K133696 · Jjgc Industria E Comercio DE Materiais Dentarios S · NHA · Feb 25, 2014 · Dental
Device Facts
Record ID
K133696
Device Name
NEODENT IMPLANT SYSTEM
Applicant
Jjgc Industria E Comercio DE Materiais Dentarios S
Product Code
NHA · Dental
Decision Date
Feb 25, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Multiple tooth applications may be rigidly splinted.
Device Story
Neodent Implant System components (Facility Universal Post and Equator Attachment CM abutments) function as endosseous dental implant abutments; interface with dental implants via Morse taper. Facility Universal Post abutments (straight, post-type) support cemented prosthetic restorations; Equator Attachment CM abutments (straight, ball-type) support overdentures or partial dentures. Used in dental clinics by licensed dentists/oral surgeons. Abutments provide mechanical connection between dental implant and prosthetic device; titanium nitride (TiN) coating increases surface hardness and wear resistance. Clinical benefit includes restoration of chewing function and support for artificial teeth. No software or automated analysis involved.
Clinical Evidence
Bench testing only; no clinical data submitted.
Technological Characteristics
Materials: Titanium alloy (ASTM F136, UNS R56401). Surface: Titanium nitride (TiN) coating. Interface: Morse taper. Components: Facility Universal Post (3.3mm diameter, 1-5mm gingival height, 4/6mm prosthetic height); Equator Attachment CM (1.5-5.5mm gingival height). Sterilization: Standard methods. No software or electronic components.
Indications for Use
Indicated for patients requiring dental restoration via endosseous implants in the upper or lower jaw to support prosthetic devices (artificial teeth) and restore chewing function; applicable for single or multiple unit restorations; suitable for immediate loading given sufficient primary stability.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
K150199 — Neodent Implant System · Jjgc Industria E Comercio DE Materiais Dentarios S · Jun 15, 2015
K172505 — MIS C1 Narrow Platform Conical Connection Implant System, MIS C1 Wide Platform Conical Connection Abutments · Mis Implants Technologies , Ltd. · Dec 28, 2017
K071919 — P.004 ABUTMENTS · Straumann USA (On Behalf of Institut Straumann Ag) · Sep 26, 2007
K173701 — Implant-One System · Implant Logistics, Inc. · Nov 29, 2018
K233167 — LW Implant System Abutment · Ossvis Co., Ltd. · Mar 25, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
# Kr33696
ﯩﻨﯩﻐ
510(k) Summary
Neodent Implant System
."
# FEB 2 5 2014
.. ""
# 510(k) Summary
# JJGC Indústria e Comércio de Materiais Dentários SA
## Neodent Implant System
December 2, 2013
| Manufacturer Name | JJGC Indústria e Comércio de Materiais Dentários SA<br>Av. Juscelino Kubitschek de Oliveira, 3291 - CIC<br>Curitiba, Paraná, 81270-200, Brazil |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| | Telephone: +55 41 2169 1003<br>Fax: +55 41 2169 1043 |
| Official Contact | Jafte Carneiro Fagundes da Silva<br>Chief Compliance, Legal and Regulatory Officer – CCLRO |
| Representative/Consultant | Kevin A. Thomas, PhD<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130 |
| | Telephone: +1 (858) 792-1235<br>Fax: +1 (858) 792-1236<br>Email: kthomas@paxmed.com |
## DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | Neodent Implant System |
|----------------------------|------------------------------------|
| Common Name | Endosseous dental implant abutment |
| Classification Regulations | Class II, 21 CFR 872.3630 |
| Product Code | NHA |
| Classification Panel | Dental Products Panel |
| Reviewing Branch | Dental Devices Branch |
#### INTENDED USE
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Multiple tooth applications may be rigidly splinted.
{1}------------------------------------------------
·
### DEVICE DESCRIPTION
The purpose of this submission is to expand the Neodent Implant System components cleared under K101945 and K123022. This submission includes Facility Universal Post and Equator Attachment CM abutments with a Morse taper interface. Facility Universal Post abutments are straight, post-type abutments for cemented prosthetic restorations. They are provided in one platform diameter (3.3 mm), gingival heights of 1 mm to 5 mm, and prosthetic heights of 4 mm and 6 mm. Facility Universal Post abutments are compatible with the Facility dental implants cleared in K123022, and with the Universal Post Cylinders cleared in K101945. Equator Attachment CM abutments are straight, ball-type abutments for the attachment of overdentures or partial dentures. They are provided in five gingival heights (1.5, 2.5, 4.5 and 5.5 mm). These abutments have a titanium nitride (TiN) coating for the purpose of creating a harder surface to increase wear resistance. This titanium nitride coating is identical to that used on the Neodent CM Mini Ball Attachment cleared in K101945, and the Facility Equator Attachment cleared in K123022. Equator Attachment CM abutments are compatible with Titamax CM, Titamax CM EX, and Alvim CM implants cleared in K101945, and with the CM Drive implants cleared in K123022. All abutments in this submission are made of titanium alloy conforming to ASTM F136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401).
### EQUIVALENCE TO MARKETED DEVICE
Neodent Implant System is substantially equivalent in indications and design principles to the following legally marketed predicate devices:
JJGC Indústria e Comércio de Materiais Dentários SA, Neodent Implant System, K101945 JJGC Indústria e Comércio de Materiais Dentários SA, Neodent Implant System, K123022
The subject device and the predicate devices have intended use, have similar technological characteristics, have similar designs and dimensions, use the same materials and the same surface treatment (titanium nitride) as abutments cleared under K101945 and K123022. The subject device and predicate devices encompass the same range of physical dimensions, including prosthetic height, gingival height, platform diameter, and angulation (straight). The subject and predicate devices are packaged in the same materials and sterilized using the same methods.
Non-clinical data submitted, referenced, or relied upon to demonstrate substantial equivalence include: biocompatibility, engineering analysis and dimensional analysis. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above. Clinical data were not submitted in this premarket notification.
The data included in this submission demonstrates substantial equivalence to the predicate devices listed above.
{2}------------------------------------------------
Image /page/2/Picture/12 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with three stripes extending from its wing, symbolizing service to the nation. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" encircles the eagle, indicating the department's name and country.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 25, 2014
JJGC Indústria e Comércio de Materiais Dentários SA C/O Dr. Kevin Thomas Regulatory Consultant 12264 El Camino Real, Suite 400 San Diego, CA 92130
Re: K133696
Trade/Device Name: Neodent Implant System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: February 3, 2014 Received: February 4, 2014
Dear Dr. Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
Page 2 - Ms. Cheng
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Image /page/3/Picture/6 description: The image shows the name "Kwame O. Ulmer" in a bold, sans-serif font. Below the name, there is some additional text that is difficult to read due to its small size and the image quality. The text appears to be related to the letters "FDA", but the details are unclear.
for
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### 510(k) Summary
Neodent Implant System
## Indications for Use
KI33696 510(k) Number:
Neodent Implant System Device Name:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Multiple tooth applications may be rigidly splinted.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Sean Runner
Mary S. Runner -5
POS 04 291462.25
06.58.26 -05'00'
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.