← Product Code [NHA](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA) · K183518

# Preat Abutments (K183518)

_Preat Corporation · NHA · Mar 18, 2019 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA/K183518

## Device Facts

- **Applicant:** Preat Corporation
- **Product Code:** [NHA](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA.md)
- **Decision Date:** Mar 18, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3630
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

Preat Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for single-unit or multi-unit prosthetic restorations. The Titanium Base abutments consists of two major parts. Specifically, the titanium base and mesostructured components make up a two-piece abutment. All digitally designed custom abutments, superstructures, and/or hybrid crowns for use with Titanium Base or Titanium Blank are to be sent to a Preat validated milling center for manufacture.

## Device Story

Preat Abutments are dental implant abutment systems used to support prosthetic restorations. The system includes ten abutment designs (temporary, esthetic angled, multi-unit straight/angled, ball, titanium base, titanium blank) compatible with twelve OEM implant systems. The titanium base abutment is a two-piece design consisting of a titanium base and a CAD/CAM zirconia superstructure. Digitally designed custom components must be manufactured at a Preat-validated milling center. The device is used by clinicians in a dental setting to connect prostheses to endosseous implants. The clinician selects the appropriate abutment based on the implant system and clinical requirements. The device provides the mechanical interface between the implant and the final restoration, facilitating functional and esthetic rehabilitation for the patient.

## Clinical Evidence

No clinical data were included in this submission. Substantial equivalence is supported by non-clinical bench testing, including sterilization validation (ISO 17665-1, ISO 14937), biocompatibility testing (ISO 10993-5, ISO 10993-12), and mechanical fatigue testing (ISO 14801).

## Technological Characteristics

Abutments and screws are manufactured from Ti-6Al-4V alloy (ASTM F136). Zirconia superstructures conform to ISO 13356. The system includes various designs (temporary, esthetic, multi-unit, ball, titanium base, titanium blank) with platform diameters from 3.0 mm to 6.5 mm. The device is provided non-sterile for end-user sterilization. It is a mechanical system; no software or energy source is involved.

## Regulatory Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

## Predicate Devices

- DESS Dental Smart Solutions ([K170588](/device/K170588.md))

## Reference Devices

- OsseoSpeed™ Plus ([K120414](/device/K120414.md))
- Multi-unit Abutment Plus ([K161416](/device/K161416.md))
- BioHorizons CAD/CAM Abutments ([K151621](/device/K151621.md))
- Inclusive® Titanium Abutments for Astra OsseoSpeed™ ([K100993](/device/K100993.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 18, 2019

Preat Corporation % Kevin Thomas Vice President and Director of Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130

Re: K183518

Trade/Device Name: Preat Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: December 17, 2018 Received: December 18, 2018

Dear Kevin Thomas:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Mary S. Mary S. Runner -S3 Runner -S3 3 Date: 2019.03.18
16:42:18 -04'00' For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K183518

Device Name

Preat Abutments

Indications for Use (Describe)

Preat Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandbular arch to provide support for single-unit or multi-unit prosthetions. The Tianium Base abutments consists of two major parts. Specifically, the titanium base and mesostructured components make up a two-piece abutment.

All digitally designed custom abutments, superstructures, and/or hybrid crowns for use with Titanium Blank are to be sent to a Preat validated milling center for manufacture.

| Compatible Implant System              | Implant Body Diameter<br>(mm) | Implant Platform Diameter<br>(mm) |
|----------------------------------------|-------------------------------|-----------------------------------|
| 3i OSSEOTITE® Certain®                 | 3.25                          | 3.4                               |
|                                        | 4.0                           | 4.1                               |
|                                        | 5.0                           | 5.0                               |
|                                        | 6.0                           | 6.0                               |
| Astra Tech OsseoSpeed™                 | 3.0                           | 3.0                               |
|                                        | 3.5, 4.0                      | 3.5/4.0                           |
|                                        | 4.5, 5.0                      | 4.5/5.0                           |
| BioHorizons Tapered Internal           | 3.0                           | 3.0                               |
|                                        | 3.5                           | 3.5                               |
|                                        | 4.0                           | 4.5                               |
| HIOSSEN ET III                         | 3.5                           | Mini                              |
|                                        | 4.0, 4.5, 5.0, 6.0, 7.0       | Regular                           |
| Implant Direct Legacy                  | 3.2                           | 3.0                               |
|                                        | 3.7, 4.2                      | 3.5                               |
|                                        | 4.7, 5.2                      | 4.5                               |
|                                        | 5.7, 7.0                      | 5.7                               |
| MegaGen AnyRidge                       | 3.5, 4.0, 4.5, 5.0, 5.5       | 3.5                               |
| Neoss                                  | 3.5, 4.0, 4.5, 5.0, 5.5       | 4.1                               |
| NobelActive®                           | 3.5                           | NP                                |
|                                        | 4.3, 5.0                      | RP                                |
| Nobel Replace™                         | 3.5                           | NP                                |
|                                        | 4.0, 4.3, 5.0                 | RP                                |
|                                        | 5.0                           | WP                                |
|                                        | 6.0                           | 6.0                               |
| Straumann® Bone Level                  | 3.3                           | NC                                |
|                                        | 4.1, 4.8                      | RC                                |
| Straumann® Tissue Level                | 3.3, 4.1, 4.8                 | RN                                |
|                                        | 4.8, 6.5                      | WN                                |
| Zimmer Screw-Vent®/Tapered Screw-Vent® | 3.3, 3.7, 4.1                 | 3.5                               |
|                                        | 4.7                           | 4.5                               |
|                                        | 6.0                           | 5.7                               |

Compatible Implant Systems

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

PSC Publishing Services (301) 443-6740 EF

{4}------------------------------------------------

#### 510(k) Summary

# K183518 – Preat Abutments

#### Preat Corporation

March 18, 2019

#### ADMINISTRATIVE INFORMATION

| Manufacturer Name | Preat Corporation<br>100 S. 4th Street<br>Grover Beach, CA 93433<br>Telephone: +1 805-202-3070<br>Fax: +1 805-202-3076 |
|-------------------|------------------------------------------------------------------------------------------------------------------------|
| Official Contact  | Chris Bormes, President and CEO                                                                                        |

Representative/Consultant

Kevin A. Thomas, PhD Floyd G. Larson, MS. MBA PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, CA 92130 Telephone: +1 858-792-1235 Fax: +1 858-792-1236 Email: kthomas@paxmed.com flarson@paxmed.com

#### DEVICE NAME AND CLASSIFICATION

| Trade/Proprietary Name | Preat Abutments                    |
|------------------------|------------------------------------|
| Common Name            | Dental implant abutment            |
| Regulation Number      | 21 CFR 872.3630                    |
| Regulation Name        | Endosseous dental implant abutment |
| Regulatory Class       | Class II                           |
| Product Code           | NHA                                |
| Classification Panel   | Dental Products Panel              |
| Reviewing Branch       | Dental Devices Branch              |

#### PREDICATE DEVICE INFORMATION

Primary Predicate K170588, DESS Dental Smart Solutions, Terrats Medical SL

Reference Devices K120414, OsseoSpeed™ Plus, Astra Tech AB K161416, Multi-unit Abutment Plus, Nobel Biocare AB K151621, BioHorizons CAD/CAM Abutments, BioHorizons Implant Systems, Inc. K100993, Inclusive® Titanium Abutments for Astra OsseoSpeed™, Prismatik Dentalcraft, Inc.

{5}------------------------------------------------

### INDICATIONS FOR USE STATEMENT

Preat Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for single-unit or multi-unit prosthetic restorations. The Titanium Base abutments consists of two major parts. Specifically, the titanium base and mesostructured components make up a two-piece abutment.

All digitally designed custom abutments, superstructures, and/or hybrid crowns for use with Titanium Base or Titanium Blank are to be sent to a Preat validated milling center for manufacture.

| Compatible Implant System              | Implant Body Diameter<br>(mm) | Implant Platform Diameter<br>(mm) |
|----------------------------------------|-------------------------------|-----------------------------------|
| 3i OSSEOTITE® Certain®                 | 3.25                          | 3.4                               |
|                                        | 4.0                           | 4.1                               |
|                                        | 5.0                           | 5.0                               |
|                                        | 6.0                           | 6.0                               |
| Astra Tech OsseoSpeed™                 | 3.0                           | 3.0                               |
|                                        | 3.5, 4.0                      | 3.5/4.0                           |
|                                        | 4.5, 5.0                      | 4.5/5.0                           |
| BioHorizons Tapered Internal           | 3.0                           | 3.0                               |
|                                        | 3.5                           | 3.5                               |
|                                        | 4.0                           | 4.5                               |
| HIOSSEN ET III                         | 3.5                           | Mini                              |
|                                        | 4.0, 4.5, 5.0, 6.0, 7.0       | Regular                           |
| Implant Direct Legacy                  | 3.2                           | 3.0                               |
|                                        | 3.7, 4.2                      | 3.5                               |
|                                        | 4.7, 5.2                      | 4.5                               |
|                                        | 5.7, 7.0                      | 5.7                               |
| MegaGen AnyRidge                       | 3.5, 4.0, 4.5, 5.0, 5.5       | 3.5                               |
| Neoss                                  | 3.5, 4.0, 4.5, 5.0, 5.5       | 4.1                               |
| NobelActive®                           | 3.5                           | NP                                |
|                                        | 4.3, 5.0                      | RP                                |
| Nobel Replace™                         | 3.5                           | NP                                |
|                                        | 4.0, 4.3, 5.0                 | RP                                |
|                                        | 5.0                           | WP                                |
|                                        | 6.0                           | 6.0                               |
| Straumann® Bone Level                  | 3.3                           | NC                                |
|                                        | 4.1, 4.8                      | RC                                |
| Straumann® Tissue Level                | 3.3, 4.1, 4.8                 | RN                                |
|                                        | 4.8, 6.5                      | WN                                |
| Zimmer Screw-Vent®/Tapered Screw-Vent® | 3.3, 3.7, 4.1                 | 3.5                               |
|                                        | 4.7                           | 4.5                               |
|                                        | 6.0                           | 5.7                               |

| Compatible Implant Systems |  |  |
|----------------------------|--|--|

### SUBJECT DEVICE DESCRIPTION

Preat Abutments is a dental implant abutment system that includes ten (10) abutment designs compatible with twelve (12) OEM implant systems. The subject device abutment platform diameters range from 3.0 mm to 6.5 mm, and the corresponding compatible implant body diameters also range from 3.0 mm to 6.5 mm. The subject device includes the following abutment designs: temporary, esthetic angled 15°, multi

{6}------------------------------------------------

unit straight, multi-unit angled 17°, ball (Clix Ball and O-Ring), titanium base, and titanium blank. The system also includes corresponding abutment screws.

The following table shows the subject device abutments platform sizes for each of the OEM implant lines and sizes.

| Preat Abutment Platform<br>Diameter /<br>Compatible Implant System | Subject Device Abutment Designs |                           |                        |                        |                          |           |        |                           |                               |                   |                    |
|--------------------------------------------------------------------|---------------------------------|---------------------------|------------------------|------------------------|--------------------------|-----------|--------|---------------------------|-------------------------------|-------------------|--------------------|
|                                                                    | Temporary<br>Engaging           | Temporary<br>Non-Engaging | Esthetic<br>Angled 15° | Multi-Unit<br>Straight | Multi-Unit<br>Angled 17° | Clix Ball | O-Ring | Titanium Base<br>Engaging | Titanium Base<br>Non-Engaging | Titanium<br>Blank | Titanium<br>Screws |
| 3i OSSEOTITE® Certain®                                             |                                 |                           |                        |                        |                          |           |        |                           |                               |                   |                    |
| 3.4                                                                | X                               | X                         |                        | X                      | X                        | X         | X      | X                         | X                             | X                 | X                  |
| 4.1                                                                | X                               | X                         |                        | X                      | X                        | X         | X      | X                         | X                             | X                 | X                  |
| 5.0                                                                | X                               | X                         |                        | X                      | X                        |           |        | X                         | X                             | X                 | X                  |
| 6.0                                                                | X                               | X                         |                        |                        |                          |           |        | X                         | X                             | X                 | X                  |
| Astra Tech OsseoSpeed™                                             |                                 |                           |                        |                        |                          |           |        |                           |                               |                   |                    |
| 3.0                                                                | X                               | X                         |                        |                        |                          |           |        | X                         | X                             | X                 | X                  |
| 3.5/4.0                                                            | X                               | X                         | X                      | X                      | X                        | X         | X      | X                         | X                             | X                 | X                  |
| 4.5/5.0                                                            | X                               | X                         | X                      | X                      | X                        | X         | X      | X                         | X                             | X                 | X                  |
| BioHorizons Tapered Internal                                       |                                 |                           |                        |                        |                          |           |        |                           |                               |                   |                    |
| 3.0                                                                | X                               | X                         |                        | X                      | X                        |           |        | X                         | X                             | X                 | X                  |
| 3.5                                                                | X                               | X                         |                        |                        | X                        | X         | X      | X                         | X                             | X                 | X                  |
| 4.5                                                                | X                               | X                         |                        |                        | X                        | X         | X      | X                         | X                             | X                 | X                  |
| HIOSSEN ET III                                                     |                                 |                           |                        |                        |                          |           |        |                           |                               |                   |                    |
| Mini                                                               | X                               | X                         |                        | X                      | X                        |           |        | X                         | X                             | X                 | X                  |
| Regular                                                            | X                               | X                         |                        | X                      | X                        |           |        | X                         | X                             | X                 | X                  |
| Implant Direct Legacy                                              |                                 |                           |                        |                        |                          |           |        |                           |                               |                   |                    |
| 3.0                                                                |                                 |                           |                        | X                      | X                        |           |        | X                         | X                             | X                 | X                  |
| 3.5                                                                | X                               | X                         |                        |                        | X                        | X         | X      | X                         | X                             | X                 | X                  |
| 4.5                                                                | X                               | X                         |                        |                        | X                        | X         | X      | X                         | X                             | X                 | X                  |
| 5.7                                                                | X                               | X                         |                        |                        |                          |           |        | X                         | X                             | X                 | X                  |
| MegaGen AnyRidge                                                   |                                 |                           |                        |                        |                          |           |        |                           |                               |                   |                    |
| 3.5                                                                | X                               | X                         |                        | X                      | X                        |           |        | X                         | X                             | X                 | X                  |
| Neoss                                                              |                                 |                           |                        |                        |                          |           |        |                           |                               |                   |                    |
| 4.1                                                                | X                               | X                         |                        | X                      | X                        | X         | X      | X                         | X                             | X                 | X                  |
| NobelActive®                                                       |                                 |                           |                        |                        |                          |           |        |                           |                               |                   |                    |
| NP                                                                 | X                               | X                         |                        | X                      | X                        | X         | X      | X                         | X                             | X                 | X                  |
| RP                                                                 | X                               | X                         |                        | X                      | X                        | X         | X      | X                         | X                             | X                 | X                  |
| Nobel Replace™                                                     |                                 |                           |                        |                        |                          |           |        |                           |                               |                   |                    |
| NP                                                                 | X                               | X                         |                        | X                      | X                        | X         | X      | X                         | X                             | X                 | X                  |
| RP                                                                 | X                               | X                         |                        | X                      | X                        | X         | X      | X                         | X                             | X                 | X                  |
| WP                                                                 | X                               | X                         |                        | X                      |                          | X         | X      | X                         | X                             | X                 | X                  |
| 6.0                                                                |                                 |                           |                        |                        |                          |           |        |                           |                               | X                 | X                  |
| Straumann® Bone Level                                              |                                 |                           |                        |                        |                          |           |        |                           |                               |                   |                    |
| NC                                                                 | X                               | X                         |                        | X                      | X                        | X         | X      | X                         | X                             | X                 | X                  |
| RC                                                                 | X                               | X                         |                        | X                      | X                        | X         | X      | X                         | X                             | X                 | X                  |
| Straumann® Tissue Level                                            |                                 |                           |                        |                        |                          |           |        |                           |                               |                   |                    |
| RN                                                                 | X                               | X                         |                        |                        |                          | X         | X      | X                         | X                             | X                 | X                  |
| WN                                                                 |                                 |                           |                        |                        |                          |           |        |                           |                               | X                 | X                  |
| Zimmer Screw-Vent®/<br>Tapered Screw-Vent®                         |                                 |                           |                        |                        |                          |           |        |                           |                               |                   |                    |
| 3.5                                                                | X                               | X                         |                        | X                      | X                        | X         | X      | X                         | X                             | X                 | X                  |
| 4.5                                                                | X                               | X                         |                        | X                      | X                        | X         | X      | X                         | X                             | X                 | X                  |

{7}------------------------------------------------

All abutments and screws are manufactured from Ti-6Al-4V alloy conforming to ASTM F136 and are provided non-sterile to the end user. All digitally designed custom abutments, superstructures, and/or hybrid crowns for use with Titanium Base or Titanium Blank are to be sent to a Preat validated milling center for manufacture. All superstructures are to be manufactured from zirconia conforming to ISO 13356.

The Titanium Base abutment is composed of two-piece abutment that is a titanium base at the bottom and a zirconia superstructure (CAD/CAM patient specific superstructure) at the top. The zirconia superstructure is straight only and is not to be designed to provide an angle or divergence correction.

For the Titanium Base abutment, the design parameters for the CAD/CAM zirconia superstructure are: Minimum wall thickness - 0.5 mm; Minimum post height for single-unit restorations - 4.0 mm; Maximum gingival height - 5.0 mm; and All zirconia superstructures are for straight abutments only.

The design parameters for the CAD/CAM Titanium Blank custom abutment are: Minimum wall thickness - 0.5 mm; Minimum post height for single-unit restoration - 4.0 mm; Maximum Angle - 30°; and Maximum gingival height - 1.5 mm to 2.65 mm (varies by implant line).

## PERFORMANCE DATA

Non-clinical data submitted to demonstrate substantial equivalence included: sterilization validation according to ISO 17665-1 and ISO 14937; biocompatibility according to ISO 10993-12; reverse engineering of OEM implant bodies, abutment screws to confirm compatibility; and static compression and compression fatigue testing according to ISO 14801. No clinical data were included in this submission.

### EQUIVALENCE TO MARKETED DEVICES

The subject device is substantially equivalent in indications and design principles to the primary predicate device and the reference devices listed above. Provided at the end of this summary are tables comparing the Indications for Use Statements and the technological characteristics of the subject device, the primary predicate device, and the reference devices.

Subject device abutments are substantially equivalent in intended use to the primary predicate K170588, and the reference devices K120414, K161416, K151621, and K100993. All are intended for use with endosseous dental implants in the maxilla and mandible to provide functional and esthetic rehabilitation of the edentulous maxilla and mandible.

The Indications for Use Statement (IFUS) for the subject device is substantially equivalent to that of the primary predicate K170588. Slight differences in language of the subject device and primary predicate device Indications for Use statements do not affect the intended use as an endosseous dental implant abutment for support of a prosthesis to restore chewing function. The minor differences between the IFUS for the subject device and the primary predicate include: the subject device IFUS includes the term "single-unit or multi-unit" and the primary predicate IFUS does not. The other minor differences are

{8}------------------------------------------------

related to the specific device names, validated milling centers, and the compatible OEM implant lines. None of these minor differences impact substantial equivalence because both IFUS express equivalent intended use to facilitate dental prosthetic restorations, and the indications are expressed equivalently using different specific wording.

Similarly, the differences between the subject device IFUS and that of each of the reference devices are related to the specific device names and design features, validated milling centers, and the compatible implant lines. None of these minor differences impact substantial equivalence because all IFUS express equivalent intended use to facilitate dental prosthetic restorations, and the indications are expressed equivalently using different specific wording.

The following subject device designs are substantially equivalent to the primary predicate device K 170588: temporary abutments; multi-unit straight abutments; titanium base abutments; and titanium blank abutments. The subject device and the primary predicate device K170588 are for single-unit or multi-unit restorations, have internal implant interface connections, and are made of Ti-6Al-4V alloy (abutments and abutment screws). The primary predicate includes straight abutments only. The subject device includes angled abutments up to 30° and the substantial equivalence of this angulation is supported by each of the reference devices.

The subject device includes designs for implant platforms ranging from 3.0 mm to 6.5 mm. The primary predicate K170588 includes implant platform sizes of 3.4 mm. Substantial equivalence of the smaller implant platform sizes of the subject device are supported by the reference devices K120414, K151621, and K100993.

The following subject device designs are substantially equivalent to those of the reference device K120414: temporary abutments; esthetic abutments (angled 15°); multi-unit straight abutments; and ball abutments. The subject device and the reference device K120414 abutments are for single-unit or multiunit restorations, include angulations up to 30°, have internal implant interface connections, and are made of Ti-6A1-4V alloy (abutments and abutment screws).

The subject device multi-unit angled abutments are substantially equivalent to those of the reference device K161416. The subject device and the reference device K161416 each include multi-unit abutments angled 17°, have internal implant interface connections, and are made of Ti-6AI-4V alloy (abutments and abutment screws).

The subject device titanium base abutments are substantially equivalent to those of the reference device K151621. The subject device and the reference device K151621 each include titanium base abutment designs with a cut out region to allow for angled access to the screw channel for multi-unit restorations. The subject device and the reference device K151621 titanium base designs have internal implant interface connections, and are made of Ti-6Al-4V alloy (abutments and abutment screws). The reference device K100993 is for support of substantial equivalence in terms of mechanical performance as discussed below. The subject device and the reference device K100993 titanium blank abutment designs with angulation of up to 30°, are for internal implant interface connections, and are made of Ti-6Al-4V alloy (abutments and abutment screws).

The subject device is to be sterilized by the end-user, the same as the primary predicate device K170588. Sterilization validation for the subject device was performed according to ISO 17665-1 and ISO TR 17665-2. This sterilization validation method is the same as the primary predicate device K170588.

{9}------------------------------------------------

Confirmatory biocompatibility testing of the subject device was performed according to ANSI / AAMI / ISO 10993-5 and 10993-12.

Mechanical performance testing was performed according to ISO 14801. For each compatible OEM implant line, worst-case constructs were subjected to static compression fatigue testing. The reference device K100993 is to support substantial equivalence of the fatigue limit of 133 N for the Astra Tech OsseoSpeed™ and BioHorizons Tapered Internal implant lines (each with 3.0 mm implant body and platform diameters). The fatigue limit data for all other implant lines demonstrated the construct strengths to be sufficient for their intended use.

Minor differences in the designs, dimensions, sizes, or compatible OEM implant lines among the subject device, the primary predicate device, and the reference devices do not affect substantial equivalence. These minor differences do not impact substantial equivalence because these differences are related to the compatible OEM implant designs, or are mitigated by the mechanical performance testing.

#### CONCLUSION

The subject device, the primary predicate device, and the reference devices have the same intended use, have similar technological characteristics, and are materials. The subject device, the primary predicate, and reference devices encompass the same range of physical dimensions, are packaged in similar materials, and are to be sterilized using similar methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.

{10}------------------------------------------------

# Comparison of Indications for Use Statements

| Subject Device                                                                                                                                                                                                                                                                                                                                                            |                                         |                                   | Primary Predicate                                                                                                                                                                                                                                                                                                                                                             |                       |                        | Reference Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Reference Device                                                                                                                                                                                         | Reference Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Reference Device                                                                                                                                                                                                                                                                                                                           |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Preat Abutments                                                                                                                                                                                                                                                                                                                                                           |                                         |                                   | K170588<br>DESS Dental Smart Solutions                                                                                                                                                                                                                                                                                                                                        |                       |                        | K120414<br>OsseoSpeed™ Plus                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | K161416<br>Multi-unit Abutment Plus                                                                                                                                                                      | K151621<br>BioHorizons CAD/CAM Abutments                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | K100993<br>Inclusive® Titanium Abutments for<br>Astra OsseoSpeed™                                                                                                                                                                                                                                                                          |
| Preat Corporation                                                                                                                                                                                                                                                                                                                                                         |                                         |                                   | Terrats Medical SL                                                                                                                                                                                                                                                                                                                                                            |                       |                        | Astra Tech AB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Nobel Biocare                                                                                                                                                                                            | BioHorizons Implant Systems, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Prismatik Dentalcraft, Inc.                                                                                                                                                                                                                                                                                                                |
| Preat Abutments are intended to be used in conjunction with endosseous dental implants in<br>the maxillary or mandibular arch to provide support for single-unit or multi-unit prosthetic<br>restorations.<br>All digitally designed custom abutments for use with Titanium Base or Titanium Blank are<br>to be sent to a Preat validated milling center for manufacture. |                                         |                                   | DESS Dental Smart Solutions abutments are intended to be used in conjunction<br>with endosseous dental implants in the maxillary or mandibular arch to provide<br>support for prosthetic restorations.<br>All digitally designed custom abutments for use with TiBase or Pre-milled<br>Blank are to be sent to a Terrats Medical validated milling center for<br>manufacture. |                       |                        | (Implant section of IFUS not applicable to<br>this submission)<br>Abutments:<br>Astra Tech Implant System Plus abutments<br>are intended to be used in conjunction with<br>Astra Tech Implant System Plus in fully<br>edentulous or partially edentulous<br>maxillary and/or mandibular arches to<br>provide support for crowns, bridges or<br>overdentures.<br><br>Atlantis Abutments:<br>The Atlantis™ Abutment is intended for<br>use with an endosseous implant to support<br>a prosthetic device in a partially or<br>completely edentulous patient. It is<br>intended for use to support single and<br>multiple tooth prostheses, in the mandible<br>or maxilla. The prosthesis can be<br>cemented, screw retained or friction fit to<br>the abutment. The abutment screw is<br>intended to secure the abutment to the<br>endosseous implant.<br><br>The Atlantis™ Crown Abutment in<br>Zirconia is intended for use with an<br>endosseous implant to function as a<br>substructure that also serves as the final<br>restoration, in partially or completely<br>edentulous patients. The prosthesis is<br>screw retained. The abutment screw is<br>intended to secure the crown abutment to<br>the endosseous implant. | The Multi-unit Abutment Plus is a pre-<br>manufactured prosthetic component<br>directly connected to the endosseous dental<br>implant and is intended for use as an aid in<br>prosthetic rehabilitation. | BioHorizons CAD/CAM Abutments are<br>dental abutments placed onto a dental<br>implant to provide support for dental<br>prosthetic restorations. The abutments<br>include: 1) Titanium abutment blanks with<br>a pre-machined implant connection where<br>the upper portion may be custom-milled in<br>accordance with a patient-specific design<br>using CAD/CAM techniques; and 2)<br>Titanium bases with a pre-machined<br>implant connection upon which a<br>CAD/CAM designed superstructure may<br>be fitted to complete a two-piece dental<br>abutment. The abutments include an<br>abutment screw for fixation to the<br>underlying implant. The abutments may be<br>used for single-unit (single-tooth) or<br>multiple-unit (bridges and bars)<br>restorations and are compatible for use<br>with BioHorizons Internal and Tapered<br>Internal implant systems and Zimmer®<br>Dental Screw-Vent® and Tapered Screw-<br>Vent® implants with 3.5mm, 4.5mm and<br>5.7mm internal hex-connection mating<br>platform diameters.<br>All digitally designed abutments and/or<br>copings for use with BioHorizons<br>CAD/CAM Abutments are intended to be<br>sent to a BioHorizons-validated milling<br>center for manufacture. BioHorizons<br>abutments designed using CAD/CAM<br>techniques must fulfill the BioHorizons<br>allowable range of design parameters. | The Inclusive® Titanium Abutments for<br>Astra OsseoSpeed™ Implants are<br>premanufactured prosthetic components<br>directly connected to endosseous dental<br>implants and are intended for use as an aid<br>in prosthetic rehabilitation. They are<br>compatible with the Astra Tech<br>OsseoSpeed™ 3.0, 3.5, 4.0, 4.5, 5.0<br>implants. |
| Compatible Implant System                                                                                                                                                                                                                                                                                                                                                 | Implant Body Diameter<br>(mm)           | Implant Platform Diameter<br>(mm) | Compatible Implant Systems                                                                                                                                                                                                                                                                                                                                                    | Implant Diameter (mm) | Platform Diameter (mm) |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                            |
| 3i OSSEOTITE® Certain®                                                                                                                                                                                                                                                                                                                                                    | 3.25<br>4.0<br>5.0<br>6.0               | 3.4<br>4.1<br>5.0<br>6.0          | 3i Certain®                                                                                                                                                                                                                                                                                                                                                                   | 3.25, 4.0, 5.0        | 3.4, 4.1, 5.0          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                            |
| Astra Tech OsseoSpeed™                                                                                                                                                                                                                                                                                                                                                    | 3.0<br>3.5, 4.0<br>4.5, 5.0             | 3.0<br>3.5/4.0<br>4.5/5.0         | 3i OSSEOTITE®                                                                                                                                                                                                                                                                                                                                                                 | 3.25, 3.75, 4.0, 5.0  | 3.4, 4.1, 5.0          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                            |
| BioHorizons Tapered Internal                                                                                                                                                                                                                                                                                                                                              | 3.0<br>3.5<br>4.0                       | 3.0<br>3.5<br>4.5                 | OsseoSpeed™                                                                                                                                                                                                                                                                                                                                                                   | 3.5, 4.0, 5.0         | 3.5/4.0, 4.5/5.0       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                            |
| HIOSSEN ET III                                                                                                                                                                                                                                                                                                                                                            | 3.5<br>4.0, 4.5, 5.0, 6.0, 7.0          | Mini<br>Regular                   | FRIADENT XiVE                                                                                                                                                                                                                                                                                                                                                                 | 3.4, 3.8, 4.5         | 3.4, 3.8, 4.5          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                            |
| Implant Direct Legacy…

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA/K183518](https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA/K183518)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
