K181520) |
|----------------------------------------------------------------------------------|---------------------------------------------------------------|
| CAD/CAM Software Version | |
| CEREC SW Chairside CAD/CAM Software | CEREC SW Chairside CAD/CAM Software |
| Acquisition Units | |
| CEREC AC | CEREC AC |
| CEREC AC Connect | CEREC AC Connect |
| CEREC Omnicam 3D digital intraoral scanner | CEREC Omnicam 3D digital intraoral scanner |
| Milling Unit | |
| CEREC MC | CEREC MC |
| CEREC MC X | CEREC MC X |
| CEREC MC XL | CEREC MC XL |
| CEREC MC XL Premium | CEREC MC XL Premium |
| Titanium Base Components | |
| Sirona TiBase | Sirona TiBase |
| Diameter: 3.0 mm - 7.0 mm | Diameter: 3.0 mm - 7.0 mm |
| Maximum Angulation of Finished Abutment: | Maximum Angulation of Finished Abutment: |
| 20° | 20° |
| Material (TiBase and Screw) | Material (TiBase and Screw) |
| Titanium alloy | Titanium alloy |
| Mesostructure material | |
| Sirona inCoris ZI zirconium blocks | Sirona inCoris ZI zirconium blocks |
| Block Material: | Block Material: |
| Zirconium oxide ceramic | Zirconium oxide ceramic |
| Block Dimensions: | Block Dimensions: |
| 24 mm (L) x 23 mm (W) x 21.5 mm (H) | 24 mm (L) x 23 mm (W) x 21.5 mm (H) |
| Available Shades: | Available Shades: |
| F0.5, F2 | F0.5, F2 |
{14}------------------------------------------------
#### 7. Non-Clinical Performance Data
Testing to verify the performance requirements of the modified Sirona Dental CAD/CAM System with CEREC Chairside Software was conducted and included in this premarket notification. The results of the performance testing support substantial equivalence.
Tests included in this premarket notification:
- Geometric compatibility of the implant to the TiBase connection implant connection . interface is confirmed for the subject new Sirona TiBase components via direct access to the original manufacturer's implant geometry specifications for the Dentsply Sirona, OsseoSpeed TX implant system. Because the implant to TiBase interface connection geometry of the proposed Sirona TiBase component for this system was designed through direct reference to the original implant design specifications, no reverse engineering analysis was conducted.
- . Dynamic fatigue testing of new TiBase variants in worst-case construct according to ISO 14801 (Dentistry - Implants - Dynamic loading test for endosseous dental implants) was conducted. The test design included test sample constructs fabricated at the maximum allowable abutment angulation and utilizing TiBase components exhibiting the worst-case design with respect to implant connection interface geometry.
- . New TiBase variants are composed of the identical materials and are fabricated utilizing the same methods as the components cleared under K181520. Therefore, no new biocompatibility data is included to support the substantial equivalence of the modified Sirona Dental CAD/CAM System.
- The recommended steam sterilization parameters for the new TiBase components included in the line extension as subject to this premarket notification are identical to the validated steam sterilization parameters recommended for the TiBase components included with the predicate device as cleared under premarket notification K181520. The validation of the recommended steam sterilization parameters was conducted according to ISO 17665-1: Sterilization of health care products -- Moist heat -- Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices and with reference to ANSI/AAMI ST79:2010 & A1:2010 &A2:2011 & A3:2012 and A4:2013: Comprehensive guide to steam sterilization and sterility assurance in health care facilities and the original validation of the recommended parameters is referenced in this premarket notification in support of substantial equivalence.
- . Software system integration testing was conducted to validate the system requirements associated with the introduction of the TiBase component line extension as selectable within the CEREC Chairside CAD/CAM software for the design and fabrication of two-piece CAD/CAM dental abutments. No modification to the CEREC CAD/CAM abutment design software's controls limiting the critical abutment design parameters are introduced in this premarket notification.
{15}------------------------------------------------
# 8. Clinical Performance Data
No human clinical data was included in this premarket notification to support the substantial equivalence of the modified Sirona Dental CAD/CAM System with CEREC Chairside Software.
# 9. Conclusion Regarding Substantial Equivalence
The information included in this premarket notification supports the substantial equivalence of the modified Sirona Dental CAD/CAM System with CEREC Chairside Software. The modified device which is the subject of this premarket notification has the identical intended use as the legally marketed predicate device cleared under premarket notification K181520. The modified device also has similar indications for use and incorporates the same fundamental technology as the predicate device (K181520).
Performance testing was conducted to demonstrate the performance of the modified Sirona Dental CAD/CAM System with CEREC Chairside Software with the line extension of the TiBase offerings. The results of the testing included in this premarket notification support a determination of substantial equivalence with the additional TiBases offered.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.