A.B. DENTAL DEVICES® Dental Implants System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. A.B. DENTAL DEVICES® Dental Implants System is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Device Story
System consists of endosseous dental implants and various abutment components (healing caps, straight/anatomic abutments, ball attachments, UCLA-style composed abutments, and LOC abutments). Used by dental professionals in clinical settings for surgical placement into jaw bone to support prosthetic restorations. Components function as interfaces between the endosseous implant and the final prosthetic device. Device provides mechanical support for artificial teeth; enables immediate loading protocols when clinical conditions (primary stability) are met. Benefits include restoration of chewing function and dental aesthetics. Operation involves standard surgical and restorative dental techniques; no software or electronic processing involved.
Clinical Evidence
Bench testing only. Evidence includes engineering analysis for worst-case determination, biocompatibility (ISO 10993-1, ISO 10993-5), sterilization validation (ANSI/AAMI/ISO 17665-1), and fatigue testing (ISO 14801). No clinical data presented.
Technological Characteristics
Endosseous dental implant abutments. Materials: Ti-6Al-4V ELI (ASTM F136), PEEK, Gold Alloy 6019, FWM 1058 Cobalt Chrome Alloy (ASTM F1058), Delrin. Mechanical connection: internal hex, anti-rotation, taper fit. Sterilization: non-sterile (intended for sterilization by user).
Indications for Use
Indicated for surgical and restorative applications in the upper or lower jaw to support prosthetic devices (e.g., artificial teeth) to restore chewing function. Indicated for immediate loading when primary stability and appropriate occlusal loading are achieved.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
A.B. DENTAL DEVICES® Dental Implants System (K162482)
A.B. DENTAL DEVICES® Dental Implants System (K132125)
A.B. DENTAL DEVICES® Dental Implants System (K112440)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 23, 2018
A.B Dental Device Ltd. % John J. Smith, M.D., J.D. Partner Hogan Lovells US LLP 555 13th St. NW Washington, District of Columbia 20004
Re: K181381
Trade/Device Name: A.B. Dental Devices® Dental Implants System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: October 25, 2018 Received: October 25, 2018
Dear John J. Smith, M.D., J.D.:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Andrew I. Steen -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement on last page
510(k) Number (if known)
K181381
Device Name
## A.B. Dental Devices® Dental Implants System
Indications for Use (Describe)
A.B. DENTAL DEVICES® Dental Implants System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. A.B. DENTAL DEVICES® Dental Implants System is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | □ Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1 FDA
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
#### K181381 510(k) SUMMARY A.B. DENTAL DEVICES® Dental Implants System
#### Sponsor:
A.B. Dental Devices Ltd. 19 Hayalomim St., Ashdod 7761117 Israel
### Contact Person:
Gabi Krauss RA Manager Phone: 972-8-8531388 Facsimile: 972-8-8522562 Email: gabi@ab-dent.com
Date Prepared: November 23, 2018
| Name of Device: | A.B. Dental Devices® Dental Implants System |
|-----------------------|-------------------------------------------------------|
| Common or Usual Name: | Abutment, implant, dental, endosseous |
| Classification Name: | Endosseous dental implant abutment; 21 CFR §872.3630. |
| Product Code: | NHA |
## Predicate Devices
K162482 A.B. DENTAL DEVICES® Dental Implants System (primary) K132125 A.B. DENTAL DEVICES® Dental Implants System (reference) K112440, DENTAL DEVICES® Dental Implants System (reference)
#### Indications for Use
A.B. DENTAL DEVICES® Dental Implants System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. A.B. DENTAL DEVICES® Dental Implants System is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
## Device Description and Technological Characteristics
A.B. DENTAL DEVICES® Dental Implants System consists of one and two stage endosseous form dental implants, internal hexagonal and one piece implants systems. Narrow platform implants are to be used with straight abutments only. The following components are being added:
- . P0-3 narrow platform healing caps, 4 and 7 mm length
- PON-3.75 standard platform, narrow healing caps, 4 and 7 mm length .
{4}------------------------------------------------
- P3-5 standard platform, straight abutment, 5 mm diameter, 5 mm length ●
- P3S-3 narrow platform, anatomic, straight abutment, 1, 2, 3 mm shoulder lengths
- . P3SW-3.75 standard platform, anatomic, straight abutment, wide diameter, 1, 2, 3 mm shoulder lengths
- P3W-3 narrow platform, straight abutment, wide diameter, 9 mm length
- P5-3 narrow platform, ball attachment, 4, 5, 6 mm height
- P5-3.75-20 standard platform, 20º ball attachment, 4 and 6 mm height ●
- P9HG-3.75 standard platform, composed abutment, gold alloy
- P9R and P9HR -3.75 standard platform, composed abutment, CoCr alloy
- . P25-3 narrow platform, A.B. LOC abutments, 0, 1, 2, 3, 4 5 mm height
- . P25-3.75 standard platform. A.B. LOC abutments. 0. 1. 2 mm height
- . P64-3.75 standard platform, straight adaptor abutment, 1, 2, 3, 4, 5 mm heights
- P64 plastic and Ti alloy sleeve
- P64 healing cap ●
- PK standard platform, anatomic anti-rotation abutment, 1, 2, 3, 4 mm height
- PK plastic sleeve
- . PK healing cap
## Technological Characteristics Comparison
The purpose of this 510(k) is to expand the current product line to include two new abutments listed above. No substantive changes are being made to the indications as compared to the most recent clearance, and the indications are provided below. Aside for K112440, the indications differ only in respect to indications related to the specific components contained within each submission. K112440 has the same intended use as it allows for implantation in the same anatomic location and, when appropriate, for immediate loading. Comparison tables for each modification are provided below.
| K181381 | K162482 | K132125 | K112440 |
|-----------------------------|-----------------------------|-----------------------------|-----------------------------|
| A.B. DENTAL DEVICES® | A.B. DENTAL DEVICES® | A.B. DENTAL DEVICES® | The AB Dental Devices |
| Dental Implants System is | Dental Implants System is | Dental Implants System is | implants are intended for |
| indicated for use in | indicated for use in | indicated for use in | surgical' placement in the |
| surgical and restorative | surgical and restorative | surgical and restorative | maxillary mandibular arch |
| applications for placement | applications for placement | applications for placement | to support crowns, bridges, |
| in the bone of the upper or | in the bone of the upper or | in the bone of the upper or | or overdentures in |
| lower jaw to provide | lower jaw to provide | lower jaw to provide | edentulous or partially |
| support for prosthetic | support for prosthetic | support for prosthetic | and/or edentulous patients. |
| devices, such as artificial | devices, such as artificial | devices, such as artificial | I7 Integral implant, I5 |
| teeth, in order to restore | teeth, in order to restore | teeth, in order to restore | Conical implant, P12-T,L |
| the patient's chewing | the patient's chewing | the patient's chewing | Temporary flat connection |
| function. A.B. DENTAL | function. A.B. DENTAL | function. A.B. DENTAL | abutment, connection |
| DEVICES® Dental | DEVICES® Dental | DEVICES® Dental | abutment, and P16 |
| Implants System is | Implants System is | Implants System is | Straight adaptor are |
| indicated also for | indicated also for | indicated also for | appropriate for immediate |
| immediate loading when | immediate loading when | immediate loading when | loading when good primary |
| good primary stability is | good primary stability is | good primary stability is | stability is achieved and |
| achieved and with | achieved and with | achieved and with | with appropriate occlusal |
| appropriate occlusal | appropriate occlusal | appropriate occlusal | loading. |
| loading. | loading. | loading. | |
| | Two Stage Implants: I22, | Two Stage Implants: | |
| | I5, I55, I10. | I2,I5,I6BI. | |
| | | One Stage: I6, I6b, I6B. | |
| | P4 and P14 angled | One Stage & One-Piece | |
| | abutments are to be used | 3.0 mm diameter implants: | |
| | only with standard platform | I6, I6B, I6BI, are intended | |
| | implants 3.5 mm in | for placement at the | |
| | diameter or larger. | mandibular central and | |
{5}------------------------------------------------
| | lateral incisors and<br>maxillary and lateral<br>incisors. Indicated also for<br>denture stabilization using<br>multiple implants.<br>One stage & One-Piece<br>2.4 mm diameter implants<br>for temporary use or long<br>term use: I6, I6b, permit<br>immediate splint stability<br>and long term fixation of<br>new or existing crown,<br>bridge and prosthesis.<br>P14 Angulated Abutment<br>Adapter is to be used with<br>implant diameter 4.2mm<br>and higher. |
|--|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
# 1. Healing Caps (P0)
| | Subject P0-3,(4,7) | Predicate P0-3 | Predicate P0-3.75 |
|-----------------------|------------------------------------------------|------------------------------------------------|-----------------------|
| K-number | K181381 | K132125 | K132125 |
| Material | Ti-6Al-4V ELI with gold<br>colored anodization | Ti-6Al-4V ELI with gold<br>colored anodization | Ti-6Al-4V ELI |
| Diameter, mm | 4.0 | 4.0 | 4.5 |
| Platform | Narrow | Narrow | Standard |
| Length, mm | 4, 7 | 2, 3, 5, 6 | 0.5 - 7 |
| Design | Standard | Standard | Standard |
| Allowed<br>angulation | No correction allowed | No correction allowed | No correction allowed |
| Sterility | Non-Sterile | Non-Sterile | Non-Sterile |
| | Subject P0N-3.75,(4,7) | Predicate P0N-3.75 | Predicate P0-3.75 |
|--------------------|------------------------|-----------------------|-----------------------|
| K-number | K181381 | K132125 | K132125 |
| Material | Ti-6Al-4V ELI | Ti-6Al-4V ELI | Ti-6Al-4V ELI |
| Diameter, mm | 3.75 | 3.75 | 4.5 |
| Platform | Standard | Standard | Standard |
| Length, mm | 4, 7 | 3, 5, 6 | 0.5 - 7 |
| Design | Standard | Standard | Standard |
| Allowed angulation | No correction allowed | No correction allowed | No correction allowed |
| Sterility | Non-Sterile | Non-Sterile | Non-Sterile |
# 2. Cement Retained, Straight Abutments (P3)
| | Subject P3-5,5 | Predicate P3-5,(7,9) |
|--------------------|-----------------------|-----------------------|
| K-number | K181381 | K162482 |
| Material | Ti-6Al-4V ELI | Ti-6Al-4V ELI |
| Diameter, mm | 5.0 | 5.0 |
| Platform | Standard | Standard |
| Length, mm | 5 | 7, 9 |
| Design | Anti -rotation | Anti-rotation |
| Allowed angulation | No correction allowed | No correction allowed |
{6}------------------------------------------------
| Sterility | Non-Sterile | Non-Sterile |
|-----------|-------------|-------------|
|-----------|-------------|-------------|
| | Subject P3S-3 | Predicate P3S-3.75 | Predicate P3-3 |
|-----------------------|------------------------------------------------|------------------------|------------------------------------------------|
| K-number | K181381 | K112440 | K112440 |
| Material | Ti-6Al-4V ELI with gold<br>colored anodization | Ti-6Al-4V ELI | Ti-6Al-4V ELI with gold<br>colored anodization |
| Diameter | 4.5 | 4.5 | 4.5 |
| Platform | Narrow | Standard | Narrow |
| Shoulder Height | 1, 2, 3 | 1, 2, 3 | |
| Design | Anatomic Anti-rotation | Anatomic Anti-rotation | Anti-rotation |
| Allowed<br>angulation | No correction allowed | No correction allowed | No correction allowed |
| Sterility | Non-Sterile | Non-Sterile | Non-Sterile |
| | Subject P3S-Peek-3 | Predicate P3S-Peek-3.75 |
|--------------------|-------------------------|-------------------------|
| K-number | K181381 | K132125 |
| Material | PEEK | PEEK |
| Diameter | 4.5 | 4.5 |
| Platform | Narrow | Standard |
| Shoulder Height | 1, 2, 3 | 1, 2, 3 |
| Design | Anatomic Anti -rotation | Anatomic Anti -rotation |
| Allowed angulation | No correction allowed | No correction allowed |
| Sterility | Non-Sterile | Non-Sterile |
| | Subject P3SW-3.75 | Predicate P3S-3.75 | Predicate P3W-3.75 |
|--------------------|--------------------------------|-------------------------|-----------------------|
| K-number | K181381 | K112440 | K112440 |
| Material | Ti-6Al-4V ELI | Ti-6Al-4V ELI | Ti-6Al-4V ELI |
| Diameter | 5.5 | 3.75 | 5.5 |
| Platform | Standard | Standard | Standard |
| Shoulder Height | 1, 2, 3 | 1, 2, 3 | 1, 2, 3 |
| Design | Anatomic Anti-rotation<br>Wide | Anatomic, Anti-rotation | Anti-rotation |
| Allowed angulation | No correction allowed | No correction allowed | No correction allowed |
| Sterility | Non-Sterile | Non-Sterile | Non-Sterile |
| | Subject P3W-3 | Predicate P3-3 | Predicate P3W-3.75 |
|----------|----------------------------------------------|----------------------------------------------|--------------------|
| K-number | K181381 | K112440 | K112440 |
| Material | Ti-6Al-4V ELI with gold color<br>anodization | Ti-6Al-4V ELI with gold color<br>anodization | Ti-6Al-4V ELI |
| Diameter | 5.5 | 3.75 | 5.5 |
| Platform | Narrow | Narrow | Standard |
| Length | 9 | 9, 12 | 9, 12 |
| Design | Anti-rotation | Anti-rotation | Anti-rotation |
{7}------------------------------------------------
| Allowed<br>angulation | No correction allowed | No correction allowed | No correction<br>allowed |
|-----------------------|-----------------------|-----------------------|--------------------------|
| Sterility | Non-Sterile | Non-Sterile | Non-Sterile |
# 3. Overdenture Ball Attachment Abutments (P5)
| | Subject P5-3,(4,5,6) | Predicate P5-3,(1,2,3) | Predicate P5-3.75 |
|--------------------|-------------------------------------------|-------------------------------------------|-----------------------|
| K-number | K181381 | K132125 | K132125 |
| Material | Ti-6Al-4V ELI with gold color anodization | Ti-6Al-4V ELI with gold color anodization | Ti-6Al-4V ELI |
| Platform | Narrow | Narrow | Standard |
| Length, mm | 4, 5, 6 | 1, 2, 3 | 1, 2, 3, 4, 5, 6 |
| Design | Ball Attachment | Ball Attachment | Ball Attachment |
| Allowed angulation | No correction allowed | No correction allowed | No correction allowed |
| Sterility | Non-Sterile | Non-Sterile | Non-Sterile |
| | Subject P5-3.75,20-(4,6) | Predicate P5-3.75,20-(1,2,3,5) |
|------------|--------------------------|--------------------------------|
| K-number | K181381 | K132125 |
| Material | Ti-6Al-4V ELI | Ti-6Al-4V ELI |
| Platform | Standard | Standard |
| Length, mm | 4, 6 | 1, 2, 3, 5 |
| Angle, ° | 20 | 20 |
| Design | Ball Attachment | Ball Attachment |
| Sterility | Non-Sterile | Non-Sterile |
# 4. Composed (UCLA) Abutments
| | Subject P9R-3.75,11, P9HR-3.75,11 | Predicate P9-3.75,11, P9H-3.75,11 | |
|----------------------------|---------------------------------------------|-----------------------------------|-----------------------|
| K-number | K181381 | K132125 | |
| Material | FWM 1058 Cobalt Chrome Alloy per ASTM F1058 | Ti-6Al-4V ELI | |
| Connection Method/Platform | Standard | Standard | |
| Diameter, mm | 3.75 | 3.75 | |
| Allowed angulation | No correction allowed | No correction allowed | |
| Length, mm | 11 | 11 | |
| Design | Composed Abutment | Composed Abutment | |
| Sterility | Non-Sterile | Non-Sterile | |
| | Subject P9HG-3.75,11 | Predicate P9G-3.75,11 | Predicate P9H-3.75,11 |
| K-number | K181381 | K132125 | K132125 |
| Material | Gold Alloy 6019 | Gold Alloy 6019 | Ti-6Al-4V |
| \Platform | Standard | Standard | Standard |
| Diameter, mm | 3.75 | 3.75 | 3.75 |
| Allowed | No correction allowed | No correction allowed | No correction allowed |
{8}------------------------------------------------
| angulation | | | |
|------------|--------------------------|------------------------------------|-----------------------|
| Length, mm | 11 | 11 | 11 |
| Design | Composed Hex<br>Abutment | Composed Free Rotating<br>Abutment | Composed Hex Abutment |
| Sterility | Non-Sterile | Non-Sterile | Non-Sterile |
# 5. A.B. LOC Abutments (P25)
| | Subject P25-3.75,(0,1,2) | Predicate P25-3.75,(3,4,5,6,7) |
|--------------------|----------------------------------------------|--------------------------------|
| K-number | K181381 | K132125 |
| Material | Ti-6Al-4V ELI | Ti-6Al-4V ELI |
| Platform | Standard | Standard |
| Diameter, mm | 3.75 | 3.75 |
| Allowed angulation | No correction allowed | No correction allowed |
| Length, mm | 0, 1, 2 | 3, 4, 5, 6, 7 |
| Design | Integrated screw, attachment | Integrated screw, attachment |
| Sterility | Non-Sterile | Non-Sterile |
| | Subject P25-3 | Predicate P25-3.75 |
| K-number | K181381 | K132125 |
| Material | Ti-6Al-4V ELI with gold color<br>anodization | Ti-6Al-4V ELI |
| Platform | Narrow | Standard |
| Diameter, mm | 3 | 3.75 |
| Allowed angulation | No correction allowed | No correction allowed |
| Length, mm | 0, 1, 2, 3, 4, 5 | 3,4,5,6,7 |
| Design | Integrated screw, attachment | Integrated screw, attachment |
| Sterility | Non-Sterile | Non-Sterile |
# 6. P64 Straight Adaptor
| | Subject P64-3.75 | Predicate P16-3.75 |
|-----------------------|----------------------------------|----------------------------------|
| K-number | K181381 | K112440 |
| Material | Ti-6Al-4V ELI | Ti-6Al-4V ELI |
| Platform | Standard | Standard |
| Allowed<br>angulation | No correction allowed | No correction allowed |
| Height | 1, 2, 3, 4, 5 | 1, 2, 3, 4, 5 |
| Design | Integrated screw, screw retained | Integrated screw, screw retained |
| Sterility | Non-Sterile | Non-Sterile |
| | Subject P64b | Predicate P14b |
|--------------------------------|------------------------------------|------------------------------------|
| K-number | K181381 | K112440 |
| Material | Delrin | Delrin |
| Connection<br>Method/ Platform | Taper fit+ screw /<br>P64 platform | Taper fit+ screw /<br>P16 platform |
| Allowed angulation | No correction allowed | No correction allowed |
| Length, mm | 10 | 10 |
{9}------------------------------------------------
| Design | Plastic Sleeve for Adaptor | Plastic Sleeve for Adaptor |
|-----------|----------------------------|----------------------------|
| Sterility | Non-Sterile | Non-Sterile |
| | Subject P64-bT | Predicate P14-bT |
|--------------------|-----------------------------|-----------------------------|
| K-number | K181381 | K132125 |
| Material | Ti-6Al-4V ELI | Ti-6Al-4V ELI |
| Connection | Taper fit+ screw / | Taper fit+ screw / |
| Method/ Platform | P64 platform | P16 platform |
| Allowed anqulation | No correction allowed | No correction allowed |
| Length, mm | 12 | 12 |
| Design | Titanium Sleeve for Adaptor | Titanium Sleeve for Adaptor |
| Sterility | Non-Sterile | Non-Sterile |
| | Subject P0-P64,5 | Predicate P0-P14 |
|--------------------|-----------------------|-----------------------|
| K-number | K181381 | K132125 |
| Material | Ti-6Al-4V ELI | Ti-6Al-4V ELI |
| Connection | P64 | P14/P16 |
| Method/Platform | | |
| Allowed angulation | No correction allowed | No correction allowed |
| Height | 5 | 2.5, 4, 5, 7 |
| Design | Temporary healing cap | Temporary healing cap |
| Sterility | Non-Sterile | Non-Sterile |
# 7. PK Abutment
| | Subject PK-P3-3.75 | Predicate P3-3.75 | Predicate P3S-<br>3.75,(1,2,3) |
|-------------------------------|------------------------|------------------------|---------------------------------|
| K-number | K181381 | K112440 | K112440 |
| Material | Ti-6Al-4V ELI | Ti-6Al-4V ELI | Ti-6Al-4V ELI |
| Connection<br>Method/Platform | Standard | Standard | Standard |
| Allowed<br>angulation | No correction allowed | No correction allowed | No correction allowed |
| Height/Length,<br>mm | 1, 2, 3, 4 | 5, 7, 9 | 1, 2, 3 |
| Design | Anti-Rotation Abutment | Anti Rotation Abutment | Anatomic Anti-Rotation Abutment |
| Sterility | Non-Sterile | Non-Sterile | Non-Sterile |
| | Subject PK-P0-3.75 | Predicate P0-P14 |
|--------------------|-----------------------|-----------------------|
| K-number | K181381 | K132125 |
| Material | Ti-6Al-4V ELI | Ti-6Al-4V ELI |
| Connection | PK | P14/P16 |
| Method/Platform | | |
| Allowed angulation | No correction allowed | No correction allowed |
| Height | N/A | 2.5, 4, 5, 7 |
| Design | Temporary healing cap | Temporary healing cap |
| Sterility | Non-Sterile | Non-Sterile |
{10}------------------------------------------------
| | Subject PK-P2-3.75 | Predicate P14b |
|--------------------|----------------------------------------|----------------------------|
| K-number | K181381 | K112440 |
| Material | Delrin | Delrin |
| Connection | Taper fit/ | Taper fit+ screw / |
| Method/ Platform | PK platform | P16 platform |
| Allowed angulation | No correction allowed | No correction allowed |
| Length, mm | 10 | 10 |
| Design | Plastic Sleeve for Straight<br>Adaptor | Plastic Sleeve for Adaptor |
| Sterility | Non-Sterile | Non-Sterile |
| | Subject PK-P2H-3.75 | Predicate P14b |
|--------------------------------|-------------------------------------------------|------------------------------------|
| K-number | K181381 | K112440 |
| Material | Delrin | Delrin |
| Connection Method/<br>Platform | Taper fit/<br>PK platform | Taper fit+ screw /<br>P16 platform |
| Allowed angulation | No correction allowed | No correction allowed |
| Length, mm | 10 | 10 |
| Design | Plastic Sleeve with Hex<br>for Straight Adaptor | Plastic Sleeve for<br>Adaptor |
| Sterility | Non-Sterile | Non-Sterile |
#### Performance Data
- Engineering analysis to determine if additional components constitute a new worst-case. ●
- . Biocompatibility testing per ISO 10993-1 per ISO 10993 as originally presented in K162482
- Cytotoxicity testing per ISO 10993-5 .
- Sterilization validation as per ANSI/AAMI/ISO 17665-1: 2006 as originally presented in . K162482
- Fatigue testing as per ISO 14801 as originally presented in K162482 ●
#### Substantial Equivalence
The A.B. Dental Devices® Dental Implants System has the same indications and similar technological characteristics and principles of operation as its predicate device. Performance data demonstrate that the A.B. Dental Devices® Dental Implants System is substantially equivalent.
#### Conclusions
The A.B. Dental Devices® Dental Implants System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.