The Southern Implants PEEK Abutments are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for provisional use up to 180 days as an aid in prosthetic rehabilitation.
Device Story
Southern Implants PEEK Abutments are premanufactured prosthetic components for endosseous dental implants; used as healing or temporary abutments. Healing abutments maintain tissue openings during prosthesis fabrication; temporary abutments aid in manufacturing temporary prosthetic rehabilitations. Devices are screw-retained; available in various engaging/non-engaging connections to suit Southern Implants systems. Used by dental clinicians in clinical settings. Output is a physical interface for prosthetic attachment; facilitates tissue healing and temporary restoration. Benefits include provisional support during rehabilitation period. Manufactured from white PEEK; supplied sterile.
Clinical Evidence
No clinical studies were conducted. Evidence is based on bench testing, including biocompatibility (ISO 10993-1, ISO 10993-5), sterilization validation (ISO 11137, ISO 17665), packaging validation (ISO 11607), and shelf-life testing (ASTM F1980).
Technological Characteristics
Material: Polyetheretherketone (PEEK) per ASTM F2026. Components: Healing and temporary abutments with titanium retaining screws. Connection: Various engaging/non-engaging types for Southern Implants systems. Sterilization: Gamma radiation (device) and steam sterilization (user).
Indications for Use
Indicated for patients requiring provisional prosthetic rehabilitation via endosseous dental implants for up to 180 days.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized eagle. To the right of that is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with "ADMINISTRATION" underneath.
Southern Implants (Pty) Ltd % Kevin Walls Consultant Regulatory Insight, Inc. 33 Golden Eagle Lane Littleton, Colorado 80127
February 9, 2018
Re: K172160
Trade/Device Name: Southern Implants PEEK Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: January 5, 2018 Received: January 8, 2018
Dear Kevin Walls:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mary S. Runner -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known)
### K172160
Device Name Southern Implants PEEK Abutments
Indications for Use (Describe)
The Southern Implants PEEK Abutments are premanufactured prosthetic components directly connected to endosseous dental implants implants and are intended for provisional use up to 180 days as an aid in prosthetic rehabilitation.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
PSC Publishing Services (301) 443-6740 EF
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
# 510(k) Summary
## Sponsor:
| Company Name: | Southern Implants (Pty) Ltd. |
|------------------|-------------------------------------------|
| Company Address: | 1 Albert Road<br>Irene, South Africa 0062 |
| Telephone: | + 27 12 667 1046 |
| Fax: | + 27 72 313 5715 |
| Contact Person: | Lauranda G. Breytenbach |
| Summary Preparation Date: | February 8, 2018 |
|---------------------------|---------------------------------------|
| Device Name: | |
| Trade Name: | Southern Implants PEEK Abutments |
| Common/Usual Name: | PEEK Abutments |
| Classification Name: | Abutment, Implant, Dental, Endosseous |
| Regulation Number: | 21 CFR 872.3630 |
| Product Codes: | NHA |
| Device Class: | Class II |
Predicate Devices:
| Predicate # | Manufacturer | Device Name | K Number |
|----------------------|------------------|-----------------------------|----------|
| Primary<br>Predicate | Nobel Biocare AB | NobelProcera PEEK Abutments | K120954 |
{4}------------------------------------------------
### Device Description:
The Southern Implants PEEK abutments are endosseous dental implant abutments. The PEEK abutments are available in a variety of preformed and customized forms. The PEEK abutments are available as healing abutments or temporary abutments.
The healing abutments are attached onto the implant during phase 2 surgery to maintain the tissue opening while the prosthesis is being fabricated. Temporary abutments are used as an aid in manufacturing a temporary prosthetic rehabilitation. They can either be used for direct connection to an endosseous implant or they can be used for connecting the prosthesis to a compact conical abutments. Temporary abutments are available in a 2 mm collar direct to implant and 1mm collar on abutment level. The temporary abutments for the compact conical abutments are for multi-unit use only.
These abutments are made from white PEEK (ASTM F2026).
The PEEK abutments are premanufactured and are available in a variety of connections, engaging and non-engaging, to suit the implant systems manufactured by Southern Implants.
Titanium retaining screws are also available to use with the PEEK abutments.
#### Indications for Use
The Southern Implants PEEK Abutments are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for provisional use up to 180 days as an aid in prosthetic rehabilitation.
#### Equivalence to Predicates
Southern Implants (Pty) Ltd submits the information in this Premarket Notification to demonstrate that, for the purposes of FDA's regulation of medical devices, the subject device is substantially equivalent in indications and design principles to the following legally marketed predicate devices:
K120954, Nobel Procera PEEK Abutments
Tables comparing the subject device and primary predicate devices is provided below.
| Device and Manufacturer | New device | Primary Predicate<br>(K120954) |
|-------------------------|----------------------------------------------------------------------------------------------------------------------|--------------------------------|
| | Southern Implants PEEK abutments | NobelProcera PEEK Abutments |
| Product Code | PK-MAX7-9; PK-MAX7-11;<br>PK-MAX8-11; PK-L-MAX7-9;<br>PK-L-MAX7-11; PK-L-MAX8-9;<br>PK-L-MAX8-11; PKA-Z6-9; PKAZ8-11 | 37819; 37820 |
{5}------------------------------------------------
| Indications for Use | The Southern Implants PEEK<br>Abutments are premanufactured<br>prosthetic components directly<br>connected to endosseous dental<br>implants and are intended for<br>provisional use up to 180 days as an<br>aid in prosthetic rehabilitation. | The Nobel Biocare PEEK Abutments<br>are premanufactured prosthetic<br>components directly connected to<br>endosseous dental implants and are<br>intended for provisional use up to<br>180 days as an aid in prosthetic<br>rehabilitation. |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Healing Abutment | Healing Abutment |
| Material | Polyetheretherketone (PEEK) | Polyetheretherketone (PEEK) |
| Colour | White | Natural and White |
| Attachment Method | Screw retained | Screw retained |
| Duration of use | 180 days, single use | 180 days, single use |
| Implant/Abutment<br>Connection | Southern Implants MAX<br>● External Hex<br>● Tri-Nex<br>● Internal Hex | Nobel Biocare<br>● Internal Conical |
| Height (mm) | 5 – 6 mm | 5 mm |
| Diameter (mm) | 6 × 9<br>8 × 11 | 6 x 7<br>7 × 8 |
| Abutment Profile | Various anatomical shapes intended<br>for different tooth locations | Various anatomical shapes intended<br>for different tooth locations or<br>individually designed with Procera<br>system |
| Sterility | Sterile | Sterile |
| Device and Manufacturer | New device | Primary Predicate<br>(K120954) |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | Southern Implants PEEK abutments<br>PKPN2H; PKBN2NH;<br>PKB2H; PKB2NH; PKBA2H; PKBA2NH;<br>PKBBB2H; PKBBB2NH;<br>PKMAX9-2H; PKMAX9-2NH;<br>PKC-EL-35-2; PKC-NL-35-2;<br>PKC-EL-43-2; PKC-NL-43-2;<br>PKC-EL-50-2; PKC-NL-50-2;<br>PKC-EL-60-2; PKC-NL-60-2;<br>PKC-DC3-2; PKC-NDC3-2;<br>PKC-DC4-2; PKC-NDC4-2;<br>PKC-DC5-2; PKC-NDC5-2;<br>ITS6-PKC1; ITS-PKC1;<br>PKC-MC; PKC-MCW. | NobelProcera PEEK Abutments<br>30259, 30260, 30261, 30262, 30256,<br>30257, 30258, 31350, 31351, 31352,<br>31353, 1048-0, 1050-0, 1052-0. |
| Indications for Use | The Southern Implants PEEK<br>Abutments are premanufactured<br>prosthetic components directly<br>connected to endosseous dental<br>implants and are intended for<br>provisional use up to 180 days as an<br>aid in prosthetic rehabilitation. | The Nobel Biocare PEEK Abutments<br>are premanufactured prosthetic<br>components directly connected to<br>endosseous dental implants and are<br>intended for provisional use up to<br>180 days as an aid in prosthetic<br>rehabilitation. |
| Intended Use | Temporary Abutment | Provisional Abutment |
| Material | Polyetheretherketone (PEEK) | Polyetheretherketone (PEEK) |
{6}------------------------------------------------
| Colour | White | Natural and White |
|--------------------------------|---------------------------------|---------------------------------|
| Attachment Method | Screw retained | Screw retained |
| | Engaging and Non-Engaging | Engaging and Non-Engaging |
| Duration of use | 180 days, single use | 180 days, single use |
| Implant/Abutment<br>Connection | Southern Implants MAX | Nobel Biocare |
| | • External Hex | • External Hex |
| | • Tri-Nex | • Internal Tri-lobe |
| | • Deep Conical | |
| | • Internal Hex | |
| | • Internal Taper | |
| | • Compact Conical | |
| Height (mm) | 9-14.5 | 12 |
| Collar height | 1-2 | 1-1.5 |
| Diameter (mm) | 3.35 - 7.35 | Not given |
| Abutment Profile | Cylinder with retention feature | Cylinder with retention feature |
| Sterility | Sterile | Non-Sterile |
The Southern Implants PEEK Healing Abutments minimum diameter and height is the same as the Primary Predicate. The Southern Implants range has a slightly higher version and wider version because it is used on wider implant platforms (up to Ø7 compared to the Ø5.5 platform of the predicate).
The Southern Implants PEEK Temporary Abutments have a range of lengths that falls on either side of the Primary Predicate temporary abutment lengths, by up to 3 mm. The difference in length is immaterial since all lengths fall within the range of lengths that a prosthetic can be made to. If the PEEK abutment is shorter it will just need to be built up more, and if it is longer it will just need to be cut down more.
The Southern Implants PEEK abutments are substantially equivalent to predicate device in:
- Indications for Use
- Intended Use
- Material
- Colour
- Attachment Method
- Duration of Use
- Abutment Profile
The Southern Implants PEEK Abutments are delivered to the patients sterile and is not intended for user sterilization as the predicate. The Southern Implants PEEK Abutments are also available in more connection systems as the predicate to compliment the Southern Implants' range.
## Nonclinical Testing and Performance Testing
Non-clinical testing data provided or referenced to demonstrate substantial equivalence included:
- The sterilization method is Gamma radiation and has been validated in accordance with ISO 11137.
- The proposed user sterilization method is steam sterilization and is validated in accordance with ISO 17665.
- Packaging was validated in accordance with ISO 11607. Accelerated aging per ASTM-F-1980
{7}------------------------------------------------
has been applied on the final packaging followed by validating durability to peel and dye tests conditions, in order to substantiate 5 years shelf life.
- . The subject device is manufactured from the same material using the same manufacturing method as the predicate, has the same intended use, and the same patient contact type and duration. The subject devices are biocompatible in accordance with ISO 10993-1.
- . The subject device shown no cytotoxicity when tested in accordance with ISO 10993-5.
#### Clinical Studies
No clinical studies were conducted.
#### Final Conclusion
Substantial equivalence has been shown for Southern Implants PEEK Abutments.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.