← Product Code [NHA](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA) · K153779

# Abutment for Bridges (K153779)

_Altatec GmbH · NHA · Mar 31, 2016 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA/K153779

## Device Facts

- **Applicant:** Altatec GmbH
- **Product Code:** [NHA](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA.md)
- **Decision Date:** Mar 31, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3630
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

CAMLOG® and CONELOG® Abutments are intended to be used to fabricate prosthetic restorations in conjunction with CAMLOG® and CONELOG® implants to support prostheses in the maxillary and/or mandibular arch. iSy® Abutments are intended to be used to fabricate prosthetic restorations in conjunction with iSy implants to support prostheses in the maxillary and/or mandibular arch.

## Device Story

Abutment for Bridges is a straight, titanium dental abutment for multi-unit, cement-retained restorations; used in conjunction with CAMLOG, CONELOG, and iSy implant systems. Device facilitates prosthetic restoration support in maxillary/mandibular arches. Available in various platform diameters and gingival heights. Temporary Abutment version is a one-piece titanium component for provisional restorations. Used by dental professionals in clinical settings. Device connects to implants to provide a base for crowns/bridges; lacks anti-rotation features, limiting use to multi-unit applications. Benefits include stable prosthetic support for patients undergoing dental implant procedures.

## Clinical Evidence

No clinical data were included in this submission. Substantial equivalence is supported by non-clinical bench testing, including sterilization validation (ISO 17665-1, ISO 17665-2) and biocompatibility assessment (ISO 10993-1) referencing predicate device data.

## Technological Characteristics

Materials: Ti-6Al-4V ELI. Design: Straight, multi-unit abutment; non-indexed (no anti-rotation feature). Sterilization: Moist heat (ISO 17665-1/2). Connectivity: Mechanical interface with CAMLOG, CONELOG, and iSy implant systems.

## Regulatory Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

## Predicate Devices

- CONELOG® Implant System ([K113779](/device/K113779.md))

## Reference Devices

- iSy® Implant System ([K133991](/device/K133991.md))
- CAMLOG Implant System Modified Implants and Abutments ([K083496](/device/K083496.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol representing the human form, depicted as three overlapping profiles facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 31, 2016

ALTATEC GmbH c/o Ms. Linda Schultz Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130

Re: K153779

Trade/Device Name: Abutment for Bridges Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous dental implant abutment Regulatory Class: II Product Code: NHA Dated: December 30, 2015 Received: December 31, 2015

Dear Ms. Schultz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Ms. Linda Schultz:

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tina
Kiang -S

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

K153779

#### Device Name

Abutment for Bridges

Indications for Use (Describe)

# CAMLOG® and CONELOG®

CAMLOG® and CONELOG® Abutments are intended to be used to fabricate prosthetic restorations in conjunction with CAMLOG® and CONELOG® implants to support prostheses in the maxillary and/or mandibular arch.

iSy®

iSy® Abutments are intended to be used to fabricate prosthetic restorations in conjunction with iSy implants to support prostheses in the maxillary and/or mandibular arch.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary ALTATEC GmbH

# Abutment for Bridges

March 30, 2016

# ADMINISTRATIVE INFORMATION

| Manufacturer Name         | ALTATEC GmbH<br>Maybachstrasse 5 |                            |
|---------------------------|----------------------------------|----------------------------|
|                           | D-71299 Wimsheim, Germany        |                            |
|                           | Telephone:                       | +49 7044 9445 0            |
|                           | Fax:                             | +49 7044 9445 723          |
| Official Contact          | Colleen Boswell                  |                            |
|                           | Director of RA/QA, Henry Schein  |                            |
| Representative/Consultant |                                  | Linda K. Schulz, BSDH, RDH |
|                           | Floyd G. Larson, MS, MBA         |                            |
|                           | PaxMed International, LLC        |                            |
|                           | 12264 El Camino Real, Suite 400  |                            |
|                           | San Diego, CA 92130              |                            |
|                           | Telephone:                       | +1 (858) 792-1235          |
|                           | Fax:                             | +1 (858) 792-1236          |
|                           | Email:                           | lschulz@paxmed.com         |
|                           |                                  | flarson@paxmed.com         |

# DEVICE NAME AND CLASSIFICATION

| Trade/Proprietary Name     | Abutment for Bridges               |
|----------------------------|------------------------------------|
| Common Name                | Endosseous dental implant abutment |
| Classification Name        | Endosseous dental implant abutment |
| Classification Regulations | 21 CFR 872.3630, Class II          |
| Product Code               | NHA                                |
| Classification Panel       | Dental Products Panel              |
| Reviewing Branch           | Dental Devices Branch              |

{4}------------------------------------------------

510(k) Summary Page 2 of 4

# PREDICATE INFORMATION

Primary Predicate ALTATEC GmbH, CONELOG® Implant System - K113779 Reference Predicates ALTATEC GmbH, iSy® Implant System - K133991 ALTATEC GmbH, CAMLOG Implant System Modified Implants and Abutments - K083496

INTENDED USE

CAMLOG® and CONELOG®

CAMLOG® and CONELOG® Abutments are intended to be used to fabricate prosthetic restorations in conjunction with CAMLOG® and CONELOG® implants to support prostheses in the maxillary and/or mandibular arch.

iSy®

iSy Abutments are intended to be used to fabricate prosthetic restorations in conjunction with iSy implants to support prostheses in the maxillary and/or mandibular arch.

# DEVICE DESCRIPTION

Abutment for Bridges is a straight abutment designed for multi-unit, cement-retained restorations. It is available with two gingival heights (0.8 and 2.0 mm) for CONELOG and iSy and one gingival height for CAMLOG.

Abutment for Bridges is available for CAMLOG, CONELOG and iSy Implant Systems. It is available in five CAMLOG platform diameters (3.3 mm, 4.3 mm, 5.0 mm, and 6.0 mm), four CONELOG platform diameters (3.3 mm, 3.8 mm, 4.3 mm and 5.0 mm), and one iSy interface connection.

The Temporary Abutment is a straight, one-piece titanium abutment for cement-retained provisional restorations. It is available in a non-indexed bridge version for CAMLOG, CONELOG and iSy, and an indexed crown version for CAMLOG implant system.

# PERFORMANCE DATA

Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence included: sterilization validation according to ISO 17665-1 Sterilization of health care products – Moist heat – Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices, ISO 17665-2 Sterilization of health care products – Moist heat – Part 2: Guidance on the application of ISO 17665-1, and ISO 10993-1 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process, and reference to the applicant's own predicate devices in K113779, K133991, and K0823496 for demonstration of biocompatibility.

No clinical data were included in this submission.

{5}------------------------------------------------

# EQUIVALENCE TO MARKETED DEVICE

Abutment for Bridges is equivalent to the abutments in K113779 and K083496. They are straight abutments designed for cement retained restorations. The primary design difference is that the Abutment for Bridges has no anti-rotation feature in the implant/abutment connection or on the post for restoration.

Temporary Abutment is equivalent to the temporary abutment in K113779. The primary difference is that the Temporary Abutment has no anti-rotation feature because it is a multi-unit abutment.

|                                  | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                    | Primary Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                  | ALTATEC GmbH Abutment for Bridges<br>K153779                                                                                                                                                                                                                                                                                                                                                                                                      | ALTATEC GmbH CONELOG® Implant System<br>K113779                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Indications for Use              | CAMLOG® and CONELOG®<br>CAMLOG® and CONELOG® Abutments are intended to<br>be used to fabricate prosthetic restorations in conjunction<br>with CAMLOG® and CONELOG® implants to support<br>prostheses in the maxillary and/or mandibular arch.<br><br>iSy®<br>iSy Abutments are intended to be used to fabricate<br>prosthetic restorations in conjunction with iSy® implants<br>to support prostheses in the maxillary and/or mandibular<br>arch. | Implant Indications for Use<br>CONELOG® Implant System Implants are intended for immediate<br>or delayed placement in the bone of the maxillary or mandibular<br>arch. CONELOG® Implant System Abutments are intended for<br>use as support for crowns, bridges or overdentures. When a one- stage surgical approach is applied, the implant may be<br>immediately loaded when good primary stability is achieved and<br>the functional load is appropriate.<br><br>CONELOG® Implants with 3.3 mm diameter have the following<br>additional specific indications:<br>These are an alternative in cases where the alveolar ridge width<br>is only 5 – 6 mm. Because of their lower mechanical strength<br>compared with larger diameter implants, they should only be<br>used under the following conditions:<br>As single implants, they should be used only to replace mandibular incisors and/or maxillary lateral incisors. An edentulous arch can only be restored with a bar retained superstructure with at least four implants of 3.3 mm diameter without distal extensions. Implants of Ø 3.3 mm are suitable for a partially edentulous arch when combined with implants of larger diameter for |
|                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                   | splinted superstructures. However, the limited strength of the<br>implants with Ø 3.3 mm must be taken into account.<br>Avoid excessive mechanical stressing of the implants when using ball abutments in combination with Ø 3.3 mm implants The healing time for Ø 3.3 mm implants is at least 12 weeks.<br>CONELOG® Implants with 7 mm length have the following<br>additional specific indications:<br>CONELOG® SCREW-LINE Implants should only be used<br>when there is not enough space for a longer implant. Delayed<br>loading in single tooth replacement is indicated with these<br>implants. If the ratio of crown length to implant length is<br>unfavorable the biomechanical risk factors have to be considered<br>and appropriate measures have to be taken by the dental<br>professional.                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Abutment Indications for Use<br>CONELOG® Abutments are intended to be used to fabricate<br>prosthetic restorations in conjunction with CONELOG implants to<br>support prostheses in the maxillary and/or mandibular arch.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Vario SR Indications for Use<br>CONELOG® Vario SR components for crown and bridge<br>restorations:<br>Occlusal screw-retained crown, bridge and bar constructions on CONELOG® implants in the anterior and posterior region of the maxilla and mandible<br>CONELOG® Vario SR components for bar restorations:<br>Anchorage of implant-supported full dentures for the edentulous maxilla in conjunction with 4 or more CONELOG® implants and in the edentulous mandible in conjunction with 2 or 4 or more CONELOG® implants.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                  | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                    | Primary Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                  | ALTATEC GmbH Abutment for Bridges                                                                                                                                                                                                                                                                                                                                                                                                                 | ALTATEC GmbH CONELOG® Implant System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                  | K153779                                                                                                                                                                                                                                                                                                                                                                                                                                           | K113779                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Design                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Abutment/Implant<br>Diameter, mm | 3.3, 3.8, 4.3, 4.4, 5.0, 6.0                                                                                                                                                                                                                                                                                                                                                                                                                      | 3.3, 3.8, 4.3, 5.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Abutment Angle                   | Straight                                                                                                                                                                                                                                                                                                                                                                                                                                          | Up to 20°                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Restoration                      | Cement-retained                                                                                                                                                                                                                                                                                                                                                                                                                                   | Screw-retained, Cement-retained                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Material                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Abutments and<br>Abutment Screw  | Ti-6Al-4V ELI                                                                                                                                                                                                                                                                                                                                                                                                                                     | Ti-6Al-4V ELI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |

{6}------------------------------------------------

The Indications for Use for the subject device differs from the identified primary predicate in that the submission device Indications for Use is a sub-set of the Indications for Use of the primary predicate. This does not change the intended use of the subject device as compared to the predicate because the subject device, a dental abutment, is a subset of the primary predicate, a complete implant system. The intended use of the subject device is similar to the intended use of the dental abutments cleared in the primary predicate.

Implant/abutment interfaces for CAMLOG, CONELOG and iSy subject device abutments are similar to the corresponding predicate device Implant Systems as identified in the primary and reference predicate devices. However, the lack of an anti-rotation feature does not affect the fit of the abutment to the corresponding implant and is further mitigated by the limitation of the Abutment for Bridges and non-indexed Temporary Abutment devices as multi-unit abutments.

# CONCLUSION

Overall, the subject device has the following similarities to the predicate devices:

- has the same intended use, ●
- uses the same operating principle,
- incorporates the same basic design, ●
- incorporates the same or very similar materials, and
- . has similar packaging and is to be sterilized using the same processes.

The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA/K153779](https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA/K153779)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
