Dentium Ti-Base abutments are intended for use on Dentium endosseous dental implants in the edentulous or partially edentulous maxilla or mandible, as an aid in prosthetic rehabilitation. All digitally designed abutments for use with Dentium Ti-Base abutments are intended to be sent to a Dentium-validated milling center for manufacture.
Device Story
Dentium Ti-Base is a patient-specific dental abutment; consists of unalloyed titanium base with precision implant interface and zirconia ceramic superstructure. Used for support of CAD/CAM fabricated zirconia superstructures, hybrid crowns, or bridges. Fabricated by Dentium; superstructures manufactured at Dentium-validated milling centers. Available in multiple diameters (4.5, 5.5 mm) and gingival heights (0.5-2.0 mm); indexed or non-indexed. Used by clinicians to restore chewing function in edentulous or partially edentulous patients. Provides stable connection between endosseous implant and prosthetic restoration.
Clinical Evidence
No clinical data included. Bench testing only: sterilization validation (ISO 17665-2), cytotoxicity (ISO 10993-5), and dynamic compression-bending testing (ISO 14801).
Technological Characteristics
Materials: Unalloyed titanium (ASTM F67) and zirconia ceramic (ISO 13356). Design: CAD/CAM patient-specific abutment with internal implant interface. Dimensions: 4.5-5.5 mm diameter, 0.5-2.0 mm gingival height. Sterilization: Validated to SAL 10^-6.
Indications for Use
Indicated for use on Dentium endosseous dental implants in edentulous or partially edentulous maxilla or mandible to aid in prosthetic rehabilitation. Prescription use only.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 11, 2018
Dentium Co., Ltd. % Linda Schulz Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K171622
Trade/Device Name: Dentium Ti-Base Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: December 14, 2017 Received: December 15, 2017
Dear Linda Schulz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mary S. Runner -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171622
Device Name
Dentium Ti-Base
Indications for Use (Describe)
Dentium Ti-Base abutments are intended for use on Dentium endosseous dental implants in the edentulous or partially edentulous maxilla or mandible, as an aid in prosthetic rehabilitation.
All digitally designed abutments for use with Dentium Ti-Base abutments are intended to be sent to a Dentium-validated milling center for manufacture.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary Dentium Co., Ltd. Dentium Ti-Base
January 11, 2018
#### ADMINISTRATIVE INFORMATION Manufacturer Name Dentium Co., Ltd. 150, Eondong-ro, Giheung-gu Yongin-si, Gyeonggi-do, 443-270 Republic of Korea Telephone: +82-31-207-2200 Fax: +82-31-207-3883 Official Contact Byungsun Kim, Team Manager of Regulatory Affairs Representative/Consultant Linda K. Schulz, BSDH, RDH Kevin A. Thomas, PhD PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, CA 92130 Telephone: +1-858-792-1235 Fax: +1-858-792-1236 Email: LSchulz@paxmed.com KThomas@paxmed.com
### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | Dentium Ti-Base |
|----------------------------|--------------------------------------|
| Common Name | Abutment, implant, dental, endosseou |
| Classification Name | Endosseous dental implant abutment |
| Classification Regulations | 21 CFR 872.3630 |
| Product Code | NHA |
| Classification Panel | Dental Products Panel |
| Reviewing Branch | Dental Devices Branch |
### PREDICATE DEVICE INFORMATION
Primary Predicate for Ti-Base K150367, Neodent Implant System, JJGC Indústria e Comércio de Materiais Dentários SA
Reference Predicates K161713, Dentium CAD/CAM Abutments, Dentium Co., Ltd. K041368, Dentium Company Limited Implantium®, Dentium Co., Ltd. K160965, Dentium SuperLine® Implants, Dentium Co., Ltd.
{4}------------------------------------------------
### INDICATIONS FOR USE
Dentium Ti-Base abutments are intended for use on Dentium endosseous dental implants in the edentulous or partially edentulous maxilla or mandible, as an aid in prosthetic rehabilitation.
All digitally designed abutments for use with Dentium Ti-Base abutments are intended to be sent to a Dentium-validated milling center for manufacture.
#### DEVICE DESCRIPTION
Dentium Ti-Base is a patient specific abutment fabricated by Dentium and composed of unalloyed titanium and zirconia ceramic. The Ti-Base has a titanium pre-manufactured precision abutment/implant interface and ceramic superstructure bonded to the post of the titanium base. Ti-Base is used for support of CAD/CAM fabricated zirconia superstructures or for support of hybrid crowns or bridges. Hybrid crowns or bridges and zirconia superstructures are manufactured by a Dentium validated milling center.
Ti-Base Abutments are available in two prosthetic diameters (4.5 and 5.5 mm) and four gingival heights (0.5, 1.0, 1.5, and 2.0 mm). Each of the abutments is available indexed and non-indexed. All subject device abutments are compatible with Dentium and Dentium Superline dental implants.
#### PERFORMANCE DATA
Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence included: sterilization validation according to ISO 17665-2 to an SAL of 10°, Cytotoxicity testing according to ISO 10993-5 determining the subject device is non-cytotoxic , and dynamic compression-bending testing according to ISO 14801 ensuring that the subject device is strong enough for its intended use. No clinical data were included in this submission.
### EQUIVALENCE TO MARKETED DEVICE
The subject device is substantially equivalent in indications and design principles to the predicate devices shown above. Below is a summary table showing technical comparison between the subject device abutments and the predicate device abutments.
| | Indications for Use Statement |
|---------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device | |
| Dentium Ti-Base<br>Dentium Co., Ltd. | Dentium Ti-Base abutments are intended for use on Dentium endosseous dental implants<br>in the edentulous or partially edentulous maxilla or mandible, as an aid in prosthetic<br>rehabilitation.<br>All digitally designed abutments for use with Dentium Ti-Base Abutments are intended to<br>be sent to a Dentium-validated milling center for manufacture. |
| Primary Predicate Device | |
| K150367<br>Neodent Implant System<br>JJGC Indústria e Comércio de<br>Materiais Dentários SA | Titanium Base Abutment is a titanium base placed onto Neodent dental implants to<br>provide support for customized prosthetic restorations. It is used with a coping and crown,<br>or crown alone, and is indicated for cement-retained single or multi-unit restorations, or<br>screw-retained single restorations.<br>PreFace Abutment is a titanium abutment to be used in fabricating a full custom abutment<br>and placed onto Neodent dental implants to provide support for customized prosthetic<br>restorations. PreFace Abutments are indicated for screw-retained single restorations or<br>cement-retained single or multi-unit restorations. |
| Table of Substantial Equivalence - Indications for Use Statement | | |
|------------------------------------------------------------------|--|--|
| | | |
{5}------------------------------------------------
| Reference Predicate Devices | Indications for Use Statement |
|------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K161713<br>Dentium CAD/CAM Abutments<br>Dentium Co., Ltd. | Dentium abutments are intended for use on endosseous dental implants in the edentulous<br>or partially edentulous maxilla or mandible, as an aid in prosthetic rehabilitation.<br>All digitally designed abutments for use with Dentium CAD/CAM Abutments are<br>intended to be sent to a Dentium-validated milling center for manufacture. |
| K041368<br>Dentium Company Limited<br>Implantium®<br>Dentium Co., Ltd. | The Dentium Co., Ltd Implantium is intended to be surgically placed in the bone of the<br>upper or lower jaw arches to provide support for prosthetic devices, such as artificial<br>teeth, and to restore the patient's chewing function. This may be accomplished by either a<br>two-stage surgical procedure or a single surgical procedure. If a single surgical procedure<br>is used, single or multiple implants may be inserted (type I, II or III bone) provided good<br>initial stability (> 40 Ncm) is achieved. Not intended for immediate loading. |
| K160965<br>Dentium SuperLine® Implants<br>Dentium Co., Ltd. | Superline is indicated for use in surgical and restorative applications for placement in the<br>bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial<br>teeth, in order to restore the patient's chewing function. Superline Implant System is<br>indicated also for immediate loading when good primary stability is achieved and with<br>appropriate occlusal loading. |
| Comparison | Subject Device | Primary Predicate Device | Reference Predicate Device |
|---------------------------------------|--------------------------------------|----------------------------------------------------------------------------------------------|----------------------------------------------------------|
| | Dentium Ti-Base<br>Dentium Co., Ltd. | K150367<br>Neodent Implant System<br>JJGC Indústria e Comércio de<br>Materiais Dentários SA. | K161713<br>Dentium CAD/CAM Abutments<br>Dentium Co. Ltd. |
| Design | | | |
| Abutment Design | CAD/CAM Ti-Base | CAD/CAM Blank<br>CAD/CAM TiBase | CAD/CAM Blank<br>CAD/CAM Custom Abutment |
| Prosthesis<br>Attachment | Cement-retained<br>Screw-retained | Cement-retained<br>Screw-retained | Cement-retained<br>Screw-retained |
| Restoration | Single-unit,<br>Multi-unit | Single-unit,<br>Multi-unit | Single-unit,<br>Multi-unit |
| Abutment/Implant<br>Platform Diameter | 3.6 - 7.0 mm | 3.0 - 6.0 mm | 3.6 - 5.0 mm |
| Abutment Angle | Up to 30° | Up to 30° | Up to 30° |
| Abutment/ Implant<br>Interface | Internal | Internal | Internal |
| Material | | | |
| Abutment | Unalloyed Titanium<br>ASTM F67 | Titanium Alloy<br>ASTM F136 | Unalloyed Titanium<br>ASTM F67 |
| Superstructure | Zirconia<br>ISO 13356 | Zirconia<br>ISO 13356 | NA |
The subject device Ti-Base is substantially equivalent to the predicate Neodent Titanium Base Abutment cleared in K150367 in design, function, material, manufacture and intended use. Both are titanium abutment for support of direct crowns/bridges or CAD/CAM fabricated zirconia copings manufactured by a validated milling center. Differences between the subject device and predicate device are in the design dimensions and implant/abutment interface.
{6}------------------------------------------------
All subject device abutments are by prescription order from a clinician, and have the same implant/abutment interface as the predicate K161713. Subject device abutments are compatible with Dentium Implantium (K041368) and Dentium SuperLine (K160965) dental implants.
The Indications for Use of the subject device is substantially equivalent to the previously cleared Intended Use in K150367. The Indications for Use Statement in K150367 includes two abutment designs (TiBase and Preface) and names the individual restoration options. The Indications for Use Statement for the subject device is for one device and does not list the individual restoration options. Minor differences in wording between the Indications for Use Statements do not affect the intended use of a dental abutment placed on a dental implant for prosthetic restoration.
The Indications for Use of the subject device is substantially equivalent as Dentium CAD/CAM Abutments in K161713, with the only difference being in the proprietary name of the device.
## CONCLUSION
The subject device and the predicate devices have intended use, have similar technological characteristics, and are made of similar materials. The subject device and predicate devices encompass the same range of physical dimensions, including diameter, gingival height, and angle of the abutments. The subject and predicate devices are packaged in similar materials and are to be sterilized using similar methods.
The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.