K180609 · Rodo Medical, Inc. · NHA · May 23, 2018 · Dental
Device Facts
Record ID
K180609
Device Name
Rodo Smilekey
Applicant
Rodo Medical, Inc.
Product Code
NHA · Dental
Decision Date
May 23, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Rodo Smilekey is intended to be used with the Rodo Abutment System for removing prostheses compatible with Smileloc Sleeve by using induction heating.
Device Story
Rodo Smilekey is a handheld dental device used to remove prostheses from the Rodo Abutment System. It functions by delivering induction energy (alternating electromagnetic field) to the Smileloc Sleeve, which heats the sleeve to release the prosthesis. The device consists of a tip and a handle containing a rechargeable lithium battery and an electronic control board. The operator inserts only the tip into the patient's mouth; LED indicators on the handle signal when energy is being applied and when the unlocking procedure is complete. The device is used by dental professionals to facilitate the removal of dental restorations. By providing a controlled induction heating method, it allows for the safe release of the prosthesis while maintaining implant and abutment temperatures within established safety limits.
Clinical Evidence
No clinical data. Bench testing only. Testing included electrical safety (AAMI/ANSI ES60601-1), EMC (AAMI/ANSI/IEC 60601-1-2), biocompatibility (ISO 10993-1, ISO 10993-5), cleaning/disinfection validation, software verification/validation, battery testing (IEC 62133), usability (IEC 60601-1-6), and thermal safety (ASTM F2182-11). Thermal testing confirmed that implant/abutment temperatures during activation remain below established safety limits.
Technological Characteristics
Handheld device with tip and handle; rechargeable lithium-ion battery; induction heating principle (alternating electromagnetic field); LED status indicators; internal timer for energy regulation. Materials evaluated for biocompatibility per ISO 10993-1. Thermal safety validated per ASTM F2182-11. Software-controlled energy delivery.
Indications for Use
Indicated for use with the Rodo Abutment System to remove prostheses compatible with the Smileloc Sleeve via induction heating.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and white and includes the words "U.S. Food & Drug Administration."
Rodo Medical, Inc. % Randy Prebula Partner Hogan Lovells US LLP 555 Thirteen Street NW Washington, District of Columbia 20004 May 23, 2018
Re: K180609
Trade/Device Name: Rodo Smilekey Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: March 8, 2018 Received: March 8, 2018
Dear Randy Prebula:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mary S. Runner -S
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known) K180609
Device Name
Rodo Smilekey
Indications for Use (Describe)
The Rodo Smilekey is intended to be used with the Rodo Abutment System for removing prostheses compatible with Smileloc Sleeve by using induction heating.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY Rodo Medical, Inc.'s Rodo Smilekey (K180609)
#### Submitter
| Company: | Rodo Medical, Inc. |
|-----------------|------------------------------------------------------------|
| Address: | 6399 San Ignacio Avenue<br>Suite 100<br>San Jose, CA 95119 |
| Phone: | 408-245-7636 |
| Facsimile: | 408-338-6940 |
| Contact Person: | Michael Parsons |
| Date Prepared: | May 10, 2018 |
## Subject Device
| Name of Device: | Rodo Smilekey |
|----------------------|------------------------------------|
| Classification Name: | Endosseous Dental Implant Abutment |
| Regulation: | 21 C.F.R. § 872.3630 |
| Regulatory Class: | Class II |
| Product Code: | NHA |
## Predicate Device
| Company: | Rodo Medical, Inc. |
|----------------|----------------------|
| Device Name: | Rodo Abutment System |
| 510(k) Number: | K160786 |
| Regulation: | 21 C.F.R. § 872.3630 |
| Product Codes: | NHA |
## Device Description
The Smilekey is an additional remover device intended for use with the predicate Rodo Abutment System (K160786). It is used to deliver induction energy to the Smileloc Sleeve. The Smilekey consists of a tip and handle that uses induction (alternating electromagnetic field) to deliver precise energy to the Smileloc. Only the end of the tip is inserted inside the patient's mouth while the handle is hand-held by the operator. The Smilekey is powered by a rechargeable lithium battery inside the handle. The handle also contains an electronic control board that regulates the entire operation, including LED indicators that let the operator know when energy is being applied to the Smileloc Sleeve and when the unlocking procedure has been completed.
{4}------------------------------------------------
## Intended Use / Indications for Use
The Rodo Smilekey is intended to be used with the Rodo Abutment System for removing prostheses compatible with Smileloc Sleeve by using induction heating.
#### Performance Data
The following bench testing was conducted to demonstrate the performance of the Smilekey:
- Electrical safety testing according to AAMI/ANSI ES60601-1:2005/(R)2012 And A1:2012, ● C1:2009/(R)2012 And A2:2010/(R)2012
- . EMC testing according to AAMI/ANSI/IEC 60601-1-2:2014
- Biocompatibility evaluation of the patient contacting materials .
- Identical Handle material to predicate Smileloc Activator (K160786) O
- FDA guidance document titled, "Use of International Standard ISO 10993-1, O 'Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process", issued on June 16, 2016
- o Cytotoxicity testing of the tip housing material according to ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- . Cleaning and disinfection testing with simulated challenge soiling and organism
- FDA guidance document titled, "Reprocessing Medical Devices in Health Care O Settings: Validation Methods and Labeling", issued on May 2, 2011
- AAMI TIR12:2010 Designing, testing, and labeling reusable medical devices for O reprocessing in health care facilities: A quide for medical device manufacturers
- Software verification and validation testing .
- FDA quidance document titled, "Content of Premarket Submission for Software O Contained in Medical Devices", issued on May 11, 2005
- Battery testing according to IEC 62133:2012 .
- Usability testing according to IEC 60601-1-6 Edition 3.1 2013
- . Thermal safety testing to evaluate the temperature rise on implant, abutment and restoration during the unlocking procedure using the Smilekey induction activator, to ensure that temperatures do not exceed established safety limits.
- ASTM F2182-11 testing guidelines were used which included gel media and O simulated use conditions.
- The thermal safety limits were equivalent to the predicate Smileloc Activator O (K160786).
- The results demonstrated that the temperatures were within safe limits. O
## Substantial Equivalence
The proposed Rodo Smilekey and the predicate Smileloc Activator (part of the Rodo Abutment System K160786) have a similar intended use and similar indications of removing the Smileloc Sleeve from the Rodo Abutment. Both devices are intended to activate the Smileloc Sleeve to release a compatible prosthesis with the Rodo Abutment. Although the subject device uses induction heating instead of resistance heating, this difference in technological characteristics does not raise different questions of safety or effectiveness. Both devices deliver energy to increase the temperature of the Smileloc Sleeve. The induction energy is within safe levels for
{5}------------------------------------------------
the oral application. In addition, the performance of the Rodo Abutment System and the thermal safety was verified and validated using the same scientific methodology as the predicate device. In addition, bench testing of the subject device has demonstrated substantial equivalent performance to the predicate device.
# Conclusions
The Smilekey is an additional remover device intended for use with the predicate Rodo Abutment System (K160786). Similar to the predicate Smileloc Activator, the Smilekey uses heat to activate and to remove the Smileloc Sleeve. Although the Smilekey has a different heating method when compared to the Smileloc Activator, this difference in technological characteristics does not raise different questions of safety or effectiveness. Performance bench testing of the Smilekey demonstrated equivalent performance as the predicate device. Thus, the Smilekey can be found substantially equivalent to the predicate device.
| | Subject Device<br>Rodo Smilekey | Predicate Device<br>Smileloc Activator of Rodo<br>Abutment System (K160786) | Comparison |
|----------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The Rodo Smilekey is intended to<br>be used with the Rodo Abutment<br>System for removing prostheses<br>compatible with Smileloc Sleeve<br>by using induction heating. | Rodo Abutment System is<br>intended to be used in<br>conjunction with compatible<br>implant systems in the maxillary<br>or mandibular arch to provide<br>support for crowns, bridges or<br>overdentures. | The Indications for Use of<br>the subject device is derived<br>from the labeling of the<br>Smileloc Activator<br>component of the<br>overarching Rodo Abutment<br>System. |
| Heating Method | Induction heating | Resistance heating | Difference in heating<br>method does not raise<br>different questions of safety<br>or effectiveness |
| Type of battery<br>charge | Lithium Ion | Lithium Ion | Same |
| Is battery<br>rechargeable? | Yes | No | Battery rechargeability does<br>not raise additional<br>questions of safety or<br>effectiveness |
| Time needed for<br>tip on crown to<br>activate | 5 seconds for acrylic restorations,<br>9 seconds for non-acrylic<br>restorations (e.g., lithium<br>disilicate, zirconia) | 15 seconds | Difference in activation time<br>does not raise different<br>questions of safety or<br>effectiveness. Maximum<br>implant and abutment<br>temperatures reached<br>during activation remain<br>below safety limits. |
| Maximum<br>implant/abutment<br>temperature<br>reached during<br>activation | Maximum implant and abutment<br>temperatures during activation<br>remain below safety limits<br>established by AAMI/ANSI<br>ES60601-1:2005/® 2012 and<br>A1:2012, c1:2009/® 2012 and<br>a2:2010/® 2012 and Eriksson, AR<br>et al. | Maximum implant and abutment<br>temperatures during activation<br>remain below safety limits<br>established by AAMI/ANSI<br>ES60601-1:2005/® 2012 and<br>A1:2012, c1:2009/® 2012 and<br>a2:2010/® 2012 and Eriksson, AR<br>et al. | Same |
# Table 1: Comparison of Subject and Predicate Devices
{6}------------------------------------------------
| | Subject Device<br>Rodo Smilekey | Predicate Device<br>Smileloc Activator of Rodo<br>Abutment System (K160786) | Comparison |
|-------------------------------------------------------------------|-----------------------------------|-----------------------------------------------------------------------------|------------|
| Type of<br>indicators to<br>inform the end<br>user | LED | LED | Same |
| Regulation of<br>energy delivery | Internal timer | Internal timer | Same |
| Type of barrier<br>employed for<br>infection control<br>of device | FDA-cleared dental barrier sleeve | FDA-cleared dental barrier sleeve | Same |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.