Zeramex(R) XT Dental Implant System, NobelPearl Dental Implant System
K180184 · Dentalpoint AG · NHA · Nov 16, 2018 · Dental
Device Facts
Record ID
K180184
Device Name
Zeramex(R) XT Dental Implant System, NobelPearl Dental Implant System
Applicant
Dentalpoint AG
Product Code
NHA · Dental
Decision Date
Nov 16, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
The ZERAMEX® XT Dental Implant System is intended to be surgically placed in the bone of the upper and lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore aesthetics and chewing function. The ZERAMEX® XT Dental Implant System can be used for single or multiple unit restorations. The ZERAMEX® XT Dental Implants are indicated for delayed loading. The ZERAMEX® XT dental implants are specially indicated for patients with metal allergies/intolerances and chronic illnesses due to metal allergies/ intolerances.
Device Story
The Zeramex® XT and NobelPearl Dental Implant Systems are two-piece endosseous dental implants designed for surgical placement in the jawbone to support prosthetic restorations. The system includes implants, abutments, healing caps, and gingivaformers. Implants feature a 'bolt-in tube' internal connection with four retention elements for anti-rotation. Components are manufactured from aluminum-toughened zirconia; abutments are secured to implants using carbon fiber-reinforced PEEK-Optima™ Ultra VICARBO® screws, creating a ceramic-on-ceramic connection. The device is used by dental professionals in a clinical setting. The system provides a metal-free alternative for patients with metal allergies or intolerances. Clinical benefit is derived from restoring aesthetics and chewing function through stable prosthetic support.
Clinical Evidence
Bench testing only. No clinical data provided. Performance was validated through fatigue testing per ISO 14801:2007. Biocompatibility, sterilization, and surface characteristics were leveraged from predicate devices (Zeramex® T and P6) due to identical materials and surface treatments.
Technological Characteristics
Two-piece endosseous dental implant system. Materials: Aluminum-toughened zirconia (ISO 6474-1:2010, ISO 6872:2015) and carbon fiber-reinforced PEEK-Optima™ Ultra VICARBO®. Connection: 'Bolt-in tube' internal connection with four retention elements. Sterilization: Moist heat (ISO 17665-1, ISO/TS 17665-2). Dimensions: Diameters 4.2mm/5.5mm; lengths 8mm-14mm. Abutments: Straight and 15° angled.
Indications for Use
Indicated for patients requiring dental implants for single or multiple unit restorations in the upper or lower jaw to support prosthetic devices. Specifically indicated for patients with metal allergies/intolerances and associated chronic illnesses. Implants are indicated for delayed loading.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a circular seal with an eagle emblem and text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA." To the right, there is a blue square with the letters "FDA" in white. Next to the square, the text "U.S. FOOD & DRUG ADMINISTRATION" is written in blue.
November 16, 2018
Dentalpoint AG % Elizabeth Rose Manager, Medical Devices Mapi USA, Inc. 2343 Alexandria Drive Suite 100 Lexington, Kentucky 40504
Re: K180184
Trade/Device Name: Zeramex® XT Dental Implant System, NobelPearl Dental Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: October 18, 2018 Received: October 19, 2018
Dear Elizabeth Rose:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm_identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Andrew I. Steen -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180184
Device Name Zeramex® XT Dental Implant System
### Indications for Use (Describe)
The ZERAMEX® XT Dental Implant System is intended to be surgically placed in the bone of the upper and lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore aesthetics and chewing function. The ZERAMEX® XT Dental Implant System can be used for single or multiple unit restorations. The ZERAMEX® XT Dental Implants are indicated for delayed loading. The ZERAMEX® XT dental implants are specially indicated for patients with metal allergies/intolerances and chronic illnesses due to metal allergies/ intolerances.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CER 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180184
Device Name
NobelPearl Dental Implant System
Indications for Use (Describe)
The NobelPearl Dental Implant System is intended to be surgically placed in the bone of the upper and lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore aesthetics and chewing function. The NobelPearl Dental Implant System can be used for single or multiple unit restorations.
NobelPearl Dental Implant System are intended for delayed loading.
NobelPearl Dental Implant System are specially indicated for patients with metal allergies/intolerances and chronic illness due to metal allergies/intolerances.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:14px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:14px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 3.0 510(k) Summary
#### I. Submitter
Dentalpoint AG Bodenäckerstrasse 5 Spreitenbach, 8957 Switzerland
Phone: (+41) 44 542 41 71 Fax: (+41) 44 388 36 39
Contact Person: Anja Lieberherr, Quality Management Date Prepared: November 16, 2018
#### II. Device
| Device Proprietary Name: | Zeramex® XT Dental Implant System and NobelPearl Dental<br>Implant System |
|--------------------------|---------------------------------------------------------------------------|
| Common or Usual Name: | Dental Implant System |
| Classification Name: | Endosseous Dental Implant |
| Regulation Number: | 21 CFR 872.3640 |
| Primary Product Code: | DZE |
| Secondary Product Code | NHA |
| Device Classification | II |
#### III. Predicate Device
Substantial equivalence is claimed to the following primary predicate devices:
- Zeramex® T Dental Implant System, K133255, Dentalpoint AG
The following devices are referenced within the submission:
- Zeramex® P6 Dental Implant System, K163043, Dentalpoint AG
- Zeramex® P6 Dental Implant System, K152385, Dentalpoint AG
#### IV. Device Description
The Zeramex® XT Dental Implant System is an endosseous dental implant/abutment system including various sizes of endosseous two piece dental implants, abutments, and accessories. The Zeramex® XT implants may be restored with screw retained Zeramex® XT abutments. The Zeramex® XT implants are placed using the Zeramex® XT surgical tools.
{5}------------------------------------------------
The NobelPearl Dental Implant System is an endosseous dental implant/abutment system including various sizes of endosseous two piece dental implants, abutments, and accessories. The NobelPearl Dental Implant System may be restored with screw retained NobelPearl abutments. The NobelPearl Dental Implant System are placed using the NobelPearl surgical tools.
The implants, abutments, and surgical tools for the Zeramex® XT and NobelPearl Dental Implant Systems are exactly the same; two trade names are being used for marketing purposes.
The implants, produced from aluminum toughened zirconia (conforming to ISO 6474-1:2010 Implants for Surgery - Ceramic Materials and ISO 6872:2015 Dentistry - Ceramic Materials), are provided sterile in two (2) diameters (0 4.2 mm (RB), and 5.5 mm (WB)) and three (3) lengths (8 mm, 10 mm, and 12 mm). The Ø 4.2 mm implant is also provided in 14 mm. The implants are designed with a "bolt-in tube" internal connection which provide anti-rotational features.
Straight and angular (15°) screw retained abutments, provided non-sterile in two sizes (RB; Ø 5.0 mm and SB, Ø 6.0 mm), are compatible for use with the system implants. The straight and angular abutments are made from the same zirconium materials as the system implants.
The screw retained abutments fit within four (4) retention elements and are affixed to the implant with a carbon fiber reinforced PEEK-Optima™ Ultra VICARBO® screw (straight) which fits the internal thread of the implant and provides a secure, screw retained ceramic on ceramic connection.
Healing caps and gingivaformers are also provided in the system. These components, provided in two sizes, are manufactured from PEEK and are connected to the implant using a screw. Provisional restoration components may be used to support temporary crowns after the healing period. The provisional restorations and the provisional restoration screws are made of carbon fiber reinforced PEEK.
#### V. Indications for Use
## Zeramex® XT Dental Implant System
The ZERAMEX® Dental Implant System is intended to be surgically placed in the bone of the upper and lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore aesthetics and chewing function. The ZERAMEX® Dental Implant System can be used for single or multiple unit restorations. The ZERAMEX® Dental Implants are indicated for delayed loading. The ZERAMEX® dental implants are specially indicated for patients with metal allergies/intolerances and chronic illnesses due to metal allergies/intolerances.
## NobelPearl Dental Implant System
{6}------------------------------------------------
The NobelPearl Dental Implant System is intended to be surgically placed in the bone of the upper and lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore aesthetics and chewing function. The NobelPearl Dental Implant System can be used for single or multiple unit restorations. NobelPearl Dental Implant System are intended for delayed loading. NobelPearl Dental Implant System are specially indicated for patients with metal allergies/intolerances and chronic illness due to metal allergies/intolerances.
#### VI. Comparison of Technological Characteristics
The Zeramex® XT and NobelPearl Dental Implant Systems and the predicate devices share the following characteristics:
- two-piece design (K133255 and K163043); ●
- implant and abutment material of construction (K163043);
- screw material of construction (K133255 and K163043);
- implant lengths (K133255 and K163043);
- implant diameters (K133255);
- abutment angulation (K133255 and K163043);
- screw retained implant/abutment connection (K163043);
- provision of carbon-fiber reinforced PEEK screws (K163043); and
- implant and abutment sterilization methods. ●
The Zeramex® XT and NobelPearl Dental Implant Systems are technologically different from the predicate devices with respect to the anti-rotational feature. The subject devices feature a "Bolt-In Tube" connection with four (4) retention elements, while the predicate devices have internal cylindrical connections. The Zeramex® P6 Dental Implant System also has an external hexagonal anti-rotational feature.
| | Zeramex® XT and NobelPearl<br>Dental Implant Systems | Zeramex® T Dental Implant<br>System (K133255) | Zeramex® P6 Dental Implant<br>System (K163043) |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Dentalpoint AG | Dentalpoint AG | Dentalpoint AG |
| Indications for<br>Use | The ZERAMEX® Dental<br>Implant System is intended to<br>be surgically placed in the bone<br>of the upper and lower jaw to<br>provide support for prosthetic<br>devices, such as artificial teeth,<br>in order to restore aesthetics<br>and chewing function. The<br>ZERAMEX® Dental Implant<br>System can be used for single<br>or multiple unit restorations. | The Zeramex® T Dental<br>Implant System is intended to<br>be surgically placed in the bone<br>of the upper and lower jaw to<br>provide support for prosthetic<br>devices, such as artificial teeth,<br>in order to restore chewing<br>function.<br>The Zeramex® T Dental<br>Implant System can be used for | The Zeramex® P6 Dental<br>Implant System is intended to<br>be surgically placed in the bone<br>of the upper and lower jaw to<br>provide support for prosthetic<br>devices, such as artificial teeth,<br>in order to restore chewing<br>function.<br>The Zeramex® P6 Dental<br>Implant System can be used for |
{7}------------------------------------------------
| The ZERAMEX® Dental<br>Implants are indicated for<br>delayed loading. The<br>ZERAMEX® dental implants<br>are specially indicated for<br>patients with metal<br>allergies/intolerances and<br>chronic illnesses due to metal<br>allergies/intolerances.<br><br>The NobelPearl Dental Implant<br>System is intended to be<br>surgically placed in the bone of<br>the upper and lower jaw to<br>provide support for prosthetic<br>devices, such as artificial teeth,<br>in order to restore aesthetics<br>and chewing function. The<br>NobelPearl Dental Implant<br>System can be used for single<br>or multiple unit restorations.<br>NobelPearl Dental Implant<br>System are intended for delayed<br>loading. NobelPearl Dental<br>Implant System are specially<br>indicated for patients with<br>metal allergies/intolerances and<br>chronic illness due to metal<br>allergies/intolerances. | single or multiple unit<br>restorations.<br>Zeramex® T implants are<br>intended for delayed loading.<br>The Zeramex® T dental<br>implants are specially indicated<br>for patients with metal allergies<br>and chronic illnesses due to<br>metal allergies.<br>The Zeramex® T (Ø 3.5 mm)<br>implant may only be used in the<br>anterior teeth in the lower jaw<br>and lateral teeth in the upper<br>jaw. | single or multiple unit<br>restorations.<br>The Zeramex® P6 implants are<br>intended for delayed loading.<br>The Zeramex® P6 implants are<br>specially indicated for patients<br>with metal allergies/<br>intolerances and chronic illness<br>due to metal<br>allergies/intolerances.<br>The Zeramex® P6 (Ø3.3mm<br>SN) implant may only be used<br>in the anterior teeth in the lower<br>jaw and lateral incisor in the<br>upper jaw. |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
## Discussion
The technological difference between the subject and predicate devices does not raise new questions. As seen above, the Zeramex® XT and NobelPearl Dental Implant Systems have the same indications for use, materials of construction, design, size, surface treatment, and sterilization methods as the predicate devices. The data within this submission support that the subject devices are substantially equivalent to the identified predicate devices.
#### VII. Performance Data
As the subject devices are a line extension to the Zeramex® Implant Systems, verification activities, as identified through risk analysis, were conducted to written protocols with predefined acceptance criteria.
{8}------------------------------------------------
Fatigue testing in accordance with ISO 14801 Second Edition 2007-11-15 Dentistry-Implants-Dynamic Fatigue Test for Endosseous Dental Implants was conducted on the subject devices, Zeramex® XT and NobelPearl Dental Implant Systems. The fatigue testing as conducted in saline because of the ceramic/polymer device composition.
The following performance tests, conducted on the predicate devices, are being leveraged in support of this submission:
- shelf-life, packaging, and transport validation studies:
- cleaning and sterilization validation; ●
- o ISO 17665-1 First Edition 2006-08-15 Sterilization of Health Care Products -Moist Heat - Part 1: Requirements For The Development, Validation And Routine Control Of A Sterilization Process For Medical Devices;
- ISO/TS 17665-2:2009-01-15 Sterilization Of Health Care Products Moist Heat o - Part 2: Guidance On The Application Of ISO 17665-1;
- ISO 11737-1 Second Edition 2006-04-01 Sterilization Of Medical Devices - O Microbiological Methods - Part 1: Determination Of A Population Of Microorganisms On Products [Including: Technical Corrigendum 1 (2007)]; and
- ISO 11737-2 Second Edition 2009-11-15 Sterilization Of Medical Devices o Microbiological Methods - Part 2: Tests Of Sterility Performed In The Definition, Validation and Maintenance Of A Sterilization Process.
- biocompatibility; ●
- ISO 6474-1 First Edition 2010-02-15 Implants for Surgery Ceramic Materials o Part 1: Ceramic Materials Based on High Purity Alumina, and
- o ISO 6872 Fourth Edition 2015-06-01 Dentistry Ceramic Materials.
- bacterial endotoxin testing (LAL Method); ●
- analysis of surface composition; and
- surface roughness and contact angle testing.
## Clinical Studies
Data supporting the Zeramex® T Dental Implant System and Zeramex® P6 Dental Implant System are applicable to the subject devices as the products are made from the same materials, undergo the same surface treatments, and have the same outer design. Fatigue testing supports that the subject devices can withstand clinically relevant forces.
## VIII. Conclusion
Based on the above information, the Zeramex® XT and NobelPearl Implant Systems can be considered substantially equivalent to the identified predicate devices.
{9}------------------------------------------------
Although minor differences in design and technology exist between the subject and predicate devices, substantial equivalence has been demonstrated through a comparison of intended use, design, technological characteristics, and performance evaluations. Therefore, the Zeramex® XT and NobelPearl Dental Implant Systems are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.