NobelProcera HT ML FCZ Implant Bridge and Framework
K160158 · Nobel Biocare AB · NHA · Jul 25, 2016 · Dental
Device Facts
Record ID
K160158
Device Name
NobelProcera HT ML FCZ Implant Bridge and Framework
Applicant
Nobel Biocare AB
Product Code
NHA · Dental
Decision Date
Jul 25, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NobelProcera® HT ML FCZ (full contour zirconia) and framework Implant Bridge are indicated for use as a bridge anatomically shaped and/or framework in the treatment of partially or totally edentulous jaws for the purpose of restoring chewing function.
Device Story
Individualized, screw-retained dental implant bridge/framework for partially dentate or edentulous patients. Device consists of titanium base and multi-layered zirconia; available in 6 shades. Designed via CAD; manufactured at NobelProcera facility. Available as full anatomic contour (requiring minimal lab processing) or framework (requiring veneering). Attaches to internal conical connection, external hex, or internal tri-channel implants/multi-unit abutments. Used by dental professionals to restore chewing function. Benefits include customized fit and restoration of oral function.
Clinical Evidence
No clinical data. Bench testing only. Fatigue testing performed per modified ISO 14801. Biocompatibility testing (cytotoxicity per ISO 10993-5, leachables/extractables analysis) performed on ceramic and titanium alloy materials. Sterilization validated per ISO 17665-1/2.
Technological Characteristics
Materials: Zirconium Oxide (ISO 6872) and titanium vanadium alloy (ASTM F136). Design: 2-14 unit bridges; CAD-designed. Connectivity: Compatible with internal conical, external hex, and internal tri-channel implant platforms. Sterilization: Validated per ISO 17665-1/2.
Indications for Use
Indicated for partially or totally edentulous patients requiring dental implant-supported bridges or frameworks to restore chewing function.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 25, 2016
Nobel Biocare AB % Charlemagne Chua Senior Regulatory Affairs Manager Nobel Biocare USA LLC 22715 Savi Ranch Parkway Yorba Linda. California 92887
Re: K160158
Trade/Device Name: NobelProcera HT ML FCZ Implant Bridge and Framework Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: June 14, 2016 Received: June 15, 2016
Dear Charlemagne Chua:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/8 description: The image shows a signature with the name "Tina Kiang" written above the letter "S". The signature is written in black ink and is on a white background. The signature is stylized and difficult to read.
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration<br>Indications for Use | | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| 510(k) Number (if known) | K160158 | |
| Device Name | NobelProcera HT ML FCZ Implant Bridge and framework | |
| Indications for Use (Describe) | The NobelProcera® HT ML FCZ (full contour zirconia) and framework Implant Bridge are indicated for use as a bridge anatomically shaped and/or framework in the treatment of partially or totally edentulous jaws for the purpose of restoring chewing function. | |
| Type of Use (Select one or both, as applicable) | <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | |
| PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. | | |
| FOR FDA USE ONLY | | |
| Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) | | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995.<br>*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: | | |
| Department of Health and Human Services<br>Food and Drug Administration<br>Office of Chief Information Officer<br>Paperwork Reduction Act (PRA) Staff<br>PRAStaff@fda.hhs.gov | | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | |
| FORM FDA 3881 (1/14)<br>JUNE 2016 | Page 1 of 1 | |
| | PSC Publishing Services (301) 443-6740 EF<br>K160158-S001<br>Page 25 of 412 | |
10
{3}------------------------------------------------
# A.4.
#### 510(k) Summary
#### SUBMITTER .
Nobel Biocare AB Vastra Hamngatan 1 Goteborg, SE-411 17 Sweden
Submitted by: Nobel Biocare USA LLC 22715 Savi Ranch Parkway Yorba Linda, CA 92887
Contact Person: Charlemagne Chua, Senior Regulatory Affairs Manager Phone: (714) 282-4800 x 7830 Fax: (714) 998-9348
Date Prepared: July 15, 2016
II. DEVICE
Name of Device: NobelProcera HT ML FCZ Implant Bridge and Framework Common or Usual Name: Endosseous Dental Implant Abutment Classification Name: Endosseous Dental Implant Abutment (21 CFR 872.3640) Requlatory Class: II Product Code: NHA
#### III. PREDICATE DEVICE
Nobel Biocare - NobelProcera Implant Bridge Zirconia (K091907) Nobel Biocare -Angulated Screw Channel Abutment Conical Connection (K132746)
#### IV. DEVICE DESCRIPTION
The NobelProcera HT ML FCZ Implant Bridge and Framework is an individualized implant supported screw-retained dental implant bridge for the partially dentate and edentulous patients.
The NobelProcera HT ML FCZ Implant Bridge and Framework is available as either a framework requiring veneering in a dental lab or as a full anatomic contour design requiring minimum laboratory processing. The NobelProcera HT ML FCZ Implant Bridge and Framework is available for bridges between 2 to 14
Nobel Biocare Traditional 510(k) Notification NobelProcera HT ML FCZ Implant Bridge and Framework July 2016
{4}------------------------------------------------
units. The bridge is made of a titanium base and a highly transparent multi layered zirconia material which is available in 6 shades.
The NobelProcera HT ML FCZ Implant Bridge and Framework is available for the Nobel Biocare internal conical connection, external hex, and internal tri-channel implants and the Multi-unit abutments.
### V. INDICATIONS FOR USE
The NobelProcera® HT ML FCZ (full contour zirconia) and framework Implant Bridge are indicated for use as a bridge anatomically shaped and/or framework in the treatment of partially or totally edentulous jaws for the purpose of restoring chewing function.
{5}------------------------------------------------
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
| Technological<br>characteristics | | Subject Device | Predicate |
|----------------------------------|------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| | | NobelProcera HT ML FCZ<br>Implant Bridge and Framework | NobelProcera Implant Bridge Zirconia<br>(K091907) |
| Design Features | Compatible<br>Implant<br>/Abutment<br>Platform | Nobel Biocare<br>Internal Conical Connection NP, RP, WP<br>Internal Tri-Channel NP, RP, WP, 6.0<br>External Hex NP, RP, WP<br>Multi-unit NP/RP, WP | Nobel Biocare<br>Internal Conical Connection NP, RP, WP<br>Internal Tri-Channel NP, RP, WP, 6.0<br>External Hex NP, RP, WP<br>Multi-unit NP/RP, WP |
| | | | Astra Tech<br>Yellow, Aqua, Lilac |
| | | | Straumann<br>Octagon 3.5, 4.8, 6.5<br>Bone Level 3.3, 4.1/4.8 |
| | | | Biomet 3i<br>Internal 3.4, 3.75/4.0, 5.0, 6.0<br>External 3.4, 4.1, 5.0, 6.0 |
| | | | Zimmer<br>Screw-Vent 3.5, 4.5, 5.7<br>AdVent 4.5 |
| | Device Material | Zirconium Oxide with titanium vanadium alloy implant<br>interface for Internal Conical Connection. | Zirconium Oxide |
| | Bridge Design | Individualized full anatomic contour or framework | Individualized framework |
| | Arc Length | 2 to 14 units | 2 to 14 units |
| | Design method | CAD | Wax-up or CAD design |
Nobel Biocare Traditional 510(k) Notification NobelProcera HT ML FCZ Implant Bridge and Framework July 2016
{6}------------------------------------------------
| Technological<br>characteristics | Subject Device | Predicate |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturing<br>method | NobelProcera HT ML FCZ<br>Implant Bridge and Framework | NobelProcera Implant Bridge Zirconia<br>(K091907) |
| | Industrialized manufacturing at NobelProcera<br>manufacturing facility | Industrialized manufacturing at NobelProcera<br>manufacturing facility |
| | | |
| Intended use | The NobelProcera HT ML FCZ Implant Bridge and<br>Framework are customized dental implant bridges<br>intended to restore chewing functions. The Implant<br>Bridge is intended to directly attaches to the endosseous<br>dental implants and/or onto Nobel Biocare's Multi-unit<br>Abutments with clinical screws and provides a platform<br>for restoration. | The NobelProcera Implant Bridge Zirconia is a bridge<br>framework that attaches to implants or abutments. The<br>NobelProcera Implant Bridge Zirconia is intended to be<br>finished into a dental prosthesis using standard laboratory<br>dental materials such as resin composite or porcelain<br>veneer. |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| Indication for Use | The NobelProcera® HT ML FCZ (full contour zirconia)<br>and framework Implant Bridge are indicated for use as a<br>bridge anatomically shaped and/or framework in the<br>treatment of partially or totally edentulous jaws for the<br>purpose of restoring chewing function. | The NobelProcera Implant Bridge Zirconia is indicated for<br>use as a bridge framework in the treatment of partially or<br>totally edentulous jaws for the purpose of restoring<br>chewing function. |
| | | |
| | | |
| | | |
| | | |
The difference in the Indications for Use is the addition of the term "bridge anatomically shaped" allows for an anatomically shaped bridge as compared to just a bridge framework in the primary predicate device. The addition of an anatomically shaped bridge does not change the intended use of the implant bridge dental abutment device.
The additional difference between the subject device and the primary predicate is the inclusion of a titonia material and the implant body. The reference predicate (K132746) supports the addition of a titanium interface between the zirconia material and the implant body.
Nobel Biocare Traditional 510(k) Notification NobelProcera HT ML FCZ Implant Bridge and Framework July 2016
{7}------------------------------------------------
# VII. PERFORMANCE DATA
The fatique limit of the NobelProcera HT ML FCZ Implant Bridge and Framework was determined using a modified version of ISO 14801. The modifications to ISO 14801 were done to reflect the likely worst-case clinical use of the device. Both the subject and predicate device were tested under identical conditions. The results of the testing were used to address questions related to substantial equivalence based on difference in design between the subject and predicate devices.
Validation of the recommended sterilization parameters was conducted according to ISO 17665-1 and ISO 17665-2.
Cytotoxicity testing according to ISO 10993-5 as well as BC-MS analysis for leacables/extractable was conducted on the ISO 6872 ceramic and ASTM F136 titanium alloy material composition of the device.
No clinical data was used to support the decision of substantial equivalence.
#### VIII. CONCLUSIONS
The NobelProcera HT ML FCZ Implant Bridge and Framework was evaluated for substantial equivalence using standard and/or comparative testing. Based on technological characteristics and non-clinical test data included in this submission, the NobelProcera HT ML FCZ Implant Bridge and Framework has been shown to be substantially equivalent to the NobelProcera Implant Bridge Zirconia.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.